PDF User Manual

  1. Home
  2. Manuals
  3. ISA BTS200 MKII Manual

ISA BTS200 MKII Manual

Made by: ISA
Type: Manual
Category: Test Equipment
Pages: 10
Size: 0.37 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

Doc. SIE20167 

Rev. 2 

Date 27/03/2013 

BTS200 MKII 

BATTERY TEST SET 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

3 of  11 

INDEX 

1 GENERAL DESCRIPTION _____________________________________________________________ 4

 

2 APPLICABLE STANDARDS  ___________________________________________________________ 6

 

2.1 Electromagnetic Compatibility _______________________________________________________________ 6

 

2.2 Low voltage directive _______________________________________________________________________ 6

 

3 PERFORMANCE SUMMARY ___________________________________________________________ 7

 

3.1 Introduction  ______________________________________________________________________________ 7

 

3.2 Specification  ______________________________________________________________________________ 8

 

 
 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

4 of  11 

1 GENERAL DESCRIPTION 

 

This  technical  specification  is  to  describe  the  technical  characteristics  of  the  battery  test  set 
BTS200 MKII, designed for the test of battery benches using the discharge method. This testing 
device  can  discharge  the  battery  in  different  modes,  to  measure  the  capacity  of  new  and  used 
battery. With respect to the former BTS200, the MKII uses a technological improvement, that has 
allowed to reduce weight and dimensions. 
 

BTS200  MKII  allows  the  connection  to  up  to  nine  auxiliary  modules  type  ELU200  MKII  (or 
ELU200,  or  other  BTS200  MKII,  or  other  BTS200),  that  enhance  the  discharging  capability. 
With  the  optional  current  clamp,  or  with  the  external  shunt,  it  is  also  possible  to  use  the  real 
burden as the additional burden, and to control the total discharging current in the selected mode. 
 

The device is  microprocessor based. This  allows the control  of all the discharging parameters in 
real time. It also allows communication with a PC for the control and the data representation of the 
discharging parameters. The following is the test set picture. 
 

 

 
With respect to the former version, MKII features the following improvements. 

 

The weight has been decreased of 9 kg, from 48 kg to 39 kg. 

 

Dimensions have been reduced, from 283 x 803 x 420 mm (WDH) to 283 x 700 x 420 
mm. 

 

Two wheels have been added, to ease the transport. 

 

The number of discharging steps that can be programmed have been increased from 10 to 
20: this eases drawing any discharge profile. 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

5 of  11 

 

In addition to the external clamp, the external burden current can now be measured also 
with a shunt. 

 

If a test is interrupted, for instance because of a faulty contact or cell, after the repair it is 
possible to continue the test: this was not possible with the former device. 

 

The test set firmware can be upgraded with a supplied software. 

 

 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

6 of  11 

2 APPLICABLE STANDARDS 

The  test  set  conforms  to  the  EEC  directives  regarding  Electromagnetic  Compatibility  and  Low 
Voltage instruments. 

2.1 Electromagnetic Compatibility 

Directive  no.  2004/108/EC. Applicable Standard : EN61326-1 + A1 + A2. 

EMISSION 

- EN61000-3-2: harmonic content induced into the power supply: class A. 
- EN 61000-3-3: Limitation of voltage fluctuations and flicker. Acceptable limits: basic. 
- CISPR16 (EN 55011 class A): Limits and measurement methods of radio-electric disturbances 
for industrial, medical and scientific instruments at radio-electric frequencies. 
Acceptable limits for conducted emission: 
. 0.15-0.5 MHz:  

79 dB pk; 66 dB avg. 

. 0.5-5 MHz:       

73 dB pk; 60 dB avg. 

. 5-30 MHz:   

73 dB pk; 60 dB avg. 

