Infrared Thermometer 

Operation Manual 

FIRT1000

 

                                                       

 

         

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

TABLE OF CONTENTS 

 

INTRODUCTION

……………………………………….3 

FEATURES

…………………………………………….3

 

WIDE  RANGE  APPLICATION 

………………………….4

 

1 SAFETY

……………………………………………….4

 

2.DISTANCE  &  SPOT  SIZE

……………………………5

 

3  SPECIFICATIONS

……………………………………6

 

4  FRONT  PANEL  DESCRIPTION

………………………7

 

5  INDICATOR

…………………………………………..7 

6  BUTTONS

……………………………………………8 

7  MEASURMENT  PERATION

…………………………10

 

8 BATTERY PLACEMENT

……………………………..11

 

9  NOTES

………………………………………….  …12

 

10  MAINTENANCE  &  CLEARING

…………………….14

 

 

 

 

 

 


background image

 

INTRODUCTION 

Thank you for purchase of the IR Thermometer. This    Series is capable 

of  non-contact  (infrared)  temperature  measurements  at  the  touch  of  a 

button.  The  built-in  laser  pointer  increases  target  accuracy  while  the 

backlight  LCD  and  handy  push-buttons  combine  for  convenient, 

ergonomic operation. 

The  Non-contact  Infrared  Thermometers  can  be  used  to  measure  the 

temperature  of  objects’  surface  that  is  improper  to  be  measured  by 

traditional  (contact)  thermometer  (such  as  moving  object,  the  surface 

with electricity current or the objects which are uneasy to be touched.) 

Proper use and care of this meter will provide years of reliable service. 

 
FEATURES: 

 

Precise non-contact measurements 

 

High  distance  to  target  ratio  measures  smaller  surface  areas  at 

greater distances   

 

Widest temperature range 

 

Unique flat surface, modern housing design 

 

Built-in laser pointer 

 

Automatic Data Hold   

 

@

/

  switch 

 

Emissivity digitally adjustable from 0.10 to 1.0 

 

MAX,MIN,DIF, AVG temperature displays 

 

Backlight LCD display 

 

Built-in laser pointer 

 

Automatic selection range and Display Resolution 0.1

C(0.1

F) 

 

Trigger lock 

 

Set high and low alarms 

 

 


background image

 

WIDE RANGE APPLICATION:   

Food  preparation,  Safety  and  Fire  inspectors,  Plastic  molding,  Asphalt, 

Marine  and  Screen  printing,  measure  ink  and  Dryer  temperature, 

HVAC/R, Diesel and Fleet maintenance. 

 

1. 

           

SAFETY   

 

 

Use extreme caution when the laser beam is turned on. 

 

Do  not  let the beam  enter  your  eye,  another  person’s eye  or the 

eye of an animal. 

 

Be  careful  no  to  let  the  beam  on  a  reflective  surface  strike  your 

eye. 

 

Do  not allow  the laser  light  beam  impinge  on  any  gas  which can 

explode. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

 

2 Distance & Spot Size 

As the distance (D) from the object increases, the spot size (S) of the 

area measured by the unit becomes larger. The relationship between 

distance and spot size for each unit is listed below. The focal point for 

each unit is 914mm (36”). The spot sizes indicate 90% encircled energy. 

 

Fig: 1

                                                                                                                                     

                                                 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

 

3.SPECIFICATIONS 

                    Temperature range                                    D: S 

 

                    -50 to 1000

@

(-58 to 1832

)                    50:1 

Display resolution                                              0.1

@

(0.1

Accuracy                                                for targets: 

Assumes    ambient operating    temperature of 23 to 25

@

(73 to 77

)       

-50 to -20

@*

-58 to -4

R+

:          

±5

@

(±9

-20 to 200

@

(-4 to 392

):          

± 1.5% of reading + 2

@

(±3.6

); 

200 to 538

@

  (392 to 1000

):     

± 2.0% of reading +2

@

(±3.6

); 

538 to 1000

@

(1000 to 1832

):     

± 3.5% of reading ± 5

@

(±9

Response time                Less than 1 second 

Spectral response            8~14um 

Emissivity                          Digitally adjustable from 0.10 to 1.0 

Over range indication      LCD will show “-0L”,”0L” 

Polarity        

            Automatic (no indication for positive polarity); 

Minus (-) sign for negative polarity 

Diode laser                       Output <1mW,Wavelength 630~670nm, 

                                             Class 2(

II)

 laser product 

Operating temp.               0 to 50

@

(32 to 122

Storage temp.                  –20 to 60

@

  (-4 to 140

Relative humidity            10%~90%RH operating, <80%RH storage 

Power supply                   9V battery, NEDA 1604A or IEC 6LR61,   

or equivalent 

Weight                              290g (10.2 oz.) 

