PDF User Manual

  1. Home
  2. Manuals
  3. Tektronix PPG3000 Series Declassification And Security Instructions

Tektronix PPG3000 Series Declassification And Security Instructions

PPG3002\ PPG3004\ PPG3200 Series\ PED3200 Series\ PPG3001\ PPG3201\ PPG3202\ PPG3204\ PED3201\ PED3202

Made by: Tektronix
Type: Declassification And Security Instructions
Category: Inverter
Pages: 14
Size: 0.22 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

PPG3000/PPG3200/PED3200 Series

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077077600*

077-0776-00

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by
national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published
material. Speci

fi

cations and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tektronix.com

to

fi

nd contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................................................................

iii

Clear and sanitize procedures .....................................................................................................................................

1

Memory devices ...............................................................................................................................................

1

Media and data export devices ...............................................................................................................................

5

Troubleshooting.....................................................................................................................................................

7

Clear or sanitize a nonfunctional instrument ...............................................................................................................

7

Change log ..........................................................................................................................................................

8

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

i

 


background image

Table of Contents

ii

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

 


background image

Preface

This document helps customers with data security concerns to sanitize or remove memory devices from the PPG3000 or
PPG3200 Series Programmable Pattern Generators or the PED3200 Series Programmable Error Detectors.

These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to
clear or sanitize the memory devices and disable the data output devices. The instructions also tell how to declassify a
non functioning instrument.

Reference

The procedures in this document are written to meet the requirements speci

fi

ed in:

NISPOM, DoD 5220.22–M, Chapter 8

ISFO Process Manual for Certi

fi

cation & Accreditation of Classi

fi

ed Systems under NISPOM

Products

The following Tektronix products are covered by this document:

PPG3001, PPG3002, PPG3004 Programmable Pattern Generators

PPG3201, PPG3202, PPG3204 Programmable Pattern Generators

PED3201, PED3202 Programmable Error Detectors

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to

deny access to previously stored information by standard means of access.

Erase.

This is equivalent to clear.

Instrument declassi

fi

cation.

A term that refers to procedures that must be undertaken before an instrument can be

removed from a secure environment. Declassi

fi

cation procedures include memory sanitization and memory removal,

and sometimes both.

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

iii

 


background image

Preface

Media storage/data export device.

Any of several devices that can be used to store or export data from the

instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Power off.

Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For the

purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products,
you will need to either press a rear-panel OFF switch or remove the power source from the instrument.

Remove.

This is a physical means to clear the data by removing the memory device from the instrument. Instructions

are available in the product Service Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot be recovered by other means or

technology. This is typically used when the device will be moved (temporarily or permanently) from a secured
area to a non-secured area.

Scrub.

This is equivalent to sanitize.

User accessible.

User is able to directly retrieve the memory device contents.

User-modi

fi

able.

The user can write to the memory device during normal instrument operation, using the instrument

interface or remote control.

User-modi

fi

able.

The user can write to the memory device during normal instrument operation, using the instrument

interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

iv

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and sanitize procedures

Memory devices

The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options.
Detailed procedures to clear or sanitize these devices, if any, are shown following each table.

Terminology

The following terms are used in the tables in this section:

User data – Describes the type of information stored in the device. Refers to waveforms or other measurement data
representing signals connected to the instrument by users.

User settings – Describes the type of information stored in the device. Refers to instrument settings that can be
changed by the user.

Both – Describes the type of information stored in the device. It means that both user data and user settings are
stored in the device.

None – Describes the type of information stored in the device. It means that neither user data nor user settings
are stored in the device.

Directly – Describes how data is modi

fi

ed. It means that the user can modify the data.

Indirectly – Describes how data is modi

fi

ed. It means that the instrument system resources modify the data and that

the user cannot modify the data.

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

1

 


background image

Clear and sanitize procedures

Table 1: Volatile memory devices

Type and
min. size

Function

Type of user
info stored

Backed-up
by battery

Method of
modi

fi

cation

Data input
method

Location

User
accessible

To clear

Process to sanitize

Microprocessor
RAM (32 k)

Holds programs
and active
acquisition data

System and
user data

No

Indirectly

Firmware
operations

Processor
board

No

Remove power from the
instrument for at least 20
seconds.

Remove power from the
instrument for at least 20
seconds.

