background image

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  1 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

WARRANTY 

The Gentec-EO TUNER Single Channel Laser Power Meter comes with a one-year warranty (from date 
of shipment) against material and/or workmanship defects, when used under normal operating conditions. 
The warranty does not cover damages related to battery leakage or misuse.  

Gentec-EO  Inc.  will  repair  or  replace,  optionally,  any  TUNER  that  proves  to  be  defective  during  the 
warranty period, except in the case of product misuse. 

Any attempt by an unauthorized person to alter or repair the product voids the warranty. 

The manufacturer is not liable for consequential damages of any kind. 

In case of malfunction, contact your local Gentec-EO distributor or nearest Gentec-EO Inc. office to obtain 
a return authorization number. The material should be returned to: 

 

 

 

Gentec Electro-Optics, Inc. 

445, St-Jean-Baptiste, Suite 160 

Québec, QC 

Canada   G2E 5N7 

 

Tel:  (418)  651-8003 

Fax:  (418)  651-1174 

e-mail: 

service@gentec-eo.com

 

 
 

Website: 

www.gentec-eo.com

 

 

CLAIMS 

To obtain warranty service, contact your nearest Gentec-EO agent or send the product, with a description 
of  the  problem,  prepaid  shipping  and  insurance,  to  the  nearest  Gentec-EO  agent.  Gentec-EO  Inc. 
assumes no risk for damage in transit.  Gentec-EO Inc. will, at its option, repair or replace the defective 
product free of charge or refund your purchase price.  However, if Gentec-EO Inc. deems that the failure 
is  caused  by  misuse,  alterations,  accident  or  abnormal  condition  of  operation  or  handling,  you  will  be 
billed for the repair and the repaired product will be returned to you (repetitive, see above). 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  2 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

WARRANTY 

The  Gentec-EO  TUNER  Single  Channel  Laser  Power/Energy  Meter  carries  a  one-year  warranty  (from 
date  of  shipment)  against  material  and/or  workmanship  defects,  when  used  under  normal  operating 
conditions. The warranty does not cover damages related to battery leakage or misuse.  

Gentec-EO  Inc.  will  repair  or  replace,  optionally,  any  TUNER  that  proves  to  be  defective  during  the 
warranty period, except in the case of product misuse. 

Any attempt by an unauthorized person to alter or repair the product voids the warranty. 

The manufacturer is not liable for consequential damages of any kind. 

Customers must fill in and mail the warranty card in order to activate the warranty. 

In case of malfunction, contact your local Gentec-EO distributor or nearest Gentec-EO Inc. office to obtain 
a return authorization number. The material should be returned to: 

 

 

 

Gentec Electro-Optics, Inc. 

445, St-Jean-Baptiste, Suite 160 

Québec, QC 

Canada   G2E 5N7 

 

Tel:  (418)  651-8003 

Fax:  (418)  651-1174 

e-mail: 

service@gentec-eo.com

 

 
 
 

Website: 

www.gentec-eo.com

 

 

CLAIMS 

To obtain warranty service, contact your nearest Gentec-EO agent or send the product, with a description 
of  the  problem,  transportation  and  insurance  prepaid,  to  the  nearest  Gentec-EO  agent.  Gentec-EO  Inc. 
assumes  no  risk  for  the  damage  in  transit.    Gentec-EO  Inc.  will,  at  its  option,  repair  or  replace  the 
defective product free of charge or refund  your purchase price.  However, if Gentec-EO Inc. determines 
that the failure is caused by misuse, alterations, accident or abnormal condition of operation or handling, 
you will be billed for the repair and the repaired product will be returned to you, transportation prepaid. 

 

 

 

 

 

P/N: 90043187 Rev. D 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  3 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

SAFETY INFORMATION 

Do not use the TUNER if the device or the detector looks damaged, or if you suspect that the TUNER is 
not operating properly. 

Appropriate  installation  must  be  done  for  water-cooled  and  fan-cooled  detectors.    Refer  to  the  specific 
instructions  for  more  information.    The  user  must  wait  for  a  while  before  handling  these  detectors  after 
power is applied. Surfaces of the detectors get very hot and there is a risk of injury if they are not allowed 
to cool down. 

Note: 

This equipment has been tested and found to comply with the limits for a Class A digital 
device,  pursuant  to  part  15  of  the  FCC  Rules.    These  limits  are  designed  to  provide 
reasonable  protection  against  harmful  interference  in  a  residential  installation.    This 
equipment generates, uses, and can radiate radio frequency energy and, if not installed 
and used in accordance with the instructions, may cause harmful interference to radio 
communications.    However,  there  is  no  guarantee  that  interference  will  not  occur  in  a 
particular  installation.    If  this  equipment  does  cause  harmful  interference  to  radio  or 
television reception, which can be determined by turning the equipment off and on, it is 
suggested to try to correct the interference by taking one or more of the following steps: 

•  Reorient or relocate the receiving antenna. 
•  Increase the distance between the equipment and receiver. 
•  Connect the equipment to an outlet that is on a different circuit than the receiver. 
•  Consult the dealer or an experienced radio/TV technician for help. 

 

Caution: 

Changes  or  modifications  not  expressly  approved  in  writing  by  Gentec-EO  Inc.  may 
void the user’s authority to operate this equipment. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  4 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

TABLE OF CONTENTS 
 

WARRANTY .................................................................................................................................................. 1

 

CLAIMS ......................................................................................................................................................... 1

 

WARRANTY .................................................................................................................................................. 2

 

CLAIMS ......................................................................................................................................................... 2

 

P/N: 90043187 REV. D ................................................................................................................................. 2

 

SAFETY INFORMATION .............................................................................................................................. 3

 

TABLE OF CONTENTS ................................................................................................................................ 4

 

LIST OF ILLUSTRATIONS ........................................................................................................................... 4

 

1

 

THE TUNER SINGLE CHANNEL LASER POWER METER................................................................. 5

 

1.1

 

I

NTRODUCTION

 ............................................................................................................................................... 5

 

1.2

 

S

PECIFICATIONS

 ............................................................................................................................................. 6

 

1.3

 

F

RONT 

P

ANEL 

D

ESCRIPTION

 ........................................................................................................................... 7

 

1-

 

Backlight / On-Off .......................................................................................................................................... 7

 

2-

 

Speed / Zero ................................................................................................................................................... 7

 

3-

 

Needle low-high / Att Control Key ................................................................................................................. 8

 

4-

 

Display / Needle style .................................................................................................................................... 8

 

5-

 

Range ▲ / Range ▼ Control Keys ................................................................................................................ 8

 

1.4

 

C

ONNECTORS 

D

ESCRIPTION

 ........................................................................................................................... 9

 

1.5

 

D

ISPLAY DESCRIPTION

 .................................................................................................................................. 10

 

2

 

GETTING STARTED ........................................................................................................................... 12

 

2.1

 

P

OWER MEASUREMENT PROCEDURE

 ............................................................................................................. 12

 

3

 

BATTERIES ......................................................................................................................................... 13

 

3.1

 

B

ATTERY SELECTION

 .................................................................................................................................... 13

 

3.2

 

B

ATTERY INSTALLATION

 .............................................................................................................................. 13

 

3.3

 

B

ATTERY REMOVAL

 ..................................................................................................................................... 13

 

4

 

DECLARATION OF CONFORMITY .................................................................................................... 14

 

4.1

 

I

MMUNITY LIMITATIONS

 ............................................................................................................................... 15

 

5

 

APPENDIX ........................................................................................................................................... 16

 

5.1

 

R

ECYCLING AND SEPARATION PROCEDURE

................................................................................................... 16

 

5.2

 

S

EPARATION

 ................................................................................................................................................. 16

 

5.3

 

D

ISMANTLING PROCEDURE

 ........................................................................................................................... 16

 

 

LIST OF ILLUSTRATIONS 

 

F

IG

.

 

1-1

 

TUNER

 

F

RONT 

P

ANEL

 ....................................................................................................................... 7

 

F

IG

.

 

1-2

 

TUNER

 

C

ONNECTORS

 ....................................................................................................................... 9

 

F

IG

.

 

1-3

 

TUNER

 

LCD

 

D

ISPLAY

 ..................................................................................................................... 10

 

F

IG

.

 

1-4

 

N

EEDLE 

D

ISPLAY IN 

T

AIL MODE

 ......................................................................................................... 11

 

F

IG

.

 

1-5

 

N

EEDLE 

D

ISPLAY IN 

B

AR 

G

RAPH MODE

 ............................................................................................. 11

 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  5 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

THE TUNER SINGLE CHANNEL LASER POWER METER 

1.1  Introduction 

To obtain full performance from the TUNER, we recommend that you read this manual carefully. 

The TUNER is a numerical and digital needle LCD display laser power monitor that features outstanding 
efficiency, ease of use and long battery life. 

Unpacking 

Each Gentec-EO TUNER is thoroughly tested and calibrated prior to shipment. 

Visually inspect every TUNER unit after removing it from the shipping containers. If you see any damage, 
retain all packaging materials and shipping receipts. Any damage claim should be made promptly to the 
shipping company. Notify the nearest Gentec-EO representative concerning the claim, so that any repair 
or replacement can be arranged as soon as possible.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  6 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

1.2  Specifications 

The following specifications are based on a one-year calibration cycle, an operating temperature of 18 to 
28ºC (64 to 82ºF) and a relative humidity not exceeding 80%. 

Table 1-1 List of Specifications 

 

TUNER POWER METER SPECIFICATIONS 

Power Range 

10 pW to 10 kW (detector dependent) 

Power Scales  

(photodetector)

 

Autoscale

 

Power Scales  

(thermal head) 

25 scales: 10nW, 30nW, 100nW, 300nW, 1uW, 3uW, 10uW, 30uW, 100uW, 

300uW, 1mW, 3mW, 10mW, 30mW, 100mW, 300mW, 1W, 3W, 10W, 30W, 

100W, 300W, 1kW, 3kW, 10kW 

Compatible detector 
heads 

XLP12, all UP Series, PS 1.5k, PS 3k, PS 6k, all HP Series and all 

 

PH series

 

 

Digital Resolution (XLP-
12 detector) 

1 µW 

Digital Resolution 
(UP/PS thermal 
detector) 

1 mW 

Digital Resolution (HP 
detector) 

Autoscale 

Digital Resolution 
(photodetector)  

10 pW 

Monitor Accuracy 

±1 % 

5 µV  

Response Time 
(accelerated)

1

 

< 1sec 

Sampling Frequency 

16.7 Hz 

Analog out 

0-1volt (±1%) 

Digital Display 

77 x 58 mm LCD 

LCD segments 

96 (55 segments for needle) 

Needle accuracy 

0.9% 

Display Rate 

4 Hz 

Dimensions 

210 mm(W) x 122 mm (H) x 44 mm (D) 

Weight (including 
batteries) 

0.47 kg

 

Batteries (included) 

4 Alkaline AA batteries 

Current consumption 
(without backlight) 

5mA 

Backlight current 

60mA (with external 9 VDC power supply only) 

Battery life (with 
passive thermal 

500 hours  

                                                      

1

 Varies with detector head. 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  7 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

detector) 

External Power Supply 
(included) 

Input: 100/240 VAC 50-60 Hz, Output 9 VDC 1.66 A 

 

1.3  Front Panel Description 

 

Fig. 1-1 TUNER Front Panel

 

1- 

Backlight / On-Off 

 

Switches the TUNER on and off (press for at least 2 seconds to turn off the TUNER) and 
toggles backlight on and  off when the  TUNER  is on (backlighting  is available only  when 
the  TUNER  is  powered  by  the  external  9  VDC  power  supply).    Backlight  status  is 
memorized by the TUNER. 

 

 

If  no  detector  head  is  connected  when  the  TUNER 

is  powered  on,  the  message  “No 

 

 

Detector”  will  be  displayed.    Turn  off  the  TUNER;  connect  the  detector  to  the  Probe 

 

 

Input Jack (see section 1.4) and turn the TUNER on again. 

 

If  the  detector  is  not  recognized  by  the  TUNER

,  the  message  “Bad  Detector”  will  be 

displayed.   Make sure  you use a detector  head that is compatible  with the TUNER (see 
section 1.2 for the list of compatible detector heads). 

 

2- 

Speed / Zero 

 

Press  and  release  this  key  to  toggle  between  fast  and  slow  modes.  The  fast  mode 
enables anticipation, which shortens the natural response time of the detector.  The slow 
mode disables anticipation.  By default, the TUNER will always power-up in fast mode. 

 

 5 

 5 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  8 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

Press this key for at least 2 seconds for zeroing.  This cancels electronic offset, detector 
offset  and  ambient  light  on  the  detector.    The  TUNER  should  be  zeroed  before  taking 
measurements.  Refer to p.11, 

adjusting the zero (steps 8 to 11)

.  The TUNER will redo 

the zero every time you press the zero button for two seconds, reboot the  TUNER if you 
wish to remove the zero cancellation. 

 

 

 

3- 

Needle low-high / Att Control Key 

 

This key enables memorization of the minimum and maximum readings.  When enabled, 
two  blinking  needles  will  continuously  mark  the  current  minimum  and  maximum  values 
since the memorization function has been enabled.  (See point 6 and 7) for digital display 
of  minimum  and  maximum  values).  The  minimum  and  maximum  values  for  both  the 
needles  and  the  digital  high  and  low  can  be  reset  to  the  current  measured  value  by 
pressing this key 2 times quickly (like a mouse double click). 

 

The  blinking  needles  can  be  toggled  on  and  off  by  pressing  this  key  again,  but  once 
activated,  the  minimum  and  maximum  values  are  continuously  updated,  as  long  as  the 
TUNER is on. 

 

When  pressed  for  at  least  2  seconds,  Att  is  activated.  Gentec-EO

’s  photodetectors  are 

calibrated  with  and  without  attenuator,  with  both  responsitivity  data  captured  in  the 
detector  EEPROM  and  downloaded  into  the  instrument  upon  start-up.    This  key  allows 
the  user  to  select  the  proper  configuration.    The  “ATT  ON”  LCD  annunciator  message 
indicates whether the attenuator setting is activated. Press again for at least 2 seconds to 
deactivate  ATT, 

the  message  “ATT  OFF”  will  scroll  in  the  LCD  to  confirm  button  press 

action. 

4- 

Display / Needle style 

 

Toggles  the  display  between  measurement,  wavelength  selection  and  range  selection 
modes.  (See point 5 below for wavelength and range selection). 

 

When  pressed  for  at  least  2  seconds,  toggles  the  needle  display  between  the  following 
four  modes:  single  needle,  tail  needle  (displays  a  short  tail  and  a  long  tail  following  the 
needle  movement)  and  bar  graph  (displays  a  permanent,  continuous  trail  between  zero 
and the current needle position). 

 

5- 

Range ▲ / Range ▼

 Control Keys

 

 

Allows  the  user  to  select  the  wavelength  in  wavelength  display  mode,  or  the  range  in 
range  display  mode  and  in  measurement  mode.    When  selected  by  the  user,  the 
wavelength and the range are automatically stored in non-volatile memory.  The TUNER 
returns  to  this  same  wavelength  and/or  range  when  it  is  turned  on  again  with  the  same 
detector head. 

 

As the measured power varies upward, the range is automatically adjusted to provide the 
most  efficient  reading  There  will  be  no  automatic  scale  down  as  the  power  varies 
downward. 

 

The  TUNER  has  3  physical  gains  and  it  will  always  auto-scale  the  physical  gain.  The 
TUNER will display a little variation during a physical gain change. 

 

Range down can be manually selected using the range down button. 

 

6 - 

Digital high Control Keys 

Displays the maximum recorded values.  To return to measurement display mode, press the key again. 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  9 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

7 - 

Digital low / W/dBm Control Keys 

Displays the maximum recorded values.  To return to measurement display mode, press the key again. 

The dBm function is available by pressing more than 2 seconds on the Digital-Low control key. The Tuner 
will convert the displayed number in dBm (dB referenced to 1mW) and will display “dBm” message unit 
and then no units. To get back to the previous display in Watt, press again more than 2 seconds on the 
Zero control key. The ranges are also displayed in dBm. 

8 - 

Display 

Two parts digital/analog display LCD screen.  Refer to section 1.5 for a detailed description. 

 

1.4  Connectors Description 

 

Fig. 1-2 TUNER Connectors 

1- 

EXTERNAL POWER SUPPLY INPUT JACK. 

 

Input voltage required: 9  VDC/100 mA.  The external  power supply does not charge the 
batteries; it allows the use of the monitor  without batteries, with dead batteries or simply 
to avoid discharging the batteries inside the monitor. 

