User Manual 

Blood Pressure Monitor 

Model: AES-U181   

 

 

 

 

Thank you very much for selecting our AES-U181 model Blood Pressure Monitor. 

 

To use the monitor correctly and safely, please read the manual thoroughly before operating it. 

 

Please store this manual for future reference.

 

 

 

Thank you for selecting arm type blood pressure Monitor (AES-U181). 
The  monitor  features  blood  pressure  measurement,  pulse  rate  measurement  and  auto-save  the  result.  The  design 
provides you with many years of reliable service. 
Reading taken by the AES-U181 is equivalent to those obtained by a trained observer using the cuff and stethoscope 
auscultation method. 
This manual contains important safety and care information, and provides step by step instruction for using the product. 
Read the manual thoroughly before using the product. 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

 


background image

Safety information 

The below signs might be in the user manual, labeling or other component. They are the requirement of 

standard and using. 

 

PROHIBITION 

Means Forbidden with detailed items expressed in words or figures within or beside the 
mark. Left one means General Forbidden. 

                                                   

MUST OBSERVE 

Means Obligatory with detailed items expressed in words or figures within or beside the 
mark. Left one means General Compulsory.   

 

Refer to instructions manual/booklet. 

 

IMPLICATION OF SYMBOL 

Type-B 

applied part

.     

 

Caution: Consult accompanying documents

 

Do  not  dispose  this  product  as  unsorted  municipal  waste.  Collection  of  such  waste 
separately for special treatment is necessary.

 

 

 

Stand-by button

 

 

Transport package shall be kept away from rain. 

 

Transport package shall not be exposed to sunlight. 

 

Indicates correct upright position of the transport package. 

 

Contents of the transport package are fragile therefore it shall be handled with care. 

 

Indicates temperature limits within which the transport package shall be stored and handled. 

 

Non-ionizing electromagnetic radiation 

 

On/Off

 

Of/Off button used to take a measurement

 

 

MEM

 

Memory icon, to recall memory 

 

SET 

SET the unit   

 

Manufacturer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

CAUTION 

 

This device is intended for adult use only. 

This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities 

other than the arm wrist or for functions other than obtaining a blood pressure measurement. 

Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure. Do not begin or end medical 

treatment based solely on this measurement. Consult your physician for treatment advice. 

If you are taking medication, consult your physician to determine the most appropriate time to measure your blood pressure. Never 

change a prescribed medication without consulting your physician. 

This unit is not suitable for continuous monitoring during medical emergencies or operations.   

If  the  cuff  pressure  exceeds  40  kPa  (300  mmHg),the  unit  will  automatically  deflate.  Should  the  cuff  not  deflate  when  pressures 

exceeds 40 kPa (300 mmHg), detach the cuff from the arm wrist and press the 

  button to stop inflation. 

To avoid measurement errors, carefully read this manual before using the product. 

The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with 

oxygen or nitrous oxide.   

The operator shall not touch output of AC adapter and the patient simultaneously. 

To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated 

LCD display signal   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Item  Description 

Systolic blood pressure, SYS 

Diastolic blood pressure, DIA 

Pulse 

4

 

Measurement unit of blood 

pressure 

Battery 

Arrhythmia 

Recalling memory 

Memory 

Average of blood pressure   

10 

Date 

11 

Time 

12

 

User   

 


background image

Monitor components   

 

 

Replace the batteries in following conditions 

 

1.  Slide off the battery cover. 
2.  Install the batteries by matching

 

the correct polarity, as shown. 

3.  Replace the cover. 

 

Replace the batteries whenever the below happen: 

The

  shows. 

The display dims. 
The display does not light up. 

 
 

CAUTION 

 

Remove batteries if the device is not likely to be used for some time. 

The old battery is harmful to the environment, so please do not dispose with other daily trash. 

Remove the old battery from the device and follow your local recycling guidelines. 

 

Setting User, Date, time and measurement unit   

It is important to set the clock before using your blood pressure monitor, so that a time stamp can be 
assigned to each record that is stored in the memory. 
 

1.  Switch User 

When the unit is turn off, press SET button until the 

icon is flashing.   

Press MEM to choose 

  or 

Turn off the unit. 

1 Cuff 
2 LCD 
3 SET button 
4 MEM button 
5 On/Off button 

 

 


background image

 

2.  Set Date&Time 

After chosen User, press SET button to enter the Date and Time mode. 

Press On/Off button to adjust the digital No.,   

Press MEM button to confirm and enter the next parameter. 

 

 

3.  Set the Voice On/Off 

After set the Date&Time, press SET button to enter the Voice mode. 

Press MEM button to choose the Voice On or Off. 

4. 

 

 

5. 

After the Voice set, press SET to enter the Unit mode.

