PDF User Manual

  1. Home
  2. Manuals
  3. Clinispin Horizon 755VES Operator's Manual

Clinispin Horizon 755VES Operator's Manual

Made by: Clinispin Horizon
Type: Operator's Manual
Category: Laboratory Equipment
Pages: 16
Size: 1.83 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

Operator’s Manual

clinispin hOrizOn 755VEs 

Laboratory Centrifuge

Rev. 1.3

P/N 7711037


background image

Table of contents

Model Description 

 

page 3

Supplied Equipment   

page 3

Warranty Information   

page 3

Specifications 

 

page 4

Control Panel / Parts of the Centrifuge 

page 5

Setup Location 

 

page 6

Initial Setup Procedure 

pages 6,7

Operation 

 

page 7

Rotor Configurations   

                                

                        

  Horizontal Rotor   

page 8

  6-Place 50mL Fixed-Angle Rotor 

page 8

  20-Place Fixed Angle Rotor 

page 9

Control Panel Functions 

  Time Adjustment and Timer Operation 

page 10

  Speed Adjustment / RCF Display 

page 10

  Starting and Stopping a Run 

page 10

  Unlocking the Lid   

page 10

Advanced User Settings 

  Acceleration Rate   

page 10

  Braking Rate 

 

page 11

  Countdown Delay   

page 11

  Imbalance Detection 

page 11

  Audible Setting 

 

page 11

Memory Locations / Storing and Recalling 

page 12

Mechanical Cycle Counter 

page 12

Care and Preventative Maintenance 

page 13

Troubleshooting 

 

page 14

Safety 

 

page 15

Emergency Rotor Chamber Entry 

page 15

Calibration and Earth Ground Testing 

page 15

Replacement Parts   

page 16

page 2


background image

WARNING:

   For the safety of both the operator and service 

personnel, care should be taken when handling substances that are known to be toxic, radioactive 

or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II 

materials are used (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), 

a Bio-Seal should be employed. More than one level of protection must be provided in the case of 

materials of a higher group. The use of flammable or explosive materials as well as those materials 

which chemically react vigorously is prohibited.

suppliEd AccEssOriEs (standard)

One (1) 6-Carrier

Horizontal Rotor*

Also included (not shown):

•  One (1) Mains Lead    •  One (1) Operator’s Manual

Six (6)

60mL Carriers*

page 3

Available Accessories:

•  6-Place Carrier, for tubes up to 17mm x 125mm     •  30-50mL Carrier

WArrAnTY:

Woodley Equipment Company warranties that this centrifuge is free from

defects in workmanship and parts for 12 months.

Old Station Park Buildings 

St. John’s Street, Horwich

Bolton BL6 7NY, United Kingdom

Tel: +44 (0) 1204 669033 • Fax: +44 (0) 1204 669034

www.woodleyequipment.com  Email: sales@woodleyequipment.com

www.horizoncentrifuges.com Email: sales@horizoncentrifuges.com

Made in the USA 

•The rotor and rotor accessories are rated for a rotational speed of 3500 RPM (a force of 2200 xg).

  

intended use:

General purpose laboratory centrifuge for sample separation.


background image

Maximum speed: 

 

 

 

3500 RPM 

Maximum Force (rcF): 

 

 

2200 xg

Overall dimensions  (h x W x d): 

 

23 x 37 x 43 cms 

centrifuge Motor: 

 

 

 

1/2 H.P. Brushless DC

protection Breaker:

   

 

 

4 Amp. re–settable

Timer:

 

 

 

 

electronic, with hold

 

 

 

 

or 0 to 99 minutes, +/- 1%

Weight:

 

 

 

 

17.7 kgs

permitted Environmental conditions

Ambient Temperature during Operation:

  

o

C - 35 

o

C

Maximum relative Air humidity:

 

 

 

90%

Electrical requirements:

power:

 

 

 

 

 

200 Watts

Voltage:  

 

 

North America:

 

115 Volts (+/- 10%)

         

 

Japan:

   

 

100 Volts (+/- 10%)

 

 

Europe: 

 

 

230 Volts (+/- 10%)

Frequency:

 

 

 

 

 

50/60 Hz

Any use other than those specified by the Manufacturer is explicitly prohibited.  

