M U L T I C H A N N E L

5648 Copley Drive
San Diego, CA 92111
Phone 1-800-456-7741

 


background image

M U L T I C H A N N E L

 


background image

M U L T I C H A N N E L

6

 


background image

1

2

M U L T I C H A N N E L

clockwise in relation to the handle, (Fig. 2), it is
possible  to  disable  the  ejector’s  operation 
in  order  to  avoid  accidental  use,  or  to  change
its  position  in  such  a  way  as  to  make  filling 
of microplates more convenient.

The  specifications  for  accuracy  and  precision
given  in  the  following  table  are  obtained  using

CLP

tips.  These  figures  are  only  guaranteed

when 

CLP

tips are used.

channel

channel

The  specifications  for  the  volumetric  perfor-
mance  are  derived  from  extensive  gravimetric
tests as prescribed by the Deutsches Institut für  
Normung e.V. To verify the performance of your
pipette request document DIN 12650.

Condition of measurments: distilled water at tem-
perature 20±1°C 

The pipette design enables the user to perform
the recalibration process according to the rules
presented in section 5. 

CONTENTS

1 - INTRODUCTION

2 - VOLUME SETTING

3 - METHOD OF PIPETTING

4 - RECOMMENDATIONS

5 - RECALIBRATION

6 - MAINTENANCE

7 - PIPETTE KIT AND ACCESSORIES

8 - SPARE PARTS

1 - INTRODUCTION

series  of  adjustable  multichannel 

pipettes  have  been  designed  for  the  filling  of  la-
boratory microplates. The pipettes enable precise
and simultaneous delivery of 8 or 12 preset-vol-
ume  doses  of  liquid. 

pipettes  are

produced  in  four  ranges  of  volumes:  1-10  µl, 
5-50 µl, 20-200 µl, 50-300 µl.

pipettes  are  equipped  with  a  digital

readout  counter,  built  into  the  pipette’s  handle,
for measuring doses, (Fig. 1F). Volume settings
are continously adjustable and are controlled by
means of the rotating calibration knob, (Fig 1B).
The  volume  range  of  the  pipette  is  shown  on
the pipetting pushbutton, (Fig. 1A).

pipettes  should  be  used  with  poly-

propylene disposable tips, to ensure safety and
accuracy,  (Fig.  1C).  The  shafts  without 
O-rings  ensure  the  compatibility  of  the  pipettes
with a broad assortment of tips, and the special
ejector  shape  reduces  the  force  required  for 
their  ejection.  In  order  to  protect  the  user
against  inadvertent  contact  with  used  tips,  the
pipettes  are  equipped  with  tip  ejectors, 
(Fig.  1D).  By  turning  the  multichannel  module   

8.10

Min 1

±8.0

≤ 

8.0

Max 10

±2.0

≤ 

2.0

8.50

Min 5

±4.0

≤ 

3.0

Max 50

±1.6

≤ 

1.0

8.200

Min 20

±2.0

≤ 

1.5

Max 200

±1.0

≤ 

0.8

8.300

Min 50

±1.6

≤ 

1.5

Max 300

±1.0

≤ 

0.8

Cat. No.

Volume 

Inaccuracy 

Imprecision 

[µl]

[%]

[%]

Cat. No.

Volume 

Inaccuracy 

Imprecision 

[µl]

[%]

[%]

SPECIFICATIONS

12.50

Min 5

±4.0

≤ 

3.0

Max 50

±1.6

≤ 

1.0

12.200

Min 20

±2.0

≤ 

1.5

Max 200

±1.0

≤ 

0.8

12.300

Min 50

±1.6

≤ 

1.5

Max 300

±1.0

≤ 

0.8

SPECIFICATIONS

 


background image

3

4

M U L T I C H A N N E L

Next, still holding the pipette vertically, the ope-
rator should press the pipetting button until the
first resistance point is felt, and immerse the tips
in  the  liquid  to  the  depth  of  2-4  mm,  (Fig.  3B).
Releasing  the  pipetting  button  with  a  slow  and
uniform movement during 2-3 seconds, the liquid
should  be  drawn  into  the  tips,  (Fig.  3B)  which
should then be lifted above the liquid’s surface.
The  pipette  should  next  be  positioned  at  an
angle of 10-45 degrees in relation to inner walls
of  destination  vessels  and  tips  should 
be  emptied  by  pressing  the  pipetting  button 
slowly  until  the  first  resistance  point  is  felt, 
(Fig. 3C). After waiting one second, the pipetting
button  should  be  pressed  to  the  second  resis-
tance  point  in  order  to  expel  remaining  liquid, 
(Fig. 3D). Then tips should be lifted out from the
vessels  while  maintaining  contact  between 
the  ends  of  the  tips  and  the  inner  walls  of  the
vessel  until  the  pipetting  button  has  been  rela-
sed, (Fig. 3E). Finally, pressing the ejector but-
ton,  the  operator  should  separate  the  tips  from
multichannel module’s cones, (Fig. 3F).

