AWG5200 Series

Arbitrary Waveform Generators

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077133800*

077-1338-00

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by
national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published
material. Speci

fi

cations and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tek.com

to

fi

nd contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................................................................

iii

Clear and Sanitize Procedures.....................................................................................................................................

1

Memory devices ...............................................................................................................................................

1

Media storage and data export devices ......................................................................................................................

5

AWG5200 Series Declassi

fi

cation and Security Instructions

i

 


background image

Table of Contents

ii

AWG5200 Series Declassi

fi

cation and Security Instructions

 


background image

Preface

This document helps customers with data security concerns to sanitize or remove memory devices from the Tektronix
AWG5200 Series Arbitrary Waveform Generators.

All of these products have data storage (memory) devices and data output devices (USB ports and a LAN Ethernet
connector). These instructions tell how to do the following:

Clear or sanitize the memory devices.

Clear or sanitize an instrument that is not functioning.

Reference

The procedures in this document are written to meet the requirements speci

fi

ed in:

NISPOM, DoD 5220.22–M, Chapter 8

ISFO Process Manual for Certi

fi

cation & Accreditation of Classi

fi

ed Systems under NISPOM

Products

The following Tektronix products are covered by this document:

AWG5202

AWG5204

AWG5208

Related documents

You can read more about your instrument in the following documents. These are available online at
www.tektronix.com/downloads.

AWG5200 Series Arbitrary Waveform Generators Installation and Safety Instructions.

AWG5200 Arbitrary Waveform Generators Speci

fi

cations and Performance Veri

fi

cation Technical Reference

AWG5200 Series Declassi

fi

cation and Security Instructions

iii

 


background image

Preface

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to

deny access to previously stored information by standard means of access.

Erase.

This is equivalent to clear.

Instrument declassi

fi

cation.

A term that refers to procedures that must be undertaken before an instrument can be

removed from a secure environment. Declassi

fi

cation procedures include memory sanitization and memory removal,

and sometimes both.

Media storage/data export device.

Any of several devices that can be used to store or export data from the

instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Power off.

Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For the

purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products,
you will need to either press a rear-panel OFF switch or remove the power source from the instrument.

Remove.

This is a physical means to clear the data by removing the memory device from the instrument. Instructions

are available in the product Service Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot be recovered by other means or

technology. This is typically used when the device will be moved (temporarily or permanently) from a secured
area to a non-secured area.

Scrub.

This is equivalent to sanitize.

User accessible.

User is able to directly retrieve the memory device contents.

User-modi

fi

able.

The user can write to the memory device during normal instrument operation, using the instrument

interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

iv

AWG5200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize Procedures

Memory devices

The following tables list the volatile and nonvolatile memory devices in the instrument and listed options. Detailed
procedures to clear or sanitize these devices, if any, are shown following each table.

Terminology

The following terms are used in the tables in this section:

User data – Describes the type of information stored in the device. Refers to waveforms or other measurement data
representing signals connected to the instrument by users.

User settings – Describes the type of information stored in the device. Refers to instrument settings that can be
changed by the user.

Both – Describes the type of information stored in the device. It means that both user data and user settings are
stored in the device.

None – Describes the type of information stored in the device. It means that neither user data nor user settings
are stored in the device.

Directly – Describes how data is modi

fi

ed. It means that the user can modify the data.

Indirectly – Describes how data is modi

fi

ed. It means that the instrument system resources modify the data and that

the user cannot modify the data.

AWG5200 Series Declassi

fi

cation and Security Instructions

1

 


background image

Clear and Sanitize Procedures

Table 1: Volatile memory devices

Type and

min. size

Function

Type of
info stored

(system/user)

Backed up by

battery?

Method of

modi

fi

cation

Data input

method

Location

User accessible?

To clear/sanitize

SDRAM

4 GB (std)

PC system memory

Both

No

Directly

Written by
processor system

Slot-PC board

Yes

Remove the power source from
the instrument for at least 20
seconds.

SDRAM
4 Gbit

AWG playback
memory

User

No

Indirectly

Firmware
operations

AWG board

No

Remove the power source from
the instrument for at least 20
seconds.

SRAM 64k x
16

USBTMC buffer

Both

No

Indirectly

Firmware
operations

System board

No

Remove the power source from
the instrument for at least 20
seconds.

