TLA700/TLA7000 Series Tektronix Logic Analyzer

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tek.com

*P077003702*

077-0037-02

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tek.com

to find contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................

iii

Clear and sanitize procedures .....................................................................................

1

Memory devices ...............................................................................................

1

Data export devices............................................................................................

3

Troubleshooting.....................................................................................................

5

How to sanitize or clear a nonfunctional instrument.......................................................

5

TLA700/TLA7000 Series Declassification and Security Instructions

i

 


background image

Table of Contents

ii

TLA700/TLA7000 Series Declassification and Security Instructions

 


background image

Preface

This document helps customers with data security concerns to sanitize or remove
memory devices from the TLA700 and TLA7000 Series Logic Analyzer family
modules.

These products have data storage (memory) devices and data output devices.
These instructions tell how to do the following:

Clear or sanitize the memory devices

Clear or sanitize an instrument that is not functioning

Products

The following Tektronix products are covered by this document:

TLA7AA1, TLA7AA2, TLA7AA3, TLA7AA4

TLA7AB2, TLA7AB4

TLA7AC2, TLA7AC3, TLA7AC4

TLA7D1, TLA7D2

TLA7E1, TLA7E2

TLA7NA1, TLA7NA2, TLA7NA3, TLA7NA4

TLA7N1, TLA7N2, TLA7N3, TLA7N4

TLA7P2, TLA7P4

TLA7PG2

TLA7Q2, TLA7Q4

TLA7XM

TLA700/TLA7000 Series Declassification and Security Instructions

iii

 


background image

Preface

Related documents

Refer to the following service documents available on the Tektronix Web site at
www.tektronix.com/manuals or to the TLA Documentation CD that is available
with your product:

TLA700 Series Mainframe Installation Manual

TLA7000 Series Mainframe Installation Manual

Tektronix Logic Analyzer Module (TLA7AAx, TLA7ABx & TLA7NAx) Service
Manual

TLA7ACx Logic Analyzer Module Service Manual

Tektronix Logic Analyzer Module (TLA7Nx, TLA7Px, & TLA7Qx) Service
Manual

TLA7PG2 Pattern Generator and Probes Service Manual

TLA721 Benchtop & TLA7XM Expansion Mainframe Service Manual

TLA7Dx & TLA7Ex Digitizing Oscilloscope Service Manual

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured

area. All reusable memory is cleared to deny access to previously stored
information by standard means of access.

Erase.

This is equivalent to clear.

Media storage/data export device.

Any of several devices that can be used

to store or export data from the instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Remove.

This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product Service
Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot

be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.

Scrub.

This is equivalent to sanitize.

User-modifiable.

The user can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

iv

TLA700/TLA7000 Series Declassification and Security Instructions

 


background image

Clear and sanitize procedures

Memory devices

The following tables list the volatile and nonvolatile memory devices.

Table 1: Volatile memory devices

Type and
minimum size

Function

User
modi

fi

able

Data input
method

Location

To clear

To sanitize

DRAM,
1 M X 8

Firmware
execution code

No

Written
by the
processor
system

LPU board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

DRAM,
1 M X 4

Firmware
execution code

No

Written
by the
processor
system

LPU board, Enhanced
Monitor board,
TLA7PG2 Clock &
Backplane interface
board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

DRAM,
1 M X 16

Firmware
execution code

No

Written
by the
processor
system

LPU board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

DRAM,
64 K X 16

Firmware
execution code

No

Written
by the
processor
system

LPU board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

BICMOS

Firmware
execution code

No

Written
by the
processor
system

LPU board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

SDRAM
4 M X 16

Acquisition
memory for
storing acquired
data

No

Written by
controller
ASICs

Acquisition board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

SDRAM
1 M X 16

Acquisition
memory for
storing acquired
data

No

Written by
controller
ASICs

Acquisition board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

SDRAM
16 M X 16

Acquisition
memory for
storing acquired
data

No

Written by
controller
ASICs

Acquisition board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

ASIC
16 K

Processor
execution
memory

No

Written by
ASICs

Acquisition board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

SRAM,
128 K X 8

Firmware
execution code

No

Written
by the
processor
system

LPU board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

IC memory
256 K X 16 bit

Pattern generator
data

No

Written by
ASICs

TLA7PG2 PG board

Remove power source
from the instrument for
at least 20 seconds.

Remove power source
from the instrument for
at least 20 seconds.

TLA700/TLA7000 Series Declassification and Security Instructions

1

 


background image

Clear and sanitize procedures

Table 2: Nonvolatile memory devices

Type and
minimum size

Function

User
modi

fi

able

Data input
method

Location

To clear

To sanitize

FLASH,

2 M X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

LPU board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

FLASH,

1 M X 16

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

LPU board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

FLASH,

2 M X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

Probe Flash Adapter
board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

FLASH,

256 K X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7Dx & TLA7Ex
LPU board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

FLASH,

256 K X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7XM Enhanced
Monitor board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

NVRAM,
32 K X 8

Stores instrument
serial number
and calibration
constants

No

Written by
processor

LPU board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

NVRAM,
512 K X 8

Stores instrument
serial number
and calibration
constants

No

Written by
processor

TLA7Dx & TLA7Ex
LPU board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EPROM
64 K X 16

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7XM Enhanced
Monitor board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EEPROM
1 K X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7XM Enhanced
Monitor board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

PROM

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7XM Enhanced
Monitor board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EPROM
2048 X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7PG2 Clock &
Backplane interface
board, TLA7PG2 PG
board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

2

TLA700/TLA7000 Series Declassification and Security Instructions

 


background image

Clear and sanitize procedures

Table 2: Nonvolatile memory devices (cont.)

Type and
minimum size

Function

User
modi

fi

able

Data input
method

Location

To clear

To sanitize

EPROM
512 K X 8, Flash

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7PG2 Clock &
Backplane interface
board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EPROM
256 K X 16

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7PG2 Clock &
Backplane interface
board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EPROM
1 M X 1 bit

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

TLA7PG2 PG board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

EEPROM
256 K X 8

Stores instrument

fi

rmware

No

Programmed
at the
factory, no
user data

LPU board,
Acquisition board

Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

Data export devices

The following table lists the data export devices.

Table 3: Data export devices

Type

Function

User
modi

fi

able

Input method

Location

Process to disable

Analog outputs

Provides an analog
copy of signals from the
device under test

No

From the device under
test

Front panel of
module

N/A. Outputs cannot be disabled.

Backplane

Interfaces with TLA
mainframe

No

Instrument function and
application software

Backplane
connectors

Power off the TLA mainframe power and
remove the module. (See below.)

Backplane

To locate and remove the module from the TLA mainframe, refer to the

TLA7000

Series Mainframe Installation Manual

or to the

TLA700 Series Mainframe

Installation Manual

on the Tektronix Web site at www.tektronix.com/manuals or

on the TLA Documentation CD.

TLA700/TLA7000 Series Declassification and Security Instructions

3

 


background image

Clear and sanitize procedures

4

TLA700/TLA7000 Series Declassification and Security Instructions

 


background image

Troubleshooting

How to sanitize or clear a nonfunctional instrument

If your instrument is not functioning, remove the power source from the
instrument for at least 20 seconds.

TLA700/TLA7000 Series Declassification and Security Instructions