Full Text Searchable PDF User Manual

User Manual
CLEMENTS
Duplex Adaptors
SUC89660
SUC89661
Manual No. SUC 89660 081
Issue 10

2
CLEMENTS
Safety
Thank you for purchasing this Clements Duplex Adaptor.
For your safety it is imperative that this unit only be
operated by authorised personnel in accordance with the
instructions as described in this manual. Operated in this
way, the Duplex Adaptor will provide years of service.
Due to continual improvements in product design, the
Duplex Adaptor may vary in detail from the descriptions in
this manual. In the event of further questions please
contact your local distributor or BMDi TUTA Healthcare
direct.
User Manual
Duplex Adaptor
Manual Number SUC 89660 081 Issue 10
Copyright © 2014 BMDi TUTA Healthcare Pty Ltd
The information in this manual was originated by, and is the exclusive
property of BMDi TUTA Healthcare Pty Ltd. It is furnished for customer
information only, and is not an authorisation or licence to make this
product or to furnish this information to others.
BMDi TUTA Healthcare Pty Ltd
Unit 4B, 128-130 Frances Street
Lidcombe NSW 2141 Australia
Phone:
+61 2 9466 5300
Website: www.medaust.com
C
0086
Priory Analysts Ltd
The Pinnacle, 160 Midsummer Blvd
Milton Keynes, MK9 1FF, United Kingdom
EC REP

3
CLEMENTS
Contents
Contents
Specifications
..................................................
4
Transportation and Storage
............................
5
Installation and Operation
..............................
6
Spares Parts
………………………………. ..............
7
Troubleshooting
……………………………………….
8
Cleaning
..........................................................
9
Disassembly
....................................................
10
Warranty
..........................................................
14

4
CLEMENTS
Specifications
Vacuum Range
0 to -100 kPa, 0 to -750 mmHg
Flow Rate
0 to 35 Litres of Free Air Per Minute
Lever
No Lever
Height
100 mm
40 mm
Width
75 mm
75 mm
Depth
80 mm
80 mm
Weight
300 g
250 g
Duty Cycle
Continuous
Ambient Temperature
+5°C to 35° C
Standard Conditions
25°C, Sea Level, 101 kPa
Class
Class I (EU Class IIa)
ARTG
174693
GMDN
14325
Specifications
Intended Use
Connect two pieces of medical gas equipment to a single
medical gas terminal unit.

5
CLEMENTS
Transportation and Storage
Environmental conditions for transportation and storage are
shown in the following table.
Parameter
Minimum
Maximum
Temperature
10ºC
40ºC
Humidity
60% RH
95% RH
Barometric Pressure
700 mBar
1060 mBar
Warning Symbols Legend
The warning symbols marked on the equipment and their
meanings are shown as follows.
Caution, consult accompanying documents
Transportation and Storage
Waste Materials
The contents of the collection jars, suction tubing, bacteria filter,
internal exhaust filter may contain biohazard wastes. Handle using
safe handling procedures, which may include the use of rubber
gloves and eye protection, and dispose of according to local
protocols for biohazard materials.
Recycling
At the end of their service life, the Nozzle should be dismantled if
necessary, and returned to a local materials recycling centre.

6
CLEMENTS
Description
The Clements Duplex Adaptor allows the connection of two
pieces of equipment to a single wall terminal unit.
Clements Duplex Adaptors are available to connect to most
types of wall outlet.
Installation and Operation
Carefully examine the Dupledx Adaptor for any visual signs
of damage that might have occurred during shipment.
Attach the adaptor to the wall outlet with the index
handwheel or relevant connector.
1.
Ensure any equipment is turned off before connecting to
the duplex adaptor.
2.
Select the outlet to use by moving lever (lever model only)
fully to corresponding position.
Installation and Operation
NOTE:
Only ONE piece of equipment can be in use
at any given time. Ensure that equipment
connected to the other outlet is switched
off.
Item
Description
SUC 89660
Duplex Suction Adaptor without Lever, Ring Index
SUC 89661
Duplex Suction Adaptor with Lever, Ring Index

7
CLEMENTS
Spare Parts
SUC 89771
Ring Index Connector Kit for Wall
Suction Products ( Handwheel, 1/8 BSP
Threaded Nipple, O-Ring)
SUC 89240 019
Puritan Bennett Connector with 1/8 BSP Thread
SUC 89240 074
BS MKIV Connector with 1/8 BSP Thread
SUC 89250 006
Drager Connector with 1/8 BSP Thread
SUC 89660
Duplex Suction Adaptor with Lever
SUC 89661
Duplex Suction Adaptor Without Lever
SUC 89660 081
User Manual for Duplex Adaptors
Select from a range of Medical Suction Tubing, Collection
Jars, Bung Assemblies, and Handpieces
Spare Parts