 

CAUTION 

Permanent damage to the optical meter may occur if an 
external power supply other than the Gentec-EO P/N 
200960A is used. Please call Gentec-EO or your local 
distributor if extra power supplies are needed. 

 

2- 

0 to 1 VOLT ANALOG OUTPUT: 

 

 

For  monitoring  laser  average  power  by  using  external  equipment  such  as  a  chart 

 

recorder, a computer with an analog interface, a voltmeter, etc.  

 

 

The output signal is the amplified power detector response, in fast or slow mode. 

 

 

The 1 V value corresponds to the full scale reading of the selected range. That provides 

 

the best signal-to-noise ratio.  The measured power is then related to the output voltage 

 

and to the selected range according to the following equations:  

selected

 

Range

 

of

Max 

output

V

Power

  

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  10 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

 

For example: 

 

 

0.25 V corresponds to 2.5 Watts on the 10 W range 

 

 

0.10 V corresponds to 30 milliwatts on the 300 mW range 

Specifications: 

 

 

Maximum output voltage: 

 

1 V 

 

 

Output impedance: 

 

 

2.5 k

 

 

 

Connector type:  

 

 

Female 1/8” jack  

3-  

PROBE INPUT JACK. 

 

The TUNER uses a DB-15 female connector to mate with the detector heads (probes). 

The  TUNER  works  with  all  Gentec-EO  thermal  power  detectors  currently  sold.    It 
automatically  recognizes  every  power  detector  head,  which  ensures  accurate  auto-
calibration.    More  importantly,  it  can  take  advantage  of  our 

Personal

 

wavelength 

correction

™.    It  reads  the  memory  in  the 

Smart  Interface

  connector  to  provide  a 

wavelength  correction  that  is  based  on  spectral  data  measured  from  that  specific 
detector. 

1.5  Display description 

 

Fig. 1-3 TUNER LCD Display 

The LCD provides measurement information, wavelength information, range information and other useful 
messages. 

The  upper  part  of  the  screen  displays  TUNER  setting  information  and  digital  values  from  the 
measurements.    The  numeric  display  minimum  resolution  is  0.1%  of  the  full  scale.    The  smallest  value 
that can be displayed is 0.001 

μW. 

The  lower  part  of  the  screen  displays  values  from  the  measurements  by  showing  a  needle  over  an 
analogue scale.  The two following pictures show the needle in the “Tail” and “Bar Graph” display modes, 
respectively. 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  11 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

 

Fig. 1-4 Needle Display in Tail mode 

 

 

Fig. 1-5 Needle Display in Bar Graph mode 

 

When  the  batteries  are  discharged  enough  to  compromise  the  measurement,  the  TUNER 

displays  “LO 

BATT

” instead of the measurement.  Refer to section 3 to replace the batteries. 

If  the  voltage  supplied  by  the  external  power  supply  is  lower  than  7  Volts,  the  TUNER 

displays  “LO 

JACK”. 

If  no  detector  head  is  connected  when  the  TUNER 

is  powered  on,  the  message  “No  Detector”  will  be 

displayed.  Turn off the TUNER; connect the detector to the Probe Input Jack (see section 1.4) and turn 
the TUNER on again. 

If the detector is not recognized by the TUNER

, the message “Bad Detector” will be displayed.  Make sure 

you  use  a  detector  head  that  is  compatible  with  the  TUNER  (see  section  1.2  for  the  list  of  compatible 
detector heads). 

 

 

 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  12 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

Getting Started 

This  section  contains  important  information  concerning  the  installation  and  operation  of  the  TUNER.  
The  TUNER  is  delivered  ready  to  use.    Just  install  the  4  AA  batteries,  insert  a  compatible  detector 
head in the Probe Input Jack (#3 in Figure 1-2) and press the 

I/O

 key.  

2.1  Power measurement procedure 

1-  Install the power detector on its optical stand. 

2-  First, slide the connector latch to the right to unlock the connector. 

3- 

Turn  the  TUNER  off

  and  connect  a  power  detector  head  to  the  TUNER  using  the 

PROBE 

INPUT  JACK

  (see  Fig.  1-2).    It  is  recommended  to  turn  the  TUNER  off  before  connecting  a 

new head in order to prevent any loss of information from the detector head’s EEPROM. 

