PDF User Manual

  1. Home
  2. Manuals
  3. GF Signet 4150 Manual

GF Signet 4150 Manual

Made by: GF
Type: Manual
Category: Measuring Instruments
Pages: 16
Size: 3.28 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

Signet 4150 Turbidimeter

3-4150.090  

Rev 

02/09

*3-4150.090*

Specifi cations

Measurement Range 

0 to 100.0 NTU or 0 to 1000.0 NTU

Accuracy
 

<40 NTU: 

 

±2% of reading or ±0.02 NTU  

 

 

 

whichever is greater

 

 >40 NTU: 

 

±5% of reading

 
Resolution 

 

0.0001 NTU (below 10 NTU)

Response Time   

Adjustable

Display 

 

Multi-Line LCD with backlighting

Alarm Relays (2)   

120-240 VAC, 2A Form C Relay 

Analog Output 

 

Active 4-20 mA/RS485, 600 

Ω

 drive

Wetted Materials

• Tubing: 

 

Vinyl 

•  Measuring cuvette: 

Borosilicate Glass 

•  Glass washer seal: 

Silicone 

•  Pressure regulator:   

Polypropylene

 

   

FPM

 

  

 

316 Stainless Steel 

 

 

 

 

Delrin® by Dupont™ 

•  Inlet tube:   

 

316 Stainless Steel 

Maximum Inlet Pressure 

Recommended 345 kPa (50 PSI)   

 

 

based on tubing connection

Integral Pressure Regulator   Rated 1380 kPa (200 PSI)

Flow Rate 

 

0.1 L/m to 1 L/m

 

  

 

(0.026 GPM to 0.26 GPM)

Operating Temperature  

1°C to 50°C (34°F to 122°F) 

Sample Temperature Range  1°C-50°C (34°F – 122°F)
Power Supply  

 

100 to 240 VAC, 47 to 63 Hz, 80 VA

Insulation Rating   

Double Insulated

 

  

 

Pollution Degree 2

 

  

 

Overvoltage Category II 

Altitude  

 

2000 meters (6562 ft.) maximum

Relative Humidity  

Maximum 95% RH non-condensing

Enclosure Rating   

IP 66 /NEMA 4X

Environmental Conditions 

Not recommended for outdoor use

Shipping Weight   

2.5 kg (5.5 lb)

Regulatory Compliance And Certifi cations

CE
White Light version  

Compliant to U.S. EPA 180.1

Infrared version   

Compliant to ISO 7027

ETL Listed 

 

UL 61010-1 and CSA C22.2 

 

   

No. 

61010-1

  

Chinese 

RoHS

 

  For detailed information go to www.gfsignet.com

Description

The Signet 4150 Turbidimeter provides accurate and reliable water quality monitoring for municipal and industrial applications.
The 4150 uses the Nephelometric method to calculate the turbidity of a sample as it 

fl

 ows through a viewing chamber. Models are 

available that use white light technology as required by the U.S. EPA 180.1 standard, or select the models that use infrared technology 
as required by ISO 7027.
Periodic calibration is mandatory with most turbidity systems, and the 4150 makes it fast and easy with sealed, reusable primary 
calibration standards.
Two dry contact relays serve as high or low alarms, with programmable setpoints and time-based delays to prevent false alarms.
Additional features include a bright backlight for the display, and a convenient holder for the cuvette during calibration.
An ultrasonic automatic cuvette cleaning system is standard for 0-100 NTU/FNU systems and optional for 0-1000 NTU/FNU systems.

Signet Turbidimeter

ENTER

Power Supply &
Wiring Terminals

NTU=FNU=FTU

 


background image

Signet 4150 Turbidimeter

Unpacking and Inspecting the Instrument and Accessories

The table below indicates the items in the turbidimeter shipment.

Item

Quantity

3-4150- ( __ ) Turbidimeter

1

3-4150.090 Instruction Manual

1

3-4150.380 Desiccant Pack

 (Do not open until ready to install) 

1

4150-0009 Measurement Cuvette

1

4150-0005 Tubing Kit:

1 ea. shutoff clamp
1 ea. backpressure valve
2 ea. connecting tubing with 

fi

 ttings for 

fl

 ow through assembly

1

Remove the instrument from the packing carton.  Carefully inspect all items to ensure that no visible damage has occurred during 
shipment. If the items received do not match the order, please immediately contact the local distributor or the Georg Fischer Signet 
Customer Service department.

Dimensions

13.5 in.

7.34 in.

4.0 in.

2.0 in.

8.9 in.

Top View

Front View

Side View

Cuvette

Holder

Measuring

Cell

Safety

This manual contains basic instructions that must be followed 
during the commissioning, operation, care and maintenance of 
the instrument.  The safety protection provided by this equipment 
may be impaired if it is commissioned and/or used in a manner 
not described in this manual.  Consequently, all responsible 
personnel must read this manual prior to working with this 
instrument.

