Trident Installation: Preface 

This guide will illustrate how to physically install the Trident. 

The guide below will walk through the different steps illustrated on the Get 
Started card included with your Neptune Systems Trident, but the 
recommended method is to use the Trident Initial Setup task found in Apex 

Fusion.  The guide can also be used to supplement the Trident Initial Setup 
Task and provide additional information for those looking for further details. 

DO NOT CONNECT THE TRIDENT TO YOUR APEX 

..

 

While this may seem counterintuitive, do not connect your Trident to the 

Apex until instructed to do so.  Connecting the Trident before other 
necessary connections are made will cause the Trident to go into an error 
mode and will cause the installation of the system to take additional time. 

Update Apex AOS 

.

 

On the Apex dashboard click the button with 3 gears, then click the WiFi 
button.  If there is not a large orange Update AOS button, then your Apex 
already has the latest AOS version installed.  If you see a large orange 

Update AOS button, click that button and then click the Update button 

The minimum required AOS version for the Apex is 

5.05_5D19.

 

 

 

 


background image

Review Package Contents 

Items included with the Neptune Systems Trident: (Select a link for more 
information, click it again to dismiss the pop-up) 

 

Trident Marine Aquarium Water Analyzer 

 

Two Month Reagent Kit

 

 

Trident Calibration Solution

 

 

1LINK Cable

 

 

Sample Line Tubing and Holder 

 

Waste Line Tubing 

Connect Black Sample Line Tubing 

.

 

It is suggested to make the initial connections to the Trident in an open 
space.  The first connection that should be made is to connect the black 

sample line tubing to the black connector on the back of the Trident. 

DO NOT EXTEND OR SHORTEN THE LENGTH OF THE TUBING

 

Altering the length of the sample tubing will cause significant issues with the 
proper measurement of the Alkalinity, Calcium, and Magnesium levels. 

Connect Waste Line Tubing 

.

 

Connect the clear waste line tubing to the clear connector on the back of the 
Trident. Place the other end of the waste tubing in a suitable container.   The 

minimum recommenced capacity of the waste container is 2 
liters.  Optionally, if you have a drain close by, you may place the far end of 

the waste tubing into that drain. 

Like with the sample line tubing, do not cut or extend the waste tubing. 

Connect 1LINK Cable to Trident 

.

 

Connect the 1LINK cable to the 4-pin jack on the back of the Trident. 

Customers outside the US and Canada and those without an available 1LINK 

port, connect the A

quabus cable

 and 

power supply

. 

 DO NOT CONNECT THE CABLES TO THE APEX YET 

 


background image

Install Reagent C 

Turn the Trident around so the front is accessible and open the reagent 
drawer. 

Find the reagent C bottle and give it a good shake. 

Place the bottle in the drawer and connect the reagent C intake assembly 

(green) to the bottle. 

Make sure the intake assembly tubing is pointed towards the back left corner 

of the reagent drawer 

 this prevents the tubing from kinking when the 

drawer closes. 

Install Reagent B 

Find the reagent B bottle and give it a good shake. 

Place the bottle in the drawer and connect the reagent B intake assembly 
(blue) to the bottle. 

Make sure the intake assembly tubing is pointed towards the right side of 
the reagent drawer 

 this prevents the tubing from kinking when the drawer 

closes. 

Install Reagent A 

Find the reagent A bottle and give it a good shake. 

Place the bottle in the drawer and connect the reagent A intake assembly 

(red) to the bottle. 

Make sure the intake assembly tubing is pointed towards the right side of 

the reagent drawer 

 this prevents the tubing from kinking when the drawer 

closes. 

NOTE:  It is recommended that you keep a set of three caps for reagent 
bottles just in case you need to reseal the bottles.  Rinse with tap water, let 

dry, then store in a suitable spot, such as in the Trident box. 

Close the drawer and keep the packaging 

Close the Trident reagent drawer taking special care to guide the intake 
assembly line tubing appropriately. 

Please be sure to keep the Trident packaging.  Later this packaging may 
need to be re-

used for the Trident’s routine service interval

 

(NEEDS LINK)

 


background image

Position Trident 

Place the Trident near the aquarium in a position which meets the following 
requirements: 

1.

 

Level Surface 

2.

 

Must remain upright 

3.

 

Close to the aquarium or sump 

 no more than 3-

4′ from the aquarium or 

sump.  The sample line tubing must not be extended or shortened under any 

circumstances 

Install the Sample Line 

Install the black sample line tubing in the long slot of the sample line 
holder.  The following requirements need to be met in regards of the sample 

line placement: 

1.

 

The intake should ideally be 2-

3″ from the bottom of the sump.

  When the 

sample line tubing comes in contact with the bottom of your sump it is more 
likely to get clogged with micro and macro-fauna. 

2.

 

The intake should be positioned in an area without microbubbles.  If the 

sample line is near or in an area with a high density of micro-bubbles, then 
those bubbles may be drawn into the Trident along with test sample water 

and will affect the precision and accuracy of the measurements. 

3.

 

The intake should be positioned away from dosing additive lines.  If additive 

enters the test sample, then the precision and accuracy of the measurements 

will drastically be affