background image

AcoustiCal User Manual

3M Occupational Health & Environmental Safety Division

3M

TM 

AcoustiCal AC-300 Calibrator 

Precise,Simplistic

Calibration

 


background image

       

        

 

 

 

 

 

 

                                     

                                                                                                       AcoustiCal User Manual

 

 

 

 

 

 

 

 

3M

TM

 

AcoustiCal AC-300 Calibrator

 

Dangers, warnings and cautions  

 

Danger! 

Failure to observe the following procedures may result in serious personal 

injury 

 

Not for use in explosive or hazardous locations.  This product is not intrinsically 

safe. 

 

Never place the AcoustiCal against your ear, harmful sound levels may be 

present. 

 

     

Warning! 

 

Failure to observe the following procedures could damage the instrument

 

 

Read the manual before operation.   

 

Do not store in temperatures exceeding 65°C (149°F). 

 

Do not immerse in liquids. 

 

Condensation may damage your instrument. 

 

Substitution of components may impair the accuracy of the instrument.  

Repair should be done by authorized service personnel only. 

 

    

Caution!

 

     

 General  

 

Best results are achieved when the calibrator and sound measuring 

instrument are given time to acclimate to the current environment. 

 

For proper calibration the ambient sound level should be less than 88 dBA. 

 

A non-condensing environment is required for proper measurements. 

 

An annual laboratory calibration is recommended to ensure accurate 

calibration results.  Between laboratory calibrations, or when there is an 

accuracy question, the AcoustiCal may be field checked by comparing the 

output with another AcoustiCal, or reference meter with similar 

specifications/accuracy. 

 

 

Intended Use:

   

The AcoustiCal is intended to calibrate precision Class/Type 1 and Class/Type 2 

instruments with a microphone input.  Consult your company’s safety professional 

for local standards, or call 3M at 1-800-243-4630. 

 

 

 

 

 


background image

 i         Table of Contents 

        

 

 

 

                  

 

 

AcoustiCal User Manual

 

Table of Contents

 

AcoustiCal .................................................................................................................. 1 

Models .................................................................................................................. 1 

AcoustiCal display indicators ................................................................................ 1 

Powering and settings .......................................................................................... 1 

Battery ........................................................................................................................ 2 

Low Battery ........................................................................................................... 2 

Changing the battery ............................................................................................ 2 

Performing a calibration ............................................................................................. 3 

Factory service on the calibrator ................................................................................ 4 

Supported microphones and adapters ....................................................................... 4 

Specifications ............................................................................................................. 5 

Conformance standards ....................................................................................... 5 

Physical characteristics ........................................................................................ 5 

Outputs ................................................................................................................. 5 

Reference environmental conditions .................................................................... 5 

Operating environmental conditions ..................................................................... 5 

Stabilization .......................................................................................................... 5 

Principal ................................................................................................................ 5 

Distortion............................................................................................................... 5 

Storage environmental conditions ........................................................................ 6 

Coupler volume coefficient ................................................................................... 6 

Electromagnetic emissions and susceptibility ...................................................... 6 

Background acoustic level .................................................................................... 6 

Power .................................................................................................................... 6 

3M Quest Service ...................................................................................................... 7 

Warranty ..................................................................................................................... 8 

  

 

Figures 

Figure 1:  Operating the calibrator ............................................................................. 1

 

Figure 2:  Re-insert battery door ................................................................................ 2

 

Figure 3:  Closing battery door ................................................................................... 2

 

Figure 4:  Example of performing a calibration .......................................................... 3

 

 


background image

     

/ AcoustiCal

        

 

 

 

                  

 

 

AcoustiCal User Manual

 

 

 AcoustiCal  

Models 

The 3M

TM 

AcoustiCal AC-300 is an acoustic calibrator for 

calibrating precision Class/Type 1 as well as general 

purpose Class/Type 2 sound level meters and noise 

dosimeters. 

 

The AC-300 produces two signals either 250 Hz or 1000 

Hz at a sound pressure level of 114.0 dB.   

 

Note:

  The sound pressure level and combinations of frequency outputs 

conform to the requirements of the IEC standard 60942:  2003-1 sound 

calibrators using all microphones listed (see “Supported microphones and 

adapters”).   

AcoustiCal display indicators 

One of the following indicators may appear in the upper-left corner of the display 

when the calibrator is powered on.    

 

 

 

Low Battery indicator:

  

If the low battery indicator is on,

 

the battery should be 

replaced.

 

 

 

 

Fault indicator:

  

If the fault indicator is on, the level is incorrect.  This may be 

due to the microphone not being inserted properly. 

Powering and settings  

 

1.

 

Press the          key to power on the calibrator.  