 

Press MEM button to choose PA unit or mmHg unit. 

 
 

Tie the cuff   

1. 

Tie the cuff on your upper arm, position the tube off-center toward the inner side of arm in line with the little finger.

 

 

 

 
 

 


background image

 
2. 

The cuff should not too tight. You should be able to insert one finger between the cuff and your arm.

 

 

3. 

Sit comfortably with your left arm resting on a flat surface. 

 

 
 

 

Take a rest for 5 minutes before measuring. 

Wait at least 3 minutes between measurements. This allows your blood circulation to recover. For a meaningful comparison, try to 

measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as 

directed by a physician 

 

 

 

Start the Measurement 

 

1. 

Press the On/Off button to run the BP monitor, it will begin to inflate and measure Blood Pressure automatically. 

 

Do not move when in the measuring process. 

The measurement will displayed and saved automatically.

 

 

 
Press On/Off button to Turn off after measured. 
 

 

 


background image

Recall records 

Choose the User 1 or User 2 before recall the memory. 

1.  Turn off the unit, then press MEM button the recall the memory. 

2.  It will show the average value of latest 3 measurements. There is a No. 3 on the Pulse position, after 1 second it will show the 

average pulse of latest 3 measurements.

 

 

 

3. 

Press the MEM again, it will show the latest 1 measurement, and press again the second, then the third…

 

 

 

Delete Memory 

1. 

Press MEM button to choose the memory that needs to be deleted.

 

2. 

Press SET button when the display shows dEL

 

3. 

Press MEM to confirm delete this memory.

 

4. 

When the unit is off, press MEM button for 3 seconds, then press SET button until the device shows ALL EE, that’s means the 

memories are all deleted.

 

 

 

Maintenance   

In order to get the best performance, please follow the below instructions: 
 
 
 
 
 
 
 
 
 
 
 
 
 

 


background image

 
 

 

 
 

Why the blood pressure I get from the hospital is different from home? 

The blood pressure is different even during 24 hour because of the weather, emotion, exercise etc. 

 

The attention need to pay 
When you measure you blood pressure at home: 
If the cuff is tied properly. 
If the cuff is too tight or too loose. 
If the cuff is tied on the upper arm. 
If you feel anxious pressured. 
You had better take deep breath 2-3 
times before beginning. 
Advice: adjust yourself for 4-5 minutes until you calm down. 

 
 
 
 

 


background image

Can I measure the right arm?

 

It is ok for both arms, but there will be some different results for different people, so we suggest you measure the same arm every 

time. 

 

 

 

 

 

Trouble Shooting   

 

PROBLEM 

SYMPTOM 

CHECK THIS 

REMEDY 

No power 

Display is dim or 
will not light up. 

Batteries are exhausted. 

Replace with new batteries 

Batteries are inserted 
incorrectly. 

Insert the batteries correctly 

Low 
batteries 

  Show on 

the display 

Batteries are low. 

Replace with new batteries 

Error 

massage 

E 1 shows 

The heart beat signal is too 
weak or the pressure of 
cuff descend 

Refasten the cuff and then 
measure again. 

E 2 shows 

The monitor detected 
motion while measuring. 

Refasten the cuff and then 
measure again. 

E 3 shows 

Measurement is not right. 

Relax for a moment and then 
measure again. 

E P shows 

Inflation fail 

Relax for a 
moment and then measure 
again. 

E E shows 

EEPROM incorrect 

Relax for a moment and then 
measure again. 

HI shows   

The 

Inflation

 time is too 

long 

Refasten the cuff and then 
measure again. 

 

 
 
 
 

 


background image

Specification   
 

Power supply 

6V DC 4*AAA alkaline batteries 

Display 

Digital LCD V.A.86*55mm 

Measurement Method 

Voltage reduced capacitance 

Measurement range 

Pressure: 4~37kpa(30~280mmHg) 
Pulse value:(40~199)times/minute 

Accuracy 

Pressure: 
10

~40

within±0.4kpa(3mmHg) 

pulse value:±5% 

Normal working condition 

Temperature:10

~40

   

Relative humidity: 15%~80% 

Storage & transportation 

condition 

Temperature:-20

~+55

℃ 

Relative humidity:10%~90% 

Weight 

Approx.280g(including the batteries) 

Product dimensions 

130*99*45mm 

Attachment 

4*AAA alkaline batteries, user manual 

Mode of operation 

Continuous operation 

Degree of protection 

Type B applied part 

 
The EMC declaration according to the requirement of EN 60601-1-2 
Cautions: 
User must regard EMC, please install and put in service AES-U181 according to the EMC information 
provided in the accompanying documents. 
Portable and mobile RF communications equipment can affect medical electrical equipment. 
The use of accessories, transducers and cables other than those specified, with the exception of 
transducers and cables sold by MY WINGS as replacement parts for internal components, may result in 
increased Emissions or decreased Immunity of AES-U181. 
The performance of the EQUIPMENT and SYSTEM that was determined to be essential performance. 
AES-U181 should not be used adjacent to or stacked with other equipment and that if adjacent or 
stacked use is necessary, AES-U181 should be observed to verify normal operation in the configuration 
in which it will be used. 
Table 201-Guidance and manufacturer’s declaration-electromagnetic emissions-for AES-U181,as 
followling table. 
Table 202-Guidance and manufacturer’s declaration-electromagnetic immunity -for AES-U181, as 
followling table. 
 

