Full Text Searchable PDF User Manual

TLA700/TLA7000 Series Tektronix Logic Analyzer
Declassi
fi
cation and Security
ZZZ
Instructions
xx
www.tek.com
*P077003702*
077-0037-02

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit
www.tek.com
to find contacts in your area.

Table of Contents
Preface ..............................................................................................................
iii
Clear and sanitize procedures .....................................................................................
1
Memory devices ...............................................................................................
1
Data export devices............................................................................................
3
Troubleshooting.....................................................................................................
5
How to sanitize or clear a nonfunctional instrument.......................................................
5
TLA700/TLA7000 Series Declassification and Security Instructions
i

Table of Contents
ii
TLA700/TLA7000 Series Declassification and Security Instructions

Preface
This document helps customers with data security concerns to sanitize or remove
memory devices from the TLA700 and TLA7000 Series Logic Analyzer family
modules.
These products have data storage (memory) devices and data output devices.
These instructions tell how to do the following:
Clear or sanitize the memory devices
Clear or sanitize an instrument that is not functioning
Products
The following Tektronix products are covered by this document:
TLA7AA1, TLA7AA2, TLA7AA3, TLA7AA4
TLA7AB2, TLA7AB4
TLA7AC2, TLA7AC3, TLA7AC4
TLA7D1, TLA7D2
TLA7E1, TLA7E2
TLA7NA1, TLA7NA2, TLA7NA3, TLA7NA4
TLA7N1, TLA7N2, TLA7N3, TLA7N4
TLA7P2, TLA7P4
TLA7PG2
TLA7Q2, TLA7Q4
TLA7XM
TLA700/TLA7000 Series Declassification and Security Instructions
iii

Preface
Related documents
Refer to the following service documents available on the Tektronix Web site at
www.tektronix.com/manuals or to the TLA Documentation CD that is available
with your product:
TLA700 Series Mainframe Installation Manual
TLA7000 Series Mainframe Installation Manual
Tektronix Logic Analyzer Module (TLA7AAx, TLA7ABx & TLA7NAx) Service
Manual
TLA7ACx Logic Analyzer Module Service Manual
Tektronix Logic Analyzer Module (TLA7Nx, TLA7Px, & TLA7Qx) Service
Manual
TLA7PG2 Pattern Generator and Probes Service Manual
TLA721 Benchtop & TLA7XM Expansion Mainframe Service Manual
TLA7Dx & TLA7Ex Digitizing Oscilloscope Service Manual
Terms
The following terms may be used in this document:
Clear.
This removes data on media/memory before reusing it in a secured
area. All reusable memory is cleared to deny access to previously stored
information by standard means of access.
Erase.
This is equivalent to clear.
Media storage/data export device.
Any of several devices that can be used
to store or export data from the instrument, such as a USB port.
Nonvolatile memory.
Data is retained when the instrument is powered off.
Remove.
This is a physical means to clear the data by removing the memory
device from the instrument. Instructions are available in the product Service
Manual.
Sanitize.
This eradicates the data from media/memory so that the data cannot
be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.
Scrub.
This is equivalent to sanitize.
User-modifiable.
The user can write to the memory device during normal
instrument operation, using the instrument interface or remote control.
Volatile memory.
Data is lost when the instrument is powered off.
iv
TLA700/TLA7000 Series Declassification and Security Instructions