 

 

 

 

 

 

 

 

 

 

          Figure 1:  Operating the calibrator 

 

2.

 

Press the          key to select 1000 Hz or 250 Hz at 114.0 dB.   

 

  

 

3.

 

Press and hold the          key until the calibrator powers off.  

 

Note:

  The AcoustiCal will automatically power off after 4 minutes from the 

last key press.  

 

 

 

Press 

      

to power on.  Press again to change 

frequency.  Press and hold to power off. 

Display 

 


background image

2     

  

Battery 

 

 

                    

                    

AcoustiCal User Manual

 

 

Battery  

A 9-volt battery is used to supply power to the calibrator.   

Low Battery 

At a low battery condition, the low battery indicator will appear.  If the battery 

level becomes too low, the calibrator will power off.     

Changing the battery   

The battery door is located on the top of the calibrator.  You will need a 

Phillips screwdriver to change the battery.   

 

1.

 

Loosen the screw on the end of the calibrator with a screwdriver. 

 

2.

 

Lift the battery door off of the calibrator and remove the battery. 

 

3.

 

Replace the battery it according to the polarity diagram inside the unit. 

 

 

4.

 

To re-insert the battery door, place the battery door over the battery 

compartment ensuring that the foam pad is positioned over the battery 

(see Figure 2). 

 

 

 

 

 

 

 

 

 

 

 

                    Figure 2:  Re-inserting battery door 

5.

 

Using your index finger, push the battery door into the slot of the housing.   

(See Figure 3).   

 

 

 

 

 

 

 

 

             Figure 3:  Closing battery door 

 

6.

 

Press the battery door into position and fasten the screw.  

 

 

 

Battery door 

Foam pad 

  

Screw  

 

Battery door 

 

Slot 

 

 


background image

3     

  

Performing a calibration 

 

 

                    

                    

AcoustiCal User Manual

 

 

 Performing a calibration 

A microphone is sensitive to temperature and pressure changes.  Calibrating 

before taking measurements assures that your level measurements are 

accurate for the current environment; and performing a calibration after use 

allows you to verify the integrity of both the calibrator and the sound level 

meter.  

 

Note: 

 

Please refer to the instrument’s user manual for details on calibrating.

 

 

1.

 

Ensure the ambient noise level is less than 88 dBA. 

 

2.

 

Slowly insert the appropriate adapter (if needed) and microphone into the 

calibrator cavity ensuring the microphone is fully seated (or inserted).   

 

3.

 

Power on the calibrator and allow at least 10 seconds to stabilize. 

 

4.

 

Set the appropriate frequency on the AcoustiCal by selecting 250 Hz or 

1000 Hz via the          key. 

 

5.

 

Perform the calibration with the sound level meter or dosimeter. 

 

6.

 

Power off when complete by pressing and holding the         key.  

 

Note:

  The microphone orientation is displayed below for a noise 

dosimeter and a sound level meter.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 4:  Examples of performing a calibration 

 

 


background image

4     

  

Factory service on the calibrator 

 

 

                    

                    

AcoustiCal User Manual

 

 

 

Factory service on the calibrator 

3M Quest Technologies recommends checking the performance and precision of your 

calibrator on an annual basis by sending it to a factory-authorized service center.   

 

The AcoustiCal is calibrated at Quest Technologies laboratory, a 3M company, with a 

Type L standard microphone and with special instrumentation traceable to NIST.  The 

calibrators are very stable; however, they are used to calibrate other equipment, and 

should be checked periodically against laboratory standards.     

Supported microphones and adapters 

The AcoustiCal supports the following microphones used with the listed microphones, 

calibrator adapters.  All microphones are supplied by Quest Technologies, a 3M 

company.  Listed below are the 3M ID/Quest part numbers (Quest P/N). 

 

 

 

 

Approved Microphones 

Microphone  

3M ID/Quest P/N 

Calibration adapter  

3M ID/Quest P/N 

QE4110 microphone, 1/4" free-field 

condenser, Class/Type 1  

 

Uses QE1160 ¼” to ½” preamp 

adapter, Quest P/N:  059-488 

 

 3M ID:  70-0715-8299-6 

 

 Quest P/N:  059-413 

 

 

 

 3M ID:  70-0715-8113-9 

 

 Quest P/N:  056-988  

 

QE4130 ½” free-field condenser,  

Class/Type 1 

 

 3M ID:  70-0715-8298-8 

 

 Quest P/N:  058-659 

 

3M ID:  70-0715-8115-4 

 

 Quest P/N:  056-990 

BK4936 ½” free-field electret, 

Class/Type 1  

 

3M ID:  70-0715-8300-2 

 

Quest P/N:  059-523 

 