 


background image

 
Table 204 -Guidance and manufacturer’s declaration-electromagnetic immunity -for AES-U181, as 
followling table. 
Table 206 Recommended separation distances between portable and mobile RF communications 
equipment and AES-U181, as followling table. 

 
Table 201 

Guidance and manufacturer’s declaration – electromagnetic emissions 

The AES-U181 is intended for use in the electromagnetic environment specified below. The customer 
or the user of the AES-U181 should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions   
CISPR 11 

Group 1 

The AES-U181 uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and 
are not likely to cause any interference in nearby 
electronic equipment. 

RF emissions 
CISPR 11 

Class B 

The AES-U181 is suitable for use in all establishments, 
including domestic establishments. 

Harmonic emissions 
IEC 61000-3-2 

Not applicable 

Voltage fluctuations 
/flicker emissions 
IEC 61000-3-3 

Not applicable 

 
Table 202 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The AES-U181 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the AES-U181 should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment –
guidance 

Electrostatic 
discharge (ESD) 
IEC 61000-4-2 

± (2, 4, 6) kV contact 
± (2, 4, 8) kV air 

± (2, 4, 6) kV contact 
± (2, 4, 8) kV air 

Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the relative 
humidity should be at least 
30 % 

 


background image

Electrical fast 
transient/burst 
IEC 61000-4-4 

±2 kV for power 
supply lines ±1 kV for 
input/output 
lines 

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment. 

Surge 
IEC 61000-4-5 

±1 kV differential 
mode 
±2 kV common mode 

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment. 

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 

<5 % UT (>95 % dip 
in UT) for 0,5 cycle   
 
40 % UT (60 % dip in 
UT) for 5 cycles   
 
70 % UT (30 % dip in 
UT) for 25 cycles 
 
<5 % UT (>95 % dip 
in UT) for 5 sec 

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment. If the 
user of the AES-U181 requires 
continued operation during 
power mains interruptions, it is 
recommended that the 
AES-U181 be powered from 
an uninterruptible power 
supply or a battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical commercial 
or hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

 

 

 

 

 

 

 

 

 

 

 


background image

Table 204 

 

 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The AES-U181 is intended for use in the electromagnetic environment specified below. 

The customer or the user of the AES-U181 should assure that it is used in such an environment. 

Immunity 
test 

IEC 60601 
test level 

Compliance 
level 

Electromagnetic environment –guidance 

Conducted 
RF 
IEC 61000-
4-6 
Radiated 
RF 
IEC 61000-
4-3 

3 Vrms 
150 kHz to 
80 MHz 
3 V/m 
80 MHz to 
2,5 GHz 

Not 
applicable 
3 V/m 

Portable and mobile RF communications equipment 
should be used no closer to any part of the AES-U181, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
Recommended separation distance 
d=1,2 

d=1,2 

p 80 MHz to 800 MHz 

d=2,3 

p 800 MHz to 2,5 GHz 

where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m). 
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey, a should be less than the compliance level in 
each frequency range. b 
Interference may occur in the vicinity of 
equipment marked with the following symbol:   

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. 
If the measured field strength in the location in which the AES-U181 is used exceeds the 
applicable RF compliance level above, the AES-U181 should be observed to verify normal 
operation. If abnormal performance is observed, additional measures may be necessary, such 
as reorienting or relocating the AES-U181. 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 


background image

Table 206 

 

Recommended separation distances between 
portable and mobile RF communications equipment and the AES-U181 

The AES-U181 is intended for use in an electromagnetic environment in which radiated RF 
disturbances are controlled. The customer or the user of the AES-U181 can help prevent 
electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the AES-U181 as recommended below, according 
to the maximum output power of the communications equipment. 

Rated maximum 

 

 

output power of 
transmitter 

Separation distance according to frequency of transmitter 

Electrical fast 
transient/burst 
IEC 61000-4-4 

150 kHz to 80 MHz   
d =1

2

80 MHz to 800 MHz   
d =1

2

800 MHz to 2,5 GHz   
d =2

3

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in meters (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according 
to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.