Full Text Searchable PDF User Manual

1
AFA 500
FUME HOOD ALARMS
Operating and Instruction Manual
Model AFA 500 – Mk2
•
Built-in or Remote sensor
•
2 Relay inputs
•
1 Relay output
•
Com port
Used for alarm indication and
monitoring on Fume Hoods
Issue: Jan 08

2
LED indicators
Function buttons Calibration.
ENTER – also used as
Mute button for audible alarm
OPERATOR DISPLAY PANEL
Built-in AIRFLOW SENSOR
Not visible when Remote
AIRFLOW SENSOR is used

3
Connection details :-
Com port
RS232
Night set-back
Relay input
Flying lead with plug
connection for PCB
15 V DC
Flying lead with plug
connection for PCB
and SM6 airflow sensor
12V DC
Flying lead from Proxy
switch with plug
connection for PCB
12V DC
Sash High
Proximity switch
( Optional )
Remote
SM6 Airflow Sensor
( Optional )
R1
OPTIONAL
REMOTE
SENSOR
SOCKET
15V DC
POWER
SUPPLY
INPUT
Sash Proximity
switch
OUTPUT RELAY
COMMS
PORT
INPUT 1
Bn
Bk Bl
Sash
Micro
Switch
REFER TO
INSTALLATION
INSTRUCTIONS
BEFORE
CONNECTING
MADE IN ENGLAND
www.tel-uk.com
WARNING
TEL
PSU
Relay output 2
Rated 110V
2A AC1
Sash High
Micro-switch
( Optional )
Plug – in power adapter for
mains socket.
(To suit local requirements )
Standard options :-
230V 50Hz -- UK 3-pin
220V 50Hz -- Euro 2-pin
120V 60Hz -- USA 2-pin
AFA 500
Notes:-
1.
The socket for the remote SM6 Airflow Sensor is not provided as standard but is available on
request. The unit operates with the in-built sensor
OR
the remote sensor.
2.
The Sash High alarm can operate with the Micro switch
OR
the Proximity Switch.
Both connection points are available as standard.

4
1.1 General Description
All systems comprise of the following components :-
1 – AFA 500 Alarm unit,
1 – AC power supply
If the Sash Alarm System option is ordered there will also be a sash micro switch or proximity
switch.
If the Remote Airflow Sensor is ordered there will be remote sensor provided
Operator Features
--- the alarm has the following operator features :-
Safe LED - Green LED (
Not flashing
)
will be displayed if the airflow is greater than the
Low air alarm point.
Alarm LED - Red LED (
Not flashing
)
will be displayed if the airflow is lower than the
Low air alarm point.
Sash High
-
Red LED
(Flashing
)
will be displayed when the Sash is raised above
the max safe working opening.
Audible Alarm
-- the Audible alarm will sound ( can be muted ) in the Air Fail and the
Sash High alarm condition
Night Set-back
-- when the Night Setback input is activated the Audible alarm will be muted
and the
Green LED will flash on/off
ENTER
--- the alarm has an Enter button -- this is multi-functional as follows :-
Press
Enter
momentarily when Low Air alarm is sounding will mute the alarm
Press
Enter
momentarily when Sash High alarm is sounding will mute
the alarm and initiate a repeat timer that will re-sound the alarm if the Sash is
not lowered to a safe position before the end of the time period
Press
Enter
for 5 secs will gain access to
Calibration
mode
SET
--
used during the airflow Calibration of the alarm

5
External Connections
-- the alarm unit will have the following connection points :-
Input 1
--- volt free relay input – ( close contact to activate the input )
This input is configured as :-
NIGHT SETBACK
Output R1
--- volt free relay output - ( contact closes on activation )
This input is configured as :-
LOW AIR ALARM
Sash High Input
--- a. Connection point for Sash High micro switch. ( Switch contact to
close and remain closed in Sash High condition )
b. Connection point for Sash High proximity switch. ( Switch contact to
close and remain closed in Sash High condition )
Note:- Use input a. OR input b. for the Sash High alarm
Remote Airflow Sensor --
Plug in connection for airflow sensor
Note:- The remote Airflow Sensor socket will only be available for
units ordered without the built-in Airflow Sensor
Com Port
--- to enable connection to Laptop or PC.
Power supply
--- low voltage DC power supply 15V DC

