DPO2000/B and MSO2000/B Series Oscilloscopes

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077010801*

077-0108-01

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Speci

fi

cations and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

TekSecure is a trademark of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tektronix.com

to

fi

nd contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................

iii

Clear and Sanitize Procedures.....................................................................................

1

Memory Devices...............................................................................................

1

Data Export Devices...........................................................................................

3

Built-In Security Features .........................................................................................

4

When to Use TekSecure.......................................................................................

4

What TekSecure Does .........................................................................................

4

Disable the LAN Port and Clear LAN Ethernet Settings..................................................

5

Disable the USB Device Port.................................................................................

6

Clear and Sanitize a Non-Functional Instrument................................................................

7

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

i

 


background image

Table of Contents

ii

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

 


background image

Preface

If you have data security concerns, this document helps you to sanitize or remove
memory devices from the DPO2000/B and MSO2000/B Series Digital Phosphor
and Mixed Signal Oscilloscopes.

The DPO2000/B and MSO2000/B series products have data storage (memory)
devices and data output devices (USB ports). These instructions tell you how
to clear or sanitize the memory devices, and also tell you how to declassify an
instrument that is not functioning.

Instrument code and calibration settings reside in nonvolatile

fl

ash memory.

Instrument setups and reference waveforms may also be stored in

fl

ash memory or

on USB drives connected to the instrument.

If you have any questions, contact the Tektronix Technical Support Center at
www.tektronix.com/support.

Products

The following Tektronix products are covered by this document:

DPO2002B

DPO2004B

DPO2012/B

DPO2014/B

DPO2022B

DPO2024/B

MSO2002B

MSO2004B

MSO2012/B

MSO2014/B

MSO2022B

MSO2024/B

Related Documents

The

DPO2000 and MSO2000 Series Digital Phosphor and Mixed Signal

Oscilloscopes Service Manual

, part number 071-2331-XX, is available on the

Tektronix Web site at www.tektronix.com/manuals.

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

iii

 


background image

Preface

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured

area. All reusable memory is cleared to deny access to previously stored
information by standard means of access.

Erase.

This is equivalent to clear.

Media storage/data export device.

Any of several devices that can be used

to store or export data from the instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Power off.

Some instruments have a “Standby” mode, in which power is

still supplied to the instrument. For the purpose of clearing data, putting the
instrument in Standby mode does not qualify as powering off. For these
products, you will need to either press a rear-panel OFF switch or remove the
power source from the instrument.

Remove.

This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product Service
Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot

be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.

Scrub.

This is equivalent to sanitize.

User-modi

fi

able.

The user can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

iv

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize Procedures

Memory Devices

The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.

Table 1: Volatile Memory Devices for the DPO2000/B and MSO2000/B Series

Type and minimum size

Function

User modi

fi

able

Data input
method

Location

To clear or sanitize

CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84

DDR memory: contains
the analog channel
waveform acquisition
data during operation

Only by the
processor system

N/A

Acquisition board

Remove power from the
instrument for at least
20 seconds

CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84

DDR memory: contains
the digital channel
waveform acquisition
data during operation

Only by the
processor system

N/A

Acquisition board
(MSO2000/B Series
only)

Remove power from the
instrument for at least
20 seconds

CMOS, SDRAM, 64MX16,
1G DDR2, 1.8V, 333MHZ;
MT47H64M16BT-3, FBGA84

DDR memory: contains
the instrument code and
data during operation

Only by the
processor system

N/A

Acquisition board

Remove power from the
instrument for at least
20 seconds

CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84

DDR memory: contains
the instrument code and
data during operation

Only by the
processor system

N/A

Acquisition board
(4-channel products
only)

