FCA3000, FCA3100 Series Timer/Counter/Analyzers,

MCA3000 Series Microwave Counter/Analyzers

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077049601*

077-0496-01

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by
national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published
material. Speci

fi

cations and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tek.com

to

fi

nd contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................................................................

iii

Products........................................................................................................................................................

iii

Terms...........................................................................................................................................................

iv

Clear and Sanitize Procedures.....................................................................................................................................

1

Memory Devices...............................................................................................................................................

1

Media and Data Export Devices .............................................................................................................................

3

Troubleshooting.....................................................................................................................................................

4

How to Clear or Sanitize a Nonfunctional Instrument.....................................................................................................

4

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

i

 


background image

Table of Contents

ii

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

 


background image

Preface

This document helps customers with data security concerns to sanitize or remove memory devices from the FCA3000,
FCA3100, and MCA3000 Series instruments.

These products have data storage (memory) devices and data export devices (USB port and GPIB connector). These
instructions describe how to clear or sanitize the memory devices and disable the data output devices. The instructions also
describe how to declassify an instrument that is not functioning.

Reference

The procedures in this document are written to meet the requirements speci

fi

ed in:

NISPOM, DoD 5220.22–M, Chapter 8

ISFO Process Manual for Certi

fi

cation & Accreditation of Classi

fi

ed Systems under NISPOM

Products

The following Tektronix products are covered by this document:

FCA3000, FCA3003, and FCA3020

FCA3100, FCA3103, and FCA3120

MCA3027 and MCA3040

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

iii

 


background image

Preface

Terms

This document may use the following terms:

Clear.

This eradicates data on media/memory before reusing it in a secured area. All reusable memory is cleared to

deny access to previously stored information by standard means of access.

Erase.

This is equivalent to clear.

Media.

Storage/data export device. A device that stores or exports data from the instrument, such as a USB

fl

ash

drive or USB port.

Sanitize.

This removes the data from media/memory so that the data cannot be recovered using any known technology.

This is typically used when the device is moved (temporarily or permanently) from a secured area to a nonsecured area.

Scrub.

This is equivalent to sanitize.

User-Accessible.

The user can directly retrieve the memory device contents.

User-Modi

fi

able.

The memory device can be written to by the user during normal instrument operation, using the

instrument user interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Power off.

Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For clearing

data, putting the instrument in Standby mode does not qualify as powering off. For these products, you must either
push a rear-panel OFF switch or remove the power source from the instrument.

Instrument Declassi

fi

cation.

A term that refers to procedures that must be undertaken before an instrument can be

removed from a secure environment. Declassi

fi

cation procedures include memory sanitization and memory removal,

and sometimes both.

iv

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

 


background image

Clear and Sanitize Procedures

Memory Devices

The following tables list the volatile and nonvolatile memory devices in the instrument.

Terminology

The tables in this section use the following terms:

User data – Describes the type of information stored in the device. Refers to waveforms or other measurement
information representing signals connected to the instrument by users.

User settings – Describes the type of information stored in the device. Refers to instrument settings that can be
changed by the user.

Both – Describes the type of information stored in the device. It means that both user data and user settings are
stored in the device.

None – Describes the type of information stored in the device. It means that neither user data or user settings
are stored in the device.

Directly – Describes how data is modi

fi

ed. It means that the user can modify the data.

Indirectly – Describes how data is modi

fi

ed. It means that the instrument system resources modi

fi

es the data and that

the user cannot modify the data.

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

1

 


background image

Clear and Sanitize Procedures

Table 1: Volatile memory devices

Type and
minimum size

Function

Type of user
info stored

Backed-up
by battery

Method of
modi

fi

cation

Data input
method

Location

User
accessible

To clear

To sanitize

SRAM, 16 kB
(All models)

Processor memory

None (working
processor
memory)

No

Indirect

System
resources

Main board

No

Remove power from
the instrument for at
least 30 seconds

Remove power from
the instrument for at
least 30 seconds

SRAM, 32 MB
(All FCA3000,
MCA3000
Series)

Firmware run time
storage, User settings,
Measurement buffers

User data, user
settings

No

Indirect

System
resources

Main board

No

Remove power from
the instrument for at
least 30 seconds

Remove power from
the instrument for at
least 30 seconds

SRAM,128 MB
(All FCA3100
Series)

Firmware run time
storage, User settings,
Measurement buffers

User data, user
settings

No

Indirect

System
resources

Main board

No

Remove power from
the instrument for at
least 30 seconds

Remove power from
the instrument for at
least 30 seconds

Table 2: Nonvolatile memory devices

Type and
minimum size

Function

Type of user
info stored

Method of
modi

fi

cation

Data input
method

Location

User
accessible

To clear

To sanitize

NVRAM, Flash,
8 MB
(All models)

Firmware storage, user
data, and user settings

User data
(explicitly stored
by user), User
settings

Direct

Loading a
Main board
con

fi

guration

fi

le

or by using the
user interface

Main board

Yes

Use the Total Reset
function to clear
NVRAM

1

Use the Total Reset
function to clear
NVRAM

1

1

The Total Reset function writes 0xFF to all

fl

ash sectors that hold stored user data (saved measurement data).

The Total Reset function also resets

fl

ash sectors that hold current instrument settings (including power-on settings, but excluding instrument con

fi

guration settings) to their default (factory reset) settings. However, due to

internal sector organization of the

fl

ash memory, and depending on the size of the setting data, settings may be stored in two

fl

ash sectors. Parts of previous setting data might exist in a

fl

ash sector with the current setting

data. The Total Reset function sets the current settings storage memory to the default settings, but previous con

fi

guration settings data can persist in the noncurrent-settings locations within memory sectors.

2

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

 


background image

Clear and Sanitize Procedures

Media and Data Export Devices

The following table lists the data export devices in the instrument.

Table 3: Media and data export devices

Type and
minimum size

Function

Method of
modi

fi

cation

Data input
method

Location

User accessible

To disable

USB connector

Transfer data

Directly

System

resources

Main board,
connector at rear

of instrument

Yes

Remove all connected USB

devices, remove the USB cable

GPIB connector

Transfer data

Directly

System
resources

Main board,

connector at rear
of instrument

Yes

Remove all connected GPIB
devices, remove the GPIB cable

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security

3

 


background image

Troubleshooting

Troubleshooting

How to Clear or Sanitize a Nonfunctional Instrument

If your instrument is not functioning, remove and destroy the instrument Main board using your organizations' security
procedures. Return the product to Tektronix. The instrument will then be repaired and adjusted as necessary.

In North America, contact the Tektronix Customer Care Center (1-800-833-9200) for assistance with removing the main
board and returning the instrument to a repair center. Worldwide, visit www.tektronix.com to

fi

nd contacts in your area.

Charges

Replacement of any missing hardware is charged according to the rate at the time of replacement.

4

FCA3000, FCA3100 & MCA3000 Series Declassi

fi

cation and Security