4 - RECOMMENDATIONS

To  achieve  maximum  safety,  precision  and  re-
liability,  the  following  principles  should  be  ob-
served:

• do  not  draw  liquids  without  tips  fitted  on  the

pipette cones,

• do not lay down the pipette with tips filled,

• do  not  draw  volumes  of  liquid  exceeding  the

pipette’s range,

• check if the tips are fitted properly,

• during operation, the pipette must be held ver-

tically,  tips  should  be  immersed  in  liquid 
to the depth of 2 to 4 milimeters and the pipet-
ting button should be depressed and released
slowly and evenly,

• new  tips  must  be  re-wetted  prior  to  pipetting,

by drawing and expelling the liquid to be me-
asured. This is especially important when wor-
king  with  liquids  of  viscosities  and  densities
different  from  that  of  water  and  in  temperatu-
res different from ambient temperature, 

2 - SETTING THE VOLUME

In order to set the volume of the dose to be deli-
vered, the calibration knob with counter is used.
The volume shown by the counter is represented
by  three  digits,  which  should  be  read  from  top
to bottom. Typical meter readings are shown in
the following table:

To  achieve  accurate  delivery  of  the  required 
volume, the operator should turn the calibration
knob  in  the  direction  diminishing  counter  rea-
dings. If the volume being currently set is higher
than the previous one, the operator should turn
the  calibration  knob  1/3  of  a  turn  beyond  the
required  value,  and  then  slowly  rotate  it  back-
wards, observing diminishing readings, until the
required value is achieved.

3 - METHOD OF PIPETTING

Fit  the  tips  onto  the  cones  of  the  multichannel
module. When you put the tips on the shafts the
pipette should be pressed against the tips in the
box.  The  liquid  aspirated  into  the  tips  should
not  flow  out  by  gravity  from  properly  fixed  tips.
While  holding  the  pipette  in  a  vertical  position,
tips should be immersed in the liquid to a depth
of 2-4 mm, and flushed once by drawing a dose
of  liquid  and  dispensing  it  out  with  slow  and
steady movement.

Cat.No.

Counter 

readings 

Set

volume 

Basic 

degree

0
3
5

2
5
0

0
8
5

0
6
5

8.10

8.50

12.50

8.200

12.200

8.300

12.300

3.5 µl

6.5 µl

85.0 µl

250.0 µl

0.02 µl

0.10 µl

0.20 µl

1.0 µl

 


background image

M U L T I C H A N N E L

6

The  recalibration  can  be  performed  within  one
full turn of the key to the right or to the left only.

Recalibration conditions:

• Ambient  temperature  and  the  temperature  of

the  pipette,  tips  and  liquid  should  be  within
the range 20-25°C and stabilized during weigh-
ing within ±0.5°C

• Measurements  should  be  conducted  using

distilled water

• Balance  sensitivity  should  be  suitable  for  the

volume to be controlled 

Recalibration procedure:

• Set the dose volume depending on the pipette

volume according to the following table: 

• Perform  three  aspiration  series  (each  series

should  include  the  aspirations  from  all  chan-
nels), weigh each time and calculate the aver-
age value of the aspirations.

• Calculate average aspirated volume in µl mul-

tiplying  the  average  aspiration  amount  [mg]
by  the  distilled  water  density  coefficient
[µl/mg],  which  depends  on  temperature  and
pressure according to the following table:

5

• when pipeting liquids which tend to wet walls

of tips, such as serums, proteins, organic sol-
vents,  measuring  must  be  performed  much
slower than with other liquids,

• tips  must  be  replaced  with  new  ones,  when

changing from one liquid to another, or if drops
of liquid remain inside the tips,

• after  work,  the  pipette  should  be  stored  verti-

cally in a holder with tips removed.

5 - RECALIBRATION

CLP

pipettes are calibrated by gravimetric met-

hod,  using 

CLP

tips  and  distilled  water,  at  the

temperature 20±1°C, according to DIN 12650.

If  during  pipette  operation  you  find  that  the
accuracy  error  (the  difference  between  the  real
aspirated  volume  and  the  preset  volume)
exceeds the permissible value given in the table
in section 1, the pipette recalibration procedure
should be carried out. 

