Instruction Manual

Multi Sub Electrophoresis Systems

MSMIDI96

MSMIDI96ST

MULTISUB4

 


background image

Contents:-

                 Page

1)

Safety Instructions

3

2)

Packing Lists

4

3)

Usage Guidance and restrictions            6

4)

Care and Maintenance

7

5)

Setting Up

8

6)

Gel Preparation

9

7)

Gel Pouring

10

8)

Gel Running

11

9) 

Gel Staining

12

10)

Solutions

13

11)

Combs

14

12)

Warranty

16

 


background image

SAFETY PRECAUTION

WHEN USED CORRECTLY, THESE UNITS POSE NO HEALTH RISK.

HOWEVER, THESE UNITS CAN DELIVER DANGEROUS LEVELS OF ELECTRICITY 

AND ARE TO BE OPERATED ONLY BY QUALIFIED PERSONNEL FOLLOWING THE 

GUIDELINES LAID OUT IN THIS INSTRUCTION MANUAL.

ANYONE INTENDING TO USE THIS EQUIPMENT SHOULD READ THE COMPLETE 

MANUAL THOROUGHLY.

THE UNIT MUST NEVER BE USED WITHOUT THE SAFETY LID CORRECTLY IN 

POSITION.

THE UNIT SHOULD NOT BE USED IF THERE IS ANY SIGN OF DAMAGE TO THE 

EXTERNAL TANK OR LID.

THESE UNITS COMPLY WITH THE STATUTORY CE SAFETY DIRECTIVES:

73/23/EEC: LOW VOLTAGE DIRECTIVE: IEC 1010-1:1990 plus AMENDMENT 1:1992

EN 61010-1:1993/BS EN 61010-1:1993

PACKING LISTS:

 


background image

MSMIDI96

Units include tank, lid and electrodes and include the following accessories:-

Tray

Tray Dams

Combs and 
comb block

Loading 
Guides

Cables

MSMIDI96

MS10-UV96 10 x 
12cm (W x L)

MS10-UVDAM
Pack of 2

MSMIDI96-8-1-CB

MS10-LG – Strips
MS10-WP – Platform

CSL-CAB

MSMIDI961.5

MS10-UV96 10 x 
12cm (W x L)

MS10-UVDAM
Pack of 2

MSMIDI96-8-1.5-
CB

MS10-LG – Strips
MS10-WP – Platform

CSL-CAB

MSMIDI96/2M

MS10-UV96 10 x 
12cm (W x L)

MS10-UVDAM
Pack of 2

MSMIDI96-8-
1/2M-CB

MS10-LG – Strips
MS10-WP – Platform

CSL-CAB

MSMIDI96/1.5/2M

MS10-UV96 10 x 
12cm (W x L)

MS10-UVDAM
Pack of 2

MSMIDI96-8-
1.5/2M-CB

MS10-LG – Strips
MS10-WP – Platform

CSL-CAB

MSMIDI96ST

Units include tank, lid and electrodes and include the following accessories:-

Tray

Tray 
Dams

Combs and 
comb block

Loading 
Guides

Cables

MSMIDI96ST

MS10-UV96ST 10 x 
24cm (W x L)

MS10-
UVDAM
Pack of 2

MSMIDI96ST-8-1-
CB

MS10-LG – Strips
MS10-WP – 
Platform

CSL-CAB

MSMIDI96ST1.5

MS10-UV96ST 10 x 
24cm (W x L)

MS10-
UVDAM
Pack of 2

MSMIDI96ST-8-1.5-
CB

MS10-LG – Strips
MS10-WP – 
Platform

CSL-CAB

MSMIDI96ST/2M

MS10-UV96ST 10 x 
24cm (W x L)

MS10-
UVDAM
Pack of 2

MSMIDI96ST-8-
1/2M-CB

MS10-LG – Strips
MS10-WP – 
Platform

CSL-CAB

MSMIDI96ST/1.5/2

M

MS10-UV96ST 10 x 
24cm (W x L)

MS10-
UVDAM
Pack of 2

MSMIDI96ST-8-
1.5/2M-CB

MS10-LG – Strips
MS10-WP – 
Platform

CSL-CAB

MultiSUB4

Units include tank, lid and electrodes and include the following accessories:-

 


background image

Trays

Combs

Caster

Cables

CSL-MULTISUB4

4 x MSUB4UV18 8 
x 18cm (W x L) 

8 x MSUB4-18/8-
1.5 1.5mm, 18 and 
8 sample

-

CSL-CAB

CSL-
MULTISUB4EXCAS

4 x MSUB4UV18 8 
x 18cm (W x L) 

8 x MSUB4-18/8-
1.5 1.5mm, 18 and 
8 sample

1 x MSUB4EXCASTER 
External Caster For Msub4 (4 
Trays)

CSL-CAB

The packing lists should be referred to as soon as the units are received to ensure that 

all components have been included. The unit should be checked for damage when 

received.

