Full Text Searchable PDF User Manual

UPHO
ULTIMATE HOMOGENIZER
USER MANUAL

Table of content
1. Precautions
2. Safety Information
- Proper Use
- Electrical Safety
- Environment
- Biological Safety
- Personal Protective Equipment (PPE)
3. Introduction
4. Main Technical Parameters
- Structure Overview
5. Installation Procedures
- Unpacking and Packing List
- Removal of the Transport Security System
- Sample Adapter Installation
6. Operating Procedures
-
Automatic Program
- User-defined Program
7. Cleaning & User Maintenance
8. Troubleshooting
9. Warranty Statement
10. Disposal
- End of Life Disposal
11. Relevant Standards & Regulatory Requirements
- EC Declaration of Conformity
- Regulations
12. UPHO Accessories
1
1
1
3
3
3
4
4
6
7
8
8
9
10
11
12
13
15
16
17
17
17
17
18
19
20

1. Precautions
Please read, understand and practice the following precautions when
operating instructions your UPHO.
Know the limitations and hazards associated with the use of this
equipment.
• Do not operate this equipment in an environment where other
unshielded devices are being used, that radiate electromagnetic radiation.
• Do not use sharp objects such as a pencil point or ballpoint pen to
intreact with the touch screen, as damage may result. Please use a suitable
soft point stylus or your finger tip.
• This equipment should only be operated in temperatures between 10º C
to 40º C, transported and stored in temperatures between 0º C to 50º C,
with relative humidity less than 70%.
• Do not disassemble, modify or remodel the unit or accessories.
• This device should be kept out of the reach of children.
2. Safety Information
Before using UPHO, it is essential you read this user manual carefully and
pay particular attention to the safety information.
The instructions and safety information in the user manual must be
followed at all times to ensure a safe operation and to maintain UPHO in a
safe operating condition.
The advice given in this manual is intended to supplement, not supersede,
the normal safety requirements prevailing in the user’s country.
Proper Use
This product is intended to be used for research purposes only.
1

2
Other applications are not covered by Geneye warranty. The product is not
suitable for use in diagnostic or therapeutic applications.
Improper use of UPHO may cause personal injuries or damage to the
instrument. Geneye shall not be held responsible for any injuries or
damages resulting from non-observance of the operating instructions and
safety information in this manual.
If any abnormality, such as excessive noise, excessive vibration or
over-heating is observe during operation, please immediately shut off the
power at the main switch (rear of instrument) or at the electrical power
socket and call for professional service.
• Do not attempt to move UPHO during operation.
• UPHO should only be operated by suitably qualified and trained
personnel in accordance with Good Laboratory Practice.
• Do not make any modifications of the instrument.
• Disassembly will void the warranty.
• UPHO should only be used in combination with Geneye accessories, such
as sample adapters and spare parts.
• Do not continue to use UPHO or UPHO accessory if damage is occurred.
• Damage or personal injury can occur if UPHO sample adapter set is not
securely installed before operation (see User Manual Section 5.3).
• The use of sample adapters other than the Geneye sample adapters
described in this user manual may lead to the UPHO being damaged.
• Avoid spilling water or chemicals onto or into UPHO. Damage caused by
water or chemical spillage will void the warranty.
• Do not leave UPHO unattended during operation.
• The machine must be placed on a flat laboratory bench (preferably solid

wood or stone countertops) to minimise vibration during operation.
• The machine is designed for frequent use with short run times (typical
run time is 1 minute). Do not use long run times (>5 minutes) when
operating UPHO.
• The UPHO accessories (e.g. adapters, cans, beads) may become worn,
depending on the frequency of use and the material to be homogenized.
They should be regularly checked for wear and tear and replaced if
necessary.
Electrical Safety
To ensure satisfactory and safe operation of your UPHO, please follow the
safety instructions below at all times:
• Do not attempt to adjust or replace any internal parts of your UPHO - if
there is a fault, please call for professional service.
• When UPHO is not in use, switch it off at the power switch, on the back
panel of the unit (please see picture on page 7).
• If UPHO will not be out-of-use for a long period of time, switch off the
power switch, on the back panel of the unit and unplug the power cord
from the power outlet.
• To protect the internal circuits and parts from damage, please do not use
running water to clean your UPHO: use a soft damp cloth.
Environment
• Excessive exposure to direct sunlight may bleach exposed parts of the
instrument and eventually cause damage to plastic parts. UPHO should be
located out of direct sunlight.
Biological Safety
• Samples may contain infectious agents. You should be aware of the
3