Disclaimer

Georg Fischer Signet LLC accepts no responsibility for damage 
caused by the introduction of vapors, 

fl

 uids or other materials into 

the instrument process stream which is not compatible with the 
wetted materials. A list of the wetted materials can be found in the 
speci

fi

 cations on page 1 of this manual.

Material Safety Data Sheets (MSDS) are available online at
http://www.gfsignet.com/msds.htm.

It is the responsibility of the Distributor, Dealer, or Agent to 
provide a current copy of the MSDS to the Consumers of Georg 
Fischer Piping Systems products.  The information contained 
herein is presented in good faith and has been compiled from 
sources believed to be reliable.  It represents the best information 
currently available to us.  No warranty express or implied, or 
merchantability, 

fi

 tness or otherwise is made and we assume no 

liability resulting from its use.  This information is offered for your 
consideration and users should make their own investigation 
and veri

fi

 cation to determine the suitability of the information for 

their particular purposes.  In no event shall Georg Fischer Piping 
Systems, the parent company or its subsidiaries be liable for any 
claims, losses, or damages of any third party or for lost pro

fi

 ts 

or any special, indirect, incidental, consequential or exemplary 
damages, howsoever arising, even if Georg Fischer Piping 
Systems has been advised of the possibility of such damages.  
This information relates to the material designated and may not 
be valid for such material used in combination with any other 
materials nor in any process.

 


background image

3

Signet 4150 Turbidimeter

Installation and Commissioning

Mounting & Site Selection

The 4150 is designed to be wall mounted. All the necessary hardware is included.

•  Mount the 4150 within 2-3 meters (6-10 ft) of the sampling point to ensure a 

quick response time.

•  For ease of service there should be about 20 cm (8 in.) free area above the 

instrument for calibration and cuvette maintenance.

•  Choose a location that is easily accessible for operation and service.
•  The display should be positioned at eye level.
•  The overall mounting dimensions of the instrument are shown on page 2.
•  Four pan head screws and four wall anchors are provided to mount the wall 

bracket.

Plumbing

•  Use 4.75 mm (3/16 in.) ID, 8 mm (5/16 in.) OD 

fl

 exible tubing for the 

water supply connections.

•  Opaque tubing (not supplied) should be used if the tubing will be 

exposed to sunlight, to prevent algae growth. 

•  The 4150 requires only 1 psi head pressure to operate.
• The 

fl

 ow through cuvette is rated for a 

fl

 ow of 100 mL/m to 1 L/m (0.026-

0.26 GPM).

•  The integral pressure regulator is rated for a maximum pressure of 200 

PSI.

 It is factory adjusted. Do not tamper with the regulator.

• 

Inlet water pressure should not exceed 50 PSI to avoid damage to 
the tubing connection to the regulator. 

•  Fluid temperature must not exceed 50°C (122°F).
•  The shutoff clamp is used to interrupt the 

fl

 ow during cuvette 

maintenance.

•  Route the sensor drain tubing to a suitable drain. Do not  reintroduce the 

drain sample to the process stream.

Signet Turbidimeter

ENTER

Sensor
Interconnect
Cable

Provide 8 in. clearance
for maintenance.

Water IN

from source

Measuring

Cell

Dessicant access

Water OUT
to Drain

Shutoff
Clamp

Pressure
Regulator

Backpressure
Valve

Drain Vent

DESSICANT COMPARTMENT
OPEN ONLY FOR MAINTENANCE

Adjust the Flow to Eliminate Bubbles 

The cuvette must be free of air bubbles to provide accurate measurement. The 4150 provides two tools to make this adjustment.

Open the SHUTOFF CLAMP to allow water to 

fl

 ow through the cuvette.

1. If the 

fl

 ow is gravity-fed, remove the screw that is blocking the 

DRAIN VENT

. This allows for atmospheric equalization and helps to 

eliminate bubbles. The vent may leak for a few seconds, until the 

fl

 ow is well established.

2. Adjust the 

BACKPRESSURE VALVE

 to prevent air from coming out of solution, which may be observed as tiny air bubbles in the 

cuvette:

•  Remove the measuring cuvette from the measuring cell and place it in the cuvette holder so it is visible while making these 

adjustments.

•  If bubbles are visible inside the cuvette, turn the BACKPRESSURE VALVE until the bubbles disappear.

Backpressure

Valve

Shutoff
Clamp

Backpressure

Valve

Drain vent

To Drain

DESSICANT COMPARTMENT
OPEN ONLY FOR MAINTENANCE

Pressure
Regulator

The Inlet pressure 
regulator is factory 
set. Do not adjust!

Contact the factory 
if inlet pressure 
will exceed 50 PSI! 
A modifi ed tubing 
connection may be 
required.

 


background image

Signet 4150 Turbidimeter

Electrical Connections

All of the electrical connections to the 4150 are made inside the 
power supply, located to the right of the display.
The connections are labeled within the terminal box and are self-
descriptive.

Follow all local and government recommendations and methods 
for installation of electrical connections to and between the 
instrument and other peripheral devices.

An external grounding terminal is provided for use in the most 
extreme electrical noise environments. It is not required for most 
installations.

This instrument requires AC voltages that can injure or kill.
Wiring should be done by qualifi ed personnel only.

WARNING

Normally Closed

Normally Open

Common

Relay
Coil

4150 Internal Power

V+

V-

External
Power 
Supply

Normal Condition

Normally Closed

Normally Open

Common

4150 Internal Power

V+

V-

External
Power 
Supply

Alarm Condition
or Power Failure

Relay
Coil

NC

COM

NO

NC

COM

NO

ALARM2 ALARM1

4-20 mA

RS-485

+

-

A

B

Removable

Terminals

AC Cable

Restraint

G

L1 L2

115/230 VAC

Ground

90-250 VAC

47-63 Hz

}

Power

•  Install a circuit breaker in the AC line before the 4150 power 

connection to allow for service.

•  The 4150 is not supplied with a power cord.  
•  The power cable bulkhead will accept cable diameters from 

5.8 mm (.230 in.) up to 10 mm (.395 in.).

•  All terminals are designed to accept wires in the range of 14-

28 AWG.

•  All wires should be stripped to a length of 6 mm (¼ in.).
•  A strain relief strap is provided to reduce tension on the AC 

power terminals.

RS-485

•  The RS-485 half-duplex (2-wire) digital interface operates 

with differential levels that are not susceptible to electrical 
interferences.

•  The last device on each bus may require terminating with a 

120-ohm resistor to eliminate signal re

fl

 ection on the line.

•  Do not run RS-485 cables in the same conduit as power.

Relays

Both Alarm relays are con

fi

 gured for fail-safe operation.

The normal condition is with power applied to the 4150 and in a non-alarm condition.

•  The relays are rated for 2 A maximum.
•  If power is removed from the 4150, the relays will be in an alarm state.

4-20 mA

•  The active 4-20 mA output is driven by a 15 VDC power 

source and can drive external loads up to 600 ohms.

•  Do not run 4-20 mA cables in the same conduit as power. 

Attach a wire to the small spade lug and connect the housing to a 
local Earth ground if necessary.

 


background image

5

Signet 4150 Turbidimeter

The Display

All of the elements on the LCD are identi

fi

 ed here:

The Keypad

4 keys are used to set up, calibrate and operate the 4150.

Signet Turbidimeter

Turbidity Measurement

Security Indicator

Offset Indicator

Alarm Indicators

Drying Period

Calibrate Menu

Options Menu

Units display
Error Messages

ENTER

ENTER Key

Press to access menus
Press to save changes

UP, DOWN keys

Scroll through Menu options

RIGHT key

Select item or character to edit

During normal operation, the instrument will have the Units selection displayed on the lower row of the 
display and the measured reading on the upper row of the display.

The instrument is equipped with a security access code to protect the instrument settings from 
inadvertent modi

fi

 cation.

If the CODE feature is enabled in the OPTIONS menu, the KEY icon is displayed. 

If OFFSET is displayed, the 4150 output has been adjusted to match an external reference 
measurement.

The ALM1 and ALM2 icons are displayed if the 4150 measurement has exceeded a HIGH or LOW 
alarm setpoint.

DRY indicates that the 4150 is in a drying period, following the replacement of the cuvette. The internal 
fan will circulate air to clear condensation from the cuvette.
NOTE: DRY is not an alarm condition. During a DRY period the ultrasonic cleaning system is disabled, 
but the instrument is still operating.

CAL is displayed when the 4150 is being calibrated.
•  The alarm relays are held in their NORMAL condition when CAL is displayed.
•  The 4150 will return to normal operation if no keys are pressed in the CAL menu for 6 minutes. 

OPTIONS is displayed when the 4150 is being con

fi

 gured.

•  The alarm relays are held in their NORMAL condition when OPTIONS is displayed.
•  The 4150 will return to normal operation if no keys are pressed in the OPTIONS menu for 10 

minutes. 

Display Icons and Functions

 


background image

Signet 4150 Turbidimeter

Security Access Feature

If CODE is enabled in the OPTIONS menu, the Access code (

-

-

-

) must be entered to gain access to CAL or OPTIONS menus.

•  Press and hold the 

 key for 2 seconds to access the CALIBRATE menu.

•  Press and hold the 

 key for 5 seconds to access the OPTIONS menu.

Signet Turbidimeter

ENTER

5s

ENTER

ENTER

ENTER

ENTER

Signet Turbidimeter

ENTER

2s

ENTER

ENTER

ENTER

ENTER

100 NTU Calibration for

4150-3, 4150-4 Only

1. Press the  

 key

1. Press the  

 key

2. Press the  

 key

2. Press the  

 key

3. Press the 

 key

3. Press the 

 key

4. Press the 

 key

4. Press the 

 key

5. Press the 

 and 

 keys to select the NTU standard value 

needed.

Alarm relay contacts are held at the last valid condition while the 
instrument is in the CALIBRATE or in the OPTIONS menu.

While in the CALIBRATE menu, the instrument has a time-out 
feature that automatically returns the system to normal operation 
if no keys are pressed for 6 minutes.

