Gather the following items:
  •  Coaguchek XS Pro Meter
  •  Container of Test Strips
  •  Test Strip Code Chip
  •  Safe-T-Pro Plus Lancet
  •  Alcohol Wipe
  •  Cotton Ball or Tissues

1.

Make sure you have the code chip 
ready to insert if you are opening 
a new box of strips. The code key 
needs to be inserted into the meter 
everytime you use a new lot of strips.

2.

Clean the fingertip with an alcohol 
wipe. Or have the patient wash his or 
her hands in warm soapy water.
Dry thoroughly.  

3.

Before continuing, review these 

tips for getting a good drop of 

blood.

Increasing the blood flow in the 
finger will help you to get a good 
drop of blood so keep in mind these 
tips:

Warm the hand. Have the patient 

• 

hold it under his or her arm, and/  
or wash with warm water.
Have the patient let that arm 

• 

hang by his or her side.
Massage the finger from its base.

• 

Use these techniques until the 
fingertip has increased in colour.

 

 

4.

Take a test strip out of the container.

Close the container tightly.

You have 10 minutes to use the test 
strip once you remove it from the 
container.

5.

Slide the test strip into the test strip 
guide in the direction of the arrows 
until it stops.

The meter turns on and the main 
menu appears.

If you insert a strip from a new lot 
number for the first time, the meter 
will display the number of the test 
strip code chip required.

  

Remove 

the old code chip and insert the new 
one.

6.

Touch the 

Scan 

button and hold the 

operator ID barcode approximately 
10cm from the built-in scanner. The 
meter beeps once the barcode has 
been successfully read.

7.

Press 

Patient Test 

on the main menu 

screen.

8.

Touch the 

Scan

 button and hold the 

patient ID barcode approximately 
10cm from the built-in barcode 
scanner. The meter beeps once the 
barcode has read successfully.

An hour glass symbol appears.

 

The test strip symbol appears after a 
few seconds with a flashing drop of 
blood. You have 180 seconds to apply 
the sample.

 

9.

Twist the protective cap off the 
lancet.

Massage the finger until you see 
increased colour in the fingertip.

Keeping the hand down, press the tip 
of the lancet firmly against the side of 
the fingertip. Press the purple trigger 
button.

Gently squeeze from the base of the 
finger to develop a drop of blood.

10.

Within 15 seconds of lancing the fingertip, 
apply the blood to the target area on the strip.
Find the clear target area on the test strip. You 
can dose from the side or the top.

Hold the blood drop to the test strip until you 
hear a beep. The flashing blood drop symbol 
will disappear and an hourglass symbol appears 
and the test begins.

Do not add more blood to the test strip. 

Do not touch the test strip while the test 

is in progress.

The meter performs an automatic quality 
control check on the strip before it displays 
the result.

 

QC  

  appears in the display briefly following a 

successful QC check. The result appears in less 
than a minute.

11.

Record your result as instructed by 
your facility.

Note: If during testing the meter 
displays an error message, refer to 
the Error Messages section of the 
user manual for an explanation  and 
steps on how to proceed.

12.

 Quick Reference Guide

CoaguChek  

XS Pro

®

 

P O I N T   O F   C A R E

T E S T I N G

Operator’s Manual

CoaguChek  

XS Pr

o

®

 

P O

I N

T  

O F

  C

A R

E

T E

S T

I N

G

Operator’s 

Manual

For Assistance Phone 

0508 MY LIFE | 0508 69 54 33 

 


background image

CoaguChek  

XS Pro

®

 

P O I N T   O F   C A R E

T E S T I N G

Operator’s Manual

CoaguChek  

XS Pr

o

®

 

P O

I N

T  

O F

  C

A R

E

T E

S T

I N

G

Operator’s 

Manual

 Running Liquid QC - Quick Reference Guide

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

12.

For Assistance Phone 

0508 MY LIFE | 0508 69 54 33 

Protection against infections: 

 There is a potential risk of infection.  Medical staff and other persons using the CoaguChek XS Pro 

instrument to perform measurements for more than one patient must be aware that any object coming into contact with human blood is a 
potential source of infection.

From the CoaguChek XS PT Controls 
Kit (stored in the fridge), take out a 
bottle of control containing the dried 
powder and a diluent filled dropper.  
Check the expiry date.

Carefully remove the screw cap 
and rubber stopper from the quality 
control bottle and make sure that 
none of the dried powder is removed 
with the rubber stopper.

Holding the dropper at the bottom of 
the stem use scissors to cut off the 
tip of the dropper at the upper end of 
the stem.  Do not squeeze the bulb of 
the dropper.

Insert the dropper tip first, into 
the control solution bottle.  Gently 
squeeze the bulb to dispense the 
entire contents of the dropper onto 
the dried powder.  Do not touch the 
dried material and the dropper. 

Gently swirl 2-3 times.

Minimum reconstitution time one 
minute. 

Use within 30 minutes of 

making up.

Slide the test strip into the test strip 
guide in the direction of the arrows 
until it stops.

The meter turns on and the main 
menu appears.

If you insert a strip from a new lot 
number for the first time, the meter 
will display the number of the test 
strip code chip required.

  

Remove 

the old code chip and insert the new 
one.

Touch the 

Scan 

button and hold the 

operator ID barcode approximately 
10cm from the built-in scanner. The 
meter beeps once the barcode has 
been successfully read.

Press 

Control Test 

on the main 

menu screen.

Select the corresponding code for the 
QC test displayed on the screen or a 
new code if using a new lot of control 
and insert quality control code chip 
when prompted.

The meter stores the code chip 
information in the memory for future 
use.

The dropper icon flashes to indicate 
that the meter is ready to perform the 
test. 
At the same time a 180 second 
countdown begins. You must apply 
the sample within this time.

Mix the control again by gently 
swirling.  Draw the solution into the 
dropper and apply one hanging drop 
to the clear target area and the meter 
will beep when the control sample is 
detected and the INR measurement 
begins. 

The result will appear and the QC 
range will be underneath in brackets.

If this is out of range, indicated by 
an up or down arrow, repeat the test 
mixing a fresh vial of QC material.  if 
still out of range

 contact 0508 MY 

LIFE | 0508 69 54 33.