Remove power from the
instrument for at least
20 seconds

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

1

 


background image

Clear and Sanitize Procedures

Table 2: Nonvolatile Memory Devices for the DPO2000/B and MSO2000/B Series

Type and minimum size

Function

User modi

fi

able

Data input
method

Location

To clear or sanitize

CMOS, FLASH; 16M X 8 OR
8M X 16, 128MBIT, 75NS, 3V;
PC28F128J3D-75, BGA64

FLASH: contains the
instrument

fi

rmware,

calibration constants,
reference waveforms,
and instrument setups

N/A

N/A

Acquisition board

To clear:

Not applicable

for calibration constants.
Use the TekSecure
feature to erase
reference waveforms
and instrument setups.
Use the procedure to
disable the LAN Ethernet
and to clear information,
such as IP addresses.
(See page 5,

Disable the

LAN Port and Clear LAN
Ethernet Settings

.)

To sanitize:

Remove

the acquisition board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Return the product
to Tektronix and a
new acquisition board
will be installed and
the instrument will be
repaired and adjusted as
necessary.

CMOS, CLOCK; SERIAL
REAL-TIME CLOCK,
CRYSTAL, 7 X 8 RAM, SPI,
3.3V; M41T93RMY, SO18.300

Real-time clock, with
battery backup

Only by the
processor system

N/A

Acquisition board

Not applicable; no user
data

CMOS, EEPROM; 128 X 8,
SERIAL; 24C01A, SOIC8

EEPROM: contains the
front end board serial
number and calibration
information

N/A

N/A

Front End Board

Not applicable; no user
data

CMOS, MICROCONTROLLER;
8-BIT, 32K BYTES ON-CHIP
FLASH, MC9S08AW32CPUE,
LQFP64

Microcontroller with
FLASH: contains the
front panel processor
code and power status
for soft power switch

Only by the
processor system

N/A

Front Panel Board

Not applicable; no user
data

CMOS, EEPROM; 256 X 8,
SERIAL; 24C02, SO8.15

EEPROM: contains
the application module

fi

rmware

N/A

N/A

DPO2000/B and
MSO2000/B application
module

Not applicable; no user
data

2

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize Procedures

Data Export Devices

The following table lists the data export devices.

Table 3: Data Export Devices

Type and minimum size

Function

User modi

fi

able

Data input
method

Location

To disable

USB host port

Supports the removable
USB

fl

ash drive; user

storage of reference
waveforms, screen
images, and instrument
setups

Yes

Save

USB host port on the
front of the instrument

Files can be deleted
or overwritten on the
instrument or a PC; the
USB

fl

ash drive can be

removed and destroyed.
The USB host port
cannot be disabled.

USB device port

Supports remote control
and data transfer to a
PC

Yes

Remote
control via
USBTMC

USB device port on the
rear of the instrument

Use the procedure to
disable the USB device
port. (See page 6,

Disable the USB Device
Port

.)

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

3

 


background image

Built-In Security Features

Built-In Security Features

When to Use TekSecure

You can use the TekSecure function to erase setup and reference waveform data
stored in internal

fl

ash memory.

The DPO2000/B and MSO2000/B series models have a USB host port on the
front panel. Any USB

fl

ash devices can be removed, and stored or destroyed.

What TekSecure Does

The TekSecure function does the following:

Replaces all waveforms in all reference memories with null sample values

Replaces the current front-panel setup and all stored setups with the default
setup values

Calculates the checksums of all reference waveform memory and setup
memory locations to verify successful completion of waveform and setup
erasure

Displays a dialog indicating whether the secure erase was successful or
unsuccessful

NOTE.

TekSecure does not erase or change factory calibration constants, user

protected data, or Ethernet settings.

To use TekSecure:

1.

Push the front-panel

Utility

button.

2.

Push the

Utility Page

lower-bezel button and use the

Multipurpose a

knob

to select

Con

fi

g

.

3.

Push the

TekSecure Erase Memory

lower-bezel button.

4.

Push the

OK Erase Setup & Ref Memory

side-bezel button. Wait for the

“TekSecure operation complete” dialog box to display.

5.

Push the

Menu Off

front-panel button to close the dialog box.