8
CLEMENTS
Fault
Check
Rectify
No Vacuum / Flow On / Off Lever
Ensure turned on
Wall outlet blocked Refer to engineering
dept.
Body blocked
Strip and clean
Troubleshooting
Troubleshooting

9
CLEMENTS
Cleaning
Cleaning
Clean using a pH neutral disinfectant for wiping or immersion
cleaning.
Flush after immersion using distilled water and dry before
reassembly of unit.
Cleaning Procedure
Tools and Materials Required
•
Allen key set, imperial
•
Vacuum grease
•
Can of spray, Dow Corning Molykote D-321R Sealer
•
Hand press and fixture for holding body
Cleaning

10
CLEMENTS
Disassembly Procedure
1.
Remove unit from wall.
2.
Remove the hand wheel by inserting an allen key into the
brass nipple in the centre of the hand wheel. Make sure
the allen key is pushed home into the brass nipple before
unscrewing.
4.
Remove the Sel-Lok pin , disc springs and O-ring using a
hand press and fixture.
5.
Push the plug out by gently twisting and pushing. Once
the plug is out, gently clean the plug with a soft brush
using a degreaser. Wash and blow-dry.
Use the same care when cleaning out the bore in the
nozzle body.
6.
Lightly coat the plug with Molykote surface spray and let
dry.
7.
Apply a light coat of vacuum grease to the body bore.
8.
To assemble - reverse the order
9.
To test - lever should be firm and smooth to turn, vacuum
when tested should read the same as the vacuum source,
flow should have a max of 2-3 lpm loss only.
10. Clean any vacuum grease off the body.
Cleaning Procedure

11
CLEMENTS
Notes

12
CLEMENTS
Notes

13
CLEMENTS
Notes

14
CLEMENTS
Warranty
BMDi TUTA Healthcare Pty Limited ("BMDi TUTA Healthcare") warrants
that this product is free from defects in workmanship and materials for
a period of 12 months from the date of shipment by BMDi TUTA
Healthcare or its authorised agent to the purchaser. Subject to the
conditions of this warranty, if the product fails to operate for any
reason within the warranty period and the product is returned to the
place of purchase at the purchaser's expense, BMDi TUTA Healthcare
will repair or replace the product free of charge.
If a valid warranty claim is made within 30 days from the date of
shipment, then BMDi TUTA Healthcare will also reimburse the
purchaser for reasonable freight costs in returning the product to the
place of purchase.
Conditions of Warranty
1.
The product must be returned to the place of purchase with proof
of purchase.
2.
This warranty is only available to the original purchaser of the
product.
3.
The product must not have had its serial number removed,
defaced or changed, its casing opened, its power supply altered or
have been tampered with in any other way.
4.
This warranty does not cover :
•
inadequate or incorrect site preparation;
•
improper installation;
•
connection to the wrong voltage;
•
failure of the product due to misuse;
•
the use or operation of the product outside of the physical,
electrical or environmental specifications of the product;
•
use in a manner or environment in which the product is not
designed to be used;
•
improper adjustment, calibration or operation by the purchaser;
•
the use of accessories including consumables, hardware or
software which were not manufactured or approved in writing by
BMDi TUTA Healthcare;
Warranty

15
CLEMENTS
•
any modifications of the product which were not authorised in
writing by BMDi TUTA Healthcare;
•
any contamination or leakages caused or induced by the
purchaser; and
•
inadequate or improper maintenance of the product.
5.
This warranty does not cover normal wear and tear.
6.
BMDi TUTA Healthcare will not be responsible for damage or loss
caused during shipping.
7.
In Australia, apart from any warranties implied by the Trade
Practices Act 1974 all other warranties expressed or implied and
whether arising by virtue of statute or otherwise are hereby
excluded.
8.
Outside Australia, all other warranties expressed or implied and
whether arising by virtue of statute or otherwise (including any
warranties implied by the Vienna Convention) are hereby
excluded.
9.
BMDi TUTA Healthcare' obligations under this warranty are
limited to the repair or replacement of the product, within the
terms of this warranty and the total liability of BMDi TUTA
Healthcare for loss or damage of every kind whether arising
pursuant to the terms of the sale of the product or otherwise in
connection with the product is limited to the amount paid by the
purchaser to BMDi TUTA Healthcare for the product.
10. Apart from any liability imposed by Part VA of the Trade Practices
Act, BMDi TUTA Healthcare accepts no other liability for any loss
or damage occasioned (including consequential loss or damages)
in any way as a result of the use of the product.
11. The warranty does not extend to cover damage to the following
parts as they are inherently prone to wear :
•
motor brushes
12. This warranty does not extend to cover corrosion due to any cause
nor to any damage to painted or anodised surfaces.
13. BMDi TUTA Healthcare will give the purchaser the benefit of any
manufacturer's warranty in respect of any components in the
product which were not manufactured by BMDi TUTA Healthcare,
if such a manufacturer's warranty is available.
Warranty