6
1.2 Alarm Configuration / Calibration
The alarm is supplied with a factory configuration.
The only part of the configuration that can be changed is the setting for the Sash High repeat
timer time delay and the percentage figure for the Low Air alarm point when using the two point
calibration mode-- if required these can be changed by connection to a Laptop or PC via the
RS232 com port.
The alarm has
two modes for the calibration of the airflow alarm point
. These two modes are
selectable via a
‘jumper connection’
on the back of the alarm PCB. The jumper is labelled as
CAL
and is situated on the edge of the PCB above the Com port.
The two calibration modes are :-
a. Single Point Calibration ( with the Jumper connected )
- The airflow is reduced
mechanically ( using balance damper or slowing down the extract fan) to the Low Air alarm
point and this airflow is ‘captured’ by the alarm. The airflow is then restored to the normal
operating value and the Low Air alarm will activate if the airflow subsequently falls to the
alarm point
b. Two Point Calibration ( with the Jumper not connected)
- This is a two point
calibration method. The airflow is set to the normal operating value and this value is
‘captured’ by the alarm and taken to be 100%. The airflow is then increased to one and a half
times the normal operating value and this value is ‘captured’ by the alarm and taken to be
150%. The Low Air alarm will then activate if the airflow subsequently falls to 80% of the
normal operating value.
NOTE :-
for fume hoods where it is not possible to use the side wall sensor connection or LEV
extract systems the sensor can be connected directly to the extract duct using adaptors and a
¼
inch PVC tube – see installation section below for details. Only the Single Point Calibration
method can be used for this configuration
See ‘Quick Start Installation’ below for details of the Calibration procedures
1.3 Start up
When unit is powered up the following sequence of events occur :-
1. The 12V DC power is applied to the airflow sensor and the alarm then performs a self test
on the functions , LEDs and audible alarm ( approx 2 sec ) and then initiates a delay timer to
allow the airflow sensor to stabilise.
2. During the whole of thedelay period all alarms and relay outputs are inhibited and the Red &
Green LEDs will be permanently ON.
3. At the end of the delay period the unit performs one of two options :-
a.
If the alarm calibration has been previously completed – the unit goes to
normal operating mode ( Run )
b.
If the unit has not been calibrated -- the Red & Green LEDs will flash on/off and
the audible alarm will be muted. It is then possible to press the Enter button for
5 secs and go into the calibration mode – (See ‘Quick Start Installation’ below for
details of the Calibration procedures)

7
1.4 Events / actions
Safe airflow
•
Airflow above alarm level ( eg > 80 fpm )
•
Green LED on
Low airflow
•
Airflow below alarm level for longer than the low air delay time ( 5 secs)
•
Red LED on ( Not flashing )
•
Audible alarm sounds ( ‘Beep’ on/off every 1 sec ) -- can be muted via Enter pushbutton
•
Low air relay R1 operates.
Reset : -- when airflow rises above Low air level for longer than the low air to
safe air delay time ( 2 secs ) the Low air alarm resets automatically
Sash High
•
When the input configured as Sash High is activated ( Micro switch or Proximity switch )
•
Red LED on ( Flashing )
•
Audible alarm sounds ( ‘Beep’ on/off every 1 sec )
•
Audible can be muted via Enter pushbutton -- this silences the alarm and initiates a repeat
timer ( factory set to 5 mins ) After the delay time the alarm re-sounds ( and can be re-
muted). During this time the Red LED flashes on / off.
Reset: -- when Sash lowered to safe position and input de-activated.
Night set-back
•
When input configured as Night set-back is activated
•
Green LED on ( Flashing)
•
Audible alarm muted
Airflow Sensor Error
•
The connection and each element of the airflow sensor are monitored at all times. In the
event of a problem with the sensor the audible alarm will sound using a different
and distinctive tone best described as ‘ a modulating siren effect’
•
The audible alarm can only be silenced by re-connecting the sensor ( if a remote sensor is
being used ) or switching the power to the unit off.
2.1 Quick Start Installation