4-  Slide the latch to the left to lock the connector into place. 

5-  Switch  the  TUNER  ON  using  the 

I/O

  key.    The  TUNER  displays  the  current  wavelength  for  a 

moment before displaying measurements.  If this wavelength is not the wavelength of the laser, 
go to step 6, otherwise proceed to step 8. 

6-  Select  the  proper  wavelength  using  the 

Display 

control  key  to  go  to  the  wavelength  display 

mode,  then  the 

Range  ▲  /  Range  ▼  control  keys  to  select  the  wavelength

.

  The  TUNER 

automatically  stores  the  wavelength  selection  in  non-volatile  memory  when  the  TUNER  is  set 
back to the measurement display mode, so it returns to this same wavelength when it is turned 
on again. 

7-  Return to the measurement mode by pressing the 

Display 

control key twice. 

Adjusting the zero (steps 8 to 11) 

8- 

Remove the detector’s protective cover. 

9-  Put  the  detector  into  the  laser  beam  path.    The  entire  laser  beam  must  be  within  the  sensor 

aperture.    Do  not  exceed  maximum  specified  power  densities.    For  the  most  accurate 
measurement, spread the beam across 60% to 80% of the sensor area.  Leave it there until the 
detector has reached its equilibrium temperature. 

10-  Block off laser radiation to the detector.  

The  power  read  by  the  TUNER  when  no  laser  beam  is  incident  on  the  detector  may  not  be 
exactly  zero.    This  is  because  the  detector  is  not  thermally  stabilized  OR  there  was  a  heat 
source  in  the  detect

or’s  field  of  view  when  you  turned  on  the  TUNER,  OR  the  internal 

electronic offset of the TUNER is not zero. 

11-  To reset the zero, wait until the reading has stabilized and press the 

Zero 

control key

 

on the 

front panel for at least 2 seconds.  The TUNER will 

display “----“ for a moment and then return 

to the normal measurement display.  You are now ready to make an accurate measurement.   

12-  Apply the laser beam to the detector. 

Notes: 

• 

Refer  to  the  specific  power  detector  documentation  for  complete  installation  and  operating 
instructions. 

• 

The power detectors are thermal sensors sensitive to temperature variations. 

For high-precision measurements, it is recommended to:  

• 

Allow the power detector’s temperature to stabilize before zeroing the TUNER.  

• 

Do not touch the detector itself when handling the power detector.  Touch only the stand. 

• 

Avoid forced airflow or drafts around the detector. 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  13 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

BATTERIES

 

3.1  Battery selection 

To avoid leakage and poor autonomy, it is highly recommended that only good quality, new and identical 
alkaline batteries be put into the TUNER.  Good quality rechargeable batteries can also be used with the 
TUNER and recharged with an external charger. 

The batteries need to be replaced  if the  TUNER 

displays “LO Batt” on its digital display or if it does not 

power-up when the I/O control key is pressed. 

3.2  Battery installation 

- Put the TUNER face down on a flat surface. 

- Lift the kickstand. 

- Open and remove the battery door. 

- Insert one battery in the nylon strap (to ease later removal) and insert it at one extremity of the TUNER 
battery compartment with the right polarity.  To avoid  intermittent contact and involuntary  disconnection, 
firmly  insert  the  batteries.  Insert  the  other  batteries  over  the  nylon  strap  with  the  right  polarity  so  that 
pulling on the strap will pull every battery out of the TUNER. 

       

       

 

- Put the battery door back in place. 

3.3  Battery removal 

- Put the TUNER face down on a flat surface. 

- Lift the kickstand. 

- Open and remove the battery door. 

- Hold down the TUNER and slowly pull on the nylon strap to remove the batteries.  Do not use a pointed 
tool  to  remove  the  batteries,  since  that  could  puncture  them.    The  batteries  need  a  firm  pull  to  be 
removed. 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  14 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

     

DECLARATION OF CONFORMITY

 

Application of Council Directive(s):   

2004/108/EC 

EMC Directive 

 

 

 

 

 

 

 

 

Manufacturer’s Name: 

 

 

Gentec Electro Optics, Inc. 

