TABLE OF CONTENTS

INDICATIONS AND CONTRAINDICATIONS ................................................. 1 
WARNINGS AND PRECAUTIONS ................................................................ 2
PACKAGE CONTENTS ................................................................................ 6
STEP BY STEP OPERATION GUIDE FOR TRAINING
• Preparing Skin for Running a Session ................................................... 7
• Inserting 3 AAA Batteries ...................................................................... 7

• Setting up the shorts

 ............................................................................ 7

• Turning ON the Device ........................................................................... 8 
• Turning OFF the Device .......................................................................... 8
Selecting the Program .............................................................................. 9 
Selecting the Exercise Intensity Level ...................................................... 9 
SPECIAL FEATURES................................................................................. 10
CARE AND MAINTENANCE ...................................................................... 11
TROUBLESHOOTING ................................................................................ 13
WL-2413B STIMULATOR TECHNICAL SPECIFICATIONS .......................... 13
Warranty .................................................................................................. 14

Please read this Use and Care Manual

carefully before you start using your

TekSmart EMS Device.

Careful reading of the

"contraindications/warnings/precautions/adverse

reactions/caution with electrodes" sections of this

manual is strongly recommended. 

 


background image

1

INDICATIONS AND CONTRAINDICATIONS

 

Read the operation manual before using.

 

Read instruction manual before operation. Be sure to comply with all “CAUTIONS” and 
“WARNINGS” in the manual. Failure to follow instructions can cause harm to user or 
device.

Please read the following information carefully before using the TekSmart electronic 
muscle stimulator. 

Indications for use

The TekSmart Fitness EMS device model WL-2413B is intended for use by healthy
persons to apply transcutaneous electrical muscle stimulation(EMS) through skin
contact electrode for the purpose of improvement of muscle tone of the buttocks 
muscles.

Contraindications 

Do not use this system if any of the following conditions are present: 

•  Do not use this system if you have a cardiac pacemaker, implanted 

  defibrillators or any other implanted metallic or electronic device. Such use 

  could cause electric shock, burns, electrical interference, or death. 

•  Do not use this system if you have undiagnosed chronic pain.

•  Do not use in case of hemorrhagic diseases.

•  Do not use if you have any form of implant.

•  You are pregnant. The safety of electronic muscle stimulation over the 

  pregnant uterus has not been established. 

•  You suffer from cancer. The effects of electronic stimulation on cancerous  

  tissue are unknown. 

•  You are under medical supervision for cognitive dysfunction as you may not be  

  able to comply with safety instructions.

•  The unit is in close proximity to shortwave or microwave diathermy equipment

  or you are connected to high-frequency surgical equipment, because of risk of  

  device interference.

•  Wearing the device necessitates placement over areas at which drugs/medicines are  

 

  administered (short-term or long-term) by injection (e.g. hormone treatment). 

•  You have epilepsy.

•  You have recently undergone a surgical procedure.

•  Following acute trauma or fracture.

•  In case of critical ischemia of the limbs.

•  Persons with suspected or diagnosed heart problems should not use this device.

•  Persons with suspected or diagnosed epilepsy should not use this device.

•  Do not place less than 2 inches from joints.

•  Do not use in case of neuromuscular diseases.

•  Do not use on animals.

•  The short must not be connected to other objects.

•  Do not touch with your fingers the shorts interior while the unit is switched on.

•  Keep your unit away from heat source.

• Keep your unit away from short-wave emitting equipment or high frequency surgical devices your

  unit away from heat source.
•  Do not take your device system into a magnetic resonance (MR) environment device.

 


background image

2

WARNING AND PRECAUTIONS 

•  If you are under the care of a Physician, consult with your Physician before using  

  this system.

•  The long-term effects of this system are not known.

•  Do not place the pads on or close to your heart.

•  Do not place the pads around or close to your neck. Do not

  apply stimulation over  the neck. Severe spasm of the muscles  

  may occur and the contractions may be strong enough to close  

  the airway or cause difficulty in breathing. Stimulation over the  

  neck could also have adverse effect hearing or blood pressure.

•  Do not apply stimulation across the chest because the introduction of electrical

  current into the chest may cause rhythm disturbances to the heart.

•  Do not place the pads on or around your head. The effects of stimulation of the brain 

  are unknown.

•  Do not use the pads over or close to sores.

• Do not place the pads on the front or sides of the neck across or through the heart  

  (one pad on the front of the chest and one on the back), in the genital region, or on  

  the head, because of the risk of stimulating inappropriate muscles and organs.  

•  Do not place the pads over any recent scars, broken or inflamed areas of infection or  

  susceptibility to acne, thrombosis or other vascular problems (e.g. varicose veins), or 

  any part of the body where feeling is limited.

•  Do not place the pads over areas of injury or restricted movement (e.g. fractures or 

 sprains).

•  Do not use this system while sleeping.

•  Do not use if you feel numbness.

•  Do not use this system in or close to water. 

•  Do not use the pads over or close to cancerous lesions.

•  Do not use the device in the following circumstances:

  -when there is a tendency to bleed internally following an injury of fracture.

  -following recent surgical procedures when stimulation may disrupt the healing    

 process.

  -over areas of skin that lack normal sensation.

•  Use the pads only on normal, healthy, clean and dry skin. Do not use the pads on   

  open wounds or rashes, or over swollen, red, infected or inflamed skin.

•  If you have ever had back surgery, consult your Physician before using this system.

•  Position the pads and operation the unit ONLY as indicated in the Quick Start Guide.

•  Avoid areas of injury or restricted movement (e.g. fractures or sprains)

•  Avoid placing the pads over metal implants.

•  Do not use in the bath or shower, or in an environment of elevated humidity (e.g.   

  Sauna, hydrotherapy, etc). 

• Stimulation should not be applied over the carotid sinus nerves, particularly in 

 

  patients with a known sensitivity to the carotid sinus reflex.

Wait before using next until:

•  At least 6 weeks after the birth of your baby (you must consult your doctor

  before use).

•  Do not to use this device if you have an IUD contraceptive device as the safety of the  

  use of this device when used with an IUD has not been established.

•  At least 3 months after having a caesarean section (consult your doctor before use).

•  The heavy days of your period have finished, because vigorous abdominal exercise is  
  not recommended at this time. 

 


background image

3

Precautions

 

 

•  Read User Manual before using this system for the first time.
•  Keep this manual available whenever you use your system.
•  The system is intended for personal use on healthy adult muscle only. 
•  The safety of using the system during pregnancy or birth has not been established.
•  The effectiveness of the system depends greatly on a person’s individual physical 
  condition. It may not always be effective for every user.
•  The safety of neuromuscular stimulation during pregnancy has not been established.
•  Use caution when/if:
  - User has skin areas that lack normal sensation.
  - Following surgical procedures if muscle contractions might impede healing process.
  - Over a menstruating or pregnant uterus.
  - There is a tendency to hemorrhage following acute trauma or fracture.
•  Place electrodes in accordance with illustrations in the Quick Start Guide.
•  This unit should not be used while driving, operating machinery or during any activity  
  in which involuntary muscle contractions may place the user at undue risk of injury.
•  Some users may experience skin irritation or hypersensitivity due to the electrical  
  stimulation or the conductive medium.
•  Keep the stimulator out of the reach of children.
•  Application of moderate heat (thermal wrap) to muscles as well as moistening skin  
  prior to treatment improves treatment efficacy; use of cold packs on treated muscles  
  after treatment is likewise recommended.
•  This unit should only be used with the leads, electrodes and accessories provided by  
  the manufacturer.
•  The device is not intended for medical use, for the treatment of any medical condition  
  or for any permanent physical changes. 
•  Contact TekSmart Fitness if your unit is not working correctly. Do not use in the  
 meantime. 
•  An effective session should not cause discomfort.
•  For first time users, muscle stimulation can be an unusual sensation. We recommend  
  that you begin in a seated position with low stimulation intensity settings to 
  familiarize yourself with the sensation before progressing to higher intensity settings. 
•  The leads and pads must not be connected to other objects.
•  Do not over exert yourself while using muscle stimulation. Any workout should be at a  
  comfortable level for you. 
•  Do not place pads over jewelry or body piercings.
•  Start all sessions in a sitting position 
 

 


background image

4

Use Caution and consult your physician before using the system if any of the 
following conditions apply to you:

