Consort EV1450 Manual

EV2320\ EV2650\ EV2150\ EV3020\ EV2310\ EV3330\ EV3610\ EV3620

Made by: Consort
Type: Manual
Category: Power Supply
Pages: 16
Size: 0.51 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

MANUAL

HANDLEIDING

MODE D’EMPLOI

ANLEITUNG

EV1450
EV2150
EV2310
EV2320
EV2650
EV3020
EV3330
EV3610
EV3620

February 2015

 


background image

Index

General informations.....................................................................................1
Very important.............................................................................................2

Procedure...................................................................................................3
Setup........................................................................................................3

Language..................................................................................................................... 3
Contrast...................................................................................................................... 3
Date/Time................................................................................................................... 4

Password..................................................................................................................... 4
Detection..................................................................................................................... 5
Data........................................................................................................................... 5
Info............................................................................................................................ 5

Method programming.....................................................................................6
Manual programming.....................................................................................8

Running condition.........................................................................................9

Display........................................................................................................................ 9
Buttons....................................................................................................................... 9

Adjusting parameters......................................................................................................9

End of procedure..........................................................................................................10
Alarm messages............................................................................................................10

USB Digital Port..........................................................................................10

Frequently Asked Questions............................................................................11
Warranty..................................................................................................13

CE Declaration...........................................................................................14

 


background image

Manual Power Supplies

General informations

Introduction

This instrument is manufactured with the latest technology and needs no particular maintenance. 

Consort

 certifies that this instrument was thoroughly inspected and tested at the factory prior to 

shipment and found to meet all requirements defined by contract under which it is furnished. However, 
dimensions and other physical characteristics may differ.
The normal operating temperature should be between 0° and 40°C. Never use the instrument in a room 
with high humidity (>95 %) or at very low temperatures (condensation water!).

Manufacturer

Consort bvba
Parklaan 36

Tel (+32)(14)41 12 79

B2300 Turnhout

Fax (+32)(14)42 91 79

Belgium

E-mail: info@consort.be

Warranty

This instrument (excluding all accessories) is warranted against defective material and workmanship for 
a period of thirty-six (36) months from the date of shipment ex factory. 

Consort

 will repair all defective 

equipment returned to it during the warranty period without charge, provided the equipment has been 

used under normal laboratory conditions and in accordance with the operating limitations and 

maintenance procedures in this  instruction manual and when not having been subject to accident, 

alteration, misuse or abuse. A return authorisation must be obtained from 

Consort

 before returning any 

product for warranty repair on a freight prepaid basis!

Consort

 is not liable for consequential damages arising out of the use or handling of its products.

Servicing

In the event of this instrument being returned for servicing, the owner is requested 

NOT

 to send the 

following items unless they are suspect:

Manual
Cables
Accessories

If serious malfunctioning occurs, stop using the unit immediately and consult your local 

Consort

 dealer.

Keyboard

During 

SETUP 

or programming a procedure:

éê

= Buttons for entering a value or for selecting a function.

= Button for programming and confirming the desired parameters.

Hold this button pressed to return to the main menu!

/

= Button to

start the procedure (MANUAL or METHOD),

cancel while entering a parameter (SETUP),

return to the previous menu (SETUP).

During a 

PROCEDURE:

éê

= Buttons for selecting a display configuration.

= Button to modify the parameters.

/

= Button to stop or to pause the experiment.

                                                                                  

 

                                                                              

Rev. 1605

1

 


background image

Manual Power Supplies

Very important

This electrophoresis power supply is a high technology instrument available in several versions. As it is 

capable of giving dangerous voltage levels by which high power is involved, we suggest that you take a 
few moments to read this manual thoroughly. Although this instrument is equipped with all necessary 
safety features against abuse and other accidental failures, caution should be exercised when working 
with high voltage equipment.

1.

A

void to touch the outlets with any conducting object

 and make sure there is a second person 

present for your safety in case of any severe electric shock.

2.

Never touch any part of the assembly (power supply, leads or tank) before having switched off.

3.

Never manipulate with wet hands.

4.

Do not ground any of the outputs or the buffer in the tank.

5.

Connect the outlets only to an insulated electrophoresis tank with safety cover.

6.

Never make any other connections, such as e.g. putting several power supplies in series or in 

parallel.

7.

In order to prevent electric shock, never open the back plate nor remove the cover.