 

Manufacturer’s Address:   

 

445 St-Jean Baptiste, suite 160 
Québec (Québec), Canada   G2E 5N7 

 
 

European Representative’s Name:   

Laser Components S.A.S. 

 

Representative’s Address:   

 

45 bis Route des Gardes 
92190 Meudon (France) 

 
 

Type of Equipment: 

 

 

Laser Power Meter 

 

Model No.: 

 

 

 

TUNER 

 

Year of test & manufacture: 

 

2011 

Standard(s) to which Conformity is declared: 
EN 61326-1: 2006   Emission generic standard 
 

Standard 

Description 

Performance 
Criteria 

CISPR 11 :2009 

A1 :2010

    

Industrial, scientific and medical equipment 

– Radio-frequency 

disturbance  characteristics 

–  Limits  and  methods  of 

measurement

 

Class A 

EN 61000-3-2:2006  
+A1:2009 

Electromagnetic  compatibility  (EMC)  -  Part  3-2:  Limits  -  Limits  for 
harmonic  current  emissions  (equipment  input  current  <=  16  A  per 
phase) 

Class A 

EN 61000-4-2 2009 
 

Electromagnetic  compatibility  (EMC) 

–  Part  4-2:  Testing  and 

measurement techniques- Electrostatic discharge.

 

Class B 

EN61000-4-3 
2006+A2:2010 

Electromagnetic  compatibility  (EMC) 

–  Part  4-3:  Testing  and 

measurement  techniques-  Radiated,  Radio  Frequency, 
electromagnetic field  immunity test

 

Class A 

EN61000-4-4 
2004+A1:2010 

 

Electromagnetic  compatibility  (EMC) 

–  Part  4-4:  Testing  and 

measurement  techniques-  Electrical  fast  transient/burst 
immunity test.

 

Class B 

EN 61000-4-5 2006 

Electromagnetic  compatibility  (EMC) 

–  Part  4-5:  Testing  and 

measurement techniques- Surge immunity test.

 

Class B 

EN 61000-4-6 2009 

Electromagnetic  compatibility  (EMC) 

–  Part  4-6:  Testing  and 

measurements  techniques-  Immunity  to  conducted  Radio 
Frequency.

 

Class A 

EN61000-4-11 2004 

Electromagnetic compatibility (EMC) 

– Part 4-11: Testing and 

measurement  techniques-  Voltage  dips,  short  interruptions 
and voltage variations immunity tests

 

Class B 
Class B 
Class C 
Class C

 

I, the undersigned, hereby declare that the equipment specified above 

conforms to the above Directive(s) and Standard(s). 

 

Place: 

 Québec (Québec)

 

 

 

 

 

Date : 

  May 4, 2011

 

 

 

 

 

                                                                

 

 

(President) 

 

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  15 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

4.1  Immunity limitations 

The TUNER  will have small measurement deviation (<  60  uV)  when  a cellular phone is  within  15 cm of 
the TUNER and detector cable.  

 


background image

TUNER User’s Manual                                             Revision 4.2

 

  16 

 Gentec Electro 

– Optics, Inc.  All rights reserved                                                            

5  APPENDIX 

5.1  Recycling and separation procedure 

This  section  is  used  by  the  recycling  center  when  the  monitor  reaches  its  end  of  life.    Breaking  the 
calibration seal or opening the monitor will void the TUNER warranty.  

The complete Monitor contains 

1 Monitor 

1 9 volt AC/DC adaptor (not manufactured by Gentec-eo) 

4 AA Alkaline Batteries 

1 CD Instruction manual 

1 Calibration certificate 

5.2  Separation 

Paper : certificate. 

Plastic: stand, connector plate, battery door, monitor enclosure, keypad. 

Wires. 

AA batteries: inside battery compartment. 

Metal battery clips. 

Printed circuit board: inside the monitor featuring a liquid crystal display less than 100 cm

2

5.3  Dismantling procedure 

- Remove batteries. 

- Remove the posts on each side of the DB15 connector using pliers. 

- Open monitor by removing the Phillips head screws in the 4 corners. 

- Cut the wires on the PCB side and battery clips side with cutters. 

- Remove battery clips with pliers. 

 

 

 

 

Internal #103142 Rev. C 

 

 


background image