GERATHERM GP-6220 Instructions For Use Manual

Made by: GERATHERM
Type: Instructions For Use
Category: Blood Pressure Monitor
Pages: 15
Size: 0.59 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

GB

1 - GB

DIGITAL WRIST BLOOD PRESSURE MONITOR

INSTRUCTION FOR USE

GP-6220

 


background image

GB

2 - GB

DESIGNATED USE

This unit uses an oscillometric measurement method in order to measure systolic and di-
astolic blood pressure, as well as the heart rate.
The measurement is conducted on the wrist.
All values can be read on an LCD screen.
This unit has been developed for home use and should only be used by adults over 18 years 
of age with a wrist diameter between 13.5 and 19.5 cm.

PRECAUTIONS

 •  This manual and the product are not substitutes for visiting the doctor.
 

Neither the information contained herein nor this product may be used to diagnose or 
treat health problems, or to prescribe drugs. If you have or suspect that you have a 
medical problem, please seek immediate advice from your doctor.

 •  Do not conduct any measurements if the temperature is low (below +5 °C) or high 

(over +40 °C), or if the relative humidity is beyond the range of 15 % to 93 %, as this 
can lead to inaccurate readings.

 •  Wait 30 to 45 minutes, before taking a measurement if you have just had a caffeinated 

drink or a cigarette.

 •  Relax for at least 5 to 10 minutes before taking a measurement.
 •  Please wait 3 to 5 minutes between measurements, so that your blood vessels can 

return to the state they were in prior to measurement. You may have to adapt the wait-
ing time to your personal physiology.

 


background image

GB

3 - GB

PRECAUTIONS

 •  It is recommended that you use the same wrist for each measurement (preferably the 

left) and take the measurement at about the same time every day.

 •  Sit down comfortably with your elbows placed on the table and both feet on the ground. 

Please do not interlock your legs during the measurement. 

 •  Place the product on level of the heart. Relax your hand. Your palm should be facing 

upwards. 

 •  Take the measurement at room temperature in a quiet and stress-free environment.
 •  The unit should not be moved or shaken during the measurement. Please do not 

speak during the measurement.

 •  Please keep in mind that blood pressure naturally varies depending on the time of day 

and is affected by many different factors. Blood pressure is usually highest at work and 
reaches its lowest level during the sleep phase.

 •  Blood pressure measurements should be assessed by a doctor or trained healthcare 

professional who is familiar with your medical history. If you use the unit and regu-
larly record the results, please keep your doctor informed with regard to the ongoing 
changes in your blood pressure.

 •  If you suffer from a cardiovascular disease (such as atherosclerosis), diabetes, a liver 

or kidney disease, severe hypertension or peripheral circulatory disorders, etc., please 
consult your doctor before using this unit.

 •  The blood pressure measurements conducted with this unit are equivalent to meas-

urements obtained by a trained observer in accordance with the values achieved us-
ing the cuff/stethoscope auscultation method and are within the speci

fi

 ed EN 1060-4 

standard limits.

 •  If the cuff causes any discomfort during the measurement, press the “POWER” button 

to turn off the unit immediately.

 •  If the pressure is over 300 mmHg and the cuff does not de

fl

 ate automatically, pull off 

the strap to detach the cuff.

 •  Do not use this appliance on infants, children or persons who cannot express their own 

intentions.

 


background image

GB

4 - GB

PRECAUTIONS

 •  Measuring too frequently may result in circulatory disorders, which can cause unpleas-

ant sensations such as localised bleeding under the skin or temporary numbness in 
your arm. These symptoms do not usually last long. However, if you have not recov-
ered after some time, please consult your doctor.

 •  The unit is not waterproof. Never immerse this instrument in any liquids.
 •  Do not use the instrument if you think it is damaged or if you notice anything unusual.

WARRANTY

The warranty for this blood pressure monitor is valid for any error on the part of the manu-
facturer under normal use for 3 years from the date of purchase. If your unit does not 
function properly due to defective parts or assembly, we will repair it free of charge. 
With the exception of the battery and cuff, all parts of the unit are subject to this warranty. 
Damage caused by improper handling of your unit is not guaranteed. 
We recommend that the accuracy of the unit be checked after 2 years from manufacturing 
date by an authorized laboratory.
This checking procedure is not a service provided under the warranty.

 


background image

GB

5 - GB

INFORMATION YOU SHOULD KNOW BEFORE OPERATING 
THE DEVICE

What is blood pressure?

A force is created by the heart as the ventricle forcibly ejects blood into the blood vessels 
and through the vascular system. Another force is created by the arteries as they resist 
the blood 

fl

 ow. Blood pressure is the result of these two forces.

What are systolic and diastolic pressures?

