Clear and Sanitize Procedures

Table 2: Nonvolatile Memory Devices (cont.)

Type and
minimum size

Function

User
modi

fi

able

1

Data input
method

Location

To clear

To sanitize

2

Real Time Clock

Stores time set by
user

Yes

UI

Main board

Set to GMT

Set to GMT

Serial EPROM

Diagnostic logs,
Main board Pix
Mon calibration,
and factory only
Looping diagnostic
mode control

No

Programmed by
software during
calibration and
diagnostics

Main board

(See page 5,

Clear

Diagnostic Log
Procedure

.)

Recalibrate Pix mon
to overwrite existing
values.

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

Serial Flash

Future expansion

No

Tested on
power up
Not used

Main board

None

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

Serial EEPROM

MAC address

No

Set at factory

Main board

None

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

1

During normal instrument operation.

2

Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on

the availability and cost of replacement parts.

Clear Presets Procedure

1.

Press and hold the

PRESET

button to display the preset con

fi

guration menu.

Then select

Recall Preset

>

Factory Preset

>

SEL

. This restores the presets

to the factory state.

2.

Press and hold the

PRESET

button. Then select

Save Preset

>

Select Group

A

. For all presets that are not empty “<e>”, select

Save

and then press

SEL

to overwrite that preset with the factory settings.

3.

Check each group and overwrite presets that are not empty as described in
the previous step.

4.

Navigate to the top level menu and select

Rename Preset

. Rename all named

groups and presets with the label “Default”.

4

1741C Declassi

fi

cation and Security