Ambulatory Infusion Pump

Body

Guard

TM

 323

 

Clinician Quick Reference

 

Loading

 the 

Administration

 Set

1. Open pump tubing door 
by pulling down on door 
latch located on right side 
of pump.

2. When loading set, 
reservoir bag will be on 
left side of pump.

3. Locate pumping segment 
of set. It consists of softer 
segment of tubing about 4 ½ 
inches long.  At right end of 
segment is black set key.  

 

4.  Match set key to slot. Set 
key fits into slot of same 
shape located in pump 
tubing channel.

5.  Push key fully into slot.

Important: 

During loading, 

keep segment straight but 
do not stretch it.

6.  Press rest of segment into 
tubing channel.

7.  Close pump tubing door 
until you hear it click shut.

Important:

 While closing door, 

keep segment straight but do 
not stretch it.

8. Prime set (see illustrated steps on reverse side of sheet):

•  Make sure patient is not connected to administration set
•  Turn pump power ON and press PRIME/PIGGY
•  Press START/OK
•  Enter priming volume using pump keypad
•  Press START/OK to start priming
•  Display screen shows priming progress
•  Repeat these steps for additional priming

Pumping Segment

Set Key

Tubing
Channel

CME

America

TM 

  

Set Key 
Slot

 


background image

                          

 

Ambulatory Infusion Pump

Body

Guard

TM    

323

Clinician Quick Reference 

Priming

 the 

Administration

 Set

1. Make sure that patient is not connected to administration set.

2. Turn pump power ON and press   

3. Press

4. Use pump keypad to enter volume of fluid required to prime set.
    For priming volume, check tubing package.         

5. Press                     to start priming procedure.

    Display screen shows priming progress.                                        

6. For additional priming volume, repeat steps 1 – 5.

Note:

• To stop priming at any time, press

• If line must be primed, it can only be primed on pump.

• When pump is locked, patient can prime set in all operating modes.

   

For 24/7 clinical and technical support, call: 877.263.0111

   
 

 

 

          

CAUTION: Before use, the user must become thoroughly familiar with the information contained in the device operating instructions. 

Note: This Clinician Quick Reference provides a convenient information summary.  For complete information on proper use of the BodyGuard 
pump and administration set, refer to the device operating instructions supplied with the product. © 2008 by CME America, LLC. All rights 
reserved.  The following are trademarks of CME America, LLC: BodyGuard, BodyComm, TSyringePump, the CME America logotype with 
“swirl” logomark and the tag line “Drug Delivery Designed for the Way You Deliver Care.” 

SM-0015  07/2008 

PRIME

Disconnect Patient
Press OK to Start

Prime Vol     

ml

Press OK to Continue

Prime Vol ___

ml

Press OK to Continue

5

Prime

2 ml

0

5

STOP

NO

START

OK

PRIME

PIGGY

START

OK

Screen Display