Acceptable limits for radiated emission: 
. 30-230 MHz:  

40 dB (30 m) 

. 230-1000 MHz:   

47 dB (30 m) 

IMMUNITY 

- EN 61000-4-2: Immunity tests for ESD. Test values: 8 kV in air; 4 kV in contact. 
- EN 61000-4-3; Immunity tests for radio frequency interference. Test values (f= 900 

 5 MHz): 

field 10 V/m, modulated AM 80%; 1 kHz 
- EN 61000-4-4; Immunity tests for high speed transients (burst). Test values: 2 kV peak; 5/50 ns. 
-  EN 61000-4-5;  Immunity tests  for surge. Test  values: 1 kV peak differential  mode; 2 kV peak 
common mode; 1.2/50 us. 
-  EN  61000-4-6:  immunity  to  low-voltage  sinusoidal  waveform.  Test  values:  0.15-80  MHz,  3 
Vrms, 80% AM 1 kHz. 
- EN 61000-4-8: Immunity tests for low frequency magnetic fields.  Test values: 30 Arms/m. 
- EN 61000-4-11: Immunity test for power supply drops. Test value: 1 cycle; 100% drop. 

2.2 Low voltage directive 

-  Directive n.  2006/95/EC. 
Applicable standards, for a class I instrument, pollution degree 2, Installation category II:  CEI EN 
61010-1. In particular: 
- Dielectric Rigidity:  1.4 kV, 1 minute. 
- Isolation resistance: > 100 Mohm @ 500 V DC. 
- Earth resistance : < 0.1 Ohm. 
- Dispersion current: < 5 mA. 
- Inputs/outputs protection: IP 20  - IEC 60529. 
- Acoustic noise: < 75 dB, at full power. 
- Operating temperature: 0 - 40°C; storage: -25°C to 70°C. 
- Relative humidity : 5 - 95%, not condensing. 
- Vibration: IEC 68-2-6 (20 m/s^2 at 10 – 150 Hz); 
- Shock: IEC 68-2-27 (15 g; 11 ms; half-sine). 
- Altitude: less than 2000 m. 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

7 of  11 

3 PERFORMANCE SUMMARY

 

3.1 Introduction 

 
BTS200 MKII  can discharge batteries with different nominal voltages, from 24 V to 240 V DC. 
The discharging current can be up to 130 A for voltages of 24 to 120 V DC nominal, and up to 70 
A for  220 to  240 V.  It  is possible to  use in  parallel  up to  10  BTS200 MKII, thus  arriving up to 
1300 A for voltages of 24 to 120 V DC nominal, and up to 700 A for 220 to 240 V. 
 
The use of BTS200 MKII is very simple: 

 

Connect the cables to the battery to be discharged; 

 

Power-on BTS200 MKII; 

 

Choose the memory area where to save the test data; 

 

Program the discharging current (or power); 

 

Program the maximum discharge duration; 

 

Program the minimum battery voltage; 

 

Program the Ah to be discharged; 

 

Press START. 

 
During  the  discharge  actual  parameters  are  displayed  on  the  graphic  screen;  all  the  measured 
parameters are saved into the selected memory.  It is also possible to connect BTS200 MKII to a 
PC with TDMS: this allows to have on the PC display the discharging diagram. 
 
Via  PC,  it  is  possible  to  download  a  current  sequence  with  up  to  20  steps,  each  step  being 
programmable  in  current  and  duration.  This  enables  simulating  a  typical  current  profile 
encountered during the day. 
 
If,  during  the  discharge,  any  of  the  programmed  limits  is  trespassed,  the  test  is  immediately 
stopped, and the alarm contact closed. 
 
Once the test is finished it is possible to watch the discharging current and voltage on the display, 
or to download the test result to the PC. 
 
The selection of discharging parameters is performed using the encoder with confirmation switch, 
and the graphic display. The operation is menu driven: turning the encoder knob it is possible to 
select the desired operation; pressing it the selection is confirmed. 
 
Optionally,  it  is  possible  to  leave  an  external  load  connected  to  the  battery  under  test.  In  this 
situation, the externally discharged current  will be measured by means of a DC current probe or 
by  a  shunt:  BTS200  MKII  takes  into  account  this  current,  and  ensures  that  the  total  discharging 
current is the programmed one. 
 
The  device  is  equipped  with  an  emergency  pushbutton,  located  on  the  front.  When  pressed,  it 
stops  all  the  BTS200  MKII  activities.  The  pushbutton  is  mechanically  self-locking:  the  operator 
has to rotate it in order to reset the normal operation. 
 