Size                                  100 x 56 x 230mm (3.9 x 2.2 x 9.0”) 

 


background image

 

Safety                                “ CE ” Comply with EMC 

Note: 

  Accuracy: Given at 18 

 

to 28

@

  (64

 

  to 82

 

), less than 80 % RH. 

  Field of View: Make sure that the target is larger than the unit’s spot 

size. The smaller the target, the closer you should be to it. When 
accuracy is critical, make sure the target is at least twice as large as 
the spot size. 

 

4.FRONT PANEL DESCRIPTION 

p""

IR sensor                                     

q

    Laser pointer beam 

r

    LCD Display

 

 

s

    Down button 

t

    Up button 

u

    Mode button 

v

    Laser/Backlight button 

w"

Measurement Trigger 

x"

Handle Grip 

y"

Battery Cover 

 

5.INDICATOR 

p"

Data hold 

q

  Measuring indication 

r

  Emissivity symbol and value 

s

  Low  battery,  lock  and  laser  “ON” 

symbols 

t

  Backlight “ON” symbol 

u

 

@

/

  symbol 

v

  High alarm and low alarm symbol 

w

  Temperature values for the MAX, MIN, DIF, AVG HAL and LAL. 

x

  Symbols for EMS, MAX, MIN, DIF, AVG, HAL and LAL. 

 


background image

 

y

  Current temperature value 

6.BUTTONS 

p"

Up button (for EMS,HAL,LAL) 

q"

MODE button   

(for cycling through the mode loop) 

r

  Down button (for EMS,HAL,LAL) 

s

  Laser/Backlight ON/OFF button   

                    (pull trigger and press button to activate 

  laser/backlight) 

 

MODE Button Function 

The  infrared  thermometer  measures 

Maximum  (MAX),  Minimum  (MIN), 

Differential  (DIF), and  Average  (AVG) 

Temp. Each  time  you  take  a reading. 

This  data  is  stored  and  can  be 

recalled with the MODE button until a 

new measurement is taken.   

When the trigger is pulled again, the 

unit  will  begin  measuring  in  the  last 

mode selected. 

Pressing the MODE button also 

allows you to access the High 

Alarm(HAL),Low Alarm(LAL), 

Emissivity(EMS).Each time you press 

MODE ,you advance through the 

mode cycle. 

The diagram shows

 

the sequence of functions in 
the MODE cycle.   

 


background image

 

 
 

Switching   

@

@

@

@

/

, LOCK ON/OFF and SET ALARM. 

p

p

p

p

!

!!

!

@

@

@

@

/

 

q

q

q

q

!

!!

!

LOCK ON/OFF

 

r

!

SET ALARM

 

Select the temperature units (

o

C or 

o

F) 

using the 

o

C/

o

F switch 

 

To lock the unit on for continuous measurement, slide the middle switch 

LOCK ON/OFF right. If the trigger is pulled while the unit is lock on, the 

laser and backlight will turn on if they have been activated. When the unit 

is locked on, the backlight and laser will remain on unless it is turned off 

using the Laser/Backlight button on the keypad. 

To activate the alarms, please slide the bottom switch SET ALARM right. 

To set values for the High Alarm (HAL), Low Alarm (LAL) and Emissivity 

(EMS), firstly active the display by pulling the trigger or press the MODE 

button, then press the MODE button until the appropriate code appears 

in the lower left corner of the display, press the UP and DOWN buttons to 

adjust the desired values. 

 

 

 

 

 

 


background image

 

10 

 

 

 

 

7.MEASURMENT OPERATION 

p"

Hold the meter by its Handle Grip and point it toward the surface 

to be measured. 

q"

Pull  and  hold the  Trigger to turn  the meter  on and  begin testing. 

The  display  will  light  if  the  battery  is  good.  Replace  the  battery  if 

the display does not light. 

r"

While measuring, the SCAN display icon will appear ins the upper 

left hand corner of the LCD. 

s

  Release the Trigger and the HOLD display icon will appear on the 

LCD indicating that the reading is being held. 

t

  The  meter  will  automatically  power  down  after  approximately  7 

seconds after the trigger is released.(Unless the unit is locked on) 

Note: Measurement considerations 

Holding the meter by its handle, point the IR Sensor toward the object 

whose  temperature  is  to  be  measured.  The  meter  automatically 

compensates  for  temperature  deviations  from  ambient  temperature. 