CALIBRATE

The 4150 is calibrated and tested prior to leaving the factory, and 
is ready to use directly out of the box.

The EPA recommends that on-line turbidity systems be calibrated 
at least once every three months if they are used for EPA 
reporting.

The 4150 is designed to recognize the speci

fi

 c NTU values of 

certifed calibration standards to make periodic recalibration very 
safe, quick and simple. The CALIBRATE menu is dedicated to 
this method of calibration.

The OFFSET function found in the OPTIONS menu allows a 
single-point alignment with an external measurement device. 

OFFSET is not a substitute for calibration, but it may be useful if 
standard solutions are not readily available.

5. Press the 

 and 

 keys to select the OPTIONS function that 

will be modi

fi

 ed.

While in the OPTIONS menu, the instrument has a time-out 
feature that automatically returns the system to normal operation 
if no keys are pressed for 10 minutes.

Access to CALIBRATE Menu

Access to OPTIONS Menu

Output Options:

Select the 4-20 mA output or the RS485 
Digital output.

Extended Options:

Display many single-set options like 
brightness and decimal location.
(See OPTIONS: EXTENDED, pg 11)

Code option:

Turn the Access Code ON or OFF.

Offset option:

Turn the OFFSET option ON or OFF.

Alarm 1 Option:

Turn on the 1st ALARM RELAY.

Alarm 2 Option:

Turn on the 2nd ALARM RELAY.

ALM RELAY 

is now LOCKED

ALM RELAY 

is now LOCKED

 


background image

7

Signet 4150 Turbidimeter

Calibration Procedure

1. 

Access the Calibrate menu by pressing the ENTER key for 2 seconds.

2. 

Enter the access security code if necessary. (See pg. 6)

 

CAL will be illuminated on the display.

 

The upper display will show the measured NTU value.

 

The lower display will alternate the value of the 

fi

 rst NTU standard and 

3. 

Remove the measuring cuvette and place it in the cuvette holder.

If the Signet 4150 will be used over the entire range, a 3-point calibration is required. Calibration standard kits include three cuvettes 
and three different turbidity solutions.

•  For 4150-3 (range 0-100 NTU):  P/N 3822-4001 (code 159 001 585) calibration kit, turbidity, 100, 10 & 0.02 NTU
•  For 4150-1 and -5 (range 0-1000 NTU):   P/N 3822-4003 (code 159 001 586)  calibration kit, turbidity, 1000, 10 & 0.02 NTU

•  A Formazin stock solution kit is also available (P/N 3822-4002). Formazin is very unstable, so it is important to ensure that a 

fresh stock suspension of Formazin is used to achieve the accuracy quoted for the instrument. The primary calibration standards 
are much more stable than Formazin and have a shelf life of 12 months.

•  If the application is limited to measurements below 10 NTU, such as potable water, a 2-point calibration may be performed using 

only a 10 NTU and a 0.02 NTU standard.

•  New calibration standard cuvettes should be indexed to the speci

fi

 c instrument before being used the 

fi

 rst time. See pg. 14.

IMPORTANT! 

 Calibration does NOT remove OFFSET values.

ALWAYS turn the OFFSET function OFF (Options menu) before performing a full calibration.

Calibration Standards

Calibration Tips

•  Keep the measurement cell covered as much as possible during the calibration period.
•  Replace the cuvette immediately after the calibration to prevent accelerated saturation of the desiccant.
•  During calibration, the fan is turned off to extend the life of the desiccant.
•  The fan will be turned on during calibration countdowns and after returning to the AUTO mode or after 

fi

 ve minutes, which ever 

comes 

fi

 rst.

10 NTU

10 NTU

Signet Turbidimeter

ENTER

Signet Turbidimeter

ENTER

CAUTION!

Do not touch the glass surface of the cuvette!
Substances on the surface of the glass
will cause errors in the measurement.

Do not remove the glass cuvette from the cap while 
holding the assembly over the measuring cell.

Do not allow any debris to fall into the measuring cell.

Do not leave the measuring cell open longer than 
necessary. Extended exposure to the atmosphere may 
shorten the effective life of the desiccant.

 


background image

Signet 4150 Turbidimeter

Calibration Error 

If the 4150 displays Error after calibration, the calibration standards were out of calibration range, or the standard did not match the 
NTU value being requested by the 4150. 

The instrument cannot be used until the error is resolved.

1.  Check the standards and recalibrate:
 

Press and hold the ENTER key for 2 seconds and start the calibration sequence again.

2.  Restore the factory calibration:

•  Press and hold the 

 key and the 

 key for about 2 seconds.

Signet Turbidimeter

ENTER

Signet Turbidimeter

ENTER

Signet Turbidimeter

ENTER

Signet Turbidimeter

ENTER

5. 

Insert the 10 NTU Standard cuvette.

• Press 

the 

 key. The 4150 will begin the calibration process. The upper display will count down the progress.

 

When the process is complete, the upper display shows the new NTU value, while the lower display alternates the value of the NTU 
standard and  

6. 

