AFG2000 Series

Arbitrary Function Generators

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077061901*

077-0619-01

 


background image

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Speci

fi

cations and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

In North America, call 1-800-833-9200.
Worldwide, visit

www.tektronix.com

to

fi

nd contacts in your area.

 


background image

Table of Contents

Preface ..............................................................................................................

iii

Clear and sanitize procedures .....................................................................................

1

Memory devices ...............................................................................................

1

Media and data export devices ...............................................................................

3

Built-In security features ..........................................................................................

5

Secure ...........................................................................................................

5

Access Protection ..............................................................................................

5

Troubleshooting.....................................................................................................

7

How to clear or sanitize a nonfunctional instrument.......................................................

7

How to recover from clearing or removing the memory ..................................................

7

Change log ..........................................................................................................

8

AFG2000 Series Declassi

fi

cation and Security Instructions

i

 


background image

Table of Contents

ii

AFG2000 Series Declassi

fi

cation and Security Instructions

 


background image

Preface

This document helps customers with data security concerns to sanitize or
remove memory devices from the Tektronix AFG2000 Series Arbitrary Function
Generators.

All of these products have data storage (memory) devices. Some models have
data output devices (USB ports). These instructions tell how to clear or sanitize
the memory devices and disable the data output devices. The instructions also tell
how to declassify an instrument that is not functioning.

Reference

The procedures in this document are written to meet the requirements speci

fi

ed in:

NISPOM, DoD 5220.22–M, Chapter 8

ISFO Process Manual for Certi

fi

cation & Accreditation of Classi

fi

ed Systems

under NISPOM

Products

The following Tektronix arbitrary function generator models are covered by this
document:

AFG2021

AFG2021-SC

AFG2021-BR

Related documents

You can read more about your instrument in the following documents. These are
available online at www.tektronix.com/downloads.

AFG2000 Series Arbitrary Function Generator Service Manual

AFG2021 Arbitrary Function Generator Quick Start User Manual

AFG2021-SC and AFG2021-BR Arbitrary Function Generators Quick Start
User Manual

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured

area. All reusable memory is cleared to deny access to previously stored
information by standard means of access.

Erase.

This is equivalent to clear.

Instrument declassi

fi

cation.

A term that refers to procedures that must be

undertaken before an instrument can be removed from a secure environment.
Declassi

fi

cation procedures include memory sanitization and memory

removal, and sometimes both.

AFG2000 Series Declassi

fi

cation and Security Instructions

iii

 


background image

Preface

Media storage/data export device.

Any of several devices that can be used

to store or export data from the instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Power off.

Some instruments have a “Standby” mode, in which power is

still supplied to the instrument. For the purpose of clearing data, putting the
instrument in Standby mode does not qualify as powering off. For these
products, you will need to either press a rear-panel OFF switch or remove the
power source from the instrument.

Remove.

This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product Service
Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot

be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.

Scrub.

This is equivalent to sanitize.

User accessible.

User is able to directly retrieve the memory device contents.

User-modi

fi

able.

The user can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

iv

AFG2000 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and sanitize procedures

Memory devices

The following tables list the volatile and nonvolatile memory devices in the
instrument. Detailed procedures to clear or sanitize these devices, if any, are
shown following each table.

Terminology

The following terms are used in the tables in this section:

User data – Describes the type of information stored in the device. Refers to
waveforms or other measurement data representing signals connected to the
instrument by users.

User settings – Describes the type of information stored in the device. Refers
to instrument settings that can be changed by the user.

Both – Describes the type of information stored in the device. It means that
both user data and user settings are stored in the device.

None – Describes the type of information stored in the device. It means that
neither user data nor user settings are stored in the device.

Directly – Describes how data is modi

fi

ed. It means that the user can modify

the data.

Indirectly – Describes how data is modi

fi

ed. It means that the instrument

system resources modify the data and that the user cannot modify the data.

Table 1: Volatile memory devices

Type and
min. size

Function

Data Input method

Location

User accessible

To clear

SDRAM 1 M
x 64

Microprocessor system
memory, Display
memory

Written by processor
system

U202, U203 on
main board

No

Remove the power source from the
instrument for at least 20 seconds.

Table 2: Nonvolatile memory devices

Type and
min. size

Function

Data Input
method

Location

User accessible

To sanitize

Flash, 4 M x
16

Holds instrument

fi

rmware, setups,

waveforms, calibration
data, Ethernet MAC
address, and instrument
serial number

Programmed
at the factory,

fi

rmware

operations, and
user input

U201 on main board

Yes

Use Secure. See the Clear or
sanitize instrument procedure
(See page 2.).

AFG2000 Series Declassi

fi

cation and Security Instructions

1

 


background image

Clear and sanitize procedures

Clear or sanitize

instrument

Use this Secure function to erase con

fi

dential data from your instrument. This

procedure does not erase or change factory calibration constants.

1.

Push the front-panel Utility button to display the Utility menu.

2.

Push the System bezel button.

3.

Push the Secure bezel button. The following message appears on the screen:
“The Secure function will erase all setups and waveforms stored in the
internal memory. Are you sure you want to execute secure function?”

