PDF User Manual

  1. Home
  2. Manuals
  3. Telaire 7001 Manual

Telaire 7001 Manual

Made by: Telaire
Type: Manual
Category: Measuring Instruments
Pages: 2
Size: 0.19 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

Start up

Battery Operation

-

 For portable use, the monitor will operate up to 70

hours on 4 AA industrial alkaline batteries. The batteries can be installed by
removing the battery cover and following the diagram imprinted on the back
plastic cover.

LOW BATT

 flashes when there is less than 30 minutes of battery life. At

this point the batteries should be replaced or the AC adapter should be
used as a substitute. If operation continues, the unit will become
inoperable and only the 

LOW BATT

 will be blink on the LCD display.

AC Power Operation

 -

 The sensor is shipped with a 6V DC

500mA  AC/DC adapter. To use the adapter, connect the plug into the
back of the unit and plug the transformer into any standard wall outlet.

NOTE 

Use the supplied adapter. The wrong adapter may cause damage

to the unit. Battery operation will not function as a back up during a
power loss.

Power-Up

1. Press the Power Button and a 2 second delay will occur before the

display becomes visible.

2. 10 seconds will elapse before displaying current CO

2

 readings

3.

WARM-UP

 will display for approximately 1 minute. During this time,

adjustments can not be made to the sensor.

Display Features and Modes

WARM-UP 

- indicates a 1 minute warm-up

ON LINE - 

Displays when a PC is communicating to the sensor via RJ45 port.

Normal Operating Mode 

- After warm-up the sensor will stabilize and

display current conditions.

Adjustment Modes

Pressing the 

mode 

button scrolls through the adjustment modes. Once

the desired mode is displayed, press Enter to make adjustments. Press
Enter again to save and leave the adjustment mode.

ELEVATION - 

Used to compensate for elevation changes.

CALIBRATION - 

Used when calibrating.

TEMPERATURE

 - For Temperature calibration

OUTSIDE

 - To Manually input CO

2

 levels for the CFM ventilation rate

CALIBRATION IN PROGRESS 

- Displays during calibration

Display Features

CO

2

 Readings

 (Upper Display) - Remain visible at all times.

Temperature and Ventilation Rates 

(Lower Display)

The Up/Down arrows allow you to toggle through the Temperature and
Ventilation modes. When pressing the up arrow, the display will go
through the following sequence:

Temp°C

 > 

Temp°F

 > 

Vent Rate l/p/s

 > 

Vent Rate cfm/p

 > 

Blank

US Standard to Metric Conversion

The Temperature, Ventilation rates and Elevation Readings can be
viewed in US Standard or Metric readings. The Temperature is converted
from Fahrenheit (°F) to Celsius (°C), the Ventilation Rates are converted
from Cubic Feet Per Minute Per Person (cfm/p) to Liters Per Second (l/p/
s) and the Feet (ft) to Meters (m).

Operation

Elevation Correction - 

The

sensor, like any other gas
measuring device is affected
by altitude changes. The
sensor is shipped with the
elevation setting set at “zero”
or sea level. If you are at an
altitude greater than 500 feet,
an adjustment should be made
to assure maximum sensor
accuracy. Once the elevation
correction is set it will be
stored and saved in the
monitor memory. To change
the default setting follow the
steps below.
1. Press the 

Mode

 button until

ELEVATION

 begins to blink.

2. Press 

Enter

.

3. Press 

Mode

 to toggle the

elevation reading between
feet(ft) & meters(m).

4. Use the 

Up/Down

 button to

adjust the altitude in increments of 500(ft) or 100(m). Once the correct
altitude is set, press 

Enter

 to save the setting and return to normal mode.

Temperature or CFM/Person –

 The lower display will cycle through the

following units when the 

Up/Down

 button is pressed: Temp°F, Temp°C,

CFM, turn lower display off.

Stand-alone Monitoring -

 Once Elevation correction has been made (as

described in the steps above), the sensor will begin to accurately display
current room conditions.

Using an External Datalogger

 - Voltage outputs for both CO

2

 and

temperature are available via an RJ-45 jack on the rear of the unit.

The Telaire

®

 7001 is an easy-to-use hand-held CO

2

/Temperature

monitor. The unit provides stable, highly accurate readings due to
our patented dual beam NDIR technology. Equipped with a 0-4V
output, the unit is perfect for long-term monitoring/recording.

Ventilation Rate (cfm/p)

Overview

 - Press the 

up/down 

button until cfm/p is displayed. This value

represents how much outside air is being introduced on a CFM per
person basis. The reading is derived from calculating the outside air
ventilation rate to a space based on the inside/outside CO

2

 differential

readings. Current codes/standards generally require 15 to 20 CFM/
Person to be delivered to most spaces to ensure acceptable air quality.
Low values indicate low ventilation rates and potentially poor air quality.
High levels can indicate excessive ventilation and potential excessive
energy usage. To obtain accurate measurements, readings should be
taken 2 to 3 hours after occupancy has stabilized in a space or at a peak
in daily CO

2

 concentrations.