8
Follow the instructions below for installing and commissioning the unit. :-
Before connecting the unit it is important to decide on the type of calibration that is to be used on
the installation. The two methods are described in section 1.2 Alarm Configuration / Calibration
above.
1. Fit the alarm to the Fume Hood using the cut-out details provided with the unit
--- see page 11to 13
2. If using a remote airflow sensor fit to the Fume Hood using the cut out and installation
details provided --- see page 12 and connect the ‘telephone style’ airflow sensor plug-in
cable to the sensor and the back of the alarm unit --- see typical connection diagram on
page 14 & 15
3. Plug in the power adapter to a 120V AC power socket and connect the flying lead to the
alarm unit --- see typical connection diagram on page 14 & 15
4. Power up the unit and wait at least 30 secs while the sensor temperature stabilises.
During this time the Red & Green LEDs will both be on ( not flashing )
If the unit has not been previously calibrated the Red & Green LEDs will begin to flash
on/off at the end of the 30 sec start up time delay but the audible alarm will not sound.
If the alarm has been calibrated it will go into normal operation.
5.
Calibration :-
Single Point Calibration
(
with CAL jumper on PCB
connected
)
a. Open the Sash to the normal operating height and adjust the face velocity to the
Low Air alarm value
using a calibrated instrument to check the value.
b. Press and hold the Enter button for 5 secs to go into the Calibration mode. This is
indicated by both Red and Green LEDS flashing on/off together will the audible
alarm sounding ( ‘Beep’ on/off 4 times every 1 sec ).
c. To initiate the alarm point calibration press and hold the ENTER and the SET
button at the same time. The unit will then sample the airflow for a 5 sec period
during which time the Green LED goes off and the Red LED flashes on/off. The
audible alarm continues to sound during this period and if the sampling is
successful will give a two tone beep at the end of the period and the unit will then
go automatically into the Run mode.
If the buttons are released during the sampling period or if the airflow is fluctuating
more than the pre-set value the audible alarm will give a lower frequency buzzing
sound for a short period and then go back into the calibration mode. If this occurs
re-press the ENTER and SET buttons to repeat the airflow sampling.
d. When complete re-set the airflow to the normal value and the unit will go to the
Safe running condition with the Green LED on.
The unit will now function and go into the ALARM condition if the Fume
Hood face velocity falls below the alarm value.
Use this method if the alarm is connected directly to the extract
duct via the adaptors and
¼
inch tube
Calibration continued:-
Two Point Calibration
(
with CAL jumper on PCB
not
connected
)

9
a. Open the Sash to the normal operating height and adjust the face velocity to the
normal operating value
using a calibrated instrument to check the value.
b. Press and hold the Enter button for 5 secs to go into the Calibration mode. This is
indicated by both Red and Green LEDS flashing on/off together with the audible
alarm sounding ( ‘Beep’ on/off 4 times every 1 sec ).
c. To initiate the normal airflow (100%) calibration press and hold the ENTER and the
SET button at the same time. The unit will then sample the normal airflow for a
5 sec period during which time the Green LED goes off and the Red LED flashes
on/off. The audible alarm continues to sound during this period and if the sampling
is successful will give a two tone beep at the end of the period and the unit will
then go automatically into the higher airflow calibration mode.
If the buttons are released during the sampling period or if the airflow is fluctuating
more than the pre-set value the audible alarm will give a lower frequency buzzing
sound for a short period and then go back into the calibration mode. If this occurs
re-press the ENTER and SET buttons to repeat the airflow sampling.
d. The higher airflow calibration mode is indicated by both Red and Green LEDS
flashing on/off together with the audible alarm sounding ( ‘Beep’ on/off for 2 secs
every 10 secs ).
Close the Sash down until the face velocity rises to one and a half times the
normal operating value used in a. above using a calibrated instrument to check
the value
To initiate the higher airflow (150%) calibration press and hold the ENTER and the
SET button at the same time. The unit will then sample the higher airflow for a
5 sec period during which time the Green LED goes off and the Red LED flashes
on/off. The audible alarm continues to sound during this period and if the sampling
is successful will give a two tone beep at the end of the period and the unit will
then go automatically into the Run mode.
If the buttons are released during the sampling period or if the airflow is fluctuating
more than the pre-set value the audible alarm will give a lower frequency buzzing
sound for a short period and then go back into the calibration mode. If this occurs
re-press the ENTER and SET buttons to repeat the airflow sampling.
The unit will now function and go into the ALARM condition if the Fume
Hood face velocity falls below 80% of the normal operating value.
2.2 Calibration Notes :-