•  You have any serious illness or injury not mentioned in this guide.
•  You have recently undergone a surgical procedure.
•  You take insulin for diabetes.
•  You use the unit as part of a rehabilitation program.  
•  If you have suspected or diagnosed heart problem.
• If you have suspected or diagnosed epilepsy.
• If you have a tendency to bleed internally following an injury.
• If you recently had surgery, or have ever had surgery on your back.
• If areas of skin lack normal sensations, such as skin that tingles or is numb.
• During menstruation or during pregnancy.
• Some people may feel skin irritation or experience a very sensitive feeling in the skin  
  due to electrical stimulation. If this occurs, stop using your system and consult your  
 physician.
• If skin under one of more pads feels irritated after using the stimulator for a long   
  period of time, use the stimulator for a shorter period of time.
•  Minor redness at stimulation placement is a normal skin reaction. It is not considered  
  as skin irritation, and it will normally disappear within 30 minutes after the electrodes  
  are removed. If the redness does not disappear after 30 minutes from the removal of  
  electrodes, do not use the stimulator again until after the excessive redness has 
 disappeared.
•  Turn off the stimulator if the stimulation feels unpleasant or does not provide
  pain relief.
•  Keep your system out of the reach of children.
•  Use your stimulator only with the pads, cables and accessories recommended by  
  the manufacturer.
•  Do not use this System when driving, operating machinery or when swimming.
•  Before removing the garment and pads, be sure to power off device to avoid
  unpleasant stimulation.

After strenuous exercises or exertion:

•  Always use lower intensity to avoid muscle fatigue.

Important:

 

 

•  Do not use your unit at the same time as any other device which transfers an
  electrical current into the body ( e.g. another muscle  stimulator).
•  Cease using your unit if you are feeling light headed or faint. Consult doctor if this  
 happens.
•  Do not touch the pads or metal studs while the unit is switched on. 
•  Do not use unit on your abs if you are wearing a belly button ring. Remove ring  
  before session.

Note: If you are in any doubt about using device for any reason, please consult your doctor before use.

 


background image

5

Pad/Electrode Precautions

•  To reposition the pads during a session, always pause the program currently running,  
  reposition the pads and then restart the program again.  
•  Only use pads supplied by the manufacturer with your device. Any others may not  
  be compatible with your unit and could degrade the minimum safety levels. 
•  The pads are for single person use only.
•  Do not plunge the pads into water while attached to the muscle stimulator.
•  Do not apply solvents of any kind to the pads.
•  Always ensure the unit is turned OFF before removing the pads.
•  Apply the whole surface of the pads firmly to the skin. Do not use pads which are
 damaged.
•  If your skin is red under the pad after a session, do not start another session in the  
  same area until your redness has completed disappeared. 

Adverse Reactions

•  You may experience skin irritation and burns beneath the stimulation electrodes 
  applied to your skin. 
•  You may experience headache and other painful sensations during or following the  
  application of electrical stimulation near your eyes and to your head and face. 
•  You should stop using the device and should consult with your physician if you 
  experience adverse reactions from the device. 

Conditions that may affect your TekSmart EMS System 

Since the stimulator is a battery-operated electronic system, its output performance 
and safety may be affected greatly in extreme humidity. Therefore, it is very important 
to keep the stimulator dry to ensure the safety and performance of the stimulator.

Additional Warnings

 

 

•  No modification to the equipment is allowed.
•  Remove the batteries if the equipment is not to be used for some time.
•  The electrode pad is the applied part of the device.
•  Simultaneous connection of a patient to a high frequency surgical ME equipment may  
  result in burns at the site of stimulator electrodes and possible damage to the    
 stimulator.
• Do not use this device directly on the eyes, covering the mouth and the front of the  
 neck.
• This device has IP2X degree of protection(Protection against penetration of solid   
  foreign objects of 12.5mm diameter and greater).

Condition (Operating/Transportation/Storage)

Item 

Operating

Transportation

Storage

Temperature 

5-40˚ C

41-104˚ F 

10-40

50-104˚ F

0-45

C

32-113˚ F

Humidity 

30-85% 

30-75% 

20-90%

 


background image

6

Package Contents: 

Device Features

1.  Power on/adjust/increase setting key
2.  Power off/adjust/decrease setting key
3.  CH1 key
4.  CH2 key
5. Program/Pause
6.  Program number
7.  Time remaining

8.  CH1 intensity level
9.  CH2 intensity level
10. Lock status indicator
11.  Battery status indicator
12. Pause indicator
13. Battery compartment
14. Battery cover

88

88

88

Pi8

PAUSE

CH1

CH2

MIN

MODE

CH1

CH2

ON

+

OFF

3- AAA Batteries

EMS Device

Lead-Wires

4 - Silicone Electrodes

3.4 oz. Tube

Conductive Gel

Conductive

Gel

3.4   oz.

The Curvassure™ Butt Toning Shorts 

25

25

30

7

 


background image

7

Inserting/Changing the Batteries

1.  Open the battery compartment at the back of the device by pushing the battery  

 

  cover labeled “Open” downward (this area features  

 

  raised marks for easy identification).

2.  Insert 3 AAA (1.5 V) batteries in the 

 

  battery compartment; make sure to match 

  up the symbols (+/–).

3.  Close the battery cover by carefully placing the   

  stud into the slot in the rear area and  

  sliding it upward, applying slight pressure.

4.  Follow the same procedure when replacing the batteries as needed.

  Always read and follow the specific instructions of the battery manufacturers.

Note:

 Always use only 3x1.5V(AAA) batteries.

•  Keep away from children.

•  Do not recharge.

•  Do not short-circuit.

•  Do not throw into a fire.

•  Please recycle. Do not dispose of old batteries with your household waste; dispose of   

them safely at your recycling centre or business where the batteries were purchased.

Preparing the Skin for a Session

Proper preparation of the skin covered by the electrodes allows more stimulation to reach 

targeted tissues, prolongs electrode life, and reduces the risk of skin irritation. Wash the area 

with mild soap and water (don't use alcohol). Rinse and dry thoroughly.

Make sure that your skin is clean and free of oil

Setting-up the Shorts

1. Turn the short inside out and lay on a flat surface.

2. Bend the electrode slightly and insert into the pockets with the gray side down and black  

  side up. Insert one electrode in each of the electrode holders,sliding the electrode under the  

  pocket lip.

3. Insert pin-end of the four lead-wires through the eyelet hole on the outside of the 

  shorts. This can be done on the left-side or right-side depending on which pocket you

  prefer to put the device in during your session.

4.  Slide one Blue tip wire over and out of the upper right eyelet hole. Slide the other Blue tip   

  wire down and out of the lower right eyelet hole. Slide one Black tip wire out of the  

  upper left eyelet hole. Slide the other Black tip wire down and out of the lower left  

  eyelet hole.

5. Insert the pin-end of the lead-wire snugly into the electrode. Repeat for all wires. 

  Make sure the wires are completely inserted into the electrode. 

6. Apply a small ( dime size) amount of the conductive gel on the electrodes with your    

  finger, smear the gel around covering entire surface of the electrode.

7.  Turn the short right side out and slide onto your body, pulling them up from the waistband.  

  Try to keep the electrodes away from contacting your gluteus until the shorts are 

  completely on. (In the event that some gel gets on your leg in the process of pulling the  

  shorts up, simply wipe it off with a cloth.)

The electrodes will need to be touching your bare skin in order to create a contraction. You may 

want to wear a thong if you do not feel comfortable wearing the compression shorts without 

undergarments. You might need to adjust the position of the garments slightly until you find 

the most comfortable position. However the positioning will become much easier after a few 

sessions. For reasons of Hygiene, please do not share Well Life Buttock Muscle Stimulator and 

its accessories with others, for single person use only !

 


background image

8

Turning On the Device

1.  Press and hold the ON+ button for one 
  (1) second to turn on the device. 
2. The most recently selected program will
  flash when the unit is turned on. 

 

Note: Do not turn the unit on until the

          shorts are set up and ready to use.

 

Turning Off the Device

1.  The device turns off automatically after 
  the exercise session time has elapsed.
2.  To turn the unit off manually, press 
  the OFF- button for three (3) seconds. 
  The display will go blank and the device 
  will turn off.
3.  In an emergency you may also pull the 
  connector wire(s) from the device. 

 

Note: To prevent unpleasant electric 

  shocks, never remove the device while 
  it is still turned on.

  

Selecting the Program

The Device offers three different preset training programs. The programs differ with 
respect to varying pulse widths, frequencies and ON/OFF time. 

1.  Press the MODE button. The "P" # will  
  be flashing. Press ON + or OFF –
  repeatedly until the desired program  
  flashes on the display. Press MODE to  
  save your selection.

25

25

30

25

25

30

Programs

Program  

Mode

Pulse Width 

(uS) 

Frequency 

(Hz) 

Ramp Up 

(sec)

On Time 

(sec)

Ramp Down 

(sec)

Off Time

(sec)

P1

Beginner

300

50

2

5

2

9

P2

Intermediate

300

60

2

4

2

8

P3

Advanced

300

70

2

3

2

7

Selecting the Contraction Intensity Level

The Device offers maximum 25 intensity level

1. Intensity is adjustable .To increase or decrease the
  intensity, press ON + (to increase) or OFF – (to
  decrease) repeatedly until the desired intensity
  level flashes on the display.

Note: You will feel the intensity increase or decrease as 
you select the intensity level. You can use this as a guide
to select a level that is comfortable for you.

2. The intensity level can only be adjusted when the
  system is set and the electrodes are in contact with
  the skin.

3. The intensity level will reset to 0 each time the
  device turns off, including after training sessions ,  
  or changing the Program during training session.

Your Contraction Intensity Level

For those not familiar with the sensation of ems it can 
be a little strange. As you increase the intensity you 
will initially feel a tingle in the area under the pads. 
Gradually increase the intensity until you feel your 
muscles contracting. As the intensity is increased more muscle fibers are stimulated 
and you will feel a strong contraction and be able to see muscle movement. The 
intensity should be set where you feel a good contraction and it is not uncomfortable. 
As your muscles adapt you will be able to increase the intensity more in order to get a 
deeper contraction.

Important Information:
The Intensity Level for this muscle stimulator can only be adjusted when the electrodes 
with conductive gel have good contact with the skin.

25

25

30

 


background image

9

Programs

Program  

Mode

Pulse Width 

(uS) 

Frequency 

(Hz) 

Ramp Up 

(sec)

On Time 

(sec)

Ramp Down 

(sec)

Off Time

(sec)

P1

Beginner

300

50

2

5

2

9

P2

Intermediate

300

60

2

4

2

8

P3

Advanced

300

70

2

3

2

7

Selecting the Contraction Intensity Level

The Device offers maximum 25 intensity level

1. Intensity is adjustable .To increase or decrease the
  intensity, press ON + (to increase) or OFF – (to
  decrease) repeatedly until the desired intensity
  level flashes on the display.

Note: You will feel the intensity increase or decrease as 
you select the intensity level. You can use this as a guide
to select a level that is comfortable for you.

2. The intensity level can only be adjusted when the
  system is set and the electrodes are in contact with
  the skin.

3. The intensity level will reset to 0 each time the
  device turns off, including after training sessions ,  
  or changing the Program during training session.

Your Contraction Intensity Level

For those not familiar with the sensation of ems it can 
be a little strange. As you increase the intensity you 
will initially feel a tingle in the area under the pads. 
Gradually increase the intensity until you feel your 
muscles contracting. As the intensity is increased more muscle fibers are stimulated 
and you will feel a strong contraction and be able to see muscle movement. The 
intensity should be set where you feel a good contraction and it is not uncomfortable. 
As your muscles adapt you will be able to increase the intensity more in order to get a 
deeper contraction.

Important Information:
The Intensity Level for this muscle stimulator can only be adjusted when the electrodes 
with conductive gel have good contact with the skin.

25

25

30

25

25

30

25

25

30

 


background image

10

Special Features

Exercise Time

 

30 minutes. 
Time will countdown on the display in 1-minute increments for the duration of
your session.

•  The device turns off automatically when the exercise time has elapsed.
•  The last exercise program you used will appear on the display, when you turn on 
  the device.
•  To change the program, press ON + or OFF – repeatedly until the desired program  
  appears on the display.
•  Press MODE to save your selection. The program you selected will appear on the  
  display the next time you turn on the program.

Lock Function

 

Press and hold the ON + and OFF – keys simultaneously for 1 second to lock/unlock 
the device.

Pause

Press MODE for a two (2) seconds and PAUSE will flash on the display.
Press MODE again, PAUSE will disappear on the display and session will continue.
The intensity level will stay at your current setting and time will continue to count down.

Automatic Shut-off

•  The device automatically turns off when no button is pressed for 60 seconds.
•  The device automatically turns off when the time for your exercise session 
  has elapsed.

Intensity Level Reset

For your safety and comfort, the intensity level will reset to 0 each time the device turns 
off, including after exercise sessions.

The session will discontinue if the wires are not correctly connected to the electrodes. 
The intensity level will reset to “0” any changes are  made the mode setting during 
exercise session.

Low Battery Status Indicator

The battery status indicator will light whenever the battery is low.  
This means that you will have to replace the batteries soon.

 


background image

11

Care and Maintenance

Muscle Stimulator

 

The stimulator may be wiped clean with a small amount of soapy water on a clean cloth. 

Do not submerge the stimulator in liquids or expose it to large amounts of water.

•  Never use aggressive cleaning products or stiff brushes to clean the device.

•  Remove the batteries before cleaning the device.

•  Do not use the device again until it is completely dry.

•  Do not expose the device to direct sunlight.

•   Protect the device from dirt and moisture.   

Short

•  For hygiene reason, the short is reusable for single person use.

Cleaning instruction:

•  Wash at 86º (30º C), low intensity wash, normal rinse and reduced spin cycles.

•  Do not bleach.

•  Do not iron or steam.

•  Do not dry clean or use stain removal solvents.

•  No tumble drying. The shorts should be air-dried, lying flat or hanging.

Cables

Disconnect the cables from the stimulator when not in use. 

Do not pull on the cables but rather on the plugs attached to the ends of the cables. 

Store the stimulator with the cables in a clean, dry place. 

Electrodes

 

•  The electrode pads are conductive silicone pad, reusable for single person use .

•  Take off from the garment after each use, wipe off and dry the electrodes with a  

  soft cloth dampened with soap and water.

How to Store Your System

1.  Store your system at room temperature in a dry place, out of the reach of children.

2.  If the stimulator will not be used for more than a week, remove the batteries.

 

Replacement Parts

Go to www.contouroutlet.com to order replacement parts. Be sure to use only
Curvassure Butt Toning System parts :
• Silicon Pad : Part # SP-95D ( 4 pcs/pack )
• Conductive gel : Part # JT-001 (3.4 fl oz /100 ml)
• Lead Wire : Part # LE-WS-042
• Shorts: There are 4 sizes available Small (S), Medium (M), Large(L) , Extra Large (XL).

Size

US Size

Waist

Hips

S

4-6

26”- 27”

34.5”- 36”

M

8-10

28”- 30”

38.5”- 39.5”

L

12-18

32”- 37”

41”- 46”

XL

20-22

38”- 43”

47”- 50”

 


background image

12

Troubleshooting

Always check the unit and accessories before use to prevent damage and defects.
1.  Make sure the batteries have sufficient charge and are not corroded.
2.  Make sure the cables fit tightly into the connection sockets of the device.
 

If you cannot remedy the defects  as described below, contact Customer Service.

Problem

Possible Cause

Solutions

The device does not 

turn on

No battery or bad battery

Replace batteries

The device turns on

and then off again

Battery not inserted properly

Battery life expired

Insert batteries again or 

replace batteries

The device turns on,

but does not generate 

electric pulses

Intensity is too low

Increase the intensity

Lead-wire broken

Replace lead-wires

Lead-wires not connected 

properly

Reinsert lead-wires

correctly

Treatment time has expired

Switch unit OFF and  back ON

The battery power is low

Replace batteries

Unpleasant feeling beneath the pads

Poor pad contact

Ensure the pads have a thin layer 

of gel on them and they are flat 

against your skin

The unit does not turn on even though

new batteries have been inserted

Device may be faulty

Contact customer service

13

Guidance and manufacturer’s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environment specified below. The customer 

or the user of the device should assure that it is used in such an environment.

Emissions

Compliance

Electromagnetic environment-- guidance

RF emissions CISPR 11

Group 1

The EQUIPMENT uses RF energy only for its internal 

function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

RF emissions CISPR 11

Class B

The EQUIPMENT is suitable for use in all establish-

ments including domestic establishments and those 

directly connected to the public low-voltage power 

supply network that supplies buildings used for 

domestic purposes.

Harmonic emissions 

IEC 61000-3-2

Not ap-

plicable

Voltage fluctuations/ 

flicker emissions IEC 

61000-3-3

Not ap-

plicable

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the 
device should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment 

-guidance

Electrostatic 

discharge (ESD) IEC 

61000-4-2

± 6 kV contact 

± 8 kV air

± 6 kV contact 

± 8 kV air

Floors should be wood, con-

crete or ceramic tile. If floors 

are covered with synthetic 

material, the relative humidity 

should be at least 30 %.

Electrical fast 

transient/burst IEC 

61000-4-4

± 2 kV for power 

supply lines ± 1 kV 

for input/output lines

± 2 kV for power 

supply lines ± 

1 kV for input/

output lines

Mains power quality should 

be that of a typical commer-

cial or hospital environment.

Surge IEC 61000-4-5

± 1 kV line(s) and 

neutral to line(s)

± 2 kV line(s) to 

earth

Mains power quality should 

be that of a typical commer-

cial or hospital environment.

Conducted RF IEC 

61000-4-6

3 Vrms

150 kHz to 80 MHz

3 Vrms

150 kHz to 80 

MHz

Interference may occur in the 

vicinity of equipment marked 

with the following symbol:

Radiated RF IEC 

61000-4-3

3 V/m

80 MHz to

2.5 GHz

3 V/m

80 MHz to

2.5 GHz

NOTE: U

T

 is the a.c. mains voltage prior to application of the test level

 


background image

Troubleshooting

Always check the unit and accessories before use to prevent damage and defects.
1.  Make sure the batteries have sufficient charge and are not corroded.
2.  Make sure the cables fit tightly into the connection sockets of the device.
 

If you cannot remedy the defects  as described below, contact Customer Service.

Problem

Possible Cause

Solutions

The device does not 

turn on

No battery or bad battery

Replace batteries

The device turns on

and then off again

Battery not inserted properly

Battery life expired

Insert batteries again or 

replace batteries

The device turns on,

but does not generate 

electric pulses

Intensity is too low

Increase the intensity

Lead-wire broken

Replace lead-wires

Lead-wires not connected 

properly

Reinsert lead-wires

correctly

Treatment time has expired

Switch unit OFF and  back ON

The battery power is low

Replace batteries

Unpleasant feeling beneath the pads

Poor pad contact

Ensure the pads have a thin layer 

of gel on them and they are flat 

against your skin

The unit does not turn on even though

new batteries have been inserted

Device may be faulty

Contact customer service

13

WL-2413B STIMULATOR TECHNICAL SPECIFICATIONS

Channel:

 Dual, isolated between channels.

Pulse amplitude:

 Adjustable 0 – 80mA peak into 500

Ω

 load each channel.

Pulse Rate:

 As pre-programming operation mode.

Pulse Width:

 As pre-programming operation mode.

Software ramp up feature:

 Pulse width ramp up when change mode.

Timer:

 30 minutes.

LCD:

 Show modes, pulse rate, pulse width, timer, CH1/CH2, intensity level, pause.

Function

Mode

Pulse Width 

(uS) 

Frequency 

(Hz) 

Ramp Up 

(sec)

On Time 

(sec)

Ramp Down 

(sec)

Off Time

(sec)

P1

300

50

2

5

2

9

P2

300

60

2

4

2

8

P3

300

70

2

3

2

7

Wave Form:

 Symmetrical Bi-Phasic square pulse.

Max charge per pulse:

 24 micro-coulombs maximum.

Power Source:

  3x1.5V(AAA) / 4.5 Volt batteries 

All electrical specifications are ±20% at 500

Ω

 load.

This symbol means “ Serial 
number “
      
  

These symbol mean 
“ Attention, consult 
the accompanying 
documents 

 

    

 

This symbol means
“ Manufacturer “ 
 

  

 

 

This symbol means type BF 
equipment; this device offers 
protection against electrical 

shock by stand compliance to leakage 
currents.
 
  

This symbol means please apply 
regulations enforced in your country   
when disposing your batteries and 
product at its end of life Ensuring it is  

disposed in an environmentally friendly 
manner.

Description of symbols : There are a number of technical symbols on your ems device explained as follows:

 


background image

WARRANTY

This TekSmart muscle stimulator device carries a one-year warranty from the date of 
purchase.

The warranty does not apply to damage resulting from failure to follow the operating 
instructions, accidents, abuse, alterations or disassembly by unauthorized individuals.

The warranty applies to the main device and necessary parts and labor relating thereto. 
Battery, lead wires, electrodes, and other accessories are warranted to be free from 
defects in workmanship and materials at the time of delivery.

The system accessories - short and conductive gel, for reasons of hygiene, they can 
neither be taken back nor exchanged once used. 

The distributor reserves the right to replace or repair the unit at their discretion.

HOW TO GET HELP

To get assistance or answers to

your questions, please contact:

TekSmart Fitness, LLC

Visit our website for updates and more information.

www.curvassure.com

Distributed by:

TekSmart Fitness LLC

Minneapolis, MN 55402

Made in Taiwan 

© 2013 TekSmart Fitness LLC, All Right Reserved

Printed 12/2013          WL2400APX Rev B          IM-24-033 RevB 13