Please contact your supplier if there are any problems or missing items.

Usage Guidance and restrictions:

• Maximum altitude 2,000m.

• Temperature range between 4°C and 65°C.

 


background image

• Maximum relative humidity 80% for temperatures up to 31

O

C decreasing linearly to 50% 

relative humidity at 40

O

C.

• Not for outdoor Use.

This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.

POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution occurs.

Occasionally, however, a temporary conductivity caused by condensation must be expected”.

 


background image

Care and Maintenance:-

Cleaning Horizontal Units

Units are best cleaned using warm water and a mild detergent. 

Water at temperatures above 

60

C can cause damage to the unit and components.

The tank should be thoroughly rinsed with warm water or distilled water to prevent build up 

of salts but care should be taken not to damage the enclosed electrode and vigorous cleaning 

is not necessary or advised.

Air drying is preferably before use.

The units should only be cleaned with the following:-

Warm water with a mild concentration of soap or other mild detergent.

Compatible detergents include dishwashing liquid, Hexane and Aliphatic hydrocarbons

The units should not be left to in detergents for more than 30 minutes. 

The units should never come into contact with the following cleaning agents, these will 

cause irreversible and accumulative damage:-

Acetone, Phenol, Chloroform, Carbon tetrachloride, Methanol, Ethanol, Isopropyl alcohol

Alkalis.

Rnase Decontamination

This can be performed using the following protocol:-

Clean the units with a mild detergent as described above.

Wash with 3% hydrogen peroxide (H2O2) for 10 minutes. 

Rinsed with 0.1% DEPC- (diethyl pyrocarbonate) treated distilled water, 

Caution: 

DEPC is a suspected carcinogen. Always take the necessary precautions when 

using.

RNaseZAP™ (Ambion) can also be used. Please consult the instructions for use with acrylic 

gel tanks.

Setting up the Horizontal Gel Tanks:-

Instructions for fitting Electrode Cables 

 


background image

1. Note the position of the lid on the unit. This shows the correct polarity and the correct 

orientation of the cables, black is negative and red positive. 

2. Remove the lid from the unit. Note if the lid is not removed, fitting the cables may 

result in un-tightening of the gold plug and damage to the electrode.

3. Screw the cables into the tapped holes as fully as possible so that there is no gap 

between the lid and the leading edge of the cable fitting.

4. Refit the lid.

Instructions for fitting Loading Guides 

These can be fitted to enhance visibility of the wells if desired. They can be fitted to 

the white vinyl platform sheet or to the unit itself. 

1. Seat the tray in the unit and note the position of the comb grooves. The samples run 

black to red but the trays can be used frontward or backwards so ensure that the comb 

grooves closest to the black electrode are marked.

2. Remove the tray.

3. Peel the back off of the loading guide and carefully apply the loading guide directly to 

the gel platform. 

The unit is now ready to be used.

Gel Preparation:-

1. Table 1 below shows the volume of agarose solution required to make the desired agarose 

gel for each unit tray size. For a standard 0.7% agarose gel, add 0.7 grammes of agarose to 

100 ml of 1x TAE or TBE solution. The same 1 x solution should be used in the tank buffer 

solution.

Multi Sub Midi 96

Multi Sub Midi 96 

Stretch

Multi Sub-4

 


background image

Gel volume for a 5mm thick 
gel

Gel volume for a 5mm thick 
gel

  Tray

Gel volume for a 4x 5mm thick gels

60ml

120ml

8 x 6cm

96ml

8 x 12cm

192ml

8 x 18cm

288ml

2. Add the agarose powder to a conical flask.

3. Add the appropriate amount of 1x TAE or TBE solution from the table above. To prevent 

evaporation during the dissolving steps below, the conical flask should be covered with 

parafilm.

4. Dissolve the agarose powder by heating the agarose either on a magnetic hot plate with 

stirring bar or in a microwave oven. If using the microwave method, the microwave should 

be set at around a 400 watt or medium setting and the flask swirled every minute. The 

solution should be heated until all crystals are dissolved. This is best viewed against a light 

background. Crystals appear as translucent crystals. These will interfere with sample 

migration if not completely dissolved.

