background image

 

 Worldwide Manufacturer of Gas Detection Solutions 

 

Remote Control 

RC2 

Operations Manual 

 

 

1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108  USA

 

(800) 959-0329 • (734) 769-0573 • www.goodforgas.com

 

 


background image

 

 

 


background image

 

 

Table of Contents 

 

 

 

 

For your safety ………………………………………………………………………….

 2 

Operational hints ……………………………………………………………………….

  2 

General description …………………………………………………………………….

 2 

Operation ………………………………………………………………………………..

 4 

Trouble shooting ………………………………………………………………………..

 4 

Replacement of battery ………………………………………………………………..

 5 

Signal transmission …………………………………………………………………….

 5 

Technical data ………………………………………………………………………….

  5 

  EC-Type Examination Certificate ……………………………………………………. 

 

6   

 

 


background image

 

 

For your Safety 

As with any piece of complex equipment, the GfG RC2 will do the job it is designed to do only 

if it is used and serviced in accordance with the manufacturer's instructions. Please protect 

yourself and your employees by following the instructions in this manual. All individuals who 

have or will have the responsibility for using and servicing this product must carefully read this 

manual. The warranties made by GfG with respect to the product are voided if functions or 

parameters are changed without the permission of GfG. They are also voided if the product is 

not used and serviced in accordance with the instructions in this manual. The above does not 

alter any statements by GfG regarding warranties, conditions of sale and/or delivery. 

 

Operational Hints 

The remote control RC2 is connected to a GfG transmitter by a connection cable and is used 

for operating the transmitter or as a display.  When connected to a GfG transmitter it provides 

a display and a keypad, the readout and the keypad functions of the remote control are 

identical with those of the GfG transmitter. 

 

The remote control RC2 is approved for use in explosive areas and is subject to an EC-Type 

examination Certificate issued by DEKRA EXAM GmbH, according to directive 94/9/EG (ATEX 

100a). 

 

Certificate: 

BVS 04 ATEX E 212 

 

Labeling:   

 II 2G Ex ia IIC T4 Gb -20°C

T

a

+50°C 

 

The remote control may be connected to or disconnected from the transmitter even in an 

explosive area. 

 

 

Do NOT replace the battery of the remote control in potentially explosive areas. 

 

Only use 9V block batteries type DURACELL-PROCELL  Alkaline,  6LR61 / 

MN1604, the plastic bottoms of which prevent the leaking of electrolyte. 

 

General Description 

The Remote Control RC2 is used for control or adjustment of the zeropoint and sensitivity 

when testing the transmitter. 

 

Should the Series 28 transmitters be mounted in hard to  access locations (i.e. close to the 

ceiling) you may, even in explosive areas, keep the cable procured permanently for connecting 

the remote control to the transmitter (maximum cable length is 10m  /  33 feet).  You can 

connect this cable to the remote control for display or service purposes. 

 

 

The maximum cable length between transmitter and remote control is 10m / 33 

feet.  Only use cables approved by GfG Instrumentation, Inc. 

 

 

 

 


background image

 

The remote control RC2 is supplied with a helix cable for connection to GfG transmitters.  On 

one end of the cable is plug A (see figure 1), which plugs into the socket on the transmitter.  

The other end of the cable is connected to the remote control and locked in place by slightly 

turning the sleeve ring of the plug (plug B). 

 

 

 

 

 

 

 

 

 

Fig. 1 

 

The On/Off switch for the remote is located at the top of the case near the plug connection. 

 

When the unit is turned on, the green operational LED above the display is illuminated. 

 

The keypad functions and the display of the remote control RC2 are absolutely identical with 

those of the transmitter. 

Nose of plug 

Plug A 

Plug B 

On/Off Switch 

Operational LED  

(green) 

Display 

Keypad 

 


background image

 

 

Operation 

The function  of the keypad  in the different operational  and  special  modes  are  described in 

detail in the operations manual of the relevant transmitter. 

 

 

For servicing a GfG transmitter without display, the remote control RC2 is 

required. 

 

The remote control RC2 is allowed to be used in explosive areas. 