Insert the 0.02 NTU Standard cuvette.

• Press 

the 

 key. The 4150 will begin the calibration process. The upper display will count down the progress.

 

When the process is complete, the upper display shows the new NTU value, while the lower display alternates the value of the NTU 
standard and  

• Press 

the 

 key.  The lower display will show the next NTU Standard value and  

 

When calibration is complete, press the 

 and 

 keys 

together to exit the CAL menu and return to normal operation.

Signet Turbidimeter

ENTER

Signet Turbidimeter

ENTER

4. 

Insert the Standard cuvette requested by the 4150.

 The 

fi

 rst NTU standard requested will be either 1000 or 100, depending on the range of the 4150.

 

If the application will be limited to low Turbidity values (less than 10 NTU), press the 

 key to go to the 10 NTU calibration.

• Press 

the 

 key. The 4150 will begin the calibration process. The upper display will count down the progress.

 

When the process is complete, the upper display shows the new NTU value, while the lower display alternates the value of the NTU 
standard and  

• Press 

the 

 key.  The lower display will show the next NTU Standard value and  

 

IMPORTANT!

This action will reset all of the 

CALIBRATE

 and 

OPTION

 

menus back to factory settings.

 


background image

9

Signet 4150 Turbidimeter

Instrument Offset

The OFFSET function allows the 4150 to be calibrated to match a process grab sample.  This procedure is not recommended in lieu of 
regular instrument calibration.  This function is only useful at turbidity levels in the immediate vicinity of the grab sample and not in the 
full range of the instrument. The OFFSET icon is illuminated whenever an offset is applied.

The maximum offset is ± 1.00 NTU. If the 4150 varies from the sample by more than 1 NTU a full calibration is required.

IMPORTANT! 

 Offset values will remain in the memory even after a complete calibration.

ALWAYS turn the OFFSET function OFF (Options menu) before performing a full calibration.

1.  Insert a cuvette 

fi

 lled with the sample water and record the turbidity reported by the 4150.

2.  Measure the turbidity of the sample using a laboratory turbidimeter.
3.  Compare the turbidity reported by the 4150 to that obtained in the laboratory. Subtract the Lab value from the 4150 value.

8.  If the OFFSET value is a POSITIVE number, press the 

 key to scroll to a 

positive offset value.

Press the 

 and 

 keys simultaneously to exit the OPTIONS menu and return to normal operation.

Example:
Laboratory sample measures 4.20 NTU, while the 4150 reads 4.50 NTU, or 0.3 NTU high. The OFFSET needs to be set to -0.3

4.  Press and hold the ENTER key until OPTIONS is illuminated on the display.
5. Push the 

 key until OFST is displayed on the lower row.

6.  If the upper display says ON, go to step 7.
 

If the upper display says OFF:

• Press 

the 

 key and then 

 to turn the OFFSET function ON.

•  Press the ENTER key to save the change.
• Press 

the 

 key to display the OFFSET value in the upper display.

7. Press the 

 key. The upper display will begin blinking.

 

If the OFFSET value is a NEGATIVE number, press the 

 key to scroll to a 

negative offset value.

9.  Press the ENTER key to save the changes.

Scroll UP for positive 
Offset.

Scroll DOWN for 
negative Offset.

 


background image

10 

Signet 4150 Turbidimeter

Options menu 

OPTIONS: Output (O/P)

1.  The lower display should show O/P. If it does not, Push the 

 

key until the lower display shows O/P.

2. Press the 

 key. The upper display will begin blinking.

3. Press the 

 or 

 keys to scroll to the 4-20 mA output or the 

RS485 Output.

4.  Press the ENTER key to save the selection.
5.  See instructions for 4-20 mA settings below.
6.  See instructions for configuring the RS-485 port below.

Setting the 4-20 mA Output

1. Press the 

 key. The lower display shows LOLM (Low Limit).

2. Press the 

 key. The upper display will begin blinking.

3. Press the 

 and 

 keys to scroll to the NTU value at 4 mA.

4.  Press the ENTER key to save the setting.
5. Press the 

 key. The lower display shows UPLM (Upper 

Limit).

6. Press the 

 and 

 keys to scroll to the NTU value at 20 mA.

7.  Press the ENTER key to save the setting.

Press the 

 and 

 keys simultaneously to exit the OPTIONS 

menu and return to normal operation.

Confi guring the RS–485 Port Output

1. Press the 

 key. The lower display shows BAUD.

2. Press the 

 key. The upper display will begin blinking.

3. Press the 

 and 

 keys to scroll to the baud rate required 

(1200, 2400, 4800, 9600, or 19200)

4.  Press the ENTER key to save the setting.
5. Press the 

 key. The lower display shows ADDR (Address).

6. Press the 

 key. The upper display will begin blinking.

7. Press the 

 and 

 keys to scroll to the correct address.

 

Address options are from 1 to 255.

 

Press the ENTER key to save the setting.

8. Press the 

 key. The lower display shows MBUS.

9. Press the 

 key. The upper display will begin blinking.

10. Press the 

 and 

 keys to select ASCII or RTU.

