PDF User Manual

  1. Home
  2. Manuals
  3. Tektronix 2445B Technical Manual

Tektronix 2445B Technical Manual

Made by: Tektronix
Type: Technical Manual
Category: Test Equipment
Pages: 18
Size: 0.59 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

• 

,  , 

, ;  

,:. 

T.O. 

33K3-4-2S90-1 

TECHNICAL MANUAL 

CALIBRATION  PROCEDURE 

FOR 

OSCILLOSCOPE 

2445B 

(TEKT  ONIXj 

CALlBRA.TION  LAB 

INSTRUCTiON  MANUAL FOR 

Model 

Cal Tech: 

DISTRIBUTION WITHIN GIDEP  IS AUTHORIZED  BY 

PEPARTMENT OF THE  AIR FORCE,HQ AEROSPACE 

GUIDANCE AND METROLOGY CENTER (AFMC) NEWARK 

AFB OHIO,  LETTER DATED 

AUGUST 

1992 

Distribution Statement·  Distribution  authorized to  U.S.  Government agencies and their contractors for 
official use  or 

for 

administrative or  operational  purposes only, 

April 

1990. 

Other  requests  for this 

document shall  be referred  to AGMC/MLEP,  Newark AFB OH 

43057-5475. 

15 APRIL 1994 

CHANGE 2 

-

30 SEPTEMBER  1995 

 


background image

T.O.  33K3-4-2590-1 

LIST OF EFFECTIVE PAGES 

INSERT LATEST 

C

HANGED PAGES.  DESTROY SUPERSEDED PAGES. 

linll in  the lluterJlUl1'!:'ins  of the page,  Changea toilluutrationo are 

indies.ted by miniature pointing handa. Changes to wiring 

diagramll 

are 

indlcated by shoded  areas. 

Dates of issue for original and changed pages are: 

Original 

..

.

.

.......

.

...

.

.

.

.

..

..

..

...

....

.

.

.

....... 

15 

April 

1994 

Change 

..................... 1 

.......

.

.

.... 

30 

September 

1994 

Change 

..................... 2 

........

.

..

.

30 

September 

1995 

TOTAL NUMBER OF PAGES 

1N 

THIS PUBLICATION  IS  16,  CONSISTING  OF THE  FOLLOWING: 

Page 

No. 

Change 

No. 

Title 

.............

.

.

.

.

.

.

.

..

.

.....

..

.

.

.

... 

A ......................................... 2 

1·2 

...

.

.....

.

...

..

.

.

..

..

..

.

.

.

....

.

....

. 0 

.......................................... 

.......

...

.

.

.

.

.

.

. .

.

.

.

.......

.

....

.

.....

.

.

. 2 

.

.

...

.

.

.

.

..

...

.

...

.

...

.

.......

.

.........

.

.

.

...

..

.

...

..

.

....

.

.

.

.

.

.

..

.

..

.

.

.

....

.

..... 

7·14 

.

.

............

.

.

.

....

.

....

.

......

.

-- �---o--

Page 

No. 

'" Change 

No. 

'"  Zero iD this culumn:indicnteo an  original page 

Page 

No. 

*' 

Change 

No. 

USAF 

 


background image

OSCILLOSCOPE 

2445B 

(TEKTRONIX) 

Test Instrument 

(TI) 

Characteristics 

Calibrator 

Vertical 

Deflection 

Delta  Volts 

Trigger  Level 

Rea dout 

Vertical  Ban dwi dth 

Channell an d  2 

compare d 

Chaonel  3 and 4 

Triggering 
(A&B) 

Table 1. 

Performance 

Range:  .4 V p-p into 1 M  Q 

stander d voltage 

Accuracy:  ±1 % 

Range: 

2 m V to 5 Vl div, CH 1 

and CH 2; .1 V an d  .5 V/ div, 
CH3an d CH4 

Accuracy: 

±2%, CH 1 an d CH 2; 

±10% , CH 3 an d  CH 4 

Range: 

Vl div setting 

Accuracy:  ± ( 1.25% ofr dg  +0.03 div) 

Range: 

18 

Vl div setting 

Accuracy:  ± (3% ofr dg  +  3% of p -p 

sigoai +0.2 div  +0.5 m 

V) 

+  (0.5 m VXProbe attenuation  factor) 

Range:  DC to  150 

MHz 

(200 MHz, 

SIN 

B 030000 and above) 

frequency. Vertical 

Accuracy: 

