Full Text Searchable PDF User Manual

Ambulatory Infusion Pump
Body
Guard
TM
323
Clinician Quick Reference
Loading
the
Administration
Set
1. Open pump tubing door
by pulling down on door
latch located on right side
of pump.
2. When loading set,
reservoir bag will be on
left side of pump.
3. Locate pumping segment
of set. It consists of softer
segment of tubing about 4 ½
inches long. At right end of
segment is black set key.
4. Match set key to slot. Set
key fits into slot of same
shape located in pump
tubing channel.
5. Push key fully into slot.
Important:
During loading,
keep segment straight but
do not stretch it.
6. Press rest of segment into
tubing channel.
7. Close pump tubing door
until you hear it click shut.
Important:
While closing door,
keep segment straight but do
not stretch it.
8. Prime set (see illustrated steps on reverse side of sheet):
• Make sure patient is not connected to administration set
• Turn pump power ON and press PRIME/PIGGY
• Press START/OK
• Enter priming volume using pump keypad
• Press START/OK to start priming
• Display screen shows priming progress
• Repeat these steps for additional priming
Pumping Segment
Set Key
Tubing
Channel
CME
America
TM
Set Key
Slot

Ambulatory Infusion Pump
Body
Guard
TM
323
Clinician Quick Reference
Priming
the
Administration
Set
1. Make sure that patient is not connected to administration set.
2. Turn pump power ON and press
3. Press
4. Use pump keypad to enter volume of fluid required to prime set.
For priming volume, check tubing package.
5. Press to start priming procedure.
Display screen shows priming progress.
6. For additional priming volume, repeat steps 1 – 5.
Note:
• To stop priming at any time, press
• If line must be primed, it can only be primed on pump.
• When pump is locked, patient can prime set in all operating modes.
For 24/7 clinical and technical support, call: 877.263.0111
CAUTION: Before use, the user must become thoroughly familiar with the information contained in the device operating instructions.
Note: This Clinician Quick Reference provides a convenient information summary. For complete information on proper use of the BodyGuard
pump and administration set, refer to the device operating instructions supplied with the product. © 2008 by CME America, LLC. All rights
reserved. The following are trademarks of CME America, LLC: BodyGuard, BodyComm, TSyringePump, the CME America logotype with
“swirl” logomark and the tag line “Drug Delivery Designed for the Way You Deliver Care.”
SM-0015 07/2008
PRIME
Disconnect Patient
Press OK to Start
Prime Vol
ml
Press OK to Continue
Prime Vol ___
ml
Press OK to Continue
5
Prime
2 ml
0
5
STOP
NO
START
OK
PRIME
PIGGY
START
OK
Screen Display
CME
America
TM