Clear and sanitize procedures
Memory devices
The following tables list the volatile and nonvolatile memory devices.
Table 1: Volatile memory devices
Type and
minimum size
Function
User
modi
fi
able
Data input
method
Location
To clear
To sanitize
DRAM,
1 M X 8
Firmware
execution code
No
Written
by the
processor
system
LPU board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
DRAM,
1 M X 4
Firmware
execution code
No
Written
by the
processor
system
LPU board, Enhanced
Monitor board,
TLA7PG2 Clock &
Backplane interface
board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
DRAM,
1 M X 16
Firmware
execution code
No
Written
by the
processor
system
LPU board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
DRAM,
64 K X 16
Firmware
execution code
No
Written
by the
processor
system
LPU board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
BICMOS
Firmware
execution code
No
Written
by the
processor
system
LPU board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
SDRAM
4 M X 16
Acquisition
memory for
storing acquired
data
No
Written by
controller
ASICs
Acquisition board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
SDRAM
1 M X 16
Acquisition
memory for
storing acquired
data
No
Written by
controller
ASICs
Acquisition board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
SDRAM
16 M X 16
Acquisition
memory for
storing acquired
data
No
Written by
controller
ASICs
Acquisition board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
ASIC
16 K
Processor
execution
memory
No
Written by
ASICs
Acquisition board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
SRAM,
128 K X 8
Firmware
execution code
No
Written
by the
processor
system
LPU board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
IC memory
256 K X 16 bit
Pattern generator
data
No
Written by
ASICs
TLA7PG2 PG board
Remove power source
from the instrument for
at least 20 seconds.
Remove power source
from the instrument for
at least 20 seconds.
TLA700/TLA7000 Series Declassification and Security Instructions
1

Clear and sanitize procedures
Table 2: Nonvolatile memory devices
Type and
minimum size
Function
User
modi
fi
able
Data input
method
Location
To clear
To sanitize
FLASH,
2 M X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
LPU board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
FLASH,
1 M X 16
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
LPU board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
FLASH,
2 M X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
Probe Flash Adapter
board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
FLASH,
256 K X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7Dx & TLA7Ex
LPU board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
FLASH,
256 K X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7XM Enhanced
Monitor board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
NVRAM,
32 K X 8
Stores instrument
serial number
and calibration
constants
No
Written by
processor
LPU board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
NVRAM,
512 K X 8
Stores instrument
serial number
and calibration
constants
No
Written by
processor
TLA7Dx & TLA7Ex
LPU board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EPROM
64 K X 16
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7XM Enhanced
Monitor board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EEPROM
1 K X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7XM Enhanced
Monitor board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
PROM
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7XM Enhanced
Monitor board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EPROM
2048 X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7PG2 Clock &
Backplane interface
board, TLA7PG2 PG
board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
2
TLA700/TLA7000 Series Declassification and Security Instructions

Clear and sanitize procedures
Table 2: Nonvolatile memory devices (cont.)
Type and
minimum size
Function
User
modi
fi
able
Data input
method
Location
To clear
To sanitize
EPROM
512 K X 8, Flash
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7PG2 Clock &
Backplane interface
board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EPROM
256 K X 16
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7PG2 Clock &
Backplane interface
board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EPROM
1 M X 1 bit
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
TLA7PG2 PG board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
EEPROM
256 K X 8
Stores instrument
fi
rmware
No
Programmed
at the
factory, no
user data
LPU board,
Acquisition board
Not applicable, does
not contain user data
or settings. Clearing
would disable instrument
functionality.
Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.
Data export devices
The following table lists the data export devices.
Table 3: Data export devices
Type
Function
User
modi
fi
able
Input method
Location
Process to disable
Analog outputs
Provides an analog
copy of signals from the
device under test
No
From the device under
test
Front panel of
module
N/A. Outputs cannot be disabled.
Backplane
Interfaces with TLA
mainframe
No
Instrument function and
application software
Backplane
connectors
Power off the TLA mainframe power and
remove the module. (See below.)
Backplane
To locate and remove the module from the TLA mainframe, refer to the
TLA7000
Series Mainframe Installation Manual
or to the
TLA700 Series Mainframe
Installation Manual
on the Tektronix Web site at www.tektronix.com/manuals or
on the TLA Documentation CD.
TLA700/TLA7000 Series Declassification and Security Instructions
3

Clear and sanitize procedures
4
TLA700/TLA7000 Series Declassification and Security Instructions

Troubleshooting
How to sanitize or clear a nonfunctional instrument
If your instrument is not functioning, remove the power source from the
instrument for at least 20 seconds.
TLA700/TLA7000 Series Declassification and Security Instructions
5