Systolic pressure represents the highest pressure at the height of the heart‘s contraction. 
Diastolic pressure is the lowest pressure when the heart is resting.

Is my blood pressure normal?

See the following blood pressure classi

fi

 cation chart released by the WHO (World Health 

Organization) for evaluation of your blood pressure level.

diastolic pr

essur

e mm Hg

systolic pressure mm Hg

Hypertension

High - Normal

Normal

Lower

 


background image

GB

6 - GB

What about low blood pressure?

Generally speaking, a lower blood pressure reading is better unless it causes some un-
comfortable symptoms such as fainting and/or lightheadedness.

Fluctuation and variation in blood pressure

The following chart shows possible blood pressure 

fl

 uctuations during a 24-hour period.

The following factors will in

fl

 uence blood pressure measurement results and cause vari-

ations:

 

• Bathing 

• Breathing

 

• Conversation 

• Meals

 

• Smoking 

• Exercise

 • 

Drinking 

Alcohol 

• 

Moving

 

• Thoughts 

• Mental tension

 

•  Temperature change, etc.

INFORMATION YOU SHOULD KNOW BEFORE OPERATING 
THE DEVICE

Systolic

Diastolic

 


background image

GB

7 - GB

ABOUT THE DEVICE

The cuff is designed to 

fi

 t wrist sizes between 13.5 and 19.5 cm

 (5.35 to 7.68 inch). 

Memory Botton

LCD display

Battery Cover

Power Button

 


background image

GB

8 - GB

EXPLANATION OF DISPLAY

PREPARATION FOR USE

Installing/Replacing batteries
1.  Press lightly on the arrow and slide the battery cover in the direction 
 

of the arrow.

2.  Insert the batteries into the battery compartment mating correct 
 

polarities “+“ and “-“.

 Note:

 • Replace both batteries if the low battery 

indicator appears.

 • Remove the batteries if the device will not 

be used for a long period of time.

 • It is recommended that the same type of 

alkaline batteries be used to avoid incom-
patibility.

Keep batteries away from small children. 
Do not throw batteries into 

fi

 res: they could explode.

DATE

TIME

Date

Memory

Low Battery

Indicator

Pulse 

Indicator

IPD (Irregular 

Pulse Detection)  

Systolic

Diastolic

Pulse Rate

 


background image

GB

9 - GB

Attaching the pressure cuff

1.  Free the left wrist of any clothing etc. and wrap the 

cuff directly around it. The display should be posi-
tioned on the palm side of the wrist.

2.  Fasten the cuff. Make sure it is not too tight. The 

cuff‘s upper edge should be approximately 1 cm 
(0.39 inch) from the palm line.

3.  If it is not possible to measure on the left wrist, 

attach the cuff on the right wrist as shown in the 

fi

 gure.

Measuring posture

1.  Sit upright and place your forearm on the storage 

case so that the device is at heart level. Relax and 
measure in a natural position.

2.  Measure and record blood pressure at the same 

time every day to establish your blood pressure 
pattern.

PREPARATION FOR USE

 


background image

GB

10 - GB

HOW TO OPERATURE THE DEVICE

1.  Setting the Clock

  Enter the clock setting mode by pressing and holding the MEMORY button for over 

three seconds. The display will show a 

fl

 ashing year. Press the POWER button to 

adjust the year when it is 

fl

 ashing. Then click the MEMORY button again, a „month„ 

will appear and 

fl

 ash to be adjusted, and then „date“, „hour“, and „minute“ will follow as 

you press and release the MEMORY button.

 

To adjust the clock: 

 

Press the POWER button to adjust the „month“, „date“, „hour“ and „minute“ when they 
are 

fl

 ashing.

2.  Measuring blood pres-

sure

  Press and release the 

POWER button to start 
measuring automatically.

  When the measurement 

is 

fi

 nished, the display will 

show the measured blood 
pressure and pulse rate. 
Press the POWER button 
to turn off the device. 

 

Otherwise the device will automatically turn off 

after 150 seconds

Note: 

If the symbol 

 appears, it means the unit has detected irregular pulse during 

measurement. If the symbol appears regularly please consult a quali

fi

 ed physi-

cian for professional advice. 

Initial

display

Pumping

Measuring and

buzzer sounds

 


background image

GB

11 - GB

HOW TO OPERATURE THE DEVICE

3.    Recalling Stored Values

3.1  Press and release the MEMO-

RY key to recall the last read-
ing. (Example: no. 35 in the 
memory). The display will show 
the stored pressure and pulse 
rate.

3.2  Press the MEMORY button to 

go to previous stored values.

4.   Erasing Stored Values

 

You may erase one or all records in the memory. 

 

To delete one record:

 

Step 1: 

Press and release MEMORY button to dis-

play stored values. Press and release MEMORY 
button again to choose the record that you wish to 
delete.

