TL-510 Optical Power Meter 

User Manual 

                                  V1.0 

 

 

 

 

 
 
 
 
 
 
 

T&L TECHNOLOGY CO., LIMITED 

 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  2

CONTENTS 

1. Model Introductions

...........................1

 

2. Features

..............................................1

 

3. Technical Specifications

....................3

 

4. Layout

..................................................4

 

5. Function Keys.......................................5

 

6. Operation..............................................7

 

6.1 Turn on/off ...................................................7 
6.2 Absolute power measurement.....................7 
6.3 Relative power measurement......................7 
6.4 User self-calibration function .......................8 
6.5 Auto off function...........................................9 
6.6 Replace battery ...........................................9 

7. Maintenance. ……………………………9

 

8. Trouble-Shootings……………………10

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  1

1. Model Introductions 

TL-510A: 

Measurement range: -70~+10dBm

calibrated 

wavelength

850nm

1300nm

1310nm

1490nm

1550nm

1625nm 

TL-510B:  

Measurement range: -50~+26dBm

calibrated 

wavelength

850nm

1300nm

1310nm

1490nm

1550nm

1625nm 

2. Features   

—

 High 

measurement 

accuracy and display resolution 

—

  Quick response and measurement: It can display 

fast-tracking and real-time measurement of optical 
power when output power changes 

—

  Wide measurement range: it can provide 80dB of 

optical power measurement range 

—

  Six calibrated wavelength: 850nm

1300nm

1310nm

1490nm

1550nm and 1625nm(other 

wavelength can be provided according to 
customer’s request) 

—

  Provide absolute optical power measurement and 

relative power measurement.   

—

 Real-time monitoring and displaying the battery 

level 
Warning notice is available when the battery is low 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  2

—

 Auto-off 

function 

—

  Backlight LCD display 

—

  AAA batteries which is more convenient to carry 

—

  User self-calibration function 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  3

3. Technical Specifications 

Measurement range dBm 

-70

+10 dBm 

-50

+26 dBm 

Wavelength range

nm

 

800

1650 

Calibrated wavelength 

850nm

1300nm

1310nm

1490nm

1550nm

1625nm

other wavelength 

will be on request

 

Detector InGaAs 

Accuracy 

<±3% 

-10dBm

22

℃)

 

Resolution 

Linearity

0.1%

Non-linearity

0.01dBm 

Working temperature 

-10

℃~

+50

 

Storage temperature 

-20

℃~

+70

 

Relative humidity 

90%

(+

30

℃),

 

connector 

Movable FC/PC SC/PC

( ST as 

Optional) 

Power supply 

3pcs AAA batteries 

DC power supply 

Optional accessory which require 

special order 

Support wide range power input work 

(AC 90-240) 

Working hours 

>120 hours

without backlight

 

Weight(g) 160g 

Dimension(mm) 130×69×29 

 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  4

4. Layout   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1. Wavelength 

Display the wavelength, such as  

850nm

1300nm

1310nm

1490nm

1550nm and 

1625nm. 

2. Power value

 

—

  Displays the current output power under 

absolute 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  5

test mode 

—

  Displays the difference between the current output 

power and reference power under 

relative test 

mode 

3. Unit

mW

uW

nW

dBm

dB. 

4. Battery level : I

ndicate the current battery level. 

When the icon is blank, it means the power is too low, 
please change the batteries. 

5.

 

Auto-off 

Activate the 10-minute auto off function 

6. ON/OFF 

7. Wavelength select 

8. REF Relative power

Press REF for 2 seconds to 

start relative test 

9. DBm/w: 

Absolute power 

5. Function Keys 

1) ON/OFF

press ON/OFF to turn it on.    And press it 

for 3 seconds to turn it off. Under power-on mode, 
press this key shortly to activate or deactivate the 
10-minute auto off function. The default setting is 
auto-off function ON when start the meter. 
Operators can press ON/OFF shortly to close the 
auto-off function. 

2) dBm/W

press dBm/W to enter absolute 

measurement mode. Press it repeatly to switch the 
display power unit between W and dBm 

3) 

λ

Switch the wavelength from 6 calibrated 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  6

wavelengths, the current choice of wavelength will 
appears on the LCD 

4)  

REF

press REF key to enter relative 

measurement mode and it has 3 working modes: 

A) Directly enter the relative measurement: in 
absolute measurement mode, press REF to enter 
the relative measurement. Now the Ref value shows 
the previous saved power values. 
B) Save the reference value 

when enter the 

relative measurement mode, press REF for 2 
seconds to save the current absolute power value as 
REF value and show the relative power value in dB 
unit 
C) Enter the relative measurement mode and save 
current reference value: in the absolute 
measurement mode, press REF for 2 seconds to 
enter the REF measurement mode and save the 
current value as REF value which will appears in Ref 
position. Now the value in dB unit on the LCD and 
the vale means the difference value between the 
measurement value and REF value, meanwhile, the 
REF value will displays in the REF position. 

5)  Backlit is on when turning on the unit. It will turn off 

one minute later. Operators can press any key to 
make backlit for another one minute. 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  7

6. Operation 

6.1 Turn on/off 

—

  Press ON/OFF shortly to turn on the unit. Press 

ON/OFF for a few seconds to turn it off and press 

this key slightly to activate or deactivate the auto-off 
function. 

6.2 Absolute power measurement 

—

  Insert optical signal (Insert the pigtail properly), turn 

on the meter 

—

  Choose the wavelength via 

λ

 key. If the tested 

wavelength is not exactly same to the one in the 
power meter, then choose a close one from the 
power meter. For example, 1300nm wavelength is 
near to 1310nm, we can choose 1310nm to 
substitue 1300nm. 

—

  Press dBm/W to choose the display unit. The 

default wavelength is 1310nm and working mode is 
dBm. 

—

  After the above steps, the LCD will show the test 

power value. 

6.3 Relative power measurement

 

—

  Set the wavelength 

—

  Press REF for 2 seconds to enter the relative 

measurement mode. now LCD screen display 
00.00dB 

—

 Measurement: 

Insert 

the pigtail need to be tested 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  8

and introduce another measurement light, now the 
current value shown in the LCD is the difference 
value between the tested signal and reference 
value. 

—

  The reference value shown in the Ref position 

—

  Press dBm/W to review the current absolute power 

value. Press REF to review the current reference 
value. 

6.4 User self-calibration function 

   
 
 
 
 
 
 
 
During the operation, when operators need to calibrate 
the meter to be accordance with other meters. So our 
product offers operators self-calibration function, the 
following steps will tell you how to do:   

—

  Take out the batteries and pull up the switch 1 to 

“ON” position. 

—

  Put the battery back to the battery compartment and 

turn on the unit to choose the wavelength for 
calibration. 

ON

Work p osition

1 2

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

  9

—

  Now connect it with a pigtail which we already know 

its power and press REF or dBm/W to increase or 
decrease the power value by 0.05d every time. Until 
the LCD display the pigtail’s power (Or the 
difference gap less than 0.3db), press ON/OFF to 
save the   

—

  Repeat above operation for other wavelength 

—

  After recalibrate each wavelength, please pull down 

the switch 1 (come back to the position, the 
opposition of the “ON” )and turn off the meter 

—

  End the user self calibration operation 

6.5 Auto off function 

The auto off is on when turning on the meter. Press 

ON/OFF to activate or deactivate this function. 

6.6 Replace battery 

—

  If the battery level is very low, please turn off the unit 

immediately and replace the batteries 

—

  Take out the batteries if not in use for a long period 

of time (like one month) 

—

 

NOTE: 

do not charge the non-rechargeable battery, 

dangerous! 

7. Maintenance 

1)  Keep all optical connectors and surfaces free 

from oil, dirt or other contamination to ensure 
proper operation. 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

 10

2)  Keep using the same type of connector 
3)  Please cover the dust cap when not in use to 

keep the connector clean. 

4)  Carefully plug in or out the adapter 
5)  Regularly clean the connector. 
6)  Take out the battery when not in use for a 

period of time. 

8. Trouble-Shootings 

Problem Reason 

Solution 

Faint LCD screen 

Lower power

Replace battery 

No backlight 

Lower power

Replace battery 

No display when 
turning on 

Lower 
power/other 

Turn on again or 
replace battery 

Fail to turn on 

Lower power

Replace battery 

No changing on 
LCD screen 

Lower power

Replace battery 

In sensitive LCD 
screen 

Dirty or 
polluted 
connector 

Use the correct 
connector and 
clean it. 

Warranty Period

 

12 months from the purchasing date 

Warranty clause

 

1. Under the warranty, we would repair the problems 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

 11

caused under the normal operation free of charge. 

Note: remember to show us the warranty card when 
repair 

2. For the following situation, we have to charge certain 
cost

 

1)  Out of the warranty

 

2)  Fail to provide the warranty card 
3)  Alter and omit the warranty card 
4)  Wrong operation including the problem caused 

by human actors, abnormal working 
environment and so on. 

5)  Problems or damages are not caused by the 

products quality 

6)  Do not operate it according to this manual. 

3. For the following situation, we do not repair it 

1)  Damaged seal label. 
2) 

Take apart it without permission 

3) Products 

from 

other 

manufacture 

 
 
 
 
 
 
 
 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

 12

Packing List 

1) TL-510 Optical Power Meter…………………1pc 
2) Soft 

case  ……………………………………1pc 

3)  User Manual   ………………………………… 1pc 
4)  Battery(AAA, transportation permit) ………3pcs 
5)  Quality Certificate ………………………………1pc 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 


background image

                                                           

T&L TECHNOLOGY CO., LIMITED 

                                                   

 13

Quality Certificate 

 
Product name

Optical Power Meter

 

Product model

TL-510 

Serial number

 

Date of production

 

Inspection 

 

 

Warranty Card 

 
Date of purchase

 

Customer name

 

Customer address

 

Customer phone

 

Distributor name

 

Distributor phone

 

Posted back to

 

 
Fault unit return back to manufacture for repair purpose 
should come with the warranty card with detailed 