Maximum sample density is 1.15 grams / mL, 

(water density = 1.0 grams / mL)   

*

See label on rear of centrifuge for voltage requirements.

SpecificAtionS:

Specifications for the Clinispin Horizon  755VES Centrifuge

page 4

   

 

Part No.  #  Max tube Size  SPeed raNge 

Force raNge

6-carrier Horizontal (Standard) 

7786022  comes with your choice of 6 of the holders listed

   

w/4-Place carrier 

7713023  24  17mm x 100mm  500 to 3,500 rPM  40 to 2,200 xg

   

w/4-Place carrier 

7713023  12  17mm x 125mm  500 to 3,500 rPM  40 to 2,200 xg

   

w/50mL Hor. carrier 

7713037  6  34mm x 115mm  500 to 3,500 rPM  40 to 2,200 xg

   

6-Place 50mL Fixed-angle  7786044  comes with 6 of the holders listed

   

w/50mL Fixed-angle holder 7787051 6  34mm x 125mm  500 to 4,000 rPM  40 to 2,000 xg

   

 

 

   

20-Place Fixed-angle 

7786042  comes with 20 of each of the holders listed

   

w/75mm tube holder   7713043  20  17mm x 75mm  500 to 4,000 rPM  40 to 1,700 xg 

   

w/100mm tube holder   7713040  20  17mm x 100mm  500 to 4,000 rPM  40 to 1,900 xg 

   

w/125mm tube holder   7713044  20  17mm x 125mm  500 to 4,000 rPM  40 to 2,000 xg

   

12-Place Horizontal 

7786024  comes with 20 of each of the holders listed

   

w/75mm tube holder   7713043  20  17mm x 75mm  500 to 3,200 rPM  40 to 1,240 xg 

   

w/100mm tube holder   7713040  20  17mm x 100mm  500 to 3,200 rPM  40 to 1,540 xg 

   

w/125mm tube holder   7713044  20  17mm x 125mm  500 to 3,200 rPM  40 to 1,830 xg

 

75 mm Tube 

Holder

p/n 7713043

100 mm

Tube Holder

p/n 7713040

125 mm

Tube Holder

p/n 7713044

Tube Holder Cap

p/n 7713011

50mL

tube holder

fixed-angle

p/n 7787051

4-Place Carrier

p/n 7713023

50mL Carrier

p/n 7713037

Carrier

Cap

p/n

7713035


background image

page 5

cOnTrOl pAnEl / pArTs OF ThE cEnTriFugE:

lid Knob

lid

lid safety

interlock system
control panel

(see below)

cycle counter

power Button

up/down rpM 

change Buttons

rpM display

running

indicator 

light

locked

indicator 

light

unlocked 

indicator 

light

start Button

Open / stop 

Button

up/down Time change Buttons

Time display


background image

sETup lOcATiOn:

1.   Unpack the centrifuge and verify that all of the supplied equipment is present.
2.   Choose a setup location which meets the following criteria:

    a)   A bench top clearance height of 51 cms” is required in order to open the lid.
    b)   The clearance envelope is the space around the centrifuge which is required for    

 

        safety. Choose a setup location which will allow for a clearance envelope of at least  

 

        76 x 76 cms, (with the centrifuge at the center). No person or hazardous material shall  

 

        be permitted in the clearance envelope during operation. The operator time within  

 

        the envelope shall be limited to the time necessary for loading, unloading and    

 

        centrifuge operation only. 

    c)   Proper ventilation is necessary to prevent the overheating of samples as well as    

 

        premature failure of the centrifuge. Choose an area which will allow unencumbered  

 

        air flow.
    d)   The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary  

 

        for leveling the centrifuge, however, the surface should be flat and level.
    e) 

  Be sure the outlet is always within reach as the mains lead is the means of   

 

        emergency disconnection!

iniTiAl sETup prOcEdurE:

 if any problems are found during the initial setup 

procedure, refer to the troubleshooting section.  for further assistance, contact  Woodley 

equipment company. +44 (0) 1204 669033.