health hazard presented by such agents and should use, store, and dispose
of such samples according to the required safety regulations.
• The laboratory manager must take the necessary precautions to ensure
that the surrounding workplace is safe, and that UPHO operators are
suitably trained and not exposed to hazardous levels of infectious agents.
• When working with volatile solvents, toxic substances, etc., you must
provide an efficient laboratory ventilation system to remove steams that
may be produced.
Personal Protective Equipment (PPE)
• Operators must wear the appropriate level of personal protective
equipment as required for the particular laboratory environment in which
the System will be used. This may include: laboratory coat, gloves, safety
glasses and safety shoes.
• Operators must not operate the equipment if feeling unwell.
3. Introduction
The traditional sample grinding with a Pestle & Mortar is a very slow
method, that can only process one single sample at a time and has a high
risk of causing cross-contamination. Besides that, after such a procedure, it
is necessary to clean and disinfect the equipment. Efficient sample
disruption/grinding is a prerequisite for nucleic acid, protein and other
biomolecules extraction and purification procedures.
Incomplete sample disruption can lead to significantly reduced recovery
yields of the molecules of interest and can increase the risk of clogging
when using purification columns or magnetic particles later in the
workflow.
UPHO is a versatile bead mill device designed for rapid, efficient and
simultaneous disruption of up to 64 samples. A wide range of sample
adapters are available for UPHO providing sample flexibility and leading to
reliable and quality results.
4

This bench device is one of the most advanced sample preparation
systems available today. UPHO was specifically developed for routine as
well as difficult and resistant samples, compatible with 3 types of grinding
and lyses thoroughly and quickly a wide variety of biological samples
including plant and animal tissues, cells, bacteria and yeast.
Each sample is simultaneously disrupted and homogenized by high-speed
shaking with a bead/beads in a sealed tube (bead milling), allowing an easy
and reproducible isolation of stable RNA, active proteins and full-length
genomic DNA.
The ergonomic design of the instrument ensures an easy loading of the
disposable centrifuge sample tubes that remain securely sealed during the
grinding process, effectively avoiding sample cross-contamination.
The homogenization speed (frequency) and duration run-times are
digitally controlled via UPHO’s intuitive touch screen, improving the
consistency for multi sample lots.
The different beads sizes (0,1 – 15 mm) and bead materials (wolfram steel,
metal, glass and ceramic) makes the grinding process more complete and
efficient with UPHO when compared to other available methods.
Moreover, thanks to UPHO’s design and options for the use of cryogenic
liquids to freeze samples, the sample temperature remains constant
throughout the homogenization process.
5

4. Main Technical Parameters
• Run-time setting: 1 (sec) - 9999 (min)
• Oscillation frequency setting: 1 – 70 Hz
• Acceleration: <2 seconds to reach maximum speed
• Power rating: 195 W
• Power requirement: 220 V - 240 V~, 50 Hz, 2.5 A, 375 W
• UPHO dimensions: � 460 × 460 × 530 mm (length × width × height)
• Weight: 35 kg
• Operating temperature: 10
ºC
~ 40
ºC
(relative humidity: �70%)
• Storage temperature: -20
ºC
~ 50
ºC
(relative humidity: �93%)
• Maximum noise emission: 65 dB
• Interface: Touch screen
6

Structure Overview - Front
Protective cover
Security screw
Adaptor retainer
Display screen/ Operation panel
Cover latch
Structure Overview - Back
ON/OFF button
Power supply plug
Foam
7

5. Installation Procedures
Unpacking and Packing List
• Carefully remove the UPHO mainframe, accessories and documentation
from the wooden shipping crate.
• Check that you have received all the item in the quantities on the list
below. If anything is missing, please contact your supplier. You will also find
detailed information on the contents of the package on the packing list in
the shipping crate.
UPHO mainframe
User and Cell Disruption Manuals
Fuses
Sample adaptor security screws
(NB. one screw may be in the mainframe)
Power cord
Quality control report
Warranty card
1
1
2
2
1
1
1
Quantity Name
8