 Step 

2: 

Press and hold the POWER button until 

the display shows „dEL“. 

 Step 

3: 

Press and hold the POWER button. After three beeps are heard, the selected 

record is deleted. 

 

To delete all stored values:

 

 Follow 

steps 1

 and 

above.

 Step 

3: 

Press the MEMORY button again and the display shows „dEL ALL“. 

 Step 

4:

 Press and hold the POWER button. After three beeps are heard, all records 

are deleted.

 


background image

GB

12 - GB

CARE AND MAINTENANCE

Do not drop 
the device.

Do not twist 
the wrist cuff.

Do not modify or disassem-
ble the body or the wrist cuff.

Use a cloth moistened with 
water or neutral detergent to 
clean the body of the device 
and then wipe it dry.

Avoid thinner, 
benzene, and 
other harsh 
cleaners.

Keep the device in a suitable 
place. Avoid high tempera-
tures direct sunlight, high 
humidity, and dust.

Remove the batteries 
if the device will not be 
used for a long time.

Do not press the POWER 
button if the cuff has not 
been properly 
wrapped around the wrist.

 


background image

GB

13 - GB

ERROR MESSAGES

Message

Correction

The pressure measured was lower than 20 mm Hg. 

Please measure again.

Excessive body movement is detected during measurement. 

Please measure again.

Pumping error. 

Please check cuff and try again.

The pressure can not be measured due to signal noise.

Please measure again.

The pumping pressure is higher than 300 mm Hg. 

Please measure again.

Low battery. 

Please check and, if necessary, replace the 2 batteries.

or no display 

Observe the applicable regulations when disposing of the device and batteries.
This product must not be disposed of together with domestic waste.
All users are obliged to hand in all electrical or electronic devices, regardless of whether 
or not they contain toxic substances, at a municipal or commercial collection point so that 
they can be disposed of in an environmentally acceptable manner.
Please remove the batteries before disposing of the device/unit. 

Do not dispose of old batteries with your household waste, but at a battery
collection station at a recycling site or in a shop.

DISPOSAL 

 


background image

GB

14 - GB

SPECIFICATIONS

Model number

GP-6220

Display system 

Liquid Crystal Display

Measuring method 

Oscillometric

Power Source 

2 alkaline “AAA” type batteries (1.5 V)

Measuring Range 

20 to 300 mm Hg (blood pressure); 
40 to 200 beats/minute (pulse rate)

Accuracy 

±3 mm Hg (blood pressure), ±5 % (pulse rate)

In

fl

 ation 

Automatic pump

De

fl

 ation 

Automatic exhaust valve

Memory 85 

sets

Display 

LCD (year/date/time, pressure and pulse)

Low Battery indicator 

yes

Auto Power-Off

after 150 s without any operation

Lifetime of battery

abt. 250 measurements

IP Classi

fi

 cation

IP20 (protected against foreign bodies < 12,5 mm)

Cuff size

13.5 to 19.5 cm

Operation Environment 

+5 to +40 °C; RH: 15 to 90 % R.H.

Storage Environment 

-25 to +55 °C; RH: 15 to 90 % R.H.

Dimensions

76 mm x 76 mm x 33 mm (W x D x H)

Weight 

around 136 g (inclusive batteries)

Electromagnetic compatibility 
(EMC)

The present medical product is a device with an
extremely high degree of noise immunity. There-
fore, the enclosure with the EMC tables was not 
printed out.

Speci

fi

 cations are subject to change for improvement without notice. 

 


background image

GB

15 - GB

Geratherm

®

 is certi

fi

 ed in accordance with Council Directive 93/42/EEC and EN ISO 13485 

and is entitled to af

fi

 x the CE-mark 

 (Noti

fi

 ed Body: TÜV Rheinland LGA Products 

GmbH).
The blood pressure monitor conforms to 
 •  EN 1060-1 (Non-invasive sphygmomanometers part 1: General requirements) and 
 •  EN 1060-3 (Non-invasive sphygmomanometers part 3: Supplementary requirements for 

electromechanical blood pressure measuring systems)

 •  EN 1060-4 Non-invasive sphygmomanometers, Part 4: Test procedures to determine the 

overall system accuracy of automated non-invasive sphygmomanometers

QUALITY STANDARD

SYMBOL INDEX

Follow instructions for use

Type BF Equipment

Keep dry

Batch code 
(mm/yyyy; month/year)

Store between 15 % and 
90 % R.H. 

Serial number

Store between -25 °C and 
+55 °C 

Manufacturer

The device must not be disposed 
of with household waste.

Caution! Read the instruction 
manual!

SN

Geratherm Medical AG
Fahrenheitstrasse 1
D-98716 Geschwenda
Germany
www.geratherm.com

-25 °C

+55 °C

90 %

15 %