1.   Plug the female end of the mains lead into the rear of the centrifuge. Plug the male end into    

    an approved electrical outlet. For electrical safety, the unit must always be properly    

 

    grounded.

2.   Engage the power button on the front of the centrifuge (if it is not already engaged). The  

 

    power button is engaged when it is illuminated.
3.   For operator safety, the locking system is always active; requiring power and direction  

 

    from the user to disengage it (the lid also automatically unlocks at the end of a run when  

 

    it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the    

 

    ‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ indicator light should    

 

    illuminate. If it does not, refer to the section on troubleshooting. The lid will remain  

 

 

    unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
4.   Turn the Lock counterclockwise and open the lid.  

5.   Spin the rotor by hand; check for free and level rotation.  If the rotor does not spin freely,  

 

    refer to the section on troubleshooting.
6.   Place the test tube carriers into the rotor and verify that they are seated properly.
7.   Close the lid. Rotate the lid knob clockwise to its complete stop position.  The  

 

 

    ’LOCKED’ indicator light should be illuminated.  If it is not, make sure that the lid is    

 

    latched properly. The centrifuge will not run unless the lid is latched properly and the    

 

    ’LOCKED’ indicator light is illuminated.
8.   Initiate a test run by pressing the ‘START’ button.
9.   The ‘RUNNING’ indicator light will illuminate and the cycle counter will increment by one.

10.  The test tube carriers will slide up into the horizontal position and the unit will accelerate  

 

    to the current set speed.
11.  Listen to the sound of the centrifuge. A smooth whirring sound should be heard.  If there  

 

    are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button 

  

    immediately and refer to the section on troubleshooting.

(continued on next page)

page 6


background image

OpErATiOn:   

nOTE:  Follow the initial setup procedure before initial operation.

1.  Press the ‘OPEN /  STOP’ button to unlock the lid and then open the lid.

2.  Place the test tube samples into the tube carriers. Be sure to follow the rules for  

 

 

 

balanced loads.  

3.  Close the lid and turn the lid knob clockwise to its complete stop position. The  

 

 

 

LOCKED’ indicator light should illuminate to indicate that the latch is closed    

 

 

 

properly. If the lid knob is not completely latched, the ‘LOCKED’ indicator light will not   

 

       illuminate and the centrifuge cannot be operated.

4.  Set the desired speed and run time using the appropriate up and down arrow buttons.
5.  Begin the run by pressing the ‘START’ button on the control panel.
6.  The centrifuge should begin to spin. The ‘RUNNING’ indicator light should illuminate.

 

IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE 

TO SHUT DOWN, PRESS THE ‘OPEN / STOP’ BUTTON IMMEDIATELY !

7.  After time has elapsed, the ‘RUNNING’ indicator light will extinguish and the rotor will   

 

 

decelerate to a complete stop.

8.  The ‘UNLOCKED’ indicator light will illuminate and the locking mechanism will  

 

 

 

disengage, allowing entry into

 the rotor chamber. If the automatic unlock times out    

 

 

after 60 seconds) simply press the ‘OPEN / STOP’ button.

9.  Turn the lid knob counterclockwise and open the lid.
10.  Remove the samples.
11.  The centrifuge may be used immediately.

iniTiAl sETup prOcEdurE (continued):

12.  Press the ‘OPEN / STOP’ button to terminate the test run. The rotor will decelerate to a  

 

    complete stop and the lid will then unlock automatically for sixty (60) seconds.

13.  The centrifuge is now ready for operation.

page 7


background image

rOTOr cOnFigurATiOns:

    1.  Opposing tube carriers must be identical.

    2.  Opposing tube carriers must be empty or loaded with an equal number of  

 

 

  equally weighted samples.

    3.  If an odd number of samples is to be spun, use a water-filled tube to mate    

 

 

  with the unpaired one.

Your centrifuge must contain a balanced load in order

to work properly. to ensure that the load is balanced,

keep these rules in mind when inserting test tube

samples into the four (6) position horizontal rotor:

This rotor is designed to hold up to 6 tube carriers.