Before  starting  the  recalibration  it  is  necessary
to  check  whether  the  following  requirements
have been fulfilled during error determination:

• the ambient temperature, and the temperature

of the pipette, tips and water was identical

• the  density  of  the  liquid  used  is  close  to  that 

of distilled water

• the  balance  with  appropriate  sensitivity  has

been used

• mg/µl  conversion  factor  has  been  taken  into

account

• the  requirements  given  in  sections  3  and  4

have been fulfilled

If the above conditions are satisfied and the accu-
racy error for selected volume given in section 1
exceeds  the  permissible  value,  the  pipette
recalibration procedure should be carried out. 

Volume checked [µl]

0.1 - 10

10 - 100

≥ 

100

Balance sensitivity [mg]

≤ 

0.001

≤ 

0.01

≤ 

0.1

Temperature [°C]

Pressure [hPa]

960

1013

1067

20

1.0028

1.0029

1.0029

21

1.0031

1.0032

1.0032

22

1.0032

1.0033

1.0033

23

1.0035

1.0035

1.0036

24

1.0037

1.0038

1.0038

25

1.0039

1.0040

1.0041

1 - 10

5 - 50

20 - 200

50 - 300

1

5

20

50

0.92 - 1.08

4.8 - 5.2

19.6 - 20.4

49.2 - 50.8

0.33

1.67

6.30

10.00

Range 

of the pipette
volumes [µl] 

Preset 

volume 

[µl] 

Permissible 

volumes 

[µl] 

Volume change 

V for

full turn of the calibra-

tion key [µl]

(24 increments)

 


background image

Accessories

8 - SPARE PARTS

Please see Fig.1 and Fig. 4:

1. Pushbutton 
2. Ejector pushbutton 
3. Calibration knob 
4. Ejector 
5. Piston assembly 
6. Calibration key 

These  items  can  be  ordered  from  Continental
Laboratory  Products.  (Please  specify  pipette
type and part name when ordering).

Before  returning  a  pipette  to  us  for  service,
please ensure that the pipette is completely free
of any chemical, biological or radioactive conta-
mination please provide information on the kind
of liquids that have been measured.

Warning:  The  replacement  of  the  plunger
requires  conducting  of  calibration  proce-
dure according to section 5.

7

8

M U L T I C H A N N E L

If  the  average  aspirated  volume  exceeds  the
permissible value, the following should be done: 

• Remove the pipetting pushbutton, (Fig. 4A),

• Holding  the  volume  setting  knob  to  protect 

it  against  rotation,  insert  the  calibration  key
into the cuts of the calibration screw, (Fig. 4B),

• Turn  the  key  clockwise  to  reduce  the  aspirat-

ed  volume,  or  counter-clockwise  to  increase
the volume. One full turn of the calibration key
changes  the  pipette  aspiration  volume  by  the
amount given in the table, (Fig. 4C),

• Take  out  the  key  and  fix  the  pipetting  push-

button, (Fig. 4D). 

Determine  the  average  aspirated  volume.  The
average  volume  should  be  within  the  permissi-
ble  range  given  in  the  table.  If  the  volume
exceeds the values stated, the recalibration pro-
cedure should be repeated.

6 - MAINTENANCE

Cones should be kept clean using ethyl alcohol.

7 - PIPETTE KIT AND ACCESSORIES

Pipette kit

The pipettes are delivered in the kits including: 

• Pipette

• Instruction manual 

• Calibration key

• Pipette shelf clip

• Identification labels

The pipette shelf clip diagram is shown in Fig. 5. 

B/02/2001

Pipettor

Tip Choices

2342 10 µl, XL, 10 racks of 96 
2040 10µl, 10 racks of 96 
BT10 10µl filter tip, sterile, DNase, RNase and pyrogen free, 10 racks of 96
BT10XL 10µl filter tip, extra long, sterile, DNase, RNase and pyrogen free, 
10 racks of 96

2102.C 200µl tip, graduated, slimline, 10 racks of 96 
BT100 100µ filter tip, sterile, DNase, RNase and pyrogen free, 10 racks of 96

2092 300µl tip, graduated, slimline, 10 racks of 96 
BT200 200µl filter tip, sterile, DNase, RNase and pyrogen free, 
10 racks of 96 (max volume is 200µl)

8.10

8

12

8 and  12

8.50
12.50

8.200
12.200
8.300
12.300

Pipettor

Stand Choices

9575 Stand for 8 channel

9576 Stand for 12 channel

9577 Carousel Stand for single and multichannels