Adjusting the Outside CO

2

 Concentration

The sensor is factory set to assume an outside level of 400 ppm, which
should be close to the outside concentration in most areas. The outside
level of CO

2

 can also be changed by measuring outside levels or by

manually adjusting the monitor.
1. Power up outside and wait for CO

2

 levels to stabilize (approx 5 min).

2. Save the reading by holding the 

Enter

 button on the sensor for 5 secs.

3. This value is used to calculate the ventilation rate based on the

differential of the measured outside value and into measured inside
concentration.

Manual Input of Outside Concentration

The monitor is factory set at 400 PPM. To adjust the factory setting (for the
cfm/person calculation) or to verify the current setting follow steps below:
1. Press the mode button until 

CO

2

 and 

ppm

 flash.

2. Press the enter button and the current outside value will display.
3. Use the 

Up/Down

 button to increase/decrease the CO

2

 value.

4. Press enter to save and store the value in the monitor.

Calibration (CO2)

IMPORTANT - 

The sensor has been factory calibrated and should need

calibration once every 12 months using either a zero concentration gas or a
gas with a specified concentration of CO

2

. For the most accurate field

calibration we recommend purchasing a Telaire calibration kit (P/N 2075).
Sensor drift usually occurs at the zero reference point. The manual calibra-
tion process allows the user to perform a one point calibration based on
ambient levels or by flowing a gas of a known concentration  through the
sensor. This process will adjust the zero offset of the sensor and will provide
an accurate calibration. If a two point calibration is required the Calibration
kit should be used. For manual calibration, follow the steps below.
1. The calibration procedure will last approximately 15 minutes. Before

performing the calibration procedure, remove the battery cover to

9

8

7

10

6

5

5

4

1

2

3

Features

1. Display

6. Adapter Connection

2. Enter Button

7. RJ-45 Connection

3. Power Button

8. Cal Button (under cover)

4. Up/Down Button

9. Gas Connection

5. Batter Compartment

10. Kickstand

Enter

Button

Power

Button

Mode

Button

Up/Down

Button

This drawing is for reference to identify the
location of the different modes and adjustment
features. The display won’t display all features
simultaneously (as shown).

Telaire

®

 7001

Carbon Dioxide and Temperature Monitor

6860 Cortona Drive, Suite B - Goleta, California, 93117

P

- 805-685-4000 

F

- 805-685-0015

62292-006 05/12/2004

 


background image

access the calibration activation switch. Connect the supplied AC adapter
to the back of the sensor or ensure new batteries are installed.

2. Power up the sensor and wait for the Warm-up to end.
3. Next verify the Elevation correction has been set. Refer to the steps in

Elevation Correction for procedure.

4. If you are calibrating to ambient conditions make sure the sensor is

displaying a stable reading - avoid breathing in the area of the monitor.
If you are flowing gas to the calibration port of the sensor, allow the
gas to flow for at least 10 minutes before initiating calibration.

5. Press the 

Mode

 button twice. The Calibration mode will begin blinking.

5. Press 

Enter

.

6. Use the 

Up/Down

 button to adjust reading to the current ambient

conditions or concentration of the gas being used. Pressing the button
once will change the readings in increments of 10 ppm. To increase
the speed, press and hold the button.

NOTE

 - For best accuracy, a reference or known concentration of CO

2

should be used when adjusting the reading. Bottled nitrogen can be used
to provide a zero concentration gas.
7. Next, on the backside of the unit locate the push button switch (under

the battery cover, in the small round hole to the right of the connector
jack), use a small pointed object to depress and hold the switch for 5
seconds. 

CALIBRATION

 will begin to blink.

8. Press Enter.
9.

Calibration In Progress

 will begin to blink. At this point the unit will

program itself based on the CO

2

 value that was input in Step 6. The

calibration process will take approximately 5 minutes.

When Calibration is complete, the display will return to the steady

Calibration 

mode. Press 

Enter

 to return to the normal operation mode.

Temperature Adjustment

Use this procedure to adjust the temperature output when you wish to
have the temperature output match a reference sensor. The accuracy of
a field adjustment is dependent upon the stability of the environment is
which the procedure is performed, and upon the accuracy of the refer-
ence sensor.
1. Connect the supplied AC adapter to the back of the sensor. If you do

not have the AC adapter, new batteries should be used.

2. Power up the sensor and place in a stable environment, free of drafts

or temperature changes. Wait 30 minutes for the unit to fully equili-
brate with the environment. Do not hold the unit in your hand during
this period. Press the mode button until the blinking word 

TEMPERA-

TURE

 appears.

4. Press 

Enter

. Both the word 

TEMPERATURE 

and the numeric

temperature display will begin blinking in unison.

5. Use the 

Up/Down

 button to adjust the temperature reading to match

the reference.