3M ID:  70-0715-8115-4  

 

Quest P/N:  056-990  

QE7052 microphone, 1/2” free-field 

electret Class/Type 2  

 

3M ID:  70-0715-8290-5 

 

Quest P/N:  056-317 

 

3M ID: 70-0715-8115-4 

 

Quest P/N:  056-990 

QE4150 microphone, ½” free-field 

condenser, Class/Type 1 

 

3M ID:  70-0715-8296-2 

 

Quest P/N:  058-489 

 

3M ID:  70-0715-8290-5 

 

Quest P/N:  056-990 

QE4170 microphone, 1" pressure 

condenser, Class/Type 1  

 

Uses ½” to 1” preamp adapter,  

Quest P/N:  058-376/3M ID:  70-

0715-8112-1 

 

3M ID:  70-0715-8295-4 

 

Quest P/N:  058-488 

N/A 

 

 

3M

TM

 Edge Microphone, ¼” free field 

electret Class/Type 2  

 

3M ID:  70-0715-8304-4 

 

Quest P/N:  057-309 

 

3M ID:  70-0715-8106-3 

 

Quest P/N:  053-753 

Pendant microphone, Class/Type 2   

 

3M ID:  70-0715-8304-4 

 

Quest P/N:  053-867 

 

3M ID:  70-0715-8107-1 

 

Quest P/N:  053-884 

Boom microphone, Class/Type 2 

 

Part number:  053-887 

 

3M ID:  70-0715-8289-7 

 

Quest P/N:  053-887 

 

3M ID:  70-0715-8107-1 

 

Quest P/N:  053-884 

Microphone ½” SD-200, non-

removable Class/Type 2  

 

3M ID 70-0715-6546-2 

 

3M ID 70-0715-8115-4 

 

Quest P/N:  56-990 

Microphone used for certification of 

AC-300 per IEC 61094:  B&K 4192, 

½” Pressure Condenser, Class/Type 1 

supplied by Bruel and Kjaer 

N/A 

 

3M ID 70-0715-8115-4 

 

Quest P/N:  56-990 

 


background image

5     

  

Specifications 

 

 

                    

                    

AcoustiCal User Manual

 

 

Specifications  

Conformance standards  

ANSI S1.40-1983 (R2006):  Specifications and Verification Procedures for Sound Calibrators 

IEC 60942 (2003-01):  Electroacoustics, Sound Calibrators 

CE, RoHS, WEEE, C-Tick

 

 

Physical characteristics  

Size 

(Length x Width x Thickness) 

Weight 

10 cm  x 5.8 cm x 4.6 cm  

3.9” x 2.3” x 1.8” 

0.17 kg (6oz) 

Outputs 

Frequency

 

251.2 Hz and 1000.0 Hz, +/- 0.5%

   

with expanded 

uncertainty in environmental conditions

  

Sound pressure level  

114.0 dB re 20

μ

PA (1 Pa = 1 N/m

2

Sound pressure level accuracy 

+/- 0.4 dB with expanded uncertainty in the reference 

environment (initial difference) 

+/- 0.4 dB with expanded uncertainty in the operating 

environmental conditions, added to initial difference 

Variation in output due to power 

Less than +/- 0.1 dB 

Reference environmental conditions 

Temperature range 

20°C to 26°C (68 °F to 79°F) 

Pressure range 

97kPa to 105kPa (29 to 31 inches of Hg) 

 

Relative humidity range 

40% to 65% humidity 

Large temperature settling time  

about 1 min/ °C (approximately 30 sec/°F) 

Operating environmental conditions 

Temperature range

 

-10°C to 50 °C (14°F to 122°F) 

 

Pressure range 

65kPa to 108kPa (19 to 32 inches of Hg) 

Approximate 3500m to -600m elevation relative to sea 

level (11500ft to -2000ft). 

 

Relative humidity range 

 

10% to 90%, non-condensing 

Stabilization 

 

Stabilization time after powering the calibrator on and coupling to a microphone is 10 

seconds.  

 

Principal 

 

The principal sound pressure level is 114 dB. 

 

The principal frequency is 1 kHz 

Distortion 

Less than 3% (all harmonics below 30.5 dB) within temperature and humidity at either 

operating frequencies, unless “overload” is showing. 

 

 

 

 


background image

6     

  

Specifications 

 

 

                    

                    

AcoustiCal User Manual

 

 

Storage environmental conditions 

Storage temperature range 

-20°C to 65°C (-4 °F to 149°F) with battery 

removed. 

Storage humidity range 

0% to 90% non-condensing 

 

Coupler volume coefficient 

No effect until overload occurs. 

Electromagnetic emissions and susceptibility 

Emission better than CISPR 22, Class/Type B susceptibility  better than 10 V/m, 150 kHz to 

1GHz.  50/60 Hz magnetic immunity.  No effect. 80 A/m  

 

Background acoustic level 

Requires greater than 25 dB signal-to-noise ratio, for optimum performance.  If external noise 

is too high, it may reduce output level.  

 

Power  

Battery type 

1 x  9v  alkaline battery.  

(Quest 058-176, ANSI/NEDA 1604A, IEC 6LR61) 

Battery life 

 

The battery life will be approximately 18 hours (@ 1kHz) at reference 

environmental conditions, mounted on a microphone. 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

7     

  

3M Quest Service 

 

 

                    

                    

AcoustiCal User Manual

 

 

 

3M Quest Service  

Should your Quest Technologies, a 3M company, equipment need to be 

returned for repair or for recalibration, please contact the service department 

at the following numbers or access the online form via the website. 

 

 

In the United States Contact 

Quest Technologies, a 3M company 

Service Department 

1060 Corporate Center Drive, Oconomowoc, WI 53066 

Phone:

  262.567.9157  

Toll Free:

 800-245-0779  

Fax:

  262/567-4047 

Email:

  quest.mail@mmm.com   

Website:

  http://www.questtechnologies.com/Service.aspx 

 

 

In Canada Contact 

Pylon Atlantic  

A Division of Pylon Electronics, Inc.  

31 Trider Crescent, Dartmouth, NS Canada B3B 1V6  

Telephone:

 902-468-3344 ext 235   

Fax:

 902-468-1203  

Email:

  tbourgoin@pylonelectronics.com 

 

 

 

In Europe Contact 

3M Service Centre 

3M Health Care Limited 

11 Seymour Court 

Manor Park, Runcorn, Cheshire, WA7 1SY, ENGLAND 

Office: 44 01928 532595  Triminet: 8227 2595 

Email

:  

Rnorton@mmm.com

  

(

Richard Norton, Service Centre Manager)

 

 

 

 

For Service Outside the United States, Canada and Europe 

Contact your locally authorized  Quest, a 3M company, Distributor. For 

assistance, 

email customer service:   quest.mail@mmm.com  

 


background image

       

Warranty 

      

 

 

                  

 

 

AcoustiCal User Manual

 

 

Warranty

 

Quest Technologies, a 3M company, warrants our instruments to be free 

from defects in materials and workmanship for one year under normal 

conditions of use and service.  For United States customers, we will 

replace or repair (our option) defective instruments at no charge, excluding 

batteries, abuse, misuse, alterations, physical damage, or instruments 

previously repaired by other than Quest Technologies, a 3M company.  

Microphones, sensors, printers, and chart recorders may have shorter or 

longer warranty periods.  This warranty states our total obligation in place 

of any other warranties expresses or implied.  Our warranty does not 

include any liability or obligation directly resulting from any defective 

instrument or product or any associated damages, injuries, or property 

loss, including loss of use or measurement data.   

 

For warranty outside the United States, a minimum of one year warranty, 

applies subject to the same limitation and exceptions as above with service 

provided or arranged through the authorized Quest distributor or our Quest 

European Service Laboratory.  Foreign purchases should contact the local 

Quest authorized sales agent for details. 

 

 


background image

Occupational Health &

Environmental Safety Division

Quest Technologies, a 3M Company

ISO 9001 Registered Company

ISO 17025 Accredited Calibration Lab

1060 Corporate Center Drive

Oconomowoc, WI 53066

Customer Service:  262-567-9157

Toll Free:  800-245-0779

Visit:  www.3M.com/detection

          
Please recycle. Printed in USA.

© 2012 3M  

All rights reserved.

053-591, Rev. B; 7/12

    

About Us

 

3M Quest Technologies is a world class manufacturer of rugged, reliable instrumentation and software 

systems that help monitor and evaluate occupational and environmental health and safety hazards, including 

noise dosimetry, heat stress, indoor air quality and select toxic/combustible gases. The 3M Detection brand 

of instrumentation is used by safety and industrial hygiene professionals to help comply with applicable 

occupational standards and regulations.

   

About 3M and 3M’s Occupational Health and Environmental Safety Division

3M Detection Solutions is part of 3M OH&ES Division, a global leader in respiratory, 

hearing, eye, head and fall protection, visibility and protective clothing, and noise 

detection and sound measurement products. 3M captures the spark of new ideas 

and transforms them into thousands of ingenious products. Our culture of creative 

collaboration inspires a never-ending stream of powerful technologies that make life 

better. 3M is the innovation company that never stops inventing.  With $30 billion in 

sales, 3M employs about 84,000 people worldwide and has operations in more than 65 

countries.