Follow the procedure below for proper tube loading. 

page 8

    1.  Opposing tube carriers must be identical.

    2.  Opposing tube carriers must be empty or loaded with an equal number of  

 

 

  equally weighted samples.

    3.  If an odd number of samples is to be spun,

      use a water-filled tube to mate with

      the unpaired one.

note: 

Do not load more than two (2) tubes per carrier if the tubes are greater than  

 

 

100mm long (15mL). When loading these tubes into the carriers, place them all  

 

 

on either side.  For example, to load twelve (12) 15mL tubes, place two tubes on  

 

 

the right-hand side of each carrier.

 

 

WARninG: 

When using glass test tubes, place a 7mm cushion at the bottom of each   

location in the carriers. Failure to do so may result in damage to the test tube.

Six (6) Place 50mL Fixed-angle Rotor (optional):

Your centrifuge must contain a balanced load in order to

work properly. to ensure that the load is balanced, keep

these rules in mind when inserting test tube samples into

the 6 place, 50mL fixed-angle rotor:

    1.  Opposing tube holders must be identical.

    2.  Opposing tube holders must be empty or loaded equally weighted samples.

    3.  If an odd number of samples is to be spun, use a water-filled tube to mate    

 

 

  with the unpaired one.

    4.  (Four Position Carriers Only):

      When loading tubes into opposing

      carriers, the carriers will be   

 

 

  balanced when the tubes are loaded   

 

  diagonally from one another

      (not across from one another).   

 

 

 

For example:

 six tubes are to be loaded

      into two carriers. In each carrier, two tubes    

  are placed in the bottom and one tube is  

 

  placed in the upper left-hand position. 

      See picture at right.

This rotor is designed to hold up to six (6) 50mL tube holders.

Six (6) Carrier Horizontal Rotor (standard):


background image

page 9

75mm Tube Holder

 

with

1525 cushion

75mm

Tube

Holder

 

100mm

Tube

Holder

125mm

Tube

Holder

with

1525

cushion

B1.      B2. 

or

D.

ORANGE

PURPLE

BLUE

BLUE

C1.     C2.

B.

C.

D.

A.

100mm

Tube

Holder

with

1525

cushion

A.

125mm

Tube

Holder

PURPLE

For Example: A tube is found to be as long as box B. Accordingly, we can 

use a 125 mm tube holder with a 1525 cushion or a 100mm tube holder 

with no cushion, (configurations B1 or B2).

This rotor is designed to hold up to twenty (20) tube holders. 

It comes with three tube holder types. The orange holders are 

designed to hold tubes up to 15mL (up to 17mm x 125mm). The 

purple holders are designed to hold tubes up to 10mL

(up to 17mm x 100mm). The blue holders are designed to hold 

tubes up to 5mL (up to 17mm x 75mm).

Your centrifuge must contain a balanced load in order to work properly. to ensure

that the load is balanced, keep these rules in mind when inserting test tube samples

into the 20 place fixed-angle rotor:

    1. 

Opposing tube holders must be identical.

    2. 

Opposing tube holders must be empty or loaded equally weighted samples.

    3. 

If an odd number of samples is to be spun, use a water-filled tube to mate    

 

 

with the unpaired one.

DiRectionS:

1.  Compare the tube to be spun with the four boxes shown below.

2.  Find the box that most closely matches the tube’s length. 

nOTE:

 The tube length with 

its stopper or cap must be shorter then the chosen box.

3.  Match the letter from the chosen box with one of the configurations shown.   

Twenty (20) Place Fixed-angle Rotor (optional):

ORANGE


background image

cOnTrOl pAnEl FuncTiOns:

time Adjustment and timer operation

The run time may be set from 30 seconds to 99 minutes and 30 seconds.  

Press the up and down arrow buttons next to the time display to change the run time. Adjustments 

may be made prior to a run or after a run begins. A quick tap will adjust the time by 30 seconds.   

Hold down the button for 1 minute adjustments - continue to hold for 5 minute adjustments.