• Inspect the unit for any damage that may have occurred during shipment.
If the equipment is damaged, report it to the carrier and contact your
supplier. Save all packaging material in the event a return is necessary.
• UPHO comes fully assembled, requiring a minimum set-up. Please install
the equipment on a clean, dry-stable surface far away from any compatible
electrical outlet.
• Allow some space around UPHO for proper ventilation. This unit is "FOR
INDOOR USE ONLY". Avoid operating in areas of excessive humidity or
extremes of temperature.
Removal of the Transport Security System
• A transport security system is located on the underside of UPHO:
comprised of 4 bolts (circled on picture below).
THIS SYSTEM MUST BE
REMOVED BEFORE OPERATING THE UPHO.
• Remove the 4 transport bolts from the UPHO by
TILTING THE UPHO
SLIGHTLY TO EXPOSE THE TRANSPORT BOLTS
(as seen in the picture) and
using a 13 mm x 16 mm metal wrench.
DO NOT TURN THE UPHO ON ITS
SIDE - THIS MAY LEAD TO INTERNAL DAMAGE.
• Please retain and carefully store the 4 transport bolts, as they will be
needed to reinstall if UPHO needs to be transported in the future.
• Operating UPHO with transport security system installed, or transporting
UPHO without the transport security device installed, can lead to damage
to the mechanical components of the instrument.
9

Sample Adapter Installation
• Make sure UPHO is connected to a suitable power outlet.
• Load the filled (sample, buffer and beads) and capped sample tubes into
the cavities of the sample adapter, so that they fit in the holes. It is preferred
to position the tubes symmetrically in the sample adaptor to reduce
vibration.
• When all sample tubes are loaded, place the sample cover on the sample
adaptor.
• Open the protective cover on the sample compartment cover by
releasing the cover latch, remove the security screw if present, and slide the
sample adapter (and cover) on to the adaptor retainer ensuring that the
sample adaptor fits over the rectangular metal locator on the adaptor
retainer.
• Thread the security screw on to the adaptor retainer tighten down until
the sample adaptor and cover are securely in position –
Security screw should by finger tight and not over tightened. Please note
that the sample adapter body may wear with prolonged use.
10

• Close protective cover on the sample compartment and secure with the
cover latch.
• Press PLAY button (on the touch-screen) to start running the selected
program.
• The program will not start if the protective cover is open or the cover
latch is not secured.
For hard samples, you can use the following adaptors and cans:
6. Operating Procedures
Suggestions and recommendations with the most common grinding
parameters (time, frequency, beads type and size) are listed on the “Cell
Disruption Guide”, provided with UPHO.
11

Automatic Program
Main Menu
When you power on UPHO (in the back of the device), the main menu
window will appear in the screen, together with a green light.
By touching the screen, you can execute several operations:
• Decelerate button: to decrease grinding speed
• Accelerate button: to increase grinding speed
• Play: to start the chosen program
• Stop: to stop the program
• Pause: to pause and restart a program
12

User-defined Program
UPHO allows the user to define a program. Please follow these steps:
1) On the main menu, touch the “SETUP” button. This window will appear;
2) On the “Choose” column, there are 10 boxes, each one of them
corresponding to a certain program defined by the user. When the user
touches a box, it will turn yellow.
3) After that, please touch the “Modify” button, at the bottom of the screen.
The “Modify” button will turn yellow as well, after you press it.
4) This will now allow the user to set the different parameters: Running
Time, Grinding Cycle State and Frequency.
Frequency
M - Minutes
S - Seconds
Number of grinding cycles during
running time. Recommended to always
set as “100”
13

5) To define a program name, touch the box on the “Group” column, in the
same line as the yellow “Choose” box.
6) After that, a pop-window will appear as follows:
7) Write the name of the program by touching the different keys. Click on
ENT button to set.
14

8) By touching the other boxes, the user can define other parameters such
as “Running Time - Minutes and Seconds”, “Grinding Cycle State” and
“Frequency”.
9) Once the user has defined the program, click the “Apply” button at the
bottom of the screen to save it.
10)Then, click on “Close”, the main display will appear. Click on PLAY button
to start the grinding process.
11)The user can always modify later its own programs, by choosing a
certain program on the Group column.
12)After a long period of time non-use time, the main menu will turn black.
7. Cleaning Maintenance
• Before cleaning your UPHO, be sure to switch the UPHO off at the main
power switch (back of the instrument) and disconnect your UPHO from the
power outlet.
• If solvents or saline, acidic, or alkaline solutions are spilt on UPHO wipe
them away immediately.
Run/S
Time
Frequency
15

• Do not use spray cleaning or disinfectant liquids onto surfaces of your
UPHO - use a soft cloth dampened with cleaner and / or disinfectant
• Do not autoclave any part of UPHO (except beads)
• The different adapters should be cleaned after use using mild detergents
or 70% ethanol, rinsed with distilled water, and then dried using paper
towels.
UPHO is supplied to you with a two-year Manufacturer’s warranty which
runs from the delivery date.
8. Troubleshooting
16
Common problem
No power
No display on the
screen
Touch screen shows
incomplete or ununsual
characteres
Possible cause
Action(s)
CPU needs to be reset
Turn off the UPHO using the Main Power Switch
located on the back of the instrument
Wait 30 seconds
Turn the UPHO back on
Check main voltage
Check that the voltage of the unit matches
that being delivered by the main power supply
Make sure the unit is plugged in properly
Check and replace fuses if blown
Change switch
On/Off switch broken
Faulty fuse
Others
Change fuse
Change switch
Contact supplier
Instrument does not run
Motor temperature above 65ºC
Wait for temperature to drop below 60ºC
Incorrect voltage selected
Power off the unit and select appropriate voltage.
If problem persists, contact supplier
Faulty Motor or Controller
Abnormal running noise
Security Screw not locked in place
Faulty bearing
Contact supplier
Secure Security Screw in place and lock down properly
Tube leakage
Speed is too high
High temperature or pressure
inside sample tubes
Reduce processing time or speed
Touch screen keys do not work
Plug connector loose
Faulty key
Contact supplier
Contact supplier

9. Warranty Statement
The Manufacturer’s warranty supplied with this UPHO does not cover wear
from normal use; it does not apply if the UPHO or parts of the UPHO have
been altered, repaired or replaced by anyone other than an employee of
Geneye or Geneye’s authorised Service Partners.
Warranty does not apply to any instrument which has been damaged
through accident, negligence, failure to follow operating instructions, or
from the use of electric currents or circuits other than those specified on
the plate affixed to the instrument, misuse, or abuse.
Geneye reserves the right to change, alter, modify, or improve any of its
instruments without any obligation whatever and to make corresponding
changes to any instrument previously sold or shipped.
10. Disposal
End of Life Disposal
Reuse, recycling and recovery. The presence of this label on a product
means that the product contains electrical or electronic materials and
therefore must not be disposed of as unsorted waste but instead is treated
separately. The presence of these materials may, if not disposed of properly,
have potential adverse effects on the environment and human health.
Within the European Union, operators are urged to recycle such products
when being replacedwith a newer version or when they have outlived their
useful lives.
11. Relevant Standards & Regulatory Requirements
17

11. CE CERTIFICATION
18
EC DECLARATION OF CONFORMITY
The UPHO flexible bead mill device, a product of Geneye, is designed for rapid and efficient
disruption of up to 64 samples simultaneously (Automated Samples Grinder)
The UPHO device is in conformity with the requirements of the following European
Directives (and their associated amendments):
The tests on the above designated product complies with the requirements of Council
Directive 2014/35/EU relating to Electrical Equipment designed for use within certain
voltage limits.
Warning: This is a class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate me asures.
Signed:
Date : January 15th, 2018
Mr. B. Coissac
Title : Managing Director
For and on behalf of: Biowire Ltd Suite D 6/F, Ho Lee Commercial Building
38-44 D’Aguilar Street, Central,
Hong Kong
Product Description:
UPHO device
Product Part Number:
GY-U001
Machinery Directive
2006/42/EC
Electromagnetic Compatibility Directive (EMC)
2014/30/EU
Electrical Safety Low voltage Directive (LVD)
2014/35/EU
Harmonized standards
EN ISO 12100:2010,EN 61010-1:2010,
EN 61326-1:2013,EN 61000-3-2:2014,
EN 61000-3-3:2013
Restrictions on the use of Certain Hazardous
Substances (RoHS)
2011/65/EU
Registration, Evaluation, Authorization and
Restruction of chemicals (REACH)
1907/2006

Regulations
UPHO also complies with other applicable European Directives and
International Standards, particularly:
• The Waste Electrical and Electronic Equipment (WEEE) Directive and the
Reduction of Hazardous Substances (RoHS) Directive
• EN 61010-1: ‘Safety requirements for electrical equipment for
measurement, control and laboratory use. General requirements’
19

12. UPHO Accessories
20

Thank you for
choosing
UPHO
!
We are sure it will
become an integral
part of your laboratory!
Visit us on the Web at www.geneye.com
E-mail us at info@geneye.com
Because of technological developments related to the products, systems, and/or services
described herein, the data and procedures are subject to change without notice.
Please consult your Geneye representative or visit www.geneye.com to verify that this
information remains valid.
© 2017, Biowire Ltd.
IFUGYUPHOa January 2018