The device can be powered by the battery under test itself, for batteries with a nominal voltage of 
220 or 240 V; else, on the front is provided a power supply plug for a wide range AC supply. 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

8 of  11 

3.2 Specification 

PARAMETER 

VALUE 

A) HARDWARE 

 

Maximum number of paralleling test sets 

10: 1 master, 9 slave 

Absolute maximum battery voltage 

270 V 

Minimum battery voltage 

18 V 

Maximum power that can be discharged (one module) 

16 kW 

Maximum power that can be discharged (10 modules) 

160 kW 

Max  discharging  current  from  24  V  DC  to  120  V  DC  (one 
module) 

130A 

Max  discharging  current  from  24V  DC  to  120  V  DC  (ten 
modules) 

1300A 

Max discharging current from 220 V DC to 240 V DC (one 
module) 

70 A 

Max  discharging current from  220 V DC to  240 V DC  (ten 
modules) 

700 A 

Discharging  modes:  constant  current;  constant  power; 
current profile; manual adjustment 

Discharging power resolution 

100 W 

Nominal battery ranges 

24, 48, 110, 240 V DC 

Battery range setting 

Automatic, on 4 ranges, as test starts 

24 V range limits 

20 to 34 V DC 

48 V range limits 

34 to 60 V DC 

72 V range limits 

60 to 84 V DC 

110/120 V range limits 

84 to 140 V DC 

240 V range limits 

140 to 270 V DC 

Maximum starting voltages on the 24 V range 

20 to 35 V DC 

Maximum starting voltages on the 48 V range 

34 to 65 V DC 

Maximum starting voltages on the 72 V range 

58 to 97 V DC 

Maximum starting voltages on the 110 V range 

84 to 140 V DC 

Maximum starting voltages on the 240 V range 

169 to 270 V DC 

Battery voltage measurement resolution 

± 0.1 V DC 

Battery voltage measurement accuracy 

± 1% of the maximum range = 2.7 V 

Battery  current  measurement  resolution  (direct  and  clip-on 
CT) 

1 A 

Battery current measurement accuracy 

± 1% of the maximum range = 1.3 A, 

starting from 15 A 

V and I measurements refresh period 

15 s 

Time measurement resolution 

1 s 

External clip-on CT conversion factor range (programmable)  1 to 500 mV. NOTE: the 1 mV range 

is to be used for currents greater than 

400 A. 

External shunt voltage input 

1 to 60 mV 

External clip-on CT or shunt accuracy 

± 2% 

Audible end of test alarm buzzer 

5 tones 

Alarm output contact  

1 SPDT 

Display type 

Graphic 128 x 64 points 

Display backlight 

YES 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

9 of  11 

 
 
 

PARAMETER 

VALUE 

Operator interface 

Encoder with push-button 

Test start 

ON-OFF pushbutton 

Test stop from a remote contact 

YES 

Alarms 

10-segments LED bar with current 

and alarms 

Test system LED 

Date and time 

Yes, with back-up battery 

PC communication 

USB 2.0 interface 

Paralleling interconnection 

2 connectors: in-out 

Paralleling interconnection cable type 

CAN bus; RJ11 connector 

Paralleling address selection 

By rotary switch 

Power supply modes: 
Mode 1: from the battery being discharged; 
Mode 2: from an AC voltage supply; 
Mode 3: from an external DC voltage supply. 

3 ways: 

. 100 to 240 V DC 

. 100 to 240 V AC, 50 to 60 Hz 

. 100 to 270 V DC 

Power supply mode selection 

By switch 

Power from the supply 

. 10 W in stand-by; 

. 150 W during the discharge. 

Emergency push-button 

Independent from the 

microprocessors 

Thermal protection 

Independent from the 

microprocessors 

Dimensions 

283 x 700 x 420 mm (WDH) 

Weight 

39 kg, without transport case. 