 

Press the ENTER key to save the setting.

Press the 

 and 

 keys simultaneously to exit the OPTIONS 

menu and return to normal operation.

The OPTIONS menu contains settings that are generally set once 
then rarely changed. The factory setting for all the items in the 
OPTIONS menu is OFF. When an option is turned ON, settings 
related to that option will appear in the menu.

Signet Turbidimeter

ENTER

5s

Output Options:

Select the 4-20 mA output or 
the RS485 Digital output.

Extended Options:

Enable access to performance and 
setup options like brightness and 
decimal location.
(See OPTIONS: EXTENDED, pg 11)

Code option: 

Turn the Access Code ON or OFF.

Offset option:

Turn the OFFSET option ON or OFF.

Alarm 1 Option:

Turn ALARM RELAY 1 ON or OFF.

Alarm 2 Option:

Turn ALARM RELAY 2 ON or OFF.

ENTER

ENTER

Press and hold the ENTER key until 
OPTIONS is illuminated on the display.
Enter the ACCESS CODE if the key is 
displayed.

Section 7.2

Section 7.3

 


background image

11

Signet 4150 Turbidimeter

OPTIONS: Extended: Response Time

The response time determines how quickly the 4150 responds to changes in the NTU measurement.
1 = approx. 5 seconds, 100 = approx. 500 s.
Select a high value to avoid reading air and other anomolies.
Select a low value to react to rapid changes in the measurement.
Set the response time using the 

 and 

 keys. Press the ENTER key to save the new setting.

OPTIONS: Extended: Display Resolution

The 4150 can display up to four decimal places. The factory setting is 0.01.  Set the resolution by 
pressing the 

 or 

 keys. Press ENTER key to save the selection.

OPTIONS: Extended: LCD Backlight Brightness 

The LCD backlight brightness can be adjusted. Ten levels are available. The default brightness is 8.
Set the brightness by pressing the 

 or 

 keys. Press ENTER key to save the selection.

OPTIONS: Extended: Units

All instruments are shipped from the factory set in NTU mode. The 4150 can also display in FNU 
(Formazin Nephelometric Units). Use the 

 and 

 keys to change to FNU. 

Press ENTER key to save the selection. 

OPTIONS: Extended

Press the 

 key. The lower display shows EXTD (Extended menu)

OPTIONS: Extended: RS485 Parameters

These menus will only appear if the RS485 is selected as the OUTPUT option. Factory settings are:

8 Bits

No (nOnE) Parity

1 Stop Bit. Make selections using the 

 and 

 keys.

Press the ENTER key to save the selection.

OPTIONS: Extended: Ultrasonic Cleaning (3-4150-3 and 3-4150-5 models only)

The default mode is On. Use the 

 and 

 keys to turn OFF.

Press the ENTER key to save the selection.

OPTIONS: Extended: Desiccant Alarm

Turn ON to activate the alarms and send the 4-20 mA to 2 mA when the humidity detector indicates 
that the internal environment is close to the point where humidity could cause condensation.
Press the ENTER key to save the selection.

The EXTENDED menu contains many single-set options like brightness and decimal location.
The EXTENDED option will always revert to OFF when you exit the menu.

 


background image

12 

Signet 4150 Turbidimeter

The 4150 has two relays that can be programmed to serve as High or Low alarms. The relays are factory set to OFF.
When activated, each relay can be programmed with a Setpoint (S/P),  a Delay ON time and a Delay OFF time.

Signet Turbidimeter

ENTER

ENTER

ENTER

ENTER

ENTER

OPTIONS: Alarm Relays

1. Press the 

 key to scroll to the ALM1 display.

3. Press the 

 and 

 keys to select Hi or Lo operating mode for this relay. 

Press the ENTER key to save the setting.

4. Press the 

 key. The lower display shows S/P (Set Point).

 

The ALM1 or ALM2 icons will be displayed to indicate which relay is being set.

 Press 

the 

 key. The upper display begins blinking.

 Press 

the 

 and 

 keys to scroll to the NTU value where the Alarm will be activated.

 

Press the ENTER key to save the setting.

 Press 

the 

 key to scroll to the next menu item.

5. Press the 

 key. The lower display shows DLY 

 (Delay ON time).

 Press 

the 

 key. The upper display begins blinking.

 Press 

the 

 and 

 keys to scroll to the time (from 1 second to 100 seconds) before 

the Alarm relay will be activated.

 

Press the ENTER key to save the setting.

 Press 

the 

 key to scroll to the next menu item.

6. Press the 

 key. The lower display shows DLY 

 (Delay OFF time).

 Press 

the 

 key. The upper display begins blinking.

 Press 

the 

 and 

 keys to scroll to the time (from 1 second to 30 seconds) before the 

Alarm relay will be deactivated.

 

Press the ENTER key to save the setting.

Press the 

 and 

 keys simultaneously to exit the 

OPTIONS menu and return to normal operation.

Press the 

 key again to go to ALM2 display.

Press the 

 key to begin programming ALM2. The upper display 

begins blinking. Repeat steps 3 through 6.

2. Press the 

 key to begin programming Alarm #1. The upper 

display begins blinking.

 


background image

13

Signet 4150 Turbidimeter

Vapor Purge System

If the air inside the 4150 is too humid it will cloud the optical glass cuvette surface and will interfere with the transmission of light. A 
continuous vapor purge system in the 4150 keeps the inside of the instrument dry.

•  The air inside the unit is dried by a replaceable desiccant pouch (order number 3-4150.380) located under the measuring cell.
•  System heat is used to warm the air.
•  A fan inside the instrument continuously circulates heated dry air around the measuring cell and the 

fl

 ow-through cuvette.

•  A humidity detector in the 4150 continuously monitors the desiccant condition.

 

When the internal 

environment approaches the point where condensation may occur, the instrument will display 
DESC as a warning that the desiccant pouch needs to be replaced.

•  The DESC warning does not interrupt the operation of the 4150, but the measurement is not 

reliable until the condensation is cleared from the glass cuvette.

•  If DESC ALARM is turned on (see OPTIONS: EXTENDED on pg 11), both relays will go into alarm 

condition and the 4-20 mA output will be forced to 2 mA.

Desiccant Pouch
3-4150.380

DESSICANT COMPARTMENT
OPEN ONLY FOR MAINTENANCE

Cleaning the Flow Through Cuvette

All glass cuvettes must be clean and free of marks or scratches.
•  Clean the interior and exterior with a detergent solution and then rinse several 

times with distilled or deionized water.

To replace a cuvette:

•  Shut off the 

fl

 ow using the shutoff clamp.

•  Unscrew the old cuvette and replace with a fresh clean one.
•  Do not touch the glass surface of the 

cuvette. Fingerprints can compromise the 
accuracy of the measurement.

Replacing Desiccant

•  To install or remove the desiccant pouch, open the round port 

on the side of the 4150.

•  Open the desiccant bag and place it inside the instrument.
•  Close the access door quickly to minimize exposing the 

desiccant to the atmosphere.

Operating Tip

•  To speed up recovery after a DESC alarm, disconnect the 

sensor interconnect cable (Connector on the right side of the 
power supply) for 2 seconds and then reconnect it.

Signet Turbidimeter

ENTER

Measuring Cell

Sensor Interconnect cable

CAUTION

Move the cuvette away from 
the measuring cell opening to 
prevent accidental spills into the 
instrument.

 


background image

14 

Signet 4150 Turbidimeter

Ultrasonic Cleaning

•  The 3-4150-3, 3-4150-4 and 3-4150-5 feature a special 

fl

 ow-through cuvette with an ultrasonic piezo attached to the 

base. Spring contacts inside the cuvette housing send a high 
frequency signal to the piezo transducer. The vibration helps 
prevent scaling and other suspended solids in the process 
water from adhering to the inside surface of the optical glass.

•  The ultrasonic system is not intended to clean dirty cuvettes, 

or to replace manual cleaning entirely, but it will dramatically 
increase the time between cleaning.

Flow-through Cuvette with
Ultrasonic Transducer
(4150-0004)

Ultrasonic Contact Springs

Ultrasonic cleaning feature 
in 3-4150-3, 3-4150-4 and 
3-4150-5 only.

•  The 4150 will show CLN on the lower display if a problem is 

detected with the cuvette:

•  Incorrect cuvette installed.
•  No contact between cuvette piezo and springs.

•  CLN is an alarm condition. Any active relays will be set to alarm state and the 4-20 mA will be locked to 2 mA.
•  After installing a cuvette, there will be a 30 minute period 4150 will show DRY on the lower display. During this time, the 

ultrasonic circuit will not operate to allow the Vapor Purge system to remove all moisture from the ultrasonic transducer.

•  The Vapor Purge system will NOT remove large droplets of water, only residual moisture. The DRY message is normal and is 

not considered an alarm condition; therefore no alarms will be implemented. If the cuvette is removed during this period no CLN 
alarm is posted until the 30 minute DRY period times out.

Indexing Calibration Cuvettes

To achieve the greatest accuracy, and account for normal scratches and aberrations in cuvette glass when calibrating, calibration 
cuvettes should be indexed using the indexing rings supplied with the Standard Kits.

The following steps allow repeatable indexing of calibration standards:
1.  With the instrument in AUTO mode, remove the measuring cuvette and insert the standard cuvette.
2.   Slowly rotate the cuvette inside the optical well, one complete revolution (360º).
 

While rotating the standard slowly, observe the measured turbidity and locate the position of the cuvette having the lowest reading.

3.   With the calibration standard positioned at the location having the lowest turbidity reading, install the Indexing Ring over the cap on 

the standard so that the pointer of the Indexing Ring faces directly forward.

When using the standards in future, always insert the standard so that the pointer of the indexing ring faces forward. Slowly rotate the 
standard back and forth about 5° to 

fi

 nd the lowest point. The standard is now indexed and ready for use.

 

The external surface of the ultrasonic cuvette must be completely dry before it is inserted into the sensor. If there is 
any visible moisture present on the cuvette or transducer, there is a great risk of damaging the sensor electronics and 
the transducer.

Cuvette Indexing Ring

 


background image

15

Signet 4150 Turbidimeter

Troubleshooting 

Signet 4150 Fault Detection

The Signet 4150 performs continuous diagnostic monitoring. There are three levels of fault detection: warnings, errors and failures.

Warning

A warning is simply a screen indication of a problem. The measurement is not interrupted and no alarms are activated.
•  When the desiccant becomes saturated, a screen warning of 

DESC

 will appear.

Error

An error indicates a failure or a problem that usually can be corrected. These errors include:
•  4-20 mA loop open (

MA

)

•  Bad calibration (

CAL

)

•  Desiccant alarm activated and replacement required (

DESC

)

•  Ultrasonic transducer is not making contact or the 

fl

 ow through cuvette has been removed (

CLN

)

•  Lamp out (

LAMP

)

 

The source lamps in the Signet 4150 are designed for long life. The IR lamp is rated for 10 years and the white light version is rated 
for 7 years.

 

Replacing the lamp involves signi

fi

 cant disassembly and handling delicate components.

 

We recommend returning the unit to the factory for this service. If 

fi

 eld replacement is required, contact the factory for assistance.

If any of these conditions occurs, both relays will be set to the alarm state and the 4-20 mA output will be held at 2 mA.

System FAIL Message

A failure is a system fault. This is NOT a problem that the operator can correct, and the unit must be returned to the factory for service. 
If a failure occurs, the instrument will not function properly and will display the word 

FAIL

 on the lower row. Both alarm relays will be 

activated and the 4-20 mA output will be held at 2 mA (if 4-20 mA is selected).

Troubleshooting

Symptom

Possible Cause

Solution

Lower display shows MA

4-20 mA loop open

Check Loop wiring

Lower display shows DESC

Desiccant pouch saturated

Change desiccant pouch

Lower display shows LAMP

Lamp failed

Contact factory for assistance

Lower display shows FAIL

Major system fault

Return to factory for service

Readings are higher than expected

Bubbles in solution

Condensate or leaky cuvette

Flow through cuvette dirty

Instrument out of calibration

Ensure that the drain vent is open and is 
not obstructed.
Apply backpressure. 
Check 

fl

 ow through cuvette for condensate 

or leaks.

Clean cuvette. 

Recalibrate

Readings are erratic

(1) Bubbles in solution
(2) Debris in 

fl

 ow through

(1) See above
(2) Clean debris from cuvette

Readings are lower than expected

Instrument out of calibration

Recalibrate. 

Upper display 

fl

 ashes

Sample Over-Range

Check sample. Sample may be too high to 
read.

 


background image

 

 

   

 

 

Signet Part Number

Code

Description

3-4150-1

159 001 596

Turbidimeter, White Light, 0 to 1000 NTU/FNU

3-4150-2

159 001 597

Turbidimeter, Infrared, 0 to 1000 NTU/FNU

3-4150-3

159 001 598

Turbidimeter, White Light, 0 to 100 NTU/FNU With AutoClean

3-4150-4

159 001 599

Turbidimeter, Infrared, 0 to 100 NTU/FNU With AutoClean

3-4150-5

159 001 600

Turbidimeter, White Light, 0 to 1000 NTU/FNU With AutoClean

3-4150-6

159 001 601

Turbidimeter, Infrared, 0 to 1000 NTU/FNU With AutoClean

3822-4001

159 001 585

*Calibration Kit, Turbidity, 100, 10 & 0.02 NTU/FNU

3822-4003

159 001 586

*Calibration Kit, Turbidity, 1000, 10 & 0.02 NTU/FNU

3-4150.380

159 001 588

Replacement Desiccant

3-4150.381

159 001 613

Replacement Desiccant Cap with Gasket

4150-0007

159 001 602

Replacement cuvette set (3 glass cuvettes)

4150-0004

159 001 589

Replacement cuvette with ultrasonic transducer

3822-4002

159 001 591

*Formazin Stock Kit

3822-4000

159 001 592

*Formazin Stock Solution, 4000 NTU/FNU, 500 ml

4150-0001

159 001 593

Pressure Regulator

4150-0003

159 001 587

Stilling/Bubble Chamber

4150-0005

159 001 595

Tubing Kit:
shutoff clamp
backpressure valve
2 lengths connecting tubing with 

fi

 ttings for 

fl

 ow through assembly

drain vent

Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA  91731-2882  U.S.A.  •  Tel. (626) 571-2770  •  Fax (626) 573-2057
For Worldwide Sales and Service, visit our website:  www.gfsignet.com  •  Or call (in the U.S.):  (800) 854-4090

GF Piping Systems

3-4150.090    Rev F 02/09     English 

 © Georg Fischer Signet LLC 2008 

Printed in U.S.A. on recycled paper

* Material Safety Data Sheets (MSDS) are available online at http://www.gfsignet.com/msds.htm.

Ordering Information