FPGA RAM
(~4.1 Mbits)

Holds pattern
data and
miscellaneous
FPGA state
information

System and
user data

No

Indirectly

Firmware
operations

FPGA
board

No

Remove power from the
instrument for at least 20
seconds.

Remove power from the
instrument for at least 20
seconds.

Table 2: Nonvolatile memory devices

Type and
min. size

Function

Type of

user info
stored

Method of
modi

fi

cation

Data Input
method

Location

User accessible

To clear

To sanitize

Microprocessor
ROM
(512 k)

Holds operating
system

None

Not
modi

fi

able

Programmed at
factory

Internal

No

Not accessible to user

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

FPGA
programming
ROM

Holds FPGA

fi

rmware

None

Not
modi

fi

able

Programmed at
factory

Internal

No

Not accessible to user

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

2

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and sanitize procedures

Table 2: Nonvolatile memory devices (cont.)

Type and
min. size

Function

Type of

user info
stored

Method of
modi

fi

cation

Data Input
method

Location

User accessible

To clear

To sanitize

SD Card
(1 G)

Holds system,
calibration, and user
data

Pattern data,
system setup

Indirect

Front panel,
USB remote
programming

Internal

Yes

Delete all user pattern data
and user instrument setup

fi

les.

Restore factory defaults to
instrument. Factory calibration
data remains.

Remove SD card.
Disables instrument
functionality. (Note:
opening instrument
and removing SD card
invalidates customer
warranty. Contact
Tektronix for more
information.)

EEPROM
(256 k)

Holds instrument
con

fi

guration data

Used to
con

fi

gure

instrument

Not
modi

fi

able

Programmed at
factory

Internal

No

Not accessible to user

Not applicable, does
not contain user data
or settings. Sanitizing
would disable instrument
functionality.

Clear the SD Card

NOTE.

The SD Card is the only location where user data is saved after turning the instrument off.

The following steps delete the instrument setup

fi

les and user pattern data

fi

les and loads the factory default parameters.

The steps involve touching soft keys on the touch screen through the graphical user interface.

1.

Touch the

UTIL

soft key.

2.

Touch the

SAVE SETUP

soft key.

3.

Touch the

DEFAULTS

soft key.

4.

Touch the

FILE MGMT

soft key.

Follow the on-screen procedure to delete all user instrument setup

fi

les.

5.

Touch the

PATTERN

soft key.

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

3

 


background image

Clear and sanitize procedures

6.

Touch the

FILE MGMT

soft key.

Follow the on-screen procedure to delete all user pattern data

fi

les.

The instrument should now have the factory default parameters loaded and all user instrument setup

fi

les and all user

pattern data

fi

les should be deleted.

7.

Power off the instrument.

Nonvolatile recovery

information

No recovery from a clear is needed as the nonvolatile clear procedure does not erase or change factory calibration constants
and the instrument

fi

rmware is not impacted. There is no standard sanitization procedure as it requires partial disassembly

of the instrument that is only permitted by factory authorized personnel.

4

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and sanitize procedures

Media and data export devices

The following table lists the data export devices. Procedures to disable these devices, if any, are shown following the table.

Table 3: Media and data export devices

Type and min.
size

Function

Method of
modi

fi

cation

Data input method

Location

User accessible

Process to disable

USB device port

Supports remote control
and data transfer to a
PC.

Directly

Remote control via
USBTMC

USB device port on rear
of instrument

Yes

The USB device port cannot be disabled.

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

5

 


background image

Clear and sanitize procedures

6

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

 


background image

Troubleshooting

Clear or sanitize a nonfunctional instrument

CAUTION.

Removing instrument covers to access internal components invalidates your warranty. Do not remove instrument

covers unless you are quali

fi

ed to do so. For additional information, contact your local Tektronix service representative.

If your instrument is not functioning and you need to clear or sanitize it, the SD card should be removed and the instrument
should be returned to Tektronix Inc. To remove the SD card, the chassis covers must be removed and the SD card must
be removed from the processor board. Please contact your local Tektronix service representative for more SD card
removal details.

The SD card is the only part of the instrument that will retain user data (such as pattern data or instrument setups) after
the instrument has been turned off.

Charges

Replacement of any missing hardware will be charged according to the rate at the time of replacement.

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions

7

 


background image

Change log

Change log

Document

part number

Revision date

Change description

077-0776-00

1-17-2013

Initial release

8

PPG3000/PPG3200/PED3200 Series Declassi

fi

cation and Security Instructions