Keep  in  mind  that  it  will  take  up  to  30  minutes  to  adjust  to  wide 

ambient  temperatures  are  to  be  measured  followed  by  high 

temperature measurements, some  time  (several  minutes) is required 

after  the  low  (and  before  the  high)  temperature  measurements  are 

made. 

This is a result of the cooling process, which must take place for the IR 

sensor. 

 


background image

 

11 

 

8.BATTERY REPLACEMENT 

p

  As battery power is not sufficient, LCD will display “          ”   

replacement with one new battery type 9V is required. 

q

  Open battery cover, then take out the battery from instrument and   

replace with a new 9-Volt battery and place the battery cover back.   

 

 

 

 

 

 

 

 

 

 


background image

 

12 

9. NOTES:       

  How it Works 

Infrared thermometers measure the surface temperature of an object. 

The  unit’s  optics  sense  emitted,  reflected,  and  transmitted  energy, 

which is collected and focused onto a detector. The unit’s electronics 

translate the information into a temperature reading, which is display n 

the  unit.  In  units  with  a  laser,  the  laser  is  used  for  aiming  purposes 

only. 

  Field of View 

Make  sure  that  the  target  is  larger  than  the  unit’s  spot  size.  The 

smaller  the  target,  the  closer  you  should  be  to  it.  When  accuracy  is 

critical, make sure the target is at least twice as large as the spot size. 

 

 

 

 

 

 

 

 

  Distance & Spot Size   

As the distance (D) from the object increases, the spot size (S) of the 

area measured by the unit becomes larger. See: Fig: 1

  Locating a hot Spot 

To  find  a  hot  spot  aim  the  thermometer  outside  the  area  of  interest, 

then  scan  across  with  an  up  and  down  motion  until  you  locate  hot 

spot. 

 


background image

 

13 

  Reminders 

p"

Not  recommended  for  use  in  measuring  shiny  or  polished  metal 

surfaces ( stainless steel, aluminum, etc.).See Emissivity 

q"

The  unit  cannot  measure  through  transparent  surfaces  such  as 

glass. It will measure the surface temperature of the glass instead. 

r"

Steam, dust, smoke,  etc.,  Can  prevent  accurate measurement  by 

obstructing the unit’s optics. 

  Emissivity 

Emissivity  is  a  term  used  to  describe  the  energy-emitting 

characteristics of materials. 

Most  (90%  of  typical  applications)  organic  materials  and  painted  or 

oxidized  surfaces  have  an  emissivity  of  0.95  (pre-set  in  the  unit). 

Inaccurate readings will result from measuring shiny or polished metal 

surfaces.  To  compensate,  cove  the  surface  to  be  measured  with 

masking tape or  flat black  paint. Allow  time for the tape to  reach  the 

same  temperature  as  the  material  underneath  it.  Measure  the 

temperature of the tape or painted surface. 

 

 

 

 

 

 

 

 

 

 

 


background image

 

14 

Emissivity Values   

 

Substance 

Thermal emissivity 

Substance 

Thermal emissivity 

Asphalt 

0.90 to 0.98 

Cloth (black) 

0.98 

Concrete 

0.94 

Human skin 

0.98 

Cement 

0.96 

Lather 

0.75 to 0.80 

Sand 

0.90 

Charcoal (powder) 

0.96 

Earth 

0.92 to 0.96 

Lacquer 

0.80 to 0.95 

Water 

0.92 to 0.96 

Lacquer (matt) 

0.97 

Ice 

0.96 to 0.98 

Rubber (black) 

0.94 

Snow 

0.83 

Plastic 

0.85 to 0.95 

Glass 

0.90 to 0.95 

Timber 

0.90 

Ceramic 

0.90 to 0.94 

Paper 

0.70 to 0.94 

Marble 

0.94 

Chromium oxides 

0.81 

Plaster 

0.80 to 0.90 

Copper oxides 

0.78 

Mortar 

0.89 to 0.91 

Iron oxides 

0.78 to 0.82 

Brick 

0.93 to 0.96 

Textiles 

0.90 

 

10. MAINTENANCE & CLEARING: 

  Repairs  or  serving  aren’t  covered  in  this  manual  should  only  be 

performed by qualified personnel. 

  Periodically wipe the case with a dry cloth. Don’t use abrasives or 

solvents on this instrument. 

  When serving, use only specified replacement parts.