Transport wheels 

YES 

Transport case dimension 

330 x 900 x 600 mm (WDH) 

Transport case weight 

16 Kg

 

Provided with: 

- USB connection cable, 2 m; 

- RJ11 connection cable, 2 m; 

- Battery discharge cable pair with 

crocodiles; 

- Power supply cable, 2 m; 

- TDMS software 

 

 

 

NOTES: 

1. The above specified current and power ranges apply when the ambient temperature is 25 °C or 
less. For higher temperatures, apply a de-rating factor of – 1 A/°C for the current; for the power, it 
is the product of the voltage by the current scaling factor. 
2. The absolute maximum ambient tempperature for full power operation is 40 °C. 
3. The test set cannot operate with a direct sun exposition, or located nearby an heat source. 
4. The test set generates very hot air, both above and on the rear: don’t touch it until the automatic 
cooling down procedure is completed. 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

10 of  11 

PERFORMANCE 

CHARACTERISTICS 

B) RESIDENT FIRMWARE 

 

Test settings 

Recorded in a NVM 

Number of memories 

Maximum test duration that can be recorded 

15 h, with 15 s resolution 

Maximum test duration, not recorded in memory 

infinite 

Memory profile mode: maximum number of test steps 

20 

Memory profile mode: time and current resolution 

1 minute – 1 A 

Memory profile mode: maximum time and current steps 

600 minutes – 1300 A 

Test set control mode 

Via encoder + push-button and 
menu on the display 

Measurements on the display 

. Date; 
. Time; 
. Internal discharge current; 
. External discharge current; 
. Battery voltage; 
. Ah discharged; 
. Elapsed test time. 

Test stop settings 

. Minimum voltage; 
. Maximum Ah; 
. Maximum test time; 
. System alarms; 
. Test stop from remote contact. 

Continue after an alarm 

YES 

Firmware upgrade 

YES 

System control 

Digital, by two microprocessors 

Test set protections 

. Heat sink over-temperature; 
. Fuse failure; 
. Fans failure; 
. Minimum battery voltage; 
. Maximum battery voltage; 
. Wrong battery voltage; 
. Overload; 
. Control circuit error; 
. Power circuits failure; 
. CAN-bus error; 
. External current metering error; 
. Manual emergency pressed; 
. USB interface error; 
. EEPROM memory error; 
. Firmware error. 

Reversed battery insertion protection 

YES 

Automatic cool down procedure at end of test 

YES 

 
 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

 


background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

11 of  11 

PERFORMANCE 

CHARACTERISTICS 

C) TDMS SOFTWARE 

 

Create, save, upload test plans 

 

Edit test profiles 

Current discharge profile: diagram, 
table 

Upload test profiles 

 

Download test results 

Input reference data, save, recall 

Examine test results 

Result diagram, table of values 
Zoom in-out 
Cursors 

Print test results 

Printer editor capability 
Customized report creation 

Results and settings data base 

Integrated in the TDMS Substation 
data base 

 

PERFORMANCE 

CHARACTERISTICS 

CODE 

D) OPTIONS 

 

 

Transport case, with wheels  Dimensions: 

330 x 900 x 600 mm; weight 16 kg.

 

PII18167 

DC current clamp 

Ranges: 40 A DC (10 mV/A) and 400 A DC (1 mV/A). 
Accuracy: 1.5% up to 40 A; 2% up to 400 A. 
Maximum conductor diameter: 30 mm. 

PII12167 

ELU200 MKII 

Same as BTS200 MKII, without the local control 

PII21167 

 
The  following  table  lists  a  number  of  different  types  of  batteries,  the  discharge  current  and  the 
discharge power that can be programmed. The value chosen for the Ah capacity is just a reference 
one: test current can be scaled to the actual capacity. 
 

NOMINAL 

BATTERY 

VOLTAGE 

NOMINAL 

CAPACITY 

 

Ah 

CONSTANT 

CURRENT 

 

CONSTANT 

POWER 

 

kW 

TEST 

DURATION 

 

DISCH. 

CAPACITY 

 

Ah 

END 

OF TEST 

VOLTAGE 

24 

500 

50 

10 

500 

20 

48 

500 

50 

10 

500 

40 

110 

500 

50 

4.7 

10 

500 

94 

120 

500 

50 

5.1 

10 

500 

102 

220 

500 

50 

9.4 

10 

500 

188 

240 

500 

50 

10.2 

10 

500 

204 

 
 
 
 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca