Coag-Sense

®

 

Prothrombin Time (PT)/INR

Monitoring System

Quick Reference Guide

Getting Started and Setting Up the Meter

1.  Gather test supplies such as test strips, lancets and sample transfer tubes.
2.  Warm your hands to increase circulation. 
3.  Wash your hands with warm soapy water and completely dry them.
4.  Place the meter on a flat, non-vibrating surface. 
5.  Turn the meter on by pressing the POWER button on the top left side of the meter. 

Performing a ProThrombin Time Test

Activity

Display Screen/Visual Aid

1.

  After turning on the meter, you will know it is ready 

when the display reads: “

READY INSERT STRIP

”. 

Open a test strip pouch, tearing at the notched end. 

Remove the strip and set the package aside. 

2.

  Holding the round end, gently push the strip com-

pletely into the meter. The meter will display “

PAT 

STRIP DETECTED

” and begin the strip warming  

process changing the display to: “

WARMING 

PLEASE WAIT

”. The time for completion of the  

warming process will be displayed in seconds.

3.

  The meter will beep to let you know it is ready for a 

sample and display the message “

APPLY SAMPLE

”. 

The sample application area will flash green indicating 

where the sample should be applied. You have 2.5 

minutes to apply the sample to the strip. 

PST

Intended Use

- The Coag-Sense Self-Test Prothrombin Time (PT)/INR Monitoring System is an 

in vitro

 diagnostic 

device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units.  It uses fresh 

capillary whole blood.  It is intended for use by properly selected and suitably trained patients or their caregivers on the 

order of the treating physician.  Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior 

to self-testing. The device is not intended to be used for screening purposes.

For In Vitro Diagnostic Use

 


background image

Activity

Display Screen/Visual Aid

4.

  Remove the cap from the one-time-use 21g lancet,  

place it against the skin, and push down firmly against 

the finger to lance the surface of skin. Note: smaller 

gauge lancets should not be used.

5.

  The blood should bleed freely. If it doesn’t, gently 

squeeze the finger on each side of the wound to 

get it started. Lower your hand below your heart if 

necessary to produce a “bead” sized drop of blood.

6.

  Hold sample transfer tube below bulb. DO NOT 

TOUCH OR SQUEEZE BULB. With tube horizontal 

touch tip to bead of blood.  Let capillary action fill cap-

illary portion of tube until blood flow stops.  Squeeze 

finger to generate additional blood if required to com-

pletely fill capillary portion of sample transfer tube.

7.

  Move fingers to flat sides of bulb being sure to cover 

air hole. Rest hand on meter. Insert tip into sample 

application well of test strip, touching tip down on 

strip at flashing green light. Gently squeeze bulb until 

blood leaves tube (do not produce air bubbles). Keep 

pressure on bulb and pull away from test strip (avoids 

back suction).  

You must apply blood to the test 

strip within 15 seconds of lancing the finger.

The meter will know that you have added the sample 

and the display will then read: “

TESTING, PLEASE 

WAIT

”. Testing time is patient’s actual clotting time.

8.

  When testing is complete, the meter beeps once 

and the results (

INR

 and Prothrombin Time (

PT

) in 

seconds) appear on the screen.

  Record the results. Remove the test strip and properly 

dispose along with the transfer tube and lancet. Turn 

off the meter.

NOTE: 

Read Self-Test User’s Manual for complete instructions for use, indications, warnings, precautions and 

troubleshooting tips. If any error messages are displayed, refer to the “Troubleshooting” section of the User 

Manual or call CoaguSense Technical Service at 

1-866-903-0890

.