Press the down button once while 00:30 is displayed  to access the “hold” feature. When a “hold” 

run is initiated, the timer will count up to keep track of total run time. The unit will run for a maximum 

of 99 minutes and 30 seconds and will then shut off automatically. The “hold” run may also be 

terminated by pressing the STOP/OPEN button. The total run time during a “hold” run is displayed 

on the time display until the operator unlocks the lid by pressing the OPEN button or opens the lid 

(during the automatic unlock after the run ends).

Speed Adjustment/ Rcf Display

 

Press the up and down arrow buttons next to the speed display to change the run speed.  

Adjustments may be made prior to a run or after a run begins. A quick tap will adjust the speed 

by 50 RPMs. Hold down the button for 100 RPM adjustments. Continue to hold for 500 RPM 

adjustments.

Press the RCF button at any time to convert the current speed setting to the corresponding force 

value (xg). The speed may be set by RCF by holding down the RCF button while using the up/down 

speed set buttons. The jumps made in the RCF value while holding down the RCF button and 

pressing the up and down arrow buttons correspond to 50 RPM adjustments in speed.  Since the 
RCF is a function of the square of the speed, this jump size will vary.

Starting and Stopping a Run

With the lid switch closed, press the START button to begin a run.  

Press the STOP/OPEN button at any time during a run to terminate it.

Unlocking the Lid

The lid is unlocked whenever the red UNLOCKED indicator light is illuminated. The lid unlocks 

automatically for 60 seconds at the end of a run. The lid can also be unlocked for an additional 15 

seconds by pressing the OPEN/STOP button while the unit is idle. The lid cannot be unlocked while 

the rotor is spinning.

imbalance Detection

This centrifuge is equipped with imbalance detection. If an imbalance is detected, the centrifuge will 

terminate the current run and will begin to brake to a stop. The words “BALANC” and “ERROR” will 

flash on the speed display. Once the rotor has stopped, open the lid to cancel the error reporting, 

balance the load and begin a new run. Alternately, the error reporting can be canceled by pressing 

the OPEN/STOP button.

Advanced user settings:

This centrifuge allows for total control of the run by the user. You may adjust the acceleration rate, 

adjust the braking rate, adjust the imbalance sensitivity level, select from several audible settings, 
and use the optional countdown delay.

Acceleration Rate

  - 

The user may adjust the acceleration rate from minimum (1) to maximum 

(10) using the following procedure:

 

1.   

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “ACCEL” is  

 

 

 

displayed in the speed display.

 

3. 

The current acceleration rate is displayed in the time display.  Use the ‘UP/ 

 

 

 

DOWN arrow buttons next to the time display to adjust the acceleration rate  

 

 

 

from 1 (minimum) to 10 (maximum).

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

page 10


background image

Advanced user settings (continued):

Braking Rate

  - 

The user may adjust the braking rate from minimum ( slowest deceleration, 0) to 

maximum (quickest deceleration, 9) using the following procedure:

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BRAKE” is  

 

 

 

displayed in the speed display.

 

3. 

The current braking rate is displayed in the brake display. Use the UP/DOWN  

 

 

 

arrow buttons next to the brake display to adjust the braking rate from 0   

 

 

 

(minimum) to 9 (maximum).

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

countdown Delay

  - 

When the countdown delay is turned on, the timer will not begin 

decrementing until the rotor has reached full speed as set by the user. This feature ensures that the 

samples see full force for the total run time.

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “CNTDLY” is    

 

 

displayed in the speed display.

 

3. 

The current status of this feature (ON or OFF) is displayed in the time display.     

 

 

Press UP/DOWN arrows next to the time display to toggle the status between    

 

 

ON and OFF.

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

imbalance Detection

  - 

This centrifuge is equipped with imbalance detection. This ensures that 

the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction 

occurs. This feature is intended to provide for operator safety and to extend the life of the 

centrifuge. The LOW setting will allow for a greater imbalance before shutting the centrifuge down.  

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BALANC” is    

 

 

displayed in the speed display.

 

3. 

The current status of this feature (LOW or HI) is displayed in the time display.     

 

 

Press UP/DOWN arrows next to the time display to toggle the status between    

 

 

LOW and HI.

