Full Text Searchable PDF User Manual

U S E R M A N U A L
B A R I A T R I C O P E N B A S E P O W E R T A B L E S
C o m b i n e L o a d a n d F o r c e C a p a c i t y : 5 5 0 l b s . ( 2 4 9 k g )
Model Numbers: 84100 • 84200 • 84430
Limited Warranty
ON FRAME &
ELECTRICAL
COMPONENTS
Limited
Warranty
CLI-MA-005
REV A

Important Information ................................................................................................................................ 2
Safety ......................................................................................................................................................... 3
Parts .......................................................................................................................................................... 3
Specifications .............................................................................................................................................. 4
Electrical Warnings ...................................................................................................................................... 4
Unpacking & Setup ................................................................................................................................. 5-6
Operation ................................................................................................................................................ 6-7
Maintenance, Parts Replacement & Cleaning .............................................................................................. 8
Options & Limited Warranty ....................................................................................................................... 9
FAQs .................................................................................................................................................... 10-11
IMPORTANT INFORMATION:
CONTENTS
Clinton Bariatric Open Base Power Tables Models 84100 , 84200 and 84430 are designed to be used to
accommodate patients and medical and physical therapy professionals for examinations and select procedures
where the load capacity or load and force is not greater than 550 pounds (249 kg).
It is extremely important for medical and therapy professionals to read and become familiar with all the tables
features, functions and controls BEFORE putting the table into service.
Clinton Bariatric Open Base Power Tables Models 84100, 84200 and 84430 use heavy gauge steel construction
for the base topped with a comfortable 2" padded surface covered in tough, woven-back vinyl upholstery. The
pivot points are self-lubricating and the smooth vertical adjustment is powered by a dependable 120v motor.
Models 84200 and 84430 are equipped with a pneumatic backrest that can be accessed from either side.
The foot controller on all models features an up button and a down button that controls the operation of the
table. An optional hand controller is available.
SERVICE
If your Clinton Bariatric Open Base Power Tables Models 84100, 84200 and 84430 should fail operating for
any reason, Please call Clinton Industries for assistance in diagnosing the problem and having your equipment
serviced, if necessary.
Please provide the following information before calling Clinton regarding an in or out of warranty problem.
Date of Purchase:
Serial Number:
Model Number:
Dealer/Distributor:
Dealer/Distributor Phone:
Dealer/Distributor Address:
2
Model Number and
Serial Number are
located on the
under side of the metal
frame at the head end
of the table

It is extremely important that all precautions regarding safe use of Clinton Open Base Power Tables Models
84100, 84200 and 84430 be strictly observed to insure the safety of both patient and healthcare
professional.
Identifying the part of the Clinton Bariatric Open Base Table can be very important when operating the
table or when reporting a problem.
Pay particular attention to Cautions and Warnings throughout
this manual. Potentially hazardous and harmful situations to users,
patients and property can occur if these cautions and warnings are
ignored.
PARTS
SAFETY
Premium, woven-back
upholstery
Plywood Top
Seamless
rounded
corners
HEAD-END
Adjustable Backrest
Actuator Lever
1 on each side
(models 80200, &
84430 only)
Top Frame
Center Leg
Center Axle
“H” Leg
“H” Leg Axle
BODY
Foot Drop Section
(model 84430
only)
Motor Assembly
Foot Drop Section
Lever (model 84430
only)
BASE
Foot Pedal
3

IMPORTANT ELECTRICAL WARNINGS
GENERAL SPECIFICATIONS
MOTOR SPECIFICATIONS
Models: 84100 & 84200
Table Weight .............................................. 415 lbs.
Total Load Capacity (including force) .......... 550 lbs.
Overall Length ................................................... 72"
Overall Width .............................. 34" & 40" Models
Height Range ...........................................18" to 35"
Backrest
(Model 84200)
Length ............................................................... 38"
Width ................................................................ 34"
Storage & Transportation
Temperature .................................... 20º F to 150º F
Humidity ............................................-10% to 90%
Operation
Temperature .................................... 50º F to 100º F
Humidity ............................................. 30% to 75%
Models: 84430
Table Weight .............................................. 415 lbs.
Total Load Capacity (including force) .......... 550 lbs.
Overall Length ................................................... 76"
Overall Width .............................. 34" & 40" Models
Height Range ...........................................18" to 35"
Backrest
Length ............................................................... 38"
Width .....................34" & 40" depending on model
Drop Section
Length ............................................................... 13"
Width .....................34" & 40" depending on model
Storage & Transportation
Temperature .................................... 20º F to 150º F
Humidity ............................................-10% to 90%
Operation
Temperature .................................... 50º F to 100º F
Humidity ............................................. 30% to 75%
WARNING:
When performing any procedure that requires use of electrical instruments or equipment,
the patient must be insulated fron the metal portions of the table with non conductive material. Failure
to do so mat result in electrical shock or injury to the patient or attendant.
WARNING:
Always use 110v, 60hz alternating current. Failure to use the proper electrical connection
may result in electric shock to patient and attendant and will result in damage to the equipment
WARNING:
Never use these tables in oxygen intense or other explosive environment. Failure to do
so could result in flash fire or explosion which could result in serious injury or death.
4
Clinton Industries Power Treatment Tables use motors and power supplies that meet UL 60601-1 and
CAN/CSA C22.2 601.1
.
120VAC 50-60hz 8000N push 1 min. of constant use out of every 9 minutes

UNPACKING & SETUP
All Clinton Bariatric Open Base Power Tables ship fully assembled. However, care should be taken when unpack-
ing the table so as not to damage the equipment.
BEFORE STARTING TO UNPACK THE EQUIPMENT inspect the boxes for damage that may have occurred during
the shipping process. Clinton uses every packaging option available to insure that equipment arrives at its
destination in factory condition, however, Clinton Industries has no control over third party handling once it
leaves our facilities. Report any visible damage to Clinton and the shipping carrier immediately.
After the package is inspected for visible damage, follow the unpacking steps below.
CAUTION:
Do not use a knife or other sharp object that may puncture or damage the upholstery
during the unpacking process.
CAUTION:
Clinton Open Base Power Tables are heavy. Clinton Industries recommends four
reasonably strong individuals using proper lifting techniques be used to lift or move the table.
Not using the recommended number of movers may result in injury to the persons attempting to lift
or move the equipment.
CAUTION:
Always lift the table by the metal frame. If table is equipped with stirrups, never used the
extended stirrups for lifting the table. This could result in damage to the stirrups or table, and may result in injury
to the individuals attempting to move or lift the table.
1
2
4
5
UNPACKING THE POWER TABLE
1. Remove any strapping and
cardboard lid
2. Remove wooden post supports
from each corner.
3. With help, lift main cardboard
surround off the table.
4. Pull corners of cardboard tray
apart so tray is flat.
5. With the assistance of one or
more persons, lift power table
off skid and into position.
3

LEVELING
OPERATION
For optimum performance an patient safety, all Clinton
Open Base Power Tables should be on a clean, level
non-slip surface.
Once the table is in the desired location, level the table
by adjusting the four leveling feet on the table.
Turn the feet counter clockwise to decrease the height
and clockwise to increase the height of the leg.
Make certain that the table is level in all directions and
that it does not wobble or rock.
HI-LO OPERATION
(All
Models)
Clinton Open Base Power Tables shown in this manual control the up and down functions of the table with a
standard foot controller. (Note: Optional hand controller is available at additional cost.)
Press the down button on the foot controller with slight even pressure and the table will slowly go down to
minimum level. Slight even pressure on the up button will slowly raise the table to maximum height.
CLEAN BEFORE USING
(All
Models)
It is important to thoroughly clean tables according to facility standards before it is put into service. See cleaning
suggestions on page 8 of this manual.
ADJUSTABLE BACKREST
(
Models 84200 & 84430 only)
If the table is equipped with an adjustable backrest with a pneumatic cylinder, the angle of the backrest is
controlled from either side of the table by an actuator. Rotate the actuator towards the backrest to adjust. Always
use your hand on top of the upholstered top to move the backrest up or down. The backrest cylinder is
pressurized in the up position only and requires force to return it to the down position.
Backrest Actuator
Backrest Cylinder
Adjusting Nuts
Backrest
Cylinder
6

FOOTREST/DROP SECTION
(Model 84430 only)
The drop section is located at the foot-end of the table and is shown below at left in a semi-extended position.
The friction lock drop section can be raised and lowered using the drop section levers on either side of the table.
To lower the drop section, place your hand under the upholstered top of the drop section and move the lever
forward towards the foot-end of the table. The lever must be kept in the forward position while gently lowering
the drop section. To raise the drop section, place your hand under the upholstered top and gently lift. The drop
section levers will automatically move forward and the drop section will lock into position wherever the lifting
motion is stopped. NEVER LUBRICATE THE LIFT MECHANISM.
Drop Section in the semi-raised position
Drop Section in the down position
OPTIONAL LOCKING CASTERS
(All Models)
Locking casters are an option on all models. The locking mechanism can be engaged by using your foot to
depressing the tab located at the rear of each caster. This action locks the swivel of the caster and acts as a
brake to minimize movement of the wheel.
ALL FOUR CASTERS MUST BE LOCKED TO ENSURE MINIMAL
MOVEMENT OF THE TABLE.
IMPORTANT:
Casters can not be added in the field without consulting the factory, due to the need to replace
the foot bar on the base to accommodate the new casters. Please contact the factory for necessary information
before proceeding.
Caster in the mobile position.
Caster in the lock position.
7

METAL
Metal frame and stirrups (if applicable) should be cleaned according to facility antimicrobial standards and dried
with a clean soft cloth. Never use abrasive cleaners or scrubbing implements on painted surfaces.
UPHOLSTERY
Clinton uses only premium vinyl for upholstery, however, care should be taken when cleaning to ensure that the
upholstery or color is not affected by the cleaning agent. Use only mild cleaning agents on upholstery. Clinton
recommends using a 1 to 1 mix of ammonia and water or a 1 to 4 mix of bleach and water. Rinse thoroughly
with clean water and pat surface dry. This is not a guarantee and does not relieve the user from responsibility of
the proper and safe use of all cleaning agents. Test on inconspicuous area prior to using any cleaning agent on
upholstery's visible surface. Certain cleaning agents can be harmful to the surface.
WARNING:
Always clean surfaces that come in contact with pathogens or infectious materials
according to standard procedures of your particular healthcare or therapy facility. Although Clinton
recommends a cleaning solution it is not a guarantee that it will meet with your facility’s standards
or state, federal and local standards for treatment and cleaning of areas and equipment that may be
exposed to pathogens and infectious materials.
PERIODIC INSPECTION
• Inspect all electrical cords and connections for damage or wear.
• Check that all fasteners are in place and securely fastened.
• Make certain that the hi-lo function of the table is in proper working condition.
• Check to ensure that the table has remained level.
BACKREST CYLINDER REPLACEMENT
(Models 84200 & 84430 only)
Remove cotter pin. Then remove
the slide-out pin. Position tab
located on the adjustment bar
against cylinder pin.
Align holes in cylinder bracket with holes in table bracket by
adjusting the angle until the holes line up. Install slide-out pin
through holes and replace cotter pin.
Finished view of cylinder assembly.
The backrest cylinder is replaceable in the field by following these simple directions below.
Cylinder Pin
Tab
PREVENTIVE MAINTENANCE & PARTS REPLACEMENT
CLEANING
8

9
LIMITED WARRANTY
TABLE OPTIONS
To the original purchaser only, Clinton Industries warrants manufactured products to be free of defects in materials,
workmanship or design under normal use and service conditions for five years on all mechanical products, all electric
equipment and support items, from the date of purchase. Original purchaser
MUST REGISTER THIS WARRANTY by
mail or online at www.clinton-ind.com/warranty registration
within 30 days of purchase to claim
benefits offered by this warranty. Failure to register your product with in the allotted time voids all offers and benefits
of this warranty.
If the product and/or equipment should become defective within the first year of the warranty period, Clinton Industries will repair or
replace it (at our discretion) free of charge, including return transportation to you provided you deliver the equipment or product prepaid
to Clinton Industries located at 1140 Edison St, York, PA, 17403. If replacement parts are deemed necessary by Clinton Industries within
(1) one year of the purchase date, all new parts and shipping will be covered by Clinton Industries. Thereafter in warranted years (2) two
through (5) five, all shipping charges of replacement parts or replacement product (at our discretion) covered under this warranty will be the
responsibility of the purchaser.
Please receive authorization before returning product to the factory, as Clinton Industries will not accept collect shipments. Clinton
Industries reserves the right to repair or replace component parts without return of the entire unit. Service can be conducted on Clinton
power tables by a trained, fully self-insured, professional service technician.
This warranty does not include damage resulting from accident, abuse or misuse of the product and expressly excludes normal wearing
of parts (including but not limited to vinyl) or defect caused by transportation, accident, fire, flood, improper assembly, alteration, or
negligence. In addition all representations and warranties made by any person or entity other than Clinton Industries main office is excluded
from this warranty.
Implied warranties, including those of merchantability and fitness for a particular purpose, are limited to five years from purchase and all
implied warranties are expressly excluded to the extent permitted by law and are otherwise limited to the terms of express warranties as set
forth above. This is the exclusive remedy, and law for consequential or incidental damages under any and all warranties is excluded to the
extent exclusion permits liability.
This warranty does not cover or compensate for any loss of use damages that may occur as a result of failure of equipment covered under
this warranty.
Clinton industries reserves the right to alter or change equipment without prior notice and to repair or replace warranted equipment with
new parts or new products.
Note: Exceeding the recommended weight capacity of any Clinton product with stationary weight or a total combined weight consisting of
stationary weight and applied force under normal intended or unintended use voids this and all implied or stated warranties.
Limited Warranty
ON FRAME &
ELECTRICAL
COMPONENTS
Limited
Warranty
080
082
083
084
087
Option: 080 Safety Straps
Length ............................................................... 92"
Width .................................................................. 2"
Intended Use:
Safety straps are intended to help to secure a normal
patient on the table. Not intended to restrain unruly
or mentally unstable patients. Patients should not be
left unattended on the table.
Option: 082 Chrome Safety Rail
Length ............................................................... 24"
Height ................................................................. 8"
Intended Use:
Safety rails are intended to help prevent patients
from inadvertently rolling off the table. Patients
should not be left unattended on the table.
Option: 083 2-Hook IV Pole and Mounting Bracket
Height Range ......................................................... "
Also available with 4-hook top
Intended Use:
IV pole and mounting bracket are intended to
integrate IV hook-up with the table
Option: 087 Locking Casters
Caster Diameter ................................................... 3"
Intended Use:
Locking Casters are intended to lock the casters and
prevent the individual wheels from turning or rolling.
Option: 084 Hand-Held Height Control
Intended Use:
Hand held table controller is intended to be an
alternative to the foot controller.

FAQs
PROBLEM:
Table is squeaking.
SOLUTION: Lubricate the table at designated points using Molykote G-0052Fg grease.
See drawing below for lubrication locations.
PROBLEM:
Backrest does not function properly.
SOLUTION: If the headrest does not remain flat in
the down position, loosen nut “A” and
tighten nut “B.”
If headrest does not move at all, loosen
nut “B” and tighten nut “A.”
After proper function has been achieved,
tighten both nuts.
Slide Block, 4 total (Base and Top Frame)
Lubricate using Molykote G-0052Fg
grease, top and bottom contact points.
Base
Top Frame
End Plug
Back Rest
Back Rest Actuator Arm
Back Rest Cylinder
Foot Lift Actuator
Arm.
Motor
Center Axle
Lubricate using
Molykote G-0052Fg
grease.
Base Axle
Top Frame Axle
DO NOT LUBRICATE.
Nut A
Nut B
10

11
PROBLEM:
Table height adjustment will not work properly.
SOLUTION: Make sure the table is connected to a working electrical outlet. Inspect all power cords
for cuts or breaks that would interfere with the electrical connection. Check to make certain foot pedal
cord is properly and securely plugged into motor. If after checking all these areas, the table will not
operate, or operates improperly, call a technician, or call Clinton Industries at 800-441-9131 for
assistance. NEVER USE THE TABLE IF IT IS NOT WORKING PROPERLY.
For assistance with other problems not covered in this guide, or questions about your
Clinton Power Table, please call Clinton Industries at 800-441-9131.
REMOVAL OF THE MOTOR FOR SERVICE
The motor assembly is located at the foot-end of the table on the base. Next, detach the motor
assembly. If possible, raise the table to its highest position. Raising the table will make it easier to remove
the motor mount bolts. Using help, lower the table onto its side onto a protective surface
such as a blanket or other material that will not harm the upholstered top. Remove the foot pedal
controller by unplugging the DIN plug on the side of the motor. Carefully remove the upper and
lower motor mount bolts to free the motor assembly.
Upper Motor Mount
Motor Assembly
Foot Pedal DIN Plug
Lower Motor Mount