1. Before You Begin
This guide provides basic guidelines to assist you in quickly getting started.
Installation of the transmitter in an explosive environment must be undertaken in accordance with the
appropriate local, national, and international standards, codes, and practices. Review the approvals
section of the Tek-Cor 1100A reference manual for any restrictions associated with a safe installation.
2. Unpack
Tek-Cor 1100A Coriolis Mass Flowmeter x 1
3. Dimensional View
Note:
For additional dimensions refer to the detailed manual.
Quick Start Guide
Quick Start Guide
Quick Start Guide
Quick Start Guide
796 Tek Drive, Crystal Lake, IL 60014 USA
+1 847 857 6076 | +1 847 655 7428
+1 847 655 6147
www.tek-trol.com
Micro-Bend
4"
(0.33 ft)
(3.70 ft)
(2016 ft)
L
L1
H
H1
Integrated
44.48"
8"
18"
16"
11.02"
36.02"
33.85"
Triangle-Shaped
(0.08 ft)
(1.33 ft)
(0.91 ft)
L
L1
H
H1
Integrated
(0.08 ft)
(0.66 ft)
(1.5 ft)
18"
(1.5 ft)
(0.91 ft)
1
1.02"
4"
1"
(4.25 ft)
(1.16 ft)
L
L1
H
H1
Integrated
U-Shaped
(0.33 ft)
(3 ft)
36.02"
(3 ft)
(2.82 ft)
33.85"
50.78"
14"
(2.82 ft)
Size
Size
Size
33.85"
25.98"
(2.82 ft)
14.56"
(1.21 ft)
8"
1"
(0.66 ft)
4"
(0.33 ft)
(
0.33 ft
)
4"
(
3.70 ft
)
44.48"
33.85"
14.56"
(
2.82 ft
)
(1.21 ft
)

Tek-Cor 1100A
Please use the operation panel of the transmitter to set the configuration, such as basic configuration parame-
ters, zero calibration, and cut-off value of low flow and output range of current frequency.
The panel of the transmitter is shown as below:
4. Display
No.
Notes
1
2
3
4
5
E key : Enter
key : Move Cursor or Return
key : Page Down
Light for Working Status
Two Line LED
5. Key Functions
Operation point of Photoelectric Key is located right behind the glass panel. It is better to operate the
Photoelectric Key in a vertical direction, not a horizontal direction.
Key
Menu State
Function State
Date State
Measurement State
Show the Measurement
Result and State on Page
1/2/3 of the Display.
Page Down to Menu State.
Return to the Upper-Level
Menu, Press the Key Several
Times to Return to the
Measurement State
Save the Input Select Yes
or No then Back to Function
Menu
Change Number
Change Unit
Change Character
Next Menu
Return to the Last Screen
Enter the Menu
Next Function
Select Function
Move the Cursor Right
Confirm and Save the
Function
E
-
Technology Solutions
1
2
3
5
4
Indication of Title
Indication of Units
Indication of Numeric Value

Quick Start Guide
Quick Start Guide
6. Sensor Mounting
Vertical Mounting
Recommended Horizontal Installation
FL
OW
Position A
Position B
Micro-Bend Sensor Mounting
U-Shaped Sensor Mounting
Vertical Mounting
Recommended Horizontal Installation
FL
OW
Position A
Position B
If the medium contains solid particles, mount the meter as shown in position “A”, and in all other cases mount
the meter as shown in position “B”. Vertical Mounting is recommended, if the medium contains gas bubbles and
solid particles.
Install the sensor and the transmitter on the pipeline.
The transmitter can be mounted with 90 ° revolutions
depending on the requirement of sensor installation.
Integral Mount
7. Transmitter Mounting
796 Tek Drive, Crystal Lake, IL 60014 USA
+1 847 857 6076 | +1 847 655 7428
+1 847 655 6147
www.tek-trol.com

Tek-Cor 1100A
Remote Mount
When Tek-Cor 1100A is mounted separately, the sensor and the transmitter should be connected through
special 9-core cable.
Line No.
Line Color
Function
1
2
2
Brown
Red
The Left Coil +
The Left Coil -
Blue
Driving Coil -
Temperature +
7
8
9
Grey
White
Black
Temperature -
Temperature Compensation
Orange
4
5
6
Yellow
Green
3
Cut off power before connecting cables. The power voltage must match that indicated in the junction
box of the transmitter and the earth connector must be well connected with the earth wire to ensure
its correct safety performance.
Note:
In case of remote mounting, the sensor and the transmitter should be respectively matched with junction
box for connecting the special 9-core cable.
The Left Coil +
The Left Coil -
Driving Coil +
Brown
Red
Orange
Yellow
Green
Blue
Gray
White
Black
Shielded all
wires
Brown
Red
Orange
Yellow
Green
Blue
Gray
White
Black
Technology Solutions