10
1.
The two calibration methods are intended to give the installer two options for
calibrating the alarm.
2.
Using the Alarm Point Capture method it is necessary to adjust the face velocity on
the fume hood using a mechanical damper ( or fan speed control if available) to the
desired alarm point and this is sampled by the unit. It is then necessary to re-adjust
the face velocity back to the normal operating value. This method produces a very
accurate alarm point at a fixed value but involves getting access to the ductwork or
fan speed controller
3.
The second method of Two Point Capture is slightly more involved but does not
require any access to the ductwork or fan speed controller and is generally a quicker
solution.
The principal behind this method is as follows :-
The output from the airflow sensor is linear over the normal operating range of the
face velocities on fume hoods. If we therefore capture the airflow sensor output at
two known points we can then calculate the alarm point and give an alarm when the
airflow falls below this value. However because the AFA 500 does not have a digital
display of the face velocity we can not enter particular values into the alarm. For this
reason we take the normal operating value of the face velocity with the sash open to
the max safe working opening and call this 100% -- this point is captured during the
first sample. Then by closing the sash it is possible to create a face velocity which
is one and a half times the normal operating value ie 150% -- this point is captured
during the second sample. We then use these two values to calculate the sensor
output at 80% and this becomes the Low Air alarm point.
The actual value of the normal operating face velocity is not important and in each
case the alarm point will be 80% of this value.
For example:-
A normal face velocity of 100 fpm would give an alarm point of 80fpm
A normal face velocity of 80 fpm would give an alarm point of 64 fpm
4.
The face velocity readings on the open sash may vary at different points on the
measuring grid by up to 20 fpm. This is quite acceptable in terms of the fume
cupboard performance so long as no individual point is below the designated Low
Air alarm point .The figure entered for the calibration point can be taken as the
average value of all the measuring grid readings or could be taken as the individual
lowest point on the grid. For most fume hoods this low point is on the bottom row in
the centre and is a convenient position to measure and for future reference when
checking the alarm during annual maintenance.
5.
Take time when measuring the face velocities for the calibration procedure to allow
for the velocities across the open sash to stabilise. If the velocities are changing or
are turbulent during the sampling period the alarm will detect this and give a low
frequency buzz at the end of the sample indicating that the sample must be
repeated.
.
3.0 Dimensions

11
When using the Duct Connection method the
¼
inch connection tube acts as a flow restrictor ( due to the higher
pressure in the duct ). Use a 6ft tube for pressures below 0.4 ins Wg and a 12ft tube for higher pressures

12
4.0 SM6 Airflow Sensor Installation
It is very important to position the AFA 500 or SM6 airflow sensor in the correct position to give long term stable reading
of the face velocity. Please read the INSTALLATION NOTES below and if in doubt contact us for further advice.
INSTALLATION NOTES :-
1. The AFA 500 or SM6 sensor must be positioned where it can " see " the room pressure of the laboratory.The back
connection spigot of the sensor is designed to accept a 25mm OD tube which should be connected to the inner
chamber of the fume hood. (This tube and fittings is known as the " vent kit " )
The
ideal position
for the end of the 25mm tube for most fume hoods is 100mm back from the sash glass
and 100mm higher than the normal sash opening height through the inner side wall.
2. If possible mount the sensor on the front of the fume hood and use a short length of tube. Tube lengths of
more than 1metre or smaller diameter will restrict the airflow through the sensor. This will lead to too much
sensitivity being required to calibrate the unit which can lead to some instability of the reading or incorrect
readings at low velocitiesl.
3. For fume hoods with a single skin side wall or a double skin with a small gap between them it may not be
possible to achieve the ideal sensing position using a flexible tube.
With a single skin side wall it is possible to fix the sensor on the outside of the fume hood and connect directly
to the inner chamber in the ideal position. This method can only be used for up to two fume hoods when they
are positioned side by side ( using the two outer walls ).
An alternative method is to mount the sensor on the front of the fume hood and connect using a short flexible
tube to a rigid wall tube attached to the inner side wall. The open end of this rigid wall tube should be positioned in
the ‘ideal position’ ie. 100mm back from the sash and 100mm higher than the normal sash opening.
Fume hoods with a high internal height can present a difficulty because the tube length to reach the ideal
position may be longer than 1 metre. In this case it is better to use a tube no longer than 1 metre which may result
in a sensing position higher than the ideal.
When fitting a sensor to a ’narrow wall’ fume hood for the first time it may be necessary to try various
combinations of rigid and flexible tube to find the best combination and position.
4. The sensor should not be mounted in a position were it is subject to draughts from the laboratory air input or
AFA 500 or
SM6 Sensor
Open to the
laboratory
Fume cupboard
outer skin
Fume cupboard
inner skin
25mm Tube
25mm
Female adapter
25mm
Male
bush
AFA 500
OR
SM6 Sensor