6. Press 

Enter

. The temperature offset is immediately adjusted, the

blinking stops, and the unit is now in normal operating mode.

Accessories

CO

2

View™  Real-Time Graphing/Calibration Software 2080

Allows the 7001 to log directly to a Windows™ compatible PC and graph
concentrations in real time (includes RS232 cable). Includes an interface
that allows the user to perform a zero and span calibration on the sensor.

Hobo® Data-Logger Kit CO2 – Temp & RH 2077

Small data-logger that mounts to the back of the 7001. Records CO

2

,

Temp, and Relative Humidity. Stores over 7900 data points and samples
at a user adjustable time interval. Graphing software included.

2070 Datalogging Cable

Connects the 7001 to 3rd party devices. Includes two leads for the CO

2

and temperature ouputs.

Calibration Kit 2075

Comes with necessary equipment to perform zero calibration.

Specifications

Method

Dual Beam Absorption Infrared™

Display - LCD

Independent CO2 and Temperature readings.
Calculates and Displays Ventilation Rates

Sample Method

Diffusion or flow through (50 - 100 ml/min)

Performance - CO2 Channel

Measurement Range

0-10,000 ppm display
0-4,000 ppm voltage output

Display Resolution

± 1 ppm

Accuracy

±50 ppm or ±5% of reading up to 5,000 PPM (Above 5,000 PPM not
specified)

Repeatability

±20 ppm

Temperature Dependence

±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, refer-
enced to 25°C

Pressure Dependence:

0.13% of reading per mm Hg (Corrected via user input for elevation)

Response Time

<60 seconds for 90% of step change

Warm-Up Time

<60 seconds at 22°C

Calibration Interval

12 months, offset adjustment using single gas at  0-1000 ppm CO

2

. Full

factory calibration available

Performance - Temperature Channel

Temperature Range

Voltage output

32 to 104°F (0 to 40°C)

Display

32 to 122°F (0 to 50°C)

Display Resolution

0.1°F (0.1°C)

Display Options

°F, °C, or Off. Set with panel button.

Accuracy

±2°F (±1°C)

Response Time

20-30 minutes (case must equilibrate with environment)

Calibration Interval

12 months, offset adjustment using temperature standard at 50 to 86°F
(10 to 30°C). Full factory calibration available

Outputs

(Analog) CO

2

0-4 VDC, 1mV/ppm (4,000 ppm max)

Temperature

0-4 VDC linear, 32-104°F (0-40°C)

Output Impedance

100 Ohms

(Digital)

RS232 for use with Telaire® CO2View™ Graphing Software

(Wiring Connection)

One RJ-45 Connector
Dual Analog output plus digital output

Power Supply

Battery Type

Four AA batteries, not included

Battery Operation

80 hours (alkaline)

External

6 VDC from external AC/DC adapter, included

Power Requirements

100 mA Peak, 20 mA average from 6V

General

Operating Conditions

32-122°F (0-50°C)
0-95% RH, non-condensing

Storage Temperatures

-4 to 140°F (-20 to 60°C)

Certifications

FCC Class 15 Part B
CE

Warranty

18 months parts and labor

This product is covered by one or more of the following patents:
5,650,624 / 5,721,430 / 5,444,249 / 5,747,808 / 5,834,777 / 5,163,332 /

5,340,986 / 5,502,308 / 6,344,798 / 6,023,069 / 5,370,114 / 5,601,079 /

5,691,704 / 5,767,776 / 5,966,077 / 6,107,925 / 5,798,700 / 5,945,924 /
5,592,147 / 6,255,653 / 6,250,133 / 6,285,290

Warranty

This Telaire product has been examined and tested for proper operation. Please
operate this product only in accordance with the instructions.

Telaire warrants this product against defects in workmanship and materials for a
period of 18 months from the date of purchase by the original owner. If the product
should become defective within this warranty period we will repair or exchange it.
A return authorization number must be obtained from the factory prior to returning
equipment. Items received without a return authorization number will be refused.
Product to be serviced under this warranty should be sent to Telaire, 6860 Cortona Dr,
Suite B, Goleta, CA 93117. Shipment must be prepaid, properly packed and insured.

Telaire is not and will not be liable for any consequential loss or damages that
may occur by reason of purchase and use of this product. The responsibility of
Telaire, in any event, is strictly limited to the replacement/repair of the product.

Telaire seeks to present reliable information concerning the composition,
properties and use of its products, however; (1) All advice concerning selection
and use of any product is provided at no charge and with no warranty. (2) No
warranty is made hereby. Products described herein are warranted to conform to
Telaire specifications only at the time of sale. All sales are subject to Telaire
standard terms and conditions, which are reproduced on the reverse side of each
invoice. All warranties of merchantability and fitness of purpose are disclaimed
and remedy for any breach of warranty is limited to replacement of the defective
product. (3) Telaire assumes no responsibility for any patent liability arising from
the use of any product in a process, manner or formula not designed by Telaire.

P/N 62292-006

05/12/2004