4.

Select OK to execute Secure. Secure does the following:

Erases all waveforms and stored setups in all user memories.

Replaces the current front-panel setup with the default factory setup

values.

Replaces the current GPIB and Ethernet settings with the default factory

settings for models with these ports.

2

AFG2000 Series Declassi

fi

cation and Security Instructions

 


background image

Clear and sanitize procedures

Media and data export devices

The following table lists the data export devices in the standard instrument and
listed options. Detailed procedures to disable these devices, if any, are shown
following the table.

Table 3: Media and data export devices

Type and min.
size

Function

User
modi

fi

able

Data Input method

Location

Process to disable

USB host port

Supports
removable USB
memory device to
save setups and
waveforms.

Yes

User writeable

Front panel

Files can be deleted or overwritten
on the arbitrary/function generator or
a PC; USB memory device can be
removed and destroyed. The USB host
port cannot be disabled.

USB device port

Supports remote
control and data
transfer to PC.

None

Remote control via
USB TMC

Rear panel

The USB device port cannot be
disabled.

LAN Ethernet

1

Supports remote
control and data
transfer to PC.

None

Remote control via
VXI-11

Rear panel

The Ethernet port can be disabled.
Set “0.0.0.0” as IP Address and
Subnet Mask under Utility-I/O
Interface-Ethernet.

GPIB

1

Supports remote
control and data
transfer to PC.

None

Remote control via
IEEE488.2

Rear panel

The GPIB can be disabled. Select
Off Bus as Utility-I/O Interface-
GPIB-Con

fi

guration.

1

This device is only available on AFG2021 instruments with option GL.

AFG2000 Series Declassi

fi

cation and Security Instructions

3

 


background image

Clear and sanitize procedures

4

AFG2000 Series Declassi

fi

cation and Security Instructions

 


background image

Built-In security features

The AFG2000 Series Arbitrary Function Generators have the following security
functions:

Secure

Access Protection

Secure

To erase con

fi

dential data from the instrument, use the Secure function (See

page 2,

Clear or sanitize instrument

.):

Erases all waveforms and stored setups in all user memories

Replaces the current front-panel setup with the default factory setup values

Replaces the current GPIB and Ethernet settings with the default factory
settings

Secure does not erase or change factory calibration constants

Access Protection

To protect calibration data and the instrument

fi

rmware, follow these steps:

1.

Push the front-panel Utility button to display the Utility menu.

2.

Push the -more- bezel button twice.

3.

Push the Security Menu bezel button.

4.

Select Access Protection to display the password input page.

5.

Enter your password. Use the general purpose knob to select the characters
and push the Enter Character bezel button after each selection.

As you enter the characters of your password, they are displayed as a series of
asterisks (*******) on the password input page.

AFG2000 Series Declassi

fi

cation and Security Instructions

5

 


background image

Built-In security features

If no speci

fi

c password has been previously de

fi

ned, use the default password

DEFAULT.

6.

Select OK to turn the access protection on.

NOTE.

If Access Protection is On, the Firmware Update and the Service Menu

cannot be selected. To change the Access Protection to Off, you must enter the
password.

You can change the password when Access Protection is off. First, enter the
old password and push the OK bezel button. Then enter the new password. The
password must be 7 characters.

For information about changing the password, refer to the Quick Start User
Manual for your product.

For information about the Service Menu, refer to the AFG2000 Series Arbitrary
Function Generators Service Manual.

6

AFG2000 Series Declassi

fi

cation and Security Instructions

 


background image

Troubleshooting

How to clear or sanitize a nonfunctional instrument

If your instrument is not functioning, perform the following actions and return the
instrument to Tektronix for repair. Describe the initial problem with the product.
Tektronix will install replacement parts and then repair and return the instrument.

Main board

Remove the main board and return the product to Tektronix. A new main board
will be installed, and the instrument will be repaired and adjusted as necessary.

For removal instructions, refer to the AFG2000 Series Arbitrary Function
Generators Service Manual, available on the Tektronix Web site at
www.tektronix.com/downloads.

After removal of the main board, refer to your company’s internal policies
regarding handling or disposal of the board.

USB memory device

Remove any attached USB memory devices from the instrument before returning
the instrument for repair.

After removal of USB memory devices, refer to your company's internal policies
regarding handling or disposal of the memory device.

Charges

Replacement of any missing hardware will be charged according to the rate at
the time of replacement.

How to recover from clearing or removing the memory

If you removed the main board in order to clear or sanitize your instrument,
reinstall the board and perform the speci

fi

ed adjustment procedures to verify that

the instrument is in the proper condition. For adjustment procedures, refer to the
AFG2000 Series Arbitrary Function Generators Service Manual.

AFG2000 Series Declassi

fi

cation and Security Instructions

7

 


background image

Change log

Change log

Document
part number

Revision
date

Change description

077-0619-00

April 2012

New document.

077–0619–01

September
2012

Updated with new instrument models.

8

AFG2000 Series Declassi

fi

cation and Security Instructions