The gel must be cooled to between 50°C and 60°C degrees before pouring.

 


background image

Gel Pouring:-

The CSL Multi sub range of units allows three different methods of gel casting:-

Casting Dams

Flexicaster

Traditional Tape

Using trays with Casting Dams (Note for MultiSUB4):-

1. Fit the casting dams over each end of the tray and place onto a level surface. The 

dams should be fitted so that there is no gap between the sides of the tray and the 

groove in the dams. This will ensure that there is no possibility of gel leakage.

2. Place the comb(s) in the grooves. Each tray has more than one comb grove so that 

multiple combs can be used. Using multiple combs increases sample number available 

per gel but decreases run length and care must be taken to ensure that samples from 

the first wells do not migrate into the lanes of the second comb wells.

3. Pour in the agarose carefully so as not to generate bubbles. Any bubbles that do occur 

can be smoothed to the edge of the gel and dispersed using a pipette tip.

4.  Allow the agarose to set, ensuring that the gel remains undisturbed.

5. Carefully remove the gel casting gates and comb and transfer the gel including tray to 

the main tank.

Using Traditional tape method:-

1. Autoclave or plastic backed general tape should be used.  A length 5cm longer than 

the width of each end of the tray should be cut. One length should be placed over one 

end of the tray and stuck 1cm in from the tray edge. This should then be folded and 

the edges sealed securely. Repeat for the other end and place onto a level surface for 

gel pouring. 

2. Place the comb(s) in the grooves. Each tray has more than one comb grove so that 

multiple combs can be used. Using multiple combs increases sample number available 

per gel but decreases run length and care must be taken to ensure that samples from 

the first wells do not migrate into the lanes of the second comb wells.

 


background image

3. Pour in the agarose carefully so as not to generate bubbles. Any bubbles that do occur 

can be smoothed to the edge of the gel and dispersed using a pipette tip.

4.  Allow the agarose to set, ensuring that the gel remains undisturbed.

5. Carefully remove the gel casting gates and comb and transfer the gel including tray to 

the main tank.

Running the Gel:-

1. Mix the sample to be loaded with sample buffer – see solutions for common sample 

buffers. Usually 3ul of sample buffer is adequate but less may be used with sample volumes 

of less than 10ul.

2. Fill the unit with enough buffer so that it will just cover the gels when they are immersed. 

This will give the fastest resolution times. For enhanced quality of resolution of sample, fill 

the unit to 5mm above the gel. 

3. On the bench surface, load a small amount of running buffer to flood the wells. Load the 

samples into the wells using a pipette. Multi-channel pipettes can be used for loading samples 

into wells formed by MC compatible combs, see listing in accessories for identification of 

these.

4. Once loaded, gently immerse the gels within the buffer, stacking them carefully on top of 

each other.  Replace lid and connect unit to the power supply using cables.

5. Typically gels are run at between 90 and 150 volts (50V MSUB4). However, maximum 

voltages are indicated on the serial badge of each unit. It should be noted that higher voltages 

generally give faster but poorer quality sample resolution.

Gel Staining and Viewing:-

The Multi Sub trays and the Mini Fast unit allow staining to be performed without removing 

the gel from the tray if this is preferred. 

1. Transfer the gel to a vessel containing the appropriate volume of 0.5 µg/ml ethidium 

bromide stain for 15–30 minutes, see solutions for stock stain concentration and adjust to the 

volume used accordingly. The entire gel should be covered.

 


background image

NOTE:-  

Ethidium bromide is a suspected carcinogen and the necessary safety precautions 

should be undertaken.

2. De-stain the gel for 10–30 minutes in distilled water again ensuring the gel is completely 

immersed. 

3. Rinse the gel twice for a couple of seconds with distilled water.

4. Transfer the gel to a UV Transilluminator.

5. The samples will often appear as brighter, clearer bands when photographed or viewed 

using a gel documentation system. However if the gel bands are too faint then the staining 

procedure should be adjusted so that there is less de-staining. If there is too much background 

then the staining procedure should be adjusted so that there is more de-staining. 

References

1. Sambrook, Fritsch, and Maniatis, 

Molecular Cloning A Laboratory Manual, 

Second 

Edition,

Cold Spring Harbor Laboratory Press, 1989.

2. 

Current Protocols in Molecular Biology, 

Greene Publishing Associates and Wiley-

Interscience,1989.

Solutions:-

1x TAE  

40 mM tris (pH 7.6), 20 mM acetic acid, 1 mM EDTA.