50 Q,  Down not more than  3 dB; 

1 M 

Down  not more than  4.7 dB 

Range:  DC to  150 MHz 

(200 

MHz, SIN 

B 030000 an d above) 

Accnracy:  Down not more than  4.7 dB 

without  Probe ; 3.0 

dB 

with  Probe 

Range: 

INT, 

1.5 div 

250 MHz; 

0.75  div 

250 MHz, CH 3 and CH 4 

Accuracy: 

Minimum 

T .O . 33K3-4-2590-l 

Test 

Method 

Compare d against a 

Checke d  with a stan dard 
voltage 

Apply a  constant amplitu de 

sigoai while  changing 

deflection 

against  deflection  at a 

reference d  frequency 

Apply mininonm signals an d 

check for  stable  display 

 


background image

• 

T.O.33K3-4-2590-1 

Test Instrument (TI) 

Characteristics 

Sweep Time 

(A&B) 

Delta  Time 

X-y 

agalnst 

reference d  

Parametric 
Measurements 

Table 1 (Cant.) 

Performance 

Range: 
A, 

10 

ns 

to 

.5 

sec/div 

B, 

10 

ns to 

50 

msl div 

Accuracy:  A: 

:t1.5% 

Accuracy: B: 

:t1.8% 

Magnifier: 

XlO 

Accuracy: 

:t2.0% 

Range: 

:tl0 

X A  sec!div 

Accuracy: 

:t(0.5% 

of time  interval 

:to.3% of 

full 

scale) 

Sensitivity Range: 

m V to 

V/ div 

Accuracy: 

:t2% 

Test 
Method 

Checke d  with s tan dar d 
Time  Mark signals 

Checke d  with a  stan dar d 

voltage 

Apply a  constan t  ampli tu de 

Ban dwi dth  Range:  DC 

to 

MHz 

signal while changiog 

frequency.  Horizon tal 

Accuracy:  Down not more  than 3 dB 

deflection  compare d 

Period Range: 

ns 

to 

100 

IDS 

Accuracy: 

0.5% +  0.5 

ns 

Frequency Range: 

Calculate d as  lIperiod 

Accuracy:  Same as period 

Vol ts  Range: 

:t 

p, POp  Average 

Average: 

5% 

of r dg 

+5 

m V  

+0.5 

m V  

Probe  attenuation  factor 

deflection  at a 

frequency 

Checke d  with  stan dar d 

Time  Mark signals 

Checke d  with a standar d 

vol tage 

Noun 

2.1 

CAL IBRATION 

Minimum Use 

Range: 

10 

m V 

to 

20 

V  POp 

Calibration 

Tek tronix 

Sub­
Item 

 


background image

T.O.33K3-4-2590-1 

GENERATOR 

PG506 

Accuracy:  ±0.25% 

P/O F5030A1 

Noun 

Minimum 

Use 

Calibration 

Sub-

Item 

2.2 

LEVELED SINE 

Range:  50 kHz 

to 

250 MHz 

Tektronix 

WAVE GENERATOR 

SG503 

Accuracy:  ±5% 

P/O F5030A1 

2.3 

FEEDTHROUGH 

Range:  50n 

Tektronix 

TERMINATION 

011-0049-01 

Accuracy:  N/A 

2.4 

TERMINATION 

ATTENUATOR 

Range: 

As 

Necessary 

As 

Available 

Accuracy:  N/A 

2.5 

ATTENUATOR PROBE 

Range:  XlO 

Tektronix 

P613lfP6136 

Accuracy:  N/A 

Supplied with TI 

2.6 

TIME 

MARK 

Range:  5 ns to .5 s 

Tektronix 

GENERATOR 

TG501 

Accuracy:  ±0.25% 

P/O F5030Al 

2.7 

METER  CALIBRATOR 

Range:  0 to .4 VDC 

Fluke 

5l00B 

Accuracy:  DC ±0.2% 

2.8 

PROBE 

TIP 

TO BNC 

Range:  N/A 

Tektronix 

013-0227-00 

N/A 

3.1  Review  and become familiar with the entire procedure before beginning calibration process. 

Unless otherwise designated, and prior to beginning the  Calibration 
Process, ensure that all test equipment voltage  and/or  current outputs 

are 

set to zero  (0) or turned off,  where applicable.  Ensure that all 

equipment  switches are  set to  the  proper position before making 
connections or applying power. 

3.2  Connect TI  and calibration equipment to  a  115 VAC/60 Hz power source.  Set  all POWER switches to 
ON and allow a 20 minute warm-up period.  After warm-up period, momentarily press  and hold both the 

CH 1 and CH 2 upper Input Coupling buttons 

until 

a moving dot display replaces the normal signal. 

This performs a  DC Balance of CH 1 and CH 2  and the readout indicates "DC BALANCE IN 

PROGRESS".  The signal and readout will return to normal when complete. 

3.3  Unless otherwise specified,  TI controls  should be set for several cycles  of stable display  with  ample 
amplitude. 

3.4  Set  TI controls as follows: 

BW LIMIT 

VAR  (all) 

CH 1 and CH 2  INPUT  COUPLING 

off 

cal (detent) 

IMnDC 

 


background image

• 

T.O. 33K3-4-2590-1 

VERTICAL  MODE 

CHI 

DELTA T 

off 

DELTA V 

off 

TRACKIINDEP 

INDEP 

A TRIGGER MODE 

AUTO 

SOURCE 

VERT 

SLOPE 

TRIGGER COUPLING 

DC 

A&BSECIDIV 

1 ms 

3.5  Use the Calibration  Generator positive going, fast rise, output terminated into 50 

at TI, set TI and 

Generator controls as required and  check TI high frequency compensation for abnormal  overshoot or 
aberrations.  Hquestions arise, Bee  Maintenance T.O. or Commercial 

Manual. 

Do  not attempt to 

optimize High-frequency compenstation adjustments unless the TI has failed bandwidth check. 

4  CALIBRATION 

NOTE 

Unless otherwise  specified, 

verify 

the 

results of each  test  and take corrective action  whenever  the  test 

requirement is not met,  before proceeding. 

4.1.1  Connect TI CALIBRATOR  to  the  CH 1 INPUT. 

4.1.2  Adjust TI  CH 1 VOLTSIDIV and VAR  controls for 4 divisions of vertical display centered  on the 
CRT. 

4.1.3  Disconnect TI CALIBRATOR from  the CH 1 INPUT. 

4.1.4  Connect the Meter Calibrator OUTPUT to TI CH 1 INPUT. 

4.1_5  Set the Meter Calibrator controls as follows: 

INTIEXT 

OPRJSTDBY 

LOCALIREM 

CLEAR 

INT 

STDBY 

LOCAL 

depress twice 

• 

4.1.6  Depress the Meter Calibrator switches in the following sequence:  0, V, ENTER, OPR. 

4.1.7  Adjust TI position control to align the Sweep on the second graticule below the center horizontal 

line. 

4.1.8  Depress the Meter Calibrator switches in the  following  sequence:  .4, V,  ENTER. 

4.1_9  Use the Meter Calibrator EDIT control to adjust the output for 4 divisions of vertical deflection. 

 


background image

4.1.10 

The Meter  Calibrator output readout  must indicate 

0.396 

to 

0.404. 

4.1.11 

Set the Meter Calibrator 

to 

STDBY and disconnect from TI. 

4.1.12 

Return TI CH 

VOLTSIDIV VAR control 

to 

the CAL (detent) position. 

T.O. 

33K3-4-2590-1 

 


background image

T.O. 33K3-4-2590-1 

4.2.1  Connect the Calibration  Generator to T I  CH  1 INPUT  and  switch on DELTA VOLTS.  BW LIMIT 

may be switched on for ease  of measurement. 

4.2.2  Set TI CH 1 VOLTSIDIV control to the fIrst position listed in the Range col

umn 

of Table  2. 

4.2.3 Set the Calibration  Generator  AMPLITUDE control to the fIrst position listed in the Applied 

column of Table  2. 

4.2.4  Adjust the Calibration Generator VARIABLE (OUT) control for the amount of vertical display 

listed in the Deflection col

umn 

of Table  2. 

4.2.5  The  Calibration  Generator readout  display must  indicate within  2%. 

4. 2.6  Set the Calibration  Generator VARIABLE (OUT) control to the 

IN 

position. 

4.2.7  Adjust TI Delta REF orDLY POS control to align the reference cursor to the bottom of the 

applicable waveform on the CRT. 

4.2.8  Adjust TI DELTA control  to align the delta cursor with the top of the  applicable waveform 

on 

the 

CRT. 

4.2.9  The Delta V readout must be within the values listed in the Readout Limits column of Table  2. 

4.2.10 Set TI TRIGGER  coupling to DC and SLOPE to positive.  Adjust  TI A  TRIGGER LEVEL control 

at the most positive voltage that produces a barely triggered jittering display. 

4. 2.11  The TI Trigger Level readings  must 

he 

within the values listed in the 

Peak Limits column of 

Table  2. 

4.2.12  Adjust TI A TRIGGER  LEVEL  control at the most negative  voltage  that  produces a barely 

triggered jittering display. 

4.2.13  The TI Trigger Level readings must 

he 

within 

the 

values listed in the  - Peak Limits  column of 

Table  2. 

4.2.14  Repeat steps  4.2.2 through  4.2.13 for the remaining  values listed in Table  2 .  

Thble 2. 

Range 

Applied 

Readout 

Trigger Level Readout 

(V 

Deflection 

(Limits) 

Peak 

·Peak 

2 mV 

10m 

9.81 to  10.2 mV 

8.0 to  12.0 m V 

+1.7 to -1.7 mV 

5 mV 

20m 

19.6 to  20.4 mV 

16.8 to  2 3.2 m V 

+2.6  to -2.6 m V 

10mV 

50m 

49.0 to 50.9  mV 

44. 0  to 56.0 m V  

+4.5  to  -4.5 mV 

20mV 

.1 

98.1 to  102.0 mV 

89 to  111 mV 

+8.0 to  -8.0 m V 

50mV 

.2 

196 to  204mV 

178 to  222mV 

+16.0 to -16.0 mV 

.1 

.5 

490 to 509mV 

450 to 550mV 

+ 35 to  - 35 mV 

6  Change  2 

 


background image

• 

T.O.  33K3-4-2590-1 

Table 2.  (Cont) 

Range 

Applied 

Readout 

Trigger Level Readout 

10m 

Deflection 

(Limits) 

Peak 

·Peak 

.2 

.981 

to 

1. 02  V 

.9 to  1.1 V 

+ 0.07 to - 0.07  V 

.5 

1.96 

to 

2.04 V 

1.78 to  2 .22 V 

+ 0.16 to  - 0.16 V 

4.90 

to 

5.09 V 

4.5 to 5.5  V 

+ 0.35 

to 

- 0.35  V 

10 

9.81 to 10.2 V 

9 tollV 

+O.7 to - 0.7  V 

20 

19. 6 

to 

2 0.4 V 

17.8 to  22.2 V 

+1.6 to -1.6 V 

4.2.15  Set the  Calibration Generator for a minimum  output.  Disconnect from TI CH  1 INPUT and 
connect to the CH  2 INPUT. 

4.2.16  Set  TI VERTICAL MODE to  CH 2. 

4.2.17  Repeat steps 4.2.2 through 4.2.14 utilizing CH 2 controls. 

4.2.18  Set the Calibration Generator for a minimum output.  Disconnect from TI  CH 2 INPUT and 
connect to the CH  3 INPUT. 

4.2.19  Set TI VERTICAL MODE and SOURCE to CH 3 and VOLTSIDIV (CH 3 

CH 4) to .l V. 

4.2.20  Set the  Calibration Generator AMPLITUDE control to .5 V. 

4 .2.21  The TI CRT vertical disp lay must be between 4.5 and 5.5 divisions. 

4.2.22  Set TI CH 3 VOLTSIDIV to .5  V. 

4.2.23  Set the Calibration Generator AMPLITUDE control to  2 V. 

4.2.24  The  TI CRT vertical display must be between 3.6 and  4.4 divisions . 

4.2.25  Set the Calibration Generator for a minimum  outp ut.  Disconnect from  TI CH 3 INPUT and 

connect to  CH 4 INPUT. 

4.2.26  Set TI VERTICAL MODE  and SOURCE to  CH 4. 

4.2.27  Repeat steps 4.2.20 through 4.2.24 utilizing CH 4 controls. 

4.2.28  Set the Calibration Generator for  a minimum output, disconnect from  TI, and switch oifthe Delta 

V. 

4.3.1  Set TI VERTICAL MODE to display CH  1 only and 

BW 

LIMIT to  OFF . 

• 

4.3.2  Set TI VOLTSIDIV (ALL) to .1 V, the INPUT COUPLING (both) to  1 M 

DC,  and the TRIGGER 

SOURCE to VERT. 

 


background image

T.O. 33K3-4-2590-1 

4.3.3  Connect the Leveled Sine Wave Generator #1 (2.2)  OUTPUT through the 50 n Feedthrough 

Termination to TI  CH 1 INPUT. 

4.3.4  Set the Leveled Sine Wave Generator frequency controls to  50 kHz .  Adjust the Output  controls for 
6 divisions of vertical display centered on TI CRT. 

4.3.5  Increase the Leveled Sine Wave Generator frequency controls nntil TI CRT vertical display 
decreases to  3.5 divisions. 

4.3.6  The Leveled  Sine Wave Generator frequency controls must indicate  150 MHz or more.  (200 MHz 
or more for SIN B030000 or more) 

4.3.7  Remove the  50 n Feedthrough Termination from the hook -up and set  TI CH 1 INPUT C OUPLING 
to 50nDC. 

4.3.8  Repeat step 4.3.4. 

4.3.9 Increase the Leveled Sine Wave Generator frequency controls until TI  CRT vertical display 

decreases to 4.2 divisions. 

4. 

3.10  Repeat step 4.3.6. 

4.3.11  Disconnect the  Leveled  Sine Wave Generator from TI CH 1 INPUT connector  and connect  to  TI 

CH 2 INPUT connector through the 50 n Feedthrough Termination. 

4.3.12  Set 

TI 

VERTICAL MODE to display CH 2 only. 

4.3.13  Repeat steps 4.3.4 through 4.3.10 utili zing  CH 2 controls. 

4.3.14  Disconnect the Leveled Sine Wave Generator from TI. 

4.3.15  Set TI VERTICAL M ODE to display CH 3 only. 

NOTE 

Use TI Calibration output to verify that the  Probe compensation, 

if 

supplied with TI, is properly adjusted. 

4_3.16  Remove the Grabber tip and plastic barrel from the XlO Attenuator  Probe.  Use the Probe Tip to 

BNC Adapter and connect the X10 Attenuator Probe to the Leveled Sine Wave Generator  OUTPUT. 

Connect 

the 

XlO Probe to the TI  CH 3 INPUT. 

NOTE 

If  the XI0 

is 

not available as required in step 4.3.16, then connect the 

Generator OUTPUT to the Attenuator (as required).  Connect the 

Attenuator to TI CH 3 INPUT through the 50 n Feedthrough 

Termination. 

4.3.17  Set the Leveled Sine Wave Generator frequancy controls to 50 kHz and adjust the Output controls 

for 4 divisions of verticai display centered on TI CRT. 

4.3.18  Increase the Leveled Sine Wave  Generator frequency control. until  TI CRT vertical display 
decreases to  2.8 divisions.  (2.3 divisions 

if 

the X10 Attenuator Probe is not available). 

4.3.19  The Leveled Sine Wave Generator frequency control must indicate  150 MHz or more. 

8  Change 2 

 


background image

T.O. 33K3-4-2590-1 

4.3.20  Disconnect the Leveled Sine Wave Generator from TI CH  3 INPUT and connect to TI  CH  4 

INPUT. 

4.3.21  Set TI VERTICAL MODE to display CH 4 only. 

4.3.22  Repeat steps 4.3.17 through 4.3.19 using the appropriate CH 4 contTol 

•. 

4.3.23  Set TI VOLTSiDIV (all) to  0.5 V CH 

1 & 

2 to  50 

DC, and the TRIGGER SOURCE to VERT. 

4.3.24  Remove the X10 Attenuator Probe and Probe to BNC  adapter.  Connect the Leveled Sine Wave 

Generator to the Attenuator (as required).  Connect the Attenuator through  the 50 

Feedthrough 

Termination to  TI CH  4 INPUT. 

4.4.1  Set the Leveled Sine Wave Generator frequency controls to  250 MHz.  (200 MHz  for SIN B030000 

and more).  Adjust 

the 

Output controls for 0.75 divisions (0.5 divisions for  SIN B030000  and above) of 

vertical display on TI CRT. 

4.4.2  A stable display on TI CRT must be observed by adjustiug the LEVEL and TIMEIDIV controls. 

4.4.3  Disconnect the Leveled Sine Wave Generator from TI CH 4 INPUT and connect to the CH 3 

INPUT. 

4.4.4  Set TI VERTICAL MODE and SOURCE to  CH 3. 

4.4.5  Repeat steps 4.4.1 and 4.4.2. 

4.4.6  Disconnect the Leveled Sine Wave Generator from TI  CH  3 INPUT and remove the 50 

load. 

Connect the Leveled Sine  Wave Generator to the CH 2 INPUT. 

4.4.7  Set TI VERTICAL MODE and SOURCE to  CH 2. 

4.4.8  Adjust the Leveled Sine Wave Generator output controls for 1.5 divisions of vertical  display on TI 

CRT.  (1.0 divisions for  SIN B030000 and above) 

4.4.9  Repeat step 4.4.2. 

4.4.10  Set TI Vertical MODE and SOURCE to CH  1. 

4.4.11  Disconnect the Leveled Sine Wave Generator from 

TI 

CH 2 INPUT and connect  to  CH 1 INPUT. 

4.4.12  Repeat  step  4.4.2. 

4.4.13  Disconnect the Leveled Sine Wave Generator from TI. 

4.5.1  Connect the  Time Mark  Generator  MARKER  OUT to 

TI 

CH 1 INPUT and switch on the DELTA 

TIME. 

4.5.2  Set TI A 

B SECIDIV (knobs locked) to the first position listed in the Range column of Table 3. 

4.5.3  Set the Time Mark Generator MARKER control to the first marker listed in the Applied column of 

Table  3. 

 


background image

T.O. 33K 3-4-2590-1 

4.5.4  Adjust TI hori zontal p osition to  align  the 2nd marker with  the  2nd vertical graticule line. 

4.5.5  Adjust the Time Mark Generator VARIABLE (OUT) control and TI horizontal position control for 

one marker per  division over the center eight divisions on the CRT. 

4.5.6  The Time Mark Generator readout must indicate within  1. 5 %. 

4.5.7  Push  in the Time Mark Generator VARIABLE (OUT) control button. 

4.5.8  Adjust TI REF or  DLY POS to  align the reference cursor with the  second time marker. 

4.5.9  Adjust TI DELTA control to align the delta cursor with the tenth  marker. 

4.5.10  The Delta  T readout must be within the corresponding 

limits 

listed in the Limits  col

umn 

of Table 

3. 

4.5.11  Repeat steps  4.5.2 through  4.5.10 for the remaining values  in  Table  3. 

A & B SEClDlV 

IOns 

20ns 

50ns 

.1 jlS 

_2  jlS 

.5 jlS 

IjlS 

2 ""  

IlS 

10 Ils 

2 0  "" 

50 "" 

.1ms 

.2ms 

.5ms 

Ims 

10 

Table 3. 

(marker) 

IOns 

20ns 

50ns 

.1 "" 

.2 "" 

.5 jlS 

IjlS 

2 ""  

5"" 

10 jlS 

20 Ils 

50 Ils 

.1ms 

.2  Ils 

.5ms 

1ms 

Limit. (readout) 

79. 30 to  80.70 ns 

158.6 0 to  16 1.40 ns 

396.5 to  403.5 ns 

79 3.0 to  807. 0  ns 

1586.0 to  1614. 0 ns 

3965 to  4035 ns 

7 .9 30 to  8.070 Ils 

15.86 to  16.14 

jlS 

39.65 to  40. 35 jlS 

79. 30 to  80_70 jlS 

158.60 to  161.40 jlS 

396.5 to  403.5 Ils 

7 9 3.0 to  807.0 

IlS 

1586.0 to  1614.0 "" 

396 5  to  4035 jlS 

7 .9 30 to  8 _070 

IDS 

 


background image

T .O . 33K3-4-2 590-1 

2 ms 

5mB 

10 IDB 

20ms 

A & B SeclDiv 

50ms 

'"  .1 s 

"'.2  s 

. 5  s 

A SECIDI V 

2 mB 

5ms 

lOms 

20ms 

Table 3.  (Cant.) 

(marker) 

50ms 

.1 s 

. 2  s 

.5 s 

4. 5.12  Set  TI A 

& B 

SECIDIV to  1 !Ls and  switch on the XlO MAG. 

NOTE 

15. 8 6  to 16.14 mB 

39. 65 to  40.35 

IDS 

7 9.30 to  80.70 

IDS 

158.60 to  1 61.40 

IDS 

Limits (readout) 

396.5 to  403.5 rns 

793.0 to  807 . 0  ms 

1578. 0 to  1622.0 

IDS 

3945 to  4055 

rns 

Use only that portion of the sweep that is in view after  the XI0 MAG has 

been set. 

4. 5.13 Set the Time Mark Generator MARKER control to .1 !Ls . 

4. 5.14  Adjust the Time Mark Generator 

ARlABLE (OUT) control and T I  horizontal position control for 

one marker per  division over  the  center  eight  divisions  on the CRT. 

1

4. 5. 1 5  The Time Mark Generator readout must indicate within 2.0%. 

4.5.1 6  Set TI control as  follows: 

XI0 

off 

trigger MODE 

RUNAFT DLY 

4.5.17  Repeat  steps  4. 5.2  through  4. 5. 6 utilizing the 

SECID IV (knobs unlocked) control for 

all 

SECIDI V settings  listed in the  Range column of Table  3 (the Time Mark  Generator  must indicate 

within  1.8 %  for 

sweep). 

4.5.18  Disconneat the Time Mark Generator from T I .  

4.6.1  Set T I  controls as follows: 

CH 2 

on 

ALL  OTHER CHANNELS 

off 

11 

 


background image

T.O. 33K3-4-2590-1 

VOLTSIDIV (CH  1) 

INPUT  (CH 

1) 

INPUT (CH 2) 

A&BSECIDIV 

.1 

1 MODC 

1M GND 

X-Y (knobs locked) 

4.6.2  Connect the Calibration Generator AMPL OUT to  the TI CH 

INPUT connector. 

4.6.3  Set the Calibration  Generator AMPLITUDE control to .5 V.  Adjust  the VARIABLE ( O UT) control 

for  5 divisions of horizontal display centered on TI CRT. 

4.6.4  The  Calibration  Generator Readout display must indicate within  2%. 

4.6.5  Disconnect the  Calibration Generator from  TI. 

4.6.6  Conoect the Leveled Sine Wave Generator #1 (2.2) OUTPUT through the 5 00 Feedthrough 
Termination to TI CH  1 INPUT connector. 

4.6.7  Set the Leveled Sine Wave  Generator frequency controls to  50 kHz.  Adjust the Output controls for 

8 divisions of horizontal display centered  on TI CRT. 

4.6.8  Increase  the Leveled Sine Wave  Generator frequency controls to  3 MHz. 

4.6.9  The TI CRT horizontal  display must be  5.6  divisions or more. 

4.6.10  Disconnect the Leveled Sine Wave Generator from  TI. 

4.7.1  Set 

TI 

controls  as follows: 

CH I 

ON 

ALL OTHER CHANNELS 

OFF 

CH  1 VOLTSIDIV 

200mV 

CH 

INPUT COUPLING 

5 0nDC 

A&B SECIDIV 

100 ns (knobs locked) 

4.7.2  Connect the  Time Mark  Generator MARKER OUT to the CH 1 INPUT. 

4.7.3  Set the Time  Mark  Generator controls for .1 

f1S 

markers. 

4.7.4  Set TI CH 1 VOLTSIDIV and TI CH 1 VOLTSlDIV VAR for 4 divisions or more of display. 

4.7.5  Push the 

MEASURE 

button. 

4.7.6  Select FREQ from the menu. 

4.7.7  The  TI PERIOD readout must indicate  99.0 to 1 01.0 ns. 

12 

 


background image

T.O. 33K3·4-2590-1 

4.7.8  The TI FREQUENCY readout mustmdicate  9.9 5 to 10.05 

MHz. 

4.7.9  Disconnect the Time Mark Generator from TI CH 1  INPUT. 

4.7.10  Set the TI controls as follows: 

CH 1  VOLTSIDN 

100mV 

A&B SECIDN 

1ms 

CH 1 INPUT COUPLING 

1 MODC 

4.7.11  Connect the Calibration  Generator 

to 

TI  CH 1  INPUT. 

4.7.12  Set the Calibration  Generator AMPLITUDE control for . 5  V. 

4.7.13  Push TIMEASURE button. 

4.7.14  Select  VOLTS from the menu. 

4.7.1 5  The TI CH 1  p-p  readout must mdicate  469. 5 

to 

530. 5 mY. 

4.7.16  Disconnect and secure all equipment. 

CA LIBRATION  PERFORMANCE  TABLE 

4.1  CALIBRATOR CALIBRATION: 

Limits 

.4 V p-p 

.4VDC 

0 .396 

to 

0.404 V p.p 

4.2 VERTICAL DEFLECTION CALIBRATION:  (both  channels) 

Range 

Applied 

Limi ts 

!Y:1l:P1 

Readout 

+ Peak 

- Peak 

2m 

10m 

9.81 

to 

10.20 mV 

8.0  to 12.0 m V 

+1.7 to -1.7 mV 

5mV 

20m 

19.6 to 20.4 m V 

16.8 to 23.2 mV 

+2.6  to ·2.6 m V 

10mV 

50m 

49.0 to 50.9 m V 

44.0 to  56.0 m V 

+4.5 

to 

-4.5 m V 

20mV 

.1 

98.1 

to 

102.0 m V 

89 to 111 mV 

+8.0  to -8.0 m V 

50mV 

.2 

196 

to 

204mV 

178 to  222 mV 

+16 to -16mV 

.1 

. 5  

490 

to 

509mV 

.450 

to 

. 550 V 

+.03 5 to -.03 5 V 

.2 

0.981 to 1.020 V 

.90 

to 

1 .10 V 

+.07 to - .07 V 

13 

 


background image

T.O . 33K3-4-2590-1 

.5 

1.96 

to 

2.04 V 

1. 78 to  2.22 V 

4.90 to  5.09 V 

4.50 to  5.50 V 

10 

9.81 to  10.2 V 

9.0 to  11.0 V 

20 

19.6 

to 

20.4 V 

17.8 to  22.2 V 

CALIBRATION P ERFORMANC E TABLE  (Cont) 

4.3  VERTICAL BANDWIDTH CALIBRATION: 

(both 

channels)  (Con t) 

(50n) 

DC 

to 

150 MH z  (200 MHz, 

SIN B030000 and shove) 

(MOl 

DC  to 150 M Hz  (200 MHz, 

SIN 

B030000 and shove) 

4.4  TRIGGERING CALIBRATION:  (A & B) 

CH1&CH2 

CH3&CH4 

4.5  SWEEP TIME CALIBRATION:  (A & B) 

IOns 

20ns 

50ns 

.1 

JlS 

.2 

JlS 

.5 

JlS 

14 

50kHz 
150 MHz  (200 MHz, 

SIN 

B030000 and above) 

50kHz 

150 MHz (200 

MHz, 

SIN 

B030000 and above) 

50kHz 
150 MHz  (200 MHz) 

SIN 

B030000 

and 

above) 

1.5 div 

250 MHz 

(1.0 div 

200 MH z, 

SIN B030000 and above) 

0.75 div 

250 MH z 

(0.5 div 

200 MHz, 

SIN B030000 and above) 

IOns 

20n8 

50ns 

.1 

JlS 

.2 

JlS 

.5 

JlS 

+.16 

to 

.-16 V 

+.35 

to 

-.35 

+0.7 to  -0.7 V 

+1.6 to -1. 6 V 

6 div  (referen ce) 
4.2 div or more 

6 div (reference) 

3.5 div or more 
CH 1&2 

4 div  (reference) 
2.8 div or more 
2.3 without Probe 

stable disp lay 

stable  display 

Lim its De lta  Time 

79.30 to  80.70 n s  

158.60 

to 

161.40 n s  

396.5 t o  403.5 n s  

793.0 to 807.0 n s  

1586.0 t o  1614.0 n s  

3965 

to 

4035 ns 

 


background image

1 �s 

2 �s 

5 �B 

10 �s 

10/lS 

20 fLs 

50/lS 

CALIBRA TION  PER FOR MANCE  TABLE  (Cont) 

4.5  SWEEP TIME CALIBR ATION: 

(A & 

B)  (Cont) 

.1ms 

.2mB 

.5ms 

1ms 

2ms 

5ms 

lOms 

20mB 

50 ros 

* .1 s 

*.2 s 

* .5 

1 �sX10 

SECID IV only 

4.6 X-Y CALIBRA TION: 

.1 VIDIV 

DC  to  3 MHz 

.1ms 

.2/lS 

.5 ms 

1mB 

2mB 

5ms 

10ms 

20ms 

50ms 

.1 s 

.2 s 

.5 B 

.1/lS 

.5 Vp-p 

50kHz 

3 MHz 

T .O .  33K3-4-2590-1 

7.930 to  8.070  �s 

15.86 to  16.14  /lS 

39.65  to  40.35 �s 

79.30 to  80. 70 �B 

15.86 to  161.4  /lS 

396.5 

to 

403.5 /lS 

Limits  Delta Time 

793.0 to  807.0 /lS 

1586 to  1614 /lS 

3965 to  4035  /lS 

7.930 to  8.070 

ms 

15.86 to 16.14 ms 

39.65 to  40.35 ms 

79.30 to  80.70 ms 

158.6 to  161.4 ms 

396.5 to  403.5 ms 

793.0 to  807.0 ms 

1578 

to 

1622 ms 

3945  to  4055 ms 

784.4 

to 

815.6 ns 

4.90 to  5.10  div 

8 div (reference) 

5.6  div or more 

15 

 


background image

T.O. 33K3-4 -2590-1 

4.7  PARAMETRIC  CALIDRATION: 

16 

TIME 

FREQ 

VOLTS 

.1 Ils 

.1 JlS 

.5 V 

Limits 

99.0 to  101 ns 

9.95 to  10.05 

MHz 

469.5 to  530.5 m V