To reset the instrument

RAM:

1.

Power off the instrument for at least 20 seconds.

2.

Power on the instrument.

4

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

 


background image

Built-In Security Features

To overwrite protected

user data

This procedure overwrites everything that is stored in the protected user data area.

Send the following commands to the oscilloscope:

:PASSWORD “XYZZY” (or current password if not “XYZZY”)

:NEWPASS “ANYTHING”

:PASSWORD “ANYTHING”

*PUD #3300xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx[repeat
for a total of 300 x’s]

Disable the LAN Port and Clear LAN Ethernet Settings

To disable the LAN port and clear LAN Ethernet settings, such as IP addresses,
follow these steps:

CAUTION.

If you need to restore network connectivity at a later date, make sure to

write down all setting values before clearing them.

1.

Remove the network cable from the LAN port on the rear of the instrument.

2.

Push the front-panel

Utility

button.

3.

Push the

Utility Page

lower-bezel button and use the

Multipurpose a

knob

to select

I/O

.

4.

Push the

Ethernet Network Settings

lower-bezel button.

5.

Push the

Change Instrument Settings

side-bezel button.

6.

Push the

or

arrow side-bezel button to position the cursor on the

User

Password

information.

7.

Push the

Clear

lower-bezel button.

8.

Push the

arrow side-bezel button to position the cursor on the

Instrument

IP Address

information.

9.

Push the

Clear

lower-bezel button.

10.

Push the

arrow side-bezel button to position the cursor on the

Domain

Name

information.

11.

Push the

Clear

lower-bezel button.

12.

Push the

arrow side-bezel button to position the cursor on the

DNS IP

Address

information.

13.

Push the

Clear

lower-bezel button.

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

5

 


background image

Built-In Security Features

14.

Push the

arrow side-bezel button to position the cursor on the

Gateway IP

Address

information.

15.

Push the

Clear

lower-bezel button.

16.

Push the

arrow side-bezel button to position the cursor on the

Subnet Mask

information.

17.

Push the

Clear

lower-bezel button.

18.

Push the

OK Accept

side-bezel button.

The LAN system is disabled and no longer allows data traf

fi

c in or out. The

relevant LAN Ethernet settings are also cleared.

Disable the USB Device Port

To disable the USB device port, follow these steps:

1.

Remove any USB cable or device from the USB device port on the rear of the
instrument.

2.

Push the front-panel

Utility

button.

3.

Push the

Utility Page

lower-bezel button and use the

Multipurpose a

knob

to select

I/O

.

4.

Push the

USB

lower-bezel button.

5.

Push the

Disabled (Off Bus)

side-bezel button to disable the USB device port.

The USB device port is disabled and no longer allows data traf

fi

c in or out.

6

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize a Non-Functional Instrument

Clear and Sanitize a Non-Functional Instrument

If your instrument is not functioning and you need to clear or sanitize it, proceed
as follows:

Acquisition board

Only quali

fi

ed personnel should perform service procedures. Read the

Service

Safety Summary

and the

General Safety Summary

in the

DPO2000 and MSO2000

Series Digital Phosphor and Mixed Signal Oscilloscopes Service Manual

before

performing any service procedures.

Remove the Acquisition board and return the product to Tektronix. A new
Acquisition board will be installed, and the instrument will be repaired and
adjusted as necessary.

For removal instructions, refer to the service manual, part number 071-2331-XX,
available on the Tektronix Web site at www.tektronix.com/manuals.

After removal of the Acquisition board, refer to your company's internal policies
regarding handling or disposal of the board.

USB Flash Drive

Remove the USB

fl

ash drive and return the instrument to Tektronix for repair.

After removal of the USB

fl

ash drive, refer to your company's internal policies

regarding handling or disposal of the

fl

ash drive.

Charges

Replacement of any missing hardware will be charged according to the rate at
the time of replacement.

DPO2000/B and MSO2000/B Series Declassi

fi

cation and Security Instructions

7