Table 2: Nonvolatile memory devices

Type and

min. size

Function

Type of user

info stored

Method of

modi

fi

cation

Data input

method

Location

User accessible

To clear/sanitize

Flash,
1 Gbit

FPGA Image, Calibration Constants

None

Indirect

Firmware
operations

AWG board

No

N/A

1

Flash,
512 Mbit

FPGA Image, Instrument Model &
Serial Number, Enabled Options

None

Indirect

Firmware
operations

System board

No

N/A

1

EEPROM
1 Mbit

USBTMC pro

fi

le

None

Indirect

None

System board

No

N/A

1

EEPROM
64 kbit

Calibration Constants

None

Indirect

Firmware
operations

Clock board

No

N/A

1

Flash,
1 Gbit

Calibration Constants

None

Indirect

Firmware
operations

Marker board

No

N/A

1

2

AWG5200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize Procedures

Table 2: Nonvolatile memory devices (cont.)

Type and
min. size

Function

Type of user
info stored

Method of
modi

fi

cation

Data input
method

Location

User accessible

To clear/sanitize

Flash,
1 Gbit

Calibration Constants

None

Indirect

Firmware
operations

Output / DAC board

No

N/A

1

Removable
Hard Drive

Holds instrument operating system
and application software. Holds
all user-storable data such as
waveforms, measurement results,
and instrument settings.

User data, user
settings

Indirect

Firmware
operations, user
input

Front panel

Yes

To clear, see footnote

2

To sanitize, see footnote

3

1

This device does not contain user data or settings. Clearing would disable instrument functionality.

2

To clear, reformat the hard drive using the procedure “Restore the OS and AWG application” in this document.

3

To sanitize, remove the hard drive. Store the removed hard drive in a secure area or destroy it.

AWG5200 Series Declassi

fi

cation and Security Instructions

3

 


background image

Clear and Sanitize Procedures

Clear or sanitize

instrument

Use this procedure to clear the AWG waveform cache and last setup cache from your instrument. This procedure does not
erase or change factory calibration constants.

1.

In the toolbar, click the Reset to Default Setup button.

2.

Close the AWG application.

3.

Open the AWG application.

Clear or sanitize a

non-functional instrument

If your instrument is not functioning, proceed as follows.

1.

Remove the hard disk drive before returning the instrument for repair.

2.

Remove any attached USB

fl

ash drive before returning the instrument for repair.

Refer to your organization’s policies regarding storage or disposal of any removed items.

Restore the OS and AWG

application

Use this procedure to restore the operating system and the AWG application on your instrument to the original factory
condition. This procedure does not erase or change factory calibration constants.

NOTE.

This utility reformats the hard disk drive during the restore process. All user data will be removed. (The Acronis

Recovery partition and the related data are not destroyed.)

The restore images remain, providing the ability to repeat the recovery process.

The Windows operating system provided is designed for this instrument's hardware and product software.

1.

Attach a keyboard to the instrument.

2.

Power on the instrument. During the boot-up process you will see the following message at the top of the screen:

Starting Acronis Loader...

press F5 for Acronis Startup Recovery Manager

3.

Repeatedly press the F5 key until the Acronis True Image Tool opens.

4

AWG5200 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and Sanitize Procedures

There is a 15-second time period from when the message appears until the instrument proceeds with the normal
instrument startup. If the instrument does not open the Acronis application, power off the instrument, then power
on the instrument and try again.

4.

Within the Acronis TrueImage window, click Restore.

5.

In the Con

fi

rmation dialog box, click Yes to begin the restore the process, or No to exit.

6.

When

fi

nished, the AWG product software is the version shipped with your instrument. All upgrades or applications

installed after your purchase will have to be reinstalled.

Media storage and data export devices

The following table lists the media storage and data export devices in the instrument and listed options. Detailed procedures
to disable these devices, if any, are shown following the table.

Table 3: Media and data export devices

Type

Function

Method of

modi

fi

cation

Data input method

Location

Process to disable

USB host
port (supports
removable USB

fl

ash drive)

User storage of reference
waveforms, screen images,
and instrument setups

Directly

User writable

USB host port on front and rear
of instrument

The USB host ports cannot be disabled.

USB device port

Supports remote control and
data transfer to a PC

Directly

Remote control via USBTMC
and/or GPIB

Rear panel

The USB device port cannot be disabled.

LAN Ethernet
connector

Supports remote control and
data transfer to PC.

Directly

N/A

Rear panel

N/A

AWG5200 Series Declassi

fi

cation and Security Instructions

5