 

 

 

For connection to a GfG transmitter fix plug A (see figure 1, page 3) into the transmitter.  If the 

plug is connected incorrectly, regardless of the transmitter polarity protection (nose at plug A), 

the display at the remote control drops out after a few seconds and will not function. 

 

When the unit is turned on, the green operations LED must be on. 

 

Trouble shooting 

If the operational LED does not light up, either replace the battery or check to ensure the 

battery is inserted correctly (check for correct polarity). 

 

If the display of the remote control does not show a reading when it is connected to the 

transmitter, although the operational LED is on, either the connection is wrong (check plug) or 

one of the units, either the transmitter of remote control, is faulty

 


background image

 

Replacement of battery 

Remove the two Allen screws on the top then slide the top upward, being careful of the 

position of the loose insulating plate.  Then slide the front part (with display and keypad) a 

slight amount upward and lift it off.  The printed circuit board and the battery holder are affixed 

to the front part by means of short cables

 

 

Do not replace the battery of the remote control in a potentially explosive area. 

 

Use only 9V batteries type DURACELL-PROCELL Alkaline, 6LR61 / MN1604. 

 

When replacing the 9V battery be cautious of the correct polarity (+ and – are indicated on the 

holder).  After turning the unit on, the green operational LED lights up.  If there is incorrect 

polarity of the battery, the unit cannot be turned on (the green LED does not light up). 

Signal Transmission 

The signal from the transmitter to remote control is an intrinsically safe transmission.  Both the 

transmitter and the cable of the remote control RC2 ensure the compliance of the following 

values: 

 

Maximum output voltage 

U

o

    6 V DC  Maximum output current 

I

o

 

45 mA 

Maximum output power 

P

o

 

68 m W  Maximum output capacity 

C

o

 

10 µF 

Maximum output inductivity 

L

o

 

  1 mH 

 

 

 

Technical Data 
Remote Control RC2 

For connection to transmitter: 

 

Series 28 transmitters 

Current Supply 

Battery Type: 

 

9 V battery DURACELL-PROCELL Alkaline, 6LR61 / 

MN1604 

Climate Conditions 

Short-term storage temperature: 

Recommended storage temperature: 

Operational temperature (ambient): 

Humidity range: 

Atmospheric pressure range: 

 

 -25 to +60°C 

    0 to +30°C 

 -20 to +50°C 

    5 to 90% r.h. 

800 to 1100 hPa 

Enclosure 

Casing material: 

Dimensions: 

Protection: 

Cable connection to transmitter: 

 

Compound, anti-static 

60 x 120 x 35 mm / 2.4 x 4.7 x 1.4 inches (W x H x 

D) (H with plug: 180mm / 7 inches) 

IP 54 

Maximum cable length 10m / 33 feet 

Approvals and Certifications 

Electromagnetic compatibility: 

Labeling and Ignition protection: 

EC-Type Examination Certificate: 

Production monitoring: 

 

DIN EN 50270:2006 Type class I and Type class 2 

  II 2G Ex ia IIC T4 Gb -

20°C≤T

a

≤+50°C

 

BVS 04 ATEX E 212 

  0158 (by notified body – DEKRA EXAM GmbH) 

 

 


background image

 

EC-Type Examination Certificate 

 

 


background image

 

 

 


background image

 

 

 

 


background image

 

 


background image

 

  

 

 
 

 

 

 

 

 

US/Canada:            (800) 959-0329 
US/Canada Fax:       (734) 769-1888 
International:            +1 734 769 0573 
International Fax:     +1 734 769 1888 
E-mail:                      info@goodforgas.com 
Website:            

www.goodforgas.com

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                        

 

 

 

7004-RC2 Rev. 2 (01/17/18) 

 

GfG Instrumentation, Inc.

 

1194 Oak Valley Dr. 
Suite 20 
Ann Arbor, MI 48108 
USA

  

           Worldwide Manufacturer of Gas Detection Solutions

 

 

GfG reserves the right to change part numbers, 

prices, and/or technical information without notification

.