Quick Start Guide
Quick Start Guide
8. Power Supply and Signal Output Wiring
The transmitter can be connected to 85 to 220 VAC or 18 to 28 VDC power supply.
AC Power Wiring
DC Power Wiring
Current Output Wiring
Pulse Output Wiring
RS485 Output Wiring
3 2 1
3 2 1
-DC18-30U+
Power
Wiring
±20
RS 485 - DC18-30U +
F+ F- I+ I- B A
3 2 1
3 2 1
AC85-265U
Power
Wiring
Power
Wiring
±20
RS 485 - AC85 - 265U
F+ F - I+ I- B A L N FG
L N FG
4-20
RS 485
F+ F - I+ I- B A
4-20
RS 485
F+ F - I+ I- B A
4-20
RS 485
F+ F - I+ I- B A
Power
Wiring
Before powering up the flowmeter in a dangerous area, please confirm that the flowmeter’s explo-
sion-proof class is consistent with the environmental requirements in order to avoid potential danger.
Make sure that the power is shut off to avoid electric shock when installing the transmitter. Follow the
installation and operation instruction manual to ensure safe operation.
9. Grounding
Both the sensor and the transmitter have to be grounded correctly; failure to do so may cause a measurement
error and the sensor may not work. If the pipeline is connected with the ground, the transmitter can be ground-
ed through the pipeline; if the pipeline is not connected with the ground, the transmitter should be grounded
independently.
796 Tek Drive, Crystal Lake, IL 60014 USA
+1 847 857 6076 | +1 847 655 7428
+1 847 655 6147
www.tek-trol.com

Tek-Cor 1100A
ing
10. Installation
Micro-Bend Series Installation
U-Shaped Series Installation
Picture-1
Picture-2
For Liquid or Slurry or Gas
For Liquid or Slurry
For Gas
For Liquid or Slurry or Gas
For Liquid or Slurry
For Gas
Picture-3
Picture-1
Picture-2
Picture-3
For the horizontal installation, the measuring tube should be installed on the downside of the pipeline when the
process medium is liquid or slurry (shown in Picture 1); on the upside of the pipeline when the process medium
is gas (shown in Picture 2). For vertical installation, the measuring tube should be installed alongside the pipe-
line when the process medium is liquid, slurry, or gas (shown in Picture 3).
A
B
C
The meters must not be mounted at the highest
point of the tubing (A), if gas bubbles are expected.
If solid particles expected, then it should not be
installed at the lowest point (B).
The meters must not be mounted in a drop line near
the open end (C), as this may cause the meter to run
empty.
Technology Solutions

Quick Start Guide
Quick Start Guide
There is a flow arrow which indicates the flow direction in front of the sensor; install the Tek-Cor 1100A according-
ly; failure to do may result in the transmitter not displaying the mass flow correctly. If the process medium is
liquid or slurry, the flow direction is down-up; if the process medium is gas, the flow direction can be either
down-up or up-down.
Note:
It is recommended that sensor of Tek-Cor 1100A is supported using a rubber connector as the buffer.
11. Flow Direction
12. Calibration
The Tek-Cor 1100A does not need field calibration; it has been calibrated before delivery. Each set of Tek-Cor
1100A has its own instrumental coefficient, including one flow coefficient and four density coefficients (high
density D1, high period K1, low density D2, and low period K2), which will be shown in the nameplate of the
Sensor or Calibration certificate.
Tek-Cor Mass Flowmeter
Model 1100A
Size
Accuracy
Connection
IP Class
Supply Voltage
Ambient
SN
Ex Marking
Tag No.
Flow Range
MMQ
Nominal Pressure
Temperature Range
Flow Cal
D1
K1
D2
K2
0.0012
3663.21
0.9965
3975.42
915.740 G/S/US
Class 1, Div.1,Gr.B,C,D
2-2-025A-1-1-150-1-G
1"
0.2
1.5
IP67
24 VDC
0.4A
-4F
to 131F
V17101601T5562
2017.06
www.tek-trol.com
Crystal Lake, IL60014, USA
NTEP CC NO.45-033
E360841
37 to 881 lb/min
14 lb.
928 PSI
-58F to 257F
GPS NJ,USA
GPE 1100A ENGL
ANSI 300 RF
Material (Wetted)
316 SS
796 Tek Drive, Crystal Lake, IL 60014 USA
+1 847 857 6076 | +1 847 655 7428
+1 847 655 6147
www.tek-trol.com

