1 General 

guidelines ................................................................................ 2 

2 Safety 

regulations ................................................................................. 3 

2.1 Before 

use ............................................................................................. 3 

2.2 

Field of application................................................................................. 3 

2.3 

Danger of electric shock ........................................................................ 3 

2.4 “Damage 

to 

the 

sealer” .......................................................................... 4 

2.5 Contamination 

of the sterilised goods  “Danger of patient infection”..... 4 

3 Installation.............................................................................................. 4 

4 Overview 

of 

controls............................................................................. 5 

5 Operating 

instructions.......................................................................... 6 

5.1 

Switching on the sealer.......................................................................... 6 

5.2 

Setting temperature level 1 - 9 .............................................................. 6 

5.3 

Setting temperature level 10 internally .................................................. 7 

5.4 Sealing 

procedure.................................................................................. 9 

6 Technical 

data........................................................................................ 9 

Wear and replacement parts .............................................................. 10 

 

GS-43-GB-002.doc  

1


background image

1 General 

guidelines 

The purpose of these instructions is to explain to the user how to use the 
GS 43 sealer safely and correctly. 

Anyone whose work involves using the sealer must be familiar with these 
instructions and the safety regulations. 

These instructions must always be available to personnel, to prevent incorrect 
use of the sealer and to ensure trouble-free operation. 

Copyright 

These instructions may only be given to third parties with the written consent 
of GETINGE. All documents are protected by copyright. 

The passing-on and copying of documents, including extracts and the 
communication or use for gain of their contents is not permitted, except by 
express consent. Infringements are punishable and will lead to claims for 
damages. We reserve all rights to exercise the provisions of commercial 
copyright law. 

Warranty and liability 

We accept liability within the terms of the stated warranty obligations for errors 
and omissions, to the exclusion of further claims. 

We reserve the right to make technical changes to the sealer covered by this 
instruction manual, for the purpose of further development. 

We accept no liability for damage and operational disturbances arising from 
incorrect operation, failure to observe these instructions, or improper repairs. 
Only original spare parts and accessories approved by GETINGE may be 
used. 

GETINGE accepts no liability for damage resulting from the use of non-
approved spare parts and accessories. 

The drawings and graphics are not to scale. 

 

2


background image

 

Explanations 

Important instructions given in the technical documentation are indicated as 
follows: 

 Warning 

)

Useful information 

 

2 Safety 

regulations 

These instructions must be read and strictly complied with. 

 

2.1 Before 

use 

Users must familiarise themselves with the instructions in this manual before 
using the GS 43 sealer. 

 

2.2 Field 

of 

application 

The GS 43 sealer is to be used only for sealing sterilisation packages. No 
liability will be accepted for use of the sealer for any other purpose. 

The sealer may only be used for its intended purpose, in accordance with the 
generally recognised rules of technology and the industrial health and safety 
regulations. It must not be used if it has defects that might be dangerous to 
employees or third parties. 

 

2.3 

Danger of electric shock 

Before the case is opened, the sealer must be disconnected from the electric 
power supply. To open the case, unscrew the four unmarked internal hexagon 
screws with a 4mm key.  

Only an authorised technician or a specialist dealer may undertake servicing 
work on the sealer. 

 

 

 

GS-43-GB-002.doc  

3


background image

2.4 

“Damage to the sealer” 

The sealer is not suitable for sealing PE bags. The sealer may be damaged if 
an unsuitable material is used. 

 

2.5 

Contamination of the sterilised goods 

 

“Danger of patient infection” 

1. 

If the pressure of the pressure rollers is incorrectly set, the seam will not 
be bacteria-tight. 

2. 

The user must set the correct sealing temperature for each packaging 
material and sealer. This is done by trial sealing operations. 
Sealing temperatures are usually between 120° C und 180° C. Our 
recommendation is to set the sealing temperature to 120 °C and to raise 
it gradually until a bacteria-tight seam is obtained. 

  

If the temperature is too low, the seam will be incomplete and the 
package will therefore be weak. 

  

If the temperature is too high, the package will be difficult to peel open. 
Paper fibres may be released and contaminate the sterile goods. 

  

Released paper fibres may result in patient infection. 

3 Installation 

Set up the GS 43 sealer on a clean, smooth surface as follows: 

  

Position the sealer as required. 

  

Leave at least 15 cm free space behind the sealer to ensure good cooling 
and trouble-free operation. 

  

Plug the mains cable into the socket on the back of the sealer. 

  

Check that the voltage and frequency of the supply agree with the voltage 
and frequency shown on the sealer. 

  

Plug the mains lead into the power socket. 

  

Install the feed plate. 

  

To adjust the distance to the seam, slacken the knurled nut on the feed 
plate. Move the feed plate to the required position and tighten the knurled 
nut. 
With the feed stop fully forward, the seam is about 5 mm away from the 
cut-off edge of the material; with the feed plate fully back it is about 35 
mm away from it. 

The GS 43 sealer has is now installed. 

 

4


background image

 

Paper guide accessories 

GETINGE recommend the use of a paper guide for convenient feeding and 
trouble-free transport of the material. 

  

Remove the feed plate and install the paper guide. 

  

To set the seam width, slacken the knurled nut on the paper guide to 
release the feed stop. Move the feed stop to the required position and 
tighten the knurled nut. 
With the feed stop fully forward, the seam is about 5 mm away from the 
cut-off edge of the material; with the feed plate fully back it is about 35 
mm away from it. 

Overview of controls 

 

 

1. 

Temperature selector switch 

2. 

Temperature setting scale 

3. 

Indicator lamp (temperature) 

4. 

Motor on/off switch 

5. 

Indicator lamp (motor) 

GS-43-GB-002.doc  

5


background image

5 Operating 

instructions 

The GS 43 sealer seals all sterilisation packages that are normally 
commercially available. The sealer is not suitable for sealing PE bags. 

 “Damage to the sealer” 

See Section 2.4, page 4. 

5.1 

Switching on the sealer 

Set the green switch on the back of the sealer to (

I

). 

  

The red lamp (1) shows a steady light, then goes out 
briefly. 
The sealer heats up. 

Do not

 feed in any material at this 

stage. 

  

The red indicator (1) lamps starts to flash. 
The sealing temperature has been reached and the 

sealer is ready to seal. 

5.2 

Setting temperature level 1   9 

You can set the temperature level in ten steps. 

Turn the switch (A) to the required position, as follows. 

  

Step 

1

 (B): Standby mode to 

save energy during pauses. 
Set to 50 °C. 

  

Steps 

2, 3, 4

 and 

5

 (C): 

around 120 °C. 
These settings are for 
Tyvek

®

 material. 

  

Steps 

6, 7, 8

 and 

9

 (D): around 

1
8

0 °C. 
These settings are for transparent 
material. 

Step 

10

 (E) can be set internally up to 200 °C without steps. 

This setting can only be done by an authorised technician. Instructions for 
setting step 10 are given in Section 5.3, page 

Fel! Bokmärket är inte 

definierat.

 

6


background image

 

Contamination of the sterilised goods  
“Danger of patient infection” 

See Section 2.4, page 4, item 2. 

)

 Important 

 

Only seal the packaging material when the red light 
is flashing, ie when the sealing temperature has 
been reached. 

 

If you reduce the temperature, the red light goes out 
(temperature too high). Wait until the temperature 
adjusts to the new setting and the red light starts to 
flash again. 

 

If you raise the temperature, the red light stays on 
continuously (without flashing) (temperature too low). 
Wait until the temperature adjusts to the new setting 
and the red light starts to flash again. 

5.3 Setting 

temperature 

level 10 internally 

The adjustment for temperature level 10 is on the underside of the case (1). 
This is how to set the temperature: 

  

Switch off the sealer and unplug it from the power supply. 

 

GS-43-GB-002.doc  

7


background image

  

Unscrew the four unmarked screws on the underside of the case cover 
with a 4 mm key. 

  

Remove the case cover (1). 

  

Unplug connector (2) by pulling it upwards as shown by the arrow. 

  

Using a screwdriver (3): 
turn screw (4) clockwise 

 - to raise the temperature - 

turn screw (4) anticlockwise 

 - to low er the temperature -. 

  

Refit the case cover (1). 

 

Danger of electric shock 

See Section 2.3, page 3. 

Contamination of the sterilised goods  
“Danger of patient infection” 

See Section 2.5, page 4, item 1. 

 

8


background image

 

5.4 Sealing 

procedure 

This is how to start the transport motor: 

 

  

Press button (1) (motor runs). 

  

indicator lamp (2) gives a steady green light. 

  

Press button (3) (motor stops). 

 

)

 Important 

 

When the packaging material is fed in, the paper 
side should be down. 

 

We recommend switching off the motor when the 
sealer is not in use (eg during breaks). 

6 Technical 

data 

SEALER GS 43 

Dimensions w x d x h 

570 x 280 x 240 mm approx 

Dimensions with paper 
guide 

575 x 320 x 240 mm approx 

Weight 13 

kg 

approx 

max. sealing temperature 

200 °C 

Transport speed 

9.8 m/min 

Seal seam dimensions 

4 x 2.5 mm approx 

max. power consumption 

740 W 

Power supply 

230V (115V) / 50Hz (60Hz) 

Mains fuse 

3.15 A slow (115V: 6A slow) 

 

Every GS 43 sealer is set up to produce a seal seam that complies with 
DIN 58953-7. Every sealer has a certificate of initial validation. 

GS-43-GB-002.doc  

9


background image

 

Wear and replacement parts 

Fax order to : 

 

 

 

  _________________ 

 

Quantity 

Description

 Ordering 

no. 

 

PTFE strip for heating element (70 cm) 

4427032 

 

PTFE strip for transport rails (100cm) 

4427030 

 

Pressure roller, upper, complete 

4419016 

 

Pressure roller, lower  

4425723 

 

Transport belt (one) 

4416004 

 

Toothed belts for pressure roller 

4416016 

 

Bearing bush (Ø10 x Ø12) 

4415015 

 

Ball bearing 

4415000 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

From:  
 ___________________________________________________ 

  ___________________________________________________ 

Series no.  ___________________________________________________ 

Signature:   ______________ 

  Date: ______________ 

Other parts are available from GETINGE on request. 

 

10