8.

Do not expose the unit to rain or any other liquid.

9. Do not spill liquid or insert metal objects inside the unit. 

Take care so that the power supply is 

not dropped to avoid damaging the cabinet which defeats safeguards or injuring yourself.

10.

If the unit has been dropped or the cabinet has been damaged, unplug it and have it checked by 

an authorised service technician to restore the safeguards.

11.

The fact that the unit operates satisfactorily does not imply that the unit is properly earthed or 

that it is completely safe. If in any doubt about the effective earthing of the unit, contact a 
qualified electrician.

12. Never block the ventilation holes 

or place the unit in any enclosure unless proper ventilation is 

provided.

13.

Never place 

the

 unit near or over a radiator, heat register or stove.

14.

Avoid locations where the instrument is exposed directly to the sun light.

                                                                                  

 

                                                                              

Rev. 1605

2

 


background image

Manual Power Supplies

Procedure?

      This is the main menu that is shown after powering the device and has 3 options:

PROCEDURE?

MANUAL

METHOD

SETUP

1. MANUAL 

To start a regular procedure with fixed parameter limits for Voltage, Current, Power and 

Timer. See 

Manual Programming

.

2. METHOD

To start a procedure with multiple steps to vary the parameter limits automatically while 
running. This method also allows Voltage Gradient settings. See 

Method Programming

.

3. SETUP

To modify the settings of the 

system

detections

 and 

data-logging

.

Setup

Use this menu to preset the system settings of the power supply for language, display contrast, real 

time clock, password, detections, data recording.

System: Language

The power supply has several languages built-in for the communication with the user.

1.

Follow these steps to select 

LANGUAGE 

and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

LANGUAGE

CONTRAST

DATE/TIME

PASSWORD

2.

Select the desired language and press 

.

System: Contrast

According to the location where the device is installed, it might be required to adjust the display 
contrast.

1.

Follow these steps to select 

CONTRAST 

and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

LANGUAGE

CONTRAST

DATE/TIME

PASSWORD

2.

Adjust the contrast setting and press 

 

to confirm.

                                                                                  

 

                                                                              

Rev. 1605

3

 


background image

Manual Power Supplies

System: Date/Time

This device has a precise real time clock built in. The real time clock is used for GLP reports and data 
recording purposes. The date and the time can be adjusted.

1.

Follow these steps to select 

DATE/TIME 

and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

LANGUAGE

CONTRAST

DATE/TIME

PASSWORD

DATE/TIME

MODIFY

07/12/2015

13:08:47

2.

Set consecutively the date, month

year, hour, minute and seconds. 

Confirm each with 

 

or stop entering the other values by pressing 

/

.

DATE/TIME

MODIFY

07

/12/2015

13:08:47

System: Password

When desired, a password can be set which will be requested when trying to modify the setup or the 
parameters of a procedure. The user can then only run or stop the experiment.

1.

Follow these steps to select 

PASSWORD

 and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

LANGUAGE

CONTRAST

DATE/TIME

PASSWORD

2.

A question to activate the password will appear. 
Choose YES and confirm to enable a password, set to NO to disable the password.

3.

When enabled, the device will request twice the same combination of five keystrokes. This 
combination will be required to modify any setting.

PASSWORD

Enter the new

password:

[-----]

CONFIRMATION!

Enter the new

password:

[-----]

                                                                                  

 

                                                                              

Rev. 1605

4

 


background image

Manual Power Supplies

Detection

Choose 

DETECTION

 to alter the actions when a low current or a power failure is detected.

Low Current

When a very low current is detected as soon as the Voltage limit has been reached, the power will 
shut down the procedure. An 

audible error message

 will be given.

Power Failure

When the detection of a power failure is enabled, the power supply will continue to proceed the 
procedure if the timer was still running at the moment that the power failure occurred. After 
powering it again, the power supply will automatically proceed with the experiment for the 
remaining time.

1.

Follow these steps to select 

LOW CURRENT

 and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

DETECTION

LOW CURRENT

POWER FAIL

2.

Decide whether or not the output voltage should shut down at a too low output current and 

press 

.

3.

Select 

POWER FAIL

 and press 

.

4.

Decide whether or not the instrument should proceed with the experiment for the remaining 
time after a power failure and press 

.

Data

To be added, check our website (

www.consort.be/wiki

) for firmware updates of the instrument. 

This feature will allow to log all information about the executed procedures.

Info 

Choose 

INFO 

to find specific information about the device, the parameter limits, the firmware version 

and more:

PROCEDURE?

MANUAL

METHOD

SETUP

SYSTEM

DETECTION

DATA

INFO

                                                                                  

 

                                                                              

Rev. 1605

5

 


background image

Manual Power Supplies

Method programming

Method programming permits to store up to 9 different methods, each with 9 steps, of frequently used 

parameters in the non-volatile memory for future recall. Each step is able to recall a next one, provi-
ding a flexible multiple step function for special techniques. The method mode also permits to program 
a linear voltage gradient for any step provided the limiting current or power is not attained. Parameters 
can be changed temporarily without interrupting the run.

1.

Verify if the instrument is switched off.

2.

Connect the electrophoresis tank to the instrument.

3.

Switch the instrument on.

4.

Select 

METHOD

 and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

5.

Select the desired method and press 

:

CHOOSE

METHOD

1

6.

When previously multiple steps have been

programmed, select the desired step and 
press 

:

METHOD 1|1

STEP

1

7.

Decide whether or not a linear voltage 

gradient has to be performed 
and press 

:

METHOD 1|1

GRADIENT

NO

8.

Enter the desired voltage limit 
and press 

 

:

METHOD 1|1

VOLTAGE

300

V

When performing a voltage gradient, 
enter the desired end Voltage. The start 
Voltage is the end point of the previous 
step.

9.

Enter the desired current limit 
and press 

 

:

METHOD 1|1

CURRENT

500

mA

This limit is automatically set to 
maximum and cannot be changed in case 
of a voltage gradient step.

                                                                                  

 

                                                                              

Rev. 1605

6

Method/steps

 


background image

Manual Power Supplies

10.

Enter the desired power limit 
and press 

:

METHOD 1|1

POWER

150

W

This limit is automatically set to 
maximum and cannot be changed in case 

of a voltage gradient step:

11.

Enter the desired timer unit (h or kVh) 
and press 

 :

METHOD 1|1

TIMER UNIT

h

12.

Enter the desired timer value

and press 

 :

METHOD 1|1

TIMER

00:30

h

The value format is 'hh:mm' when 

entering a value in hours, a value with 2 

decimals when entering kiloVolthours.

13.

Decide if the experiment should 

automatically proceed with a next step, 

press 

 

:

METHOD 1|1

END METHOD

NO

Choose NO to resume at point 

with the 

next step,

 

choose YES to stop after this 

step.

14.

Press 

/

 to start the experiment.

At any time, you can return to the main menu by holding 

 pressed.

At any time, you can avoid programming unnecessary parameters. To do so, simply press 

/

 instead of 

after entering the last relevant value and the instrument will immediately start the experiment keeping the

other parameters to their previous values.

To ensure that the desired parameter is kept constant, we suggest to pre-set all other parameters to its
maximum.

A blinking unit of the controlling  parameter appears when the regulation is complete.

The timer will always count down except when no timer value has been entered.

                                                                                  

 

                                                                              

Rev. 1605

7

 


background image

Manual Power Supplies

Manual programming

Manual programming permits to set voltage, current, power and time limits for a simple routine 
electrophoresis run. Parameters can be changed temporarily without interrupting the run.

1.

Verify if the instrument is switched off.

2.

Connect the electrophoresis tank to the instrument.

3.

Switch the instrument on. 

4.

Select 

MANUAL

 and press 

:

PROCEDURE?

MANUAL

METHOD

SETUP

5.

Enter the desired voltage limit and press 

 

:

MANUAL

VOLTAGE

300

V

6.

Enter the desired current 

limit

 and press 

 

:

MANUAL

CURRENT

500

mA

7.

Enter the desired power 

limit

 and press 

 

:

MANUAL

POWER

150

w

8.

Enter the desired timer unit (h or kVh) 
and press 

 :

MANUAL

TIMER UNIT

h

9.

Enter the desired timer value 
and press 

 :

MANUAL

TIMER

00:30

h

The value format is 'hh:mm' when entering a
value in hours. The format is a value with 2 
decimals when entering kiloVolthours. Set to
zero when required to continue until 

/

 is 

pressed or the power is switched off.

10.

Press 

/

 to start the experiment. See 

Running Condition

.

At any time, you can return to the main menu by holding 

 pressed.

At any time, you can avoid programming unnecessary parameters. To do so, simply press 

/

 instead of 

after entering the last relevant value and the instrument will immediately start the experiment keeping the
other parameters to their previous values.

To ensure that the desired parameter is kept constant, we suggest to pre-set all other parameters to its

maximum.

A blinking unit of the controlling  parameter appears when the regulation is complete.

The timer will always count down except when no timer value has been entered.

                                                                                  

 

                                                                              

Rev. 1605

8

 


background image

Manual Power Supplies

Running condition

Display

  

600

 

v

85 mA

51 W  

00:29:33 h  

 

1|2        17/12/15   10:36:18    

When running an experiment, the display allows to see the actual parameters. It is divided into three 
fields:

The 

upper left

 

field shows :

a blinking 

 symbol to indicate there is a possible dangerous condition on the output sockets. 

Avoid to contact the outlet sockets and cables!

the selected measured parameter. Use the 

éê

 keys to switch the parameter in this field.

The 

upper right

 

field shows the other 3 parameters.

The 

bottom

 

field shows date + time at the right and the actual method/step combination at the 

left when such a procedure is running, no indication when running a manual procedure. The 

bottom field line is also used for system messages when required.

The unit of the 

parameter

 that is being kept constant is given with blinking

 

bold

 characters.

Buttons

The keys be used as follows:

éê

Buttons for selecting another parameter in the upper left corner of the display .

Button to modify the parameters.

/

Button to stop or to pause the procedure.

Adjusting parameters

It is possible to modify the parameters while running. These modifications are temporarily and are not 
stored in the non volatile memory. It is not possible to modify the settings of a voltage gradient step.

Press 

 :

  

600

 

v

1000 mA

150 W  

00:29:33 h  

 

éê

=ADJUST, 

=NEXT

The display shows the preset values for V, mA and W. The timer remains counting down when a 
timer value has been set.

Choose the limit to adjust with the 

 button. 

Adjust the value with the 

éê

 buttons.

The display will return automatically to the normal run mode when no button is pressed for 5 
seconds.

                                                                                  

 

                                                                              

Rev. 1605

9

 


background image

Manual Power Supplies

End of procedure

The procedure will stop automatically when the timer has reached its end value. A signal will 
sound during 10 seconds. The sound signal can be ended by pressing a button.

The procedure can also be stopped by pressing 

/

The power supply will then be in a paused 

condition. It allows to modify the run parameters and/or to view the gel. When pressing 

/

 

again, the power supply will continue the procedure where paused. To stop the procedure 
completely, hold 

 pressed until the main menu appears.

The power supply will remain showing the display of the running condition until the voltage has 
lowered to a safety value.

When the procedure has finished, a report will be shown on the display. Scroll the display lines 
with the  

éê

 buttons to see it completely. Press 

 or 

/

 to return to the main menu.

Alarm messages

In case a problem or a hazardous situation is detected, the power supply will intterupt the experiment 

with one of the following errors. A sound will also be produced that can be stopped by pressing any key. 
The hazardous situations require that the power supply needs to be switched off before starting the 

experiment again.

GROUND LEAKAGE

A dangerous ground leakage is present (check thoroughly the complete assembly). Your power 
supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

SHORT-CIRCUIT

Connected tank is short-circuited or has a very low resistance (check thoroughly the complete 

assembly). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of 

this manual.

OVERLOAD

Connected tank has a too low resistance (check specifications). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

LOW CURRENT

The output current is very low (switch the alarm off in the 

SETUP

 menu). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

TECHNICAL PROBLEM

Failure in the electronics. See 

Support

 on 

www.consort.be

. You could also try a general RESET: 

switch the instrument on while holding 

 pressed. Attention! All stored programs and data will 

be erased!

USB Digital Port

This power supply has a digital USB port allowing to interface with a computer. Programs can read and 
control the power supply completely. More information can be found on our support website.

Driver

The driver can be downloaded using the link for the USB drivers on our website: 

http://www.consort.be/downloads/software/

.

                                                                                  

 

                                                                              

Rev. 1605

10

 


background image

Manual Power Supplies

Frequently Asked Questions

What are the relations between Voltage, Current, Power and Resistance?

Power (W) = Voltage (V) x Current (A)
Resistance (

W

) = Voltage (V) / Current (A)

How does a power supply react after pressing RUN?

The internal generator will start building up the high voltage at the output terminals while voltage and 
current are constantly measured and power calculated. When one of the pre-set parameters is 
exceeded, the generator stops and will keep that parameter constant. 

How important is the resistance of an electrophoresis unit?

The resistance of an electrophoresis unit depends on its size, gel thickness, amount of buffer, buffer 

conductivity and temperature. This resistance will normally decrease in time due to a slowly increasing 

temperature. Electrophoresis units which have a resistance below the minimum load resistance of a 
power supply will trigger an alarm! Read the output voltage and current during a run to measure the 

resistance and use above formula to calculate the value.

How to keep a constant voltage during a run?

Program the desired voltage and a higher current and power then the maximum expected values:
Current > Voltage / Resistance
Power > Voltage x Current

How to keep a constant current during a run?

Program the desired current and a higher voltage and power then the maximum expected values:
Voltage > Current x Resistance
Power > Voltage x Current

How to keep a constant power during a run?

Program the desired power and a higher voltage and current then the maximum expected values:
Voltage > Current x Resistance
Current > Voltage / Resistance

Why are my output values different from those of a similar experiment?

Either your programmed parameters are not equal to those described or the resistance of your 
electrophoresis unit is different (see above). It cannot be due to e.g. an other model of power supply as 
the relations between Voltage, Current, Power and Resistance are monitored in the same way by any 
instrument (the electrical laws cannot be disregarded!).

What about connecting more than one unit to the same power supply?

The outlets being in parallel each electrophoresis unit will be supplied with exactly the same voltage. 
However, current and power may differ due to differences between them even when exactly the same 
model, gel, buffers, etc... are used. Therefore, it is recommended to run several electrophoresis units 
only in the constant voltage mode on the same power supply. 
Mind also that adding tanks will increase the current and the power. These parameters may be limited 
by the power supply and cause it to switch over to constant current or constant power.

                                                                                  

 

                                                                              

Rev. 1605

11

 


background image

Manual Power Supplies

Specifications

                                                                                  

 

                                                                              

Rev. 1605

12

 


background image

Manual Power Supplies

WARRANTY CERTIFICATE

This instrument (excluding all accessories) is warranted against defective material
and workmanship for a period of thirty-six (36) months from the date of shipment
ex factory.
Accessories   and   breakable   items   such   as   electrodes   are   not   warranted   unless
proven to be defective before shipment.
The original purchase order numbers, Consort invoice numbers and serial numbers
of the products must be provided.

CONSORT  

will repair all defective equipment returned to it during the warranty

period without charge (CIF Turnhout prepaid by sender), provided the equipment
has   been   used   under   normal   laboratory   conditions   and   in   accordance   with   the
operating   limitations   and   maintenance   procedures   described   in   the   instruction
manual and when not having been subject to accident, alteration, misuse or abuse.
If the products have been used with or have come into contact with fluids, an MSDS
(material   safety   data   sheet)   must   be   supplied   prior   to   issuing   a   return
authorisation.
A  return   authorisation   must   be   obtained   from  

CONSORT

  before   returning   any

product for warranty repair on a freight prepaid basis! 

                                                                                  

 

                                                                              

Rev. 1605

13

 


background image

Manual Power Supplies

 

DECLARATION OF CONFORMITY

We declare under our sole responsibility that the product

Electrophoresis Power Supplies

content of the type numbers

EV1450  EV2150  EV2310
EV2320  EV2650  EV3020
EV3330  EV3610  EV3620

to which this declaration relates is in conformity

with the following standards of the European Directives

LOW VOLTAGE DIRECTIVE 2006/95/EC

EN61010-1

EMC DIRECTIVE 2004/108/EC

EN61326-1

ROHS DIRECTIVE 2011/65/EU

EN50581: 2012

Consort bvba

Parklaan 36, B-2300 Turnhout, Belgium

Tel: (+32) (0)14 41 12 79

Fax: (+32) (0)14 42 91 79

Name

Jan De Ceuster

Signature

Title

Director

Date

21 November 2015

The electro-magnetic susceptibility has been chosen at a level that gains proper operation in residential areas, on
business and light industrial premises and on small-scale enterprises, inside as well as outside of buildings. All
places of operation are characterised by their connection to the public low voltage power supply system.

                                                                                  

 

                                                                              

Rev. 1605