Tek-Cor 1100A
Zero Calibration
Zero Calibration provides the reference point for the flowmeter. It is necessary to conduct the zero calibration
whenever the Coriolis Meter is installed.
For Zero Calibration, follow the following procedure after installation:
• Warm-up the flowmeter for 30 minutes.
• Once the warm-up is complete, continue the flow through the flowmeter until the temperature of the
flowmeter is same as the working temperature of the fluid.
• When the flowmeter has reached the required temperature, close the downstream valve, and make the
fluid pass through the flowmeter under normal temperature, density and, pressure. Then close the
upstream valve to ensure the sensor is full of liquid during the process of Zero Calibration.
• Carry out the Zero Calibration of the flowmeter.
Zero Correction
Flow Configuration
Configuration
Zero-Cal
E
Finally, press
Input password to start zero calibration.
The default password is 000000
Note:
Each Zero Calibration lasts 30 seconds and must be repeated at least 10 times.
LED Indication
Note:
For additional information refer to the detailed manual.
LED Condition
Working Condition
Green Light
Normal Operation
Red Light
Error
13. Error Indication
The user can detect a fault in the flowmeter using the LED indicator and LCD displays. Different colors of the
LED lights indicates the working condition of the flowmeter, while the LCD displays can show the self-diagnos-
tic alarm information of the transmitter, which is useful for detecting the malfunctions.
Technology Solutions

Quick Start Guide
Quick Start Guide
14. Menu Tree
796 Tek Drive, Crystal Lake, IL 60014 USA
+1 847 857 6076 | +1 847 655 7428
+1 847 655 6147
www.tek-trol.com
Mass x.xxx t
Flow x.xxx t/h
Volm x.xxx m3
Flow x.xxx m3/h
Den x.xxx g/mL
Temp x.xxx ºC
Configuration
Press E to enter
Calibration
Press E to enter
Inner Para
Press E to enter
System Para
Press E to enter
Password?
000000
Flow Config
Press E to enter
Unit Config
Press E to enter
Output Config
Press E to enter
Other
Press E to enter
Change Password
Press E to enter
Add-ons Function
Press E to enter
Zero Calibration
x.xxx us
Low Flow Cut off
xxx.x kg/h
Reset Totalizer
Press E to enter
Mass Unit
t
Mass Flow Unit
t/h
Volume Unit
m3
Pulse Weight
x.xx
Pulse Output
Mass
20mA Value
xxx.x
Enter Old Password
000000
Enter New Password
000000
OW AnalyseS witch
OFF
20 °C Oil Den g/mL
x.xxx
20°C Water Deng/mL
x.xxx
New Password Again
000000
FlowCal g/s/us
xxx.xx
Temp Correction
x.xx
H.DensityD2 g/mL
x.xxx
H. Period K2 us
xxxxxx.x
Temperature Cal
x.xxxx
Temperature Zero
x.xx
Density Cal
x.xxxx
Density Zero
x.xx
Freq xx.xxx Hz
Time x.xxx us
Drv %
L R V
Pred xxxxx.x us
Ind xx.x °C
PUL: xxxx.x Hz
CUR: xx.x mA
E
E
Output AUTO Sim
Press E to enter
Flow Simulate t/h
x.xx
1
2
3
4
5
6
7
86
8
9
10
14
13
18
17
16
46
5
4
26
25
24
89
88
87
12
54
53
63
62
61
71
70
69
79
78
77
60
68
76
Flow Sim Start
Press E to enter
Pressure Comp
Press E to enter
Press Comp Mode
None
Flow Press Comp Coe
x.xx
Input Press MPa
xx.xx
11
34
32
33
44
52
41
40
39
Gas compensation
Press E to enter
Gas CompenSwitch
OFF
Gas Compen K1
X.XXXX
Gas Compen K2
X.XXXX
110
109
108
15
E
E
E
E
E
E
E
E
E
E
E
E
E
E

Tek-Cor 1100A
Flow Direction
Fwd
Mass Inventory t
x.xxx
VolmInventory m
3
x.xxx
Volume Flow Unit
m
3
/h
Density Unit
g/mL
Temperature Unit
°C
Current Output
Mass
MODBUS Address
xx
Baud Rate
9600
PWD Succeed
New PWD 000000
Gas Contain Corre
x.xx
W Contain Corre
x.xx
L.DensityD1 g/mL
x.xxx
L. Period K1 us
xxxxxx.x
Density Temp Comp
x.xx
Zero Correction
0
Menu Language
Eng.
Mutipoint Correct
OFF
Restore Para
Press E to enter
Correction Point
x
Water x.xxx g/mL
Oil x.xxx g/mLL
Density Range
OFF
Upper Den g/mL
x.xxxx
Lower Den g/mL
x.xxxx
21
20
19
36
48
47
37
28
27
29
90
58
57
56
55
67
66
65
64
72
83
82
81
80
59
Parity Bit
None
30
Pressure Comp Coef
73
Stop Bits
One
35
First Menu
Mass
Check System State
OFF
Flow Filtering s
1.5
31
49
YN En kg c 1. 16
9600, 8, N, 1 t 2. 20
74
No X Time Value
xxx.x
84
Other
75
Den Filtering s
3.0
50
No X Correct Value
x.xxxx
85
Expect Water Cut
X.XX
111
E
E
Technology Solutions

Quick Start Guide
Tek-Cor 1100A
796 Tek Drive
Crystal Lake, IL 60014
USA
Tel : +1 847 857 6076, +1 847 655 7428
Fax : +1 847 655 6147
Email: tektrol@tek-trol.com
www.tek-trol.com
Technology Solutions