50x (1L) dissolve in 750 ml distilled water:

242 g tris base (FW = 121)

57.1 ml glacial acetic acid

100 ml 0.5 M EDTA (pH 8.0).

Fill to 1 litre with distilled water.

1x TBE 

89 mM tris (pH 7.6), 89 mM boric acid, 2 mM EDTA

10x (1L) dissolve in 750 ml distilled water:

108 g tris base (FW = 121)

55 g boric acid (FW = 61.8)

40 ml 0.5 M EDTA (pH 8.0)

 


background image

Fill to 1 litre with distilled water.

Sample Loading Dye

10x sample buffer stock consists of 50% glycerol, 0.25% bromophenol blue, and

0.25% xylene cyanole FF in 1x TAE buffer. Only 1–10 ml of the 10x loading dye should be 

prepared.

Ethidium Bromide Solution

Add 10 mg of Ethidium Bromide

 

to 1 ml distilled water.

Combs – MC Denotes Multi Channel Pipette compatible.

Multi Sub Midi 96 Combs

Description

Part Number

MSMIDI 96 Comb 8 sample MC + 1 Marker, 1mm thick 
COMB BLOCK

 MSMIDI96-8-1-CB 

MSMIDI 96 Comb 8 sample MC + 1 Marker, 1.5mm 
thick COMB BLOCK

 MSMIDI96-8-1.5-CB 

MSMIDI 96 Comb 8 sample MC + 2 Marker, 1mm thick 
COMB BLOCK

 MSMIDI96-8-1/2M-CB 

MSMIDI 96 Comb 8 sample MC + 2 Marker, 1.5mm 
thick COMB BLOCK

 MSMIDI96-8-1.5/2M-CB 

MSMIDI 96 Comb 8 sample MC, 1mm thick.  one marker 
lane.

MSMIDI96-8-1

MSMIDI 96 Comb 8 sample MC, 1.5mm thick. one marker 
lane.

MSMIDI96-8-1.5

MSMIDI 96 Comb 8 sample MC, 1mm thick.  two marker 
lanes.

MSMIDI96-8-1/2M               

MSMIDI 96 Comb 8 sample MC, 1.5mm thick. two marker 
lanes.

MSMIDI96-8-1.5/2M           

Multi Sub Midi 96 STRETCH Combs

Description

Part Number

 


background image

MSMIDI 96 STRETCH Comb 8 sample MC + 1 Marker, 
1mm thick COMB BLOCK

 MSMIDI96ST-8-1-CB 

MSMIDI 96 STRETCH Comb 8 sample MC + 1 Marker, 
1.5mm thick COMB BLOCK

 MSMIDI96ST-8-1.5-CB 

MSMIDI 96 STRETCH Comb 8 sample MC + 2 Marker, 
1mm thick COMB BLOCK

 MSMIDI96ST-8-1/2M-CB 

MSMIDI 96 STRETCH Comb 8 sample MC + 2 Marker, 
1.5mm thick COMB BLOCK

 MSMIDI96ST-8-1.5/2M-CB 

MSMIDI 96 Comb 8 sample MC, 1mm thick.  one marker 
lane.

MSMIDI96-8-1

MSMIDI 96 Comb 8 sample MC, 1.5mm thick. one marker 
lane.

MSMIDI96-8-1.5

MSMIDI 96 Comb 8 sample MC, 1mm thick.  two marker 
lanes.

MSMIDI96-8-1/2M               

MSMIDI 96 Comb 8 sample MC, 1.5mm thick. two marker 
lanes.

MSMIDI96-8-1.5/2M           

MultiSUB 4 Combs:-

Description

Part Number

12/1 SAMPLE 1.5mm COMBS FOR multiSUB4

MSUB4-12/1-1.5

18/8 SAMPLE 1.5mm COMBS FOR multiSUB4

MSUB4-18/8-1.5

 


background image

Notes

 


background image

Warranty

The Cleaver Scientific Ltd. (CSL) Multi Sub Horizontal Electrophoresis units have a 

warranty against manufacturing and material faults of twelve months from date of customer 

receipt.

If any defects occur during this warranty period, CSL will repair or replace the defective parts 

free of charge. 

This warranty does not cover defects occurring by accident or misuse or defects caused by 

improper operation.

Units where repair or modification has been performed by anyone other than CSL or an 

appointed distributor or representative are no longer under warranty from the time the unit 

was modified.

Units which have accessories or repaired parts not supplied by CSL or it’s associated 

distributors have invalidated warranty.

CSL cannot repair or replace free of charge units where improper solutions or chemicals have 

been used. For a list of these please see the Care and Maintenance subsection.