13
ventilation system.
5.0 Typical Wiring Diagram
Standard Alarm with built-in Airflow Sensor
Com port
RS232
Night set-back
Relay input
Flying lead with plug
connection for PCB
15 V DC
Flying lead from Proxy
switch with plug
connection for PCB
12V DC
Sash High
Proximity switch
( Optional )
R1
OPTIONAL
REMOTE
SENSOR
SOCKET
15V DC
POWER
SUPPLY
INPUT
Sash Proximity
switch
OUTPUT RELAY
COMMS
PORT
INPUT 1
Bn
Bk Bl
Sash
Micro
Switch
REFER TO
INSTALLATION
INSTRUCTIONS
BEFORE
CONNECTING
MADE IN ENGLAND
www.tel-uk.com
WARNING
TEL
PSU
Relay output 2
Rated 110V
2A AC1
Sash High
Micro-switch
( Optional )
Plug – in power adapter for
mains socket.
(To suit local requirements )
Standard options :-
230V 50Hz -- UK 3-pin
220V 50Hz -- Euro 2-pin
120V 60Hz -- USA 2-pin
AFA 500

14
5.1 Typical Wiring Diagram
Alarm with REMOTE Airflow Sensor
Com port
RS232
Night set-back
Relay input
Flying lead with plug
connection for PCB
15 V DC
Flying lead with plug
connection for PCB
and SM6 airflow sensor
12V DC
Flying lead from Proxy
switch with plug
connection for PCB
12V DC
Sash High
Proximity switch
( Optional )
R1
OPTIONAL
REMOTE
SENSOR
SOCKET
15V DC
POWER
SUPPLY
INPUT
Sash Proximity
switch
OUTPUT RELAY
COMMS
PORT
INPUT 1
Bn
Bk Bl
Sash
Micro
Switch
REFER TO
INSTALLATION
INSTRUCTIONS
BEFORE
CONNECTING
MADE IN ENGLAND
www.tel-uk.com
WARNING
TEL
PSU
Relay output 2
Rated 110V
2A AC1
Sash High
Micro-switch
( Optional )
Plug – in power adapter for
mains socket.
(To suit local requirements )
Standard options :-
230V 50Hz -- UK 3-pin
220V 50Hz -- Euro 2-pin
120V 60Hz -- USA 2-pin
AFA 500
Remote SM6
Airflow Sensor

15
6.0 Limitation of Warranty and Liability
Seller warrants that this product, under normal use and service as described in the operator’s manual shall be free from
defects in workmanship and material for a period of twelve (12) months, or the length of time specified in the operator’s
manual, from the date of shipment to the customer. This limited warranty is subject to the following exclusion :-
a.
Batteries and certain other components when indicated in specifications are warranted for a period of
90 days from the date of shipment to the customer.
b.
With respect to any repair services rendered, Seller warrants that the parts repaired or replaced will
be free from defects in workmanship and material, under normal use, for a period of 90 days from the
date of shipment to the customer
c.
Seller does not provide any warranty on finished goods manufactured by others. Only the original
manufacturer’s warranty applies.
d.
Unless specifically authorised in a separate writing by Seller, Seller makes no warranty with respect
to, and shall have no liability in connection with, any goods which are incorporated into other
products or equipment by the Buyer. All goods returned under warranty shall be at the Buyer’s risk of
loss, Seller’s factory prepaid, and will be returned at Seller’s risk of loss, Buyer’s factory prepaid.
The foregoing is IN LIEU OF all other warranties and is subject to the conditions and LIMITATIONS stated herein. NO
OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS
MADE.
THE EXCLUSIVE REMEDY OF THE USER OR PURCHASER, AND THE LIMIT OF LIABILITY OF SELLER FOR ANY
AND ALL LOSSES, INJURIES, OR DAMAGES IN CONNECTION WITH THIS PRODUCT ( INCLUDING CLAIMS BASED
ON CONTRACT NEGLIGENCE, STRICT LIABILITY, OTHER TORT, OR OTHERWISE ) SHALL BE THE RETURN OF
THE PRODUCT TO THE FACTORY OR DESIGNATED LOCATION AND THE REFUND OF THE PURCHASE PRICE,
OR, AT THE OPTION OF THE SELLER, THE REPAIR OR REPLACEMENT OF THE PRODUCT. IN NO EVENT SHALL
SELLER BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING, REASSEMBLY OR REINSTALLATION COSTS OR CHARGES.
NO ACTION, REGARDLESS OF FORM, MAY BE BROUGHT AGAINST THE SELLER MORE THAN ONE YEAR AFTER
THE CAUSE OF ACTION HAS ACCRUED.
The purchaser and all users are deemed to have accepted the terms of this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND
LIABILITY may not be amended or modified nor may any of its terms be waived except by a writing signed by an
authorised representative of the Seller.
7.0 Contact us :-
For further information on our range of airflow alarms and controls please
contact us at :-