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

Audible Settings

  - 

The user may select from six different audible indicator levels:

Setting  Button Press Beep?  End-of-run audible notification

NO 

NONE

YES 

NONE

NO 

YES, 5 beeps

YES 

YES, 5 beeps

NO 

YES, continuous, user must press STOP/OPEN button to cancel

YES 

YES, continuous, user must press STOP/OPEN button to cancel

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BEEPER” is    

 

 

displayed in the speed display.

 

3. 

The current audible selection is displayed in the time display.  Use the UP/ 

 

 

 

DOWN arrow buttons next to the time display to select a setting (1 to 6).

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

 

page 11


background image

The Clinispin horizon 755VES centrifuge is capable of storing up to 10 user-defined presets.  

These memory presets contain all of the settings needed to define a specific run, (run time, speed, 

acceleration rate, braking rate, etc.). The user can use these memory locations to quickly configure 

the centrifuge for a specific test type and ensure that the test is run in the same way each time by 

recalling this same setting.

Store a configuration in a memory location

 

1. 

Adjust the various settings (speed, time, acceleration rate, etc.) for the    

 

 

 

configuration to be saved.

 

2. 

Press and hold the PROGRAM button until the preset display begins to flash.

 

3. 

Press the MEMORY button to increment the memory location by one. Continue    

 

 

to press the memory button until the desired location is displayed.

 

4. 

Press the program button to store the current configuration into the memory  

 

 

 

location that is displayed.

Recall a configuration from a memory location

 

1. 

Press the MEMORY button until the desired memory location is displayed in the    

 

 

memory display.

 

2. 

As you review the presets, the settings on the control panel will change to  

 

 

 

reflect the settings stored in the memory preset currently displayed.

 

3. 

When the desired preset is found, pause and the memory display will    

 

 

 

momentarily flash. The stored configuration is now loaded.

notes:

Once a configuration has been loaded, it will remain active until any change is made on the control 

panel. If for instance the user adjusts the time, the current configuration will no longer match the 

previously loaded configuration and the memory display will change to a “-” dash. The previous 

configuration may be recalled using the procedure described above.

Memory locations may be overwritten by selecting them during the store procedure.

cycle counter:

This centrifuge is equipped with a mechanical cycle counter to monitor machine usage and for 

routine maintenance. The cycle counter is located on the front of the centrifuge.

Memory locations / storing and recalling:

page 12


background image

cArE And prEVEnTATiVE MAinTEnAncE:

With proper care and maintenance your Horizon centrifuge will provide years of laboratory service.  

For proper care, the following steps should be taken:    

1.  

provide Adequate Ventilation:

  For cooling purposes, the Horizon draws in ambient air  

 

    through the air intake cover on the top of the lid and exhausts this air in the rear of the base.    

    The centrifuge should be placed on a hard smooth surface for good air circulation.

2.  

Always spin Balanced loads:

  This centrifuge is equipped with imbalance detection.   

 

    However, it is still possible to run loads that are minimally imbalanced. Refer to page 8 on  

 

    balanced loads for additional information on balancing the load.

3.  

Keep the Tube carriers / holders clean: 

  

nOTE:

 Always follow the safety guidelines of  

 

    your laboratory to properly clean up and/or dispose of materials in the event that a    

 

    substance known to be potentially toxic, radioactive or contaminated with a pathogenetic  

 

    microorganism is spilt in or on the centrifuge. Small glass fragments left in the tube carrier  

 

    / holder after a tube breakage may adhere to the next test tube inserted in that carrier /  

 

    holder. When this tube is handled, these fragments may puncture  protective gloves and  

 

    lacerate the operator’s fingers or hand. Remaining fragments may provide stress points  

 

    on subsequent tubes and result in additional breakage. If a tube breakage occurs,    

 

    carefully remove the tube holder. Properly dispose of the sample and tube fragments and  

 

    thoroughly clean both the inside and outside of the tube carrier / holder. Insert a new tube  

 

    cushion (if necessary) and replace the tube carrier / holder in the rotor.  

4. 

  Motor and Electrical Maintenance:  

This centrifuge uses a brushless-DC motor. There  

 

    are no brushes to replace and it should not need routine servicing for the life of the    

 

    centrifuge. The electrical components are selected for high reliability and should not need  

 

    service.

5.

   Keep the centrifuge clean

: The cabinet, rotor top and accessories shall be thoroughly  

 

    cleaned using either isopropyl alcohol, soap and water, or a mild bleach solution. The use  

 

    of Fully/Partially Halogenated Hydrocarbons, Ketones, Esters and all other chemicals not  

 

    prescribed by the manufacturer may cause damage to the rotor and tube carriers / holders  

 

    and should not be used.

    If it is necessary to remove the rotor for additional cleaning it is required that a qualified  

 

    technician remove the outside housing and rotor assembly. Contact Woodley Equipment 

    Company for additional information.

    Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or  

 

    other cleaning solutions as this will cause damage and void your warranty!

6.  

Tube carrier / holder replacement:

 Tube carriers are stamped with an expiration date and    

    should be replaced according to this stamp.

page 13


background image

TrOuBlEshOOTing:

  1.  problem: 

The rotor does not spin freely.

    solutions: 

• Make sure that nothing has fallen into the rotor chamber.

     

• If there is nothing obstructing the rotor, the motor may be damaged.   

 

       

  Contact your authorized dealer or Woodley Equipment Company for further    

       

  assistance.

  2.  problem: 

There is excessive noise when the machine is running.

    solutions: 

• Check to see that the load is balanced.

     

• Make sure that nothing has fallen into the rotor chamber.

     

• Make sure that the screw in the center of the rotor is snug.

     

• The motor may be damaged.  Contact your authorized dealer or  

 

 

        

  Woodley Equipment Company for further assistance.

  3.  problem: 

The centrifuge does not run or an error message is displayed.

    solutions: 

• Make sure that the centrifuge is getting power. Do the control panel   

 

       

  displays come on? Check the electrical outlet that the unit is plugged into.     

       

  Check the circuit breaker on the base near the rear of the centrifuge.

     

• If the unit is getting power, make sure that the lid latch is closed properly.   

 

      

  The latch is closed properly when the yellow ‘LOCKED’ indicator light is  

 

      

  illuminated.

     

• If the centrifuge stops soon after start-up and ‘ERROR’ is displayed on the    

       

  speed display, refer to the following table for fault information:

     

  ERROR /  SPEED

     

  The centrifuge cannot reach full speed due to a problem with the rotor, an  

 

       

  inadequate power supply, or other electrical problems. Press the OPEN/ 

 

       

  STOP button to cancel the error and then check the rotor and the line  

 

       

  voltage. Contact your authorized dealer or Woodley Equipment Company for  

     

  further assistance.

     

  ERROR / BALANC

     

  The centrifuge has detected an imbalance. Press the OPEN/STOP button to    

       

  cancel the error and balance the load. If the load is balanced, make sure  

 

      

  that the centrifuge is installed on a secure, level location.

  4.  problem: 

The lid knob cannot be turned / the lid cannot be unlocked.

    solutions: 

• Make sure the centrifuge has power and that the rotor is stopped. Press the    

      

  ‘OPEN/STOP’ button. The red ‘UNLOCKED’ indicator light should illuminate    

       

  and the unlocked mechanism should disengage, allowing entry into the  

 

       

  rotor chamber.

     

• If the lid is still locked, make sure that the red ‘UNLOCKED’ indicator is  

 

       

  illuminated and turn the lid knob first completely clockwise and then   

 

         

  counterclockwise.

     

• If the red ‘UNLOCKED’ indicator will not illuminate or the locking  

 

 

       

  mechanism will not disengage, the electronics or locking mechanism may  

 

       

  be damaged. Have a technician service the centrifuge or contact Woodley  

     

  Equipment Company for further assistance.

page 14

for additional assistance, services and technical support

contact Woodley equipment company LtD.


background image

sAFETY:

horizon lid safety switch:  

The Horizon lid is secured to the top of the cabinet by a latching 

knob and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the 

cabinet opening and prevents the lid from opening. A mechanical stop positions the pawl and 

prevents it from rotating completely.  When rotated to the stop position, the pawl makes contact 

with a micro–switch mounted underneath the cabinet top. The lid safety switch prevents the 

centrifuge from operating while the lid is open. The yellow ‘LOCKED’ indicator light on the front of 

the machine will illuminate when the lid has been latched properly.

horizon lid safety interlock system:  

In addition to the Lid Safety Switch, the Horizon has a 

true “0 RPM” lid locking system. The lid safety interlock system keeps the lid locked at all times, 

(even during power failure), and requires that the rotor be at rest in order to unlock the lid. The 

centrifuge will not allow entry into the rotor chamber unless the centrifuge has power and the rotor 

is stopped. To open the lid, make sure that the centrifuge is plugged in and, with the rotor stopped, 

press the ‘OPEN / EMERGENCY STOP’ button.

  Note: After the centrifuge has started spinning, it may be possible to rotate the lid knob      

  enough to cause the pawl to lose contact with the lid safety switch. If this happens, the   

  centrifuge motor may lose power, but the lid will still remain locked. If the knob is accidentally

  moved and this situation should occur, rotate the knob fully clockwise to its stop position and  

  the centrifuge will resume operation.

circuit Breaker:  

The Horizon is protected with a 4 Amp circuit breaker located at the rear of the 

machine mounted to the base. Any electrical short circuit will cause the breaker to cut power to the 

machine.

calibration and Earth ground Testing:

It is recommended that the top speed, ground continuity and line leakage be tested every two 

years for continued safe operation. Contact  Woodley Equipment Company. for further information 

or testing availability.

Emergency rotor chamber Entry:

In the event of power failure, it may be impossible to unlock 

the lid by conventional means. In this case, entry into the rotor 

chamber may be made by removing the latch label and using 

a pen to manually disengage the locking mechanism (see 

photo). Pull the mechanism towards the control panel and then 

unlatch and open the lid. If the unit is damaged, contact your 

authorized dealer or Woodley Equipment Company.

page 15

The Clinispin Horizon 755VES complies with all requirements of CE EN 61000-3-2, -3-3, EN 

61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-11, EN 61010-1, -2-020 and UL standard 3101–2–20, Can/

CSA C22.2 No. 1010.1 , Can/CSA C22.2 No. 1010.2.20 


background image

Made in the USA 

contact your authorized dealer or Woodley equipment company

to order replacement parts or accessories.

Old Station Park Buildings 

St. John’s Street, Horwich

Bolton BL6 7NY, United Kingdom

Tel: +44 (0) 1204 669033 • Fax: +44 (0) 1204 669034

www.woodleyequipment.com  Email: sales@woodleyequipment.com

www.horizoncentrifuges.com Email: sales@horizoncentrifuges.com

replacement parts:
  part no.  description     

                   

     

  7728052  Foot, rubber

  7751068  Lid Tray Micro-Switch

  7723002  Lid Tray Solenoid

  7735016  Motor, 1/2 H.P. Brushless DC

  7717041  PC Control Board

  3056001  Step Down Transformer

  7751043  Circuit Breaker

  7724149  Front Panel Label

  7760003  Mains lead

  7714101  Pawl, lock, lid

  7714103  Knob, lock, lid

  7712263   Lid

  7724071  Hinge, friction

  7732018  Gasket, cabinet opening

  7732019  Gasket, rotor chamber

125 mm

Tube Holder

p/n 7713044

Tube

Holder Cap

p/n 7713011

75 mm

Tube Holder

p/n 7713043

100 mm

Tube Holder

p/n 7713040

20-Place

Fixed-Angle Rotor

p/n 7786042

4-Place

Carrier

p/n 7713023

50mL

Carrier

p/n 7713037

Carrier Cap

p/n 7713035

6-Carrier

Horizontal

Rotor

p/n 7786022

50mL

tube holder

fixed-angle

p/n 7787051

6-Place 50mL

Fixed-Angle Rotor

p/n 7786044

250mm Tube 

Cushion

p/n 1525

7mm Tube 

Cushion

p/n 9150