Full Text Searchable PDF User Manual

Operators Manual
VM700A
Probe Adapter
063-0927-00
Please check for change information at the rear
of this manual.
First Printing August 1991
Reprint September 1993
Advanced Test Equipment Rentals
www.atecorp.com 800-404-ATEC (2832)
®
Established 1981

1
Copyright
E
1991 Tektronix, Inc. All rights reserved. Printed in the
United States of America. Contents of this publication may not be repro-
duced in any form without the written permission of Tektronix, Inc. Speci-
fication and price change privileges are reserved.
This instrument, in whole or in part, may be protected by one or more U.S.
patents or foreign patents or patent applications.
TEKTRONIX, TEK, PLOT 10, TEKTEST, and SCOPE-MOBILE are reg-
istered trademarks.
For further information, contact Tektronix, Inc., Corporate Offices, P. O.
Box 1000, Wilsonville, OR 97070–1000. Phone: (503) 627–7111; TLX:
192825; TWX: (910) 467–8708; Cable: TEKWSGT.

WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period
of one (1) year from the date of shipment. If any such product proves defective during this warranty period,
Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will
provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expi-
ration of the warranty period and make suitable arrangements for the performance of service. Customer
shall be responsible for packaging and shipping the defective product to the service center designated by
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if
the shipment is to a location within the country in which the Tektronix service center is located. Customer
shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products re-
turned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inade-
quate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to
repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair
or service the product; b) to repair damage resulting from improper use or connection to incompatible
equipment; or c) to service a product that has been modified or integrated with other products when the
effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU
OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VEN-
DORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DE-
FECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUS-
TOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT
BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAM-
AGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NO-
TICE OF THE POSSIBILITY OF SUCH DAMAGES.


Probe Adapter
1
PROBE ADAPTER
(067–1429–00 to 067–1429–05)
PURPOSE
The Tektronix Probe Adapter accessory is designed to interface a Tektronix P6109 10X probe to the
VM700A Video Measurement Set. This accessory is intended for use in manufacturing and repair
applications when internal signals of an device under test are accessible only using a high-impedance
probe. The Probe Adapter provides the impedance matching and signal gain needed to drive the 75
Ω
inputs of the VM700A.
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎ
ÎÎÎ
ÎÎ
ÎÎ
Î
Î
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎÎ
ÎÎÎ
ÎÎÎ
ÎÎÎ
ÎÎÎ
Î
Î
Front Panel
Rear Panel
ÎÎÎÎÎ
ÎÎÎÎÎ
ÎÎÎÎÎ
ÎÎÎÎÎ
Î
Î
Î
Î
Î
Î
Î
Î
Î
Î
Î
Î
ÎÎ
ÎÎ
067-1429-00
PROBE ADAPTER
INPUT
(10K PROBE)
1M
W
, 20 pF
42 V PEAK MAX
LINE VOLTAGE
SET FOR 90–132V
OUTPUT
(75 OHM)
10 VA MAX
49–66 Hz
Figure 1. The VM700A Probe Adapter.
APPLICATION
When the Probe Adapter accessory is used, accurate measurement results are obtained by using the
relative-to-reference feature of the VM700A’s Measure Mode. The relative-to-reference feature
subtracts out distortions introduced by the Probe Adapter, thereby providing measurement results
that represent distortions introduced by the device under test. This feature removes the effects of the
0.3 degree differential phase and 0.3% differential gain that are inherent in the Probe Adapter, as well
as any frequency response flatness errors.

Operating Information
Probe Adapter
2
Frequency response flatness from 10 Hz to 100 kHz is strongly dependent on the accuracy of the probe
compensation. It is recommended that each probe be compensated to match the individual Probe
Adapter that it will be used with. This adjustment should be made by observing the effect on the
mid-field transition of a field square-wave signal.
Flatness in the 100 kHz to 10 MHz range varies a great deal between probe types. The Probe Adapter
circuitry is optimized for the Tektronix P6109 10X probe, a common low-cost probe. Other 10X
oscilloscope probes capable of driving a 1 M
Ω
, 20 pF input may be used, but they can degrade the
flatness of the frequency response.
Two additional effects of making measurements through the Probe Adapter should be noted. First, the
video signal-to-noise ratio is degraded from the VM700A’s typical 82 dB to approximately 73 dB with
the Probe Adapter. Secondly, the Probe Adapter introduces some output offset voltage and some minor
tilt effects due to ac coupling. Both of these effects are normally eliminated by the clamping
capabilities of the VM700A.
The 1 M
Ω
10X probe connector provides a high impedance signal input that produces minimal circuit
loading for the device under test. The output of the probe adapter is at 75
Ω
to match the input of the
VM700A. Use high quality 75
Ω
BNC coaxial cable to connect to the VM700A input, and terminate the
signal line with a 75
Ω
termination at the last device in the signal path as shown in Figure 2.
ÎÎ
ÎÎ
Î
ÎÎ
ÎÎ
ÎÎ
ÎÎ
ÎÎ
75
W
BNC coaxial cable
75
W
Termination
P6109 10X Probe
To device under test
VM700A
Figure 2. Typical hookup for the Probe Adapter.

Probe Adapter
Operating Information
3
SPECIFICATIONS
The specifications given in Table 1 for gain, frequency response, and tilt are with a P6109 probe that
has been properly compensated for the Probe Adapter with which it is to be used.
Table 1
Specifications
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Characteristic
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Specification
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Supplemental Information
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Input Resistance
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
1 M
Ω
±
0.5%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Input Capacitance
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
20 pF
±
2 pF
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Use a compensated probe
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Gain
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
X1
±
2%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
From 10X probe tip, into 75
Ω
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Frequency Response
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
10 Hz to 100 kHz
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
±
1%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
100 kHz to 10 MHz
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
±
1%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Differential Gain
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
≤
0.3%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Measured with 1910 or VM700A
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Differential Phase
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
≤
0.3
°
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Measured with 1910 or VM700A
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Output Offset Voltage
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
≤
±
50 mV
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
≤
±
10 mV at 25
°
C
ÁÁÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁÁÁ
Video SNR
ÁÁÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁÁÁ
≥
70 dB rms
ÁÁÁÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁÁÁÁ
VM700A Noise Spectrum, un-
weighted
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Tilt (ac coupled)
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
≤
1% on a 50 Hz square wave
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
≤
5% when driven from 75
Ω
ÁÁÁÁÁÁÁÁÁÁÁ
Output Impedance
ÁÁÁÁÁÁÁÁÁÁÁ
75
Ω
±
1%
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Line Voltage Range
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
90 to 132 Vac or 180 to 250 Vac
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Internally set by a service person
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Line Frequency Range
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
48 to 62 Hz
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Power Consumption
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
≤
2W
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ

Operating Information
Probe Adapter
4
ELECTRICAL PARTS LIST
Table 2
Electrical Parts Lists
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Circuit Reference
ÁÁÁÁÁ
ÁÁÁÁÁ
Quantity
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Description
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Part Number
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C1,C2,C3,C4
ÁÁÁÁÁ
ÁÁÁÁÁ
4
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 0.1
µ
F
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
281-0775-01
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C5,C6
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 470
µ
F
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
290-0919-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C7,C8
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 0.47
µ
F
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
281-0563-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C9,C10
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 10
µ
F
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
290-0804-01
ÁÁÁÁÁÁÁÁÁÁ
C11,C17
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Not installed
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C12
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 4-40pF
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
281-0187-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C14
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 0.1
µ
F
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
285-0526-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C15
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 4700pF
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
281-0772-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
C16
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Capacitor, 10pF
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
281-0811-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
CR1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Diode Bridge
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
152-0585-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
CR2
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Diode
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
152-0246-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
F1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Fuse 0.1 A, 250 V, 3AG Medium Blow
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
159-0171-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
F2, F3
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Fuse, self-healing, RXE020, 0.2 A
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
307-1593-00
ÁÁÁÁÁÁÁÁÁÁ
FL1
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Line Filter
ÁÁÁÁÁÁÁ
119-1946-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
J1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Connector, Wire Set
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
198-5783-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
J2,J4
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Connector, BNC Female
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
131-3379-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
J3
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Connector, Sq Pin Jumper
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
131-0993-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Q1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Transistor, 151-1031-00
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
151-1031-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 464.0K
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3449-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R2
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 1.00M
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
321-0481-07
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R3
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 402.0
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3155-00
ÁÁÁÁÁÁÁÁÁÁ
R4,R6
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 200.0
Ω
ÁÁÁÁÁÁÁ
322-3126-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R5
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 374.0
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3152-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R7
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 75.0
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3085-07
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R8
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 1.15K
Ω
(0.1%)
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
321-0199-07
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R9
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 7.32K
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3276-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R10
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 50.5
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
321-1068-07
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R11
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 100.0
Ω
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
322-3097-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
R12
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 50
Ω
Variable
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
311-2226-00
ÁÁÁÁÁÁÁÁÁÁ
R13
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Resistor, 10.0
Ω
ÁÁÁÁÁÁÁ
322-3001-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
T1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Transformer, 120-14ASIG
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
120-1905-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
TP1,TP2,TP3,TP4,TP5, TP6
ÁÁÁÁÁ
ÁÁÁÁÁ
6
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Terminal, Test point
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
214-4085-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
U1
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
IC, MC7815
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
156-0312-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
U2
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
IC, MC7915
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
156-0527-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
U3
ÁÁÁÁÁ
ÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
IC, CLC404
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
156-3962-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
VR1,VR2
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Zener, 10V
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
152-0149-00
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
VR3,VR4
ÁÁÁÁÁ
ÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
Zener, 6.2V
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
152-0166-00

Probe Adapter
Operating Information
5
INTERNATIONAL POWER CORDS
Power cords are provided to match a variety of mains connections throughout the world. See Table 3 for
plug configurations, typical use, and Probe Adapter part number to order. A separate part number has
been provided to permit the customer to order the Probe Adapter with the power cord needed for their
application. Probe Adapters ordered for countries having 220V AC power are internally set for 220V
operation and have a label affixed on the rear panel to that effect.
Table 3
Probe Adapter Part Numbers
Plug Configuration
Typical Use
Probe Adapter Part Number
United States 115V
067–1429–00
European 220V
067–1429–01
United Kingdom 240V
a
067–1429–02
Australian 240V
067–1429–03
North American 240V
067–1429–04
Switzerland 220V
067–1429–05
a
A 6 Ampere, type C fuse is also installed inside the plug of the United Kingdom power cord.

&
$
$
&
)
)
%
)
$
"
&$
)
)
)
%
&
$
&
&
&
&"
)
$
&
"
!
)
"
"
"
(
(
$#$
)
"
"
%$
' "
$
#$
"
!
"
$#$
&
%
)
&
&"
"
"
(
"
$
%$ %$
"
"
&
"
)
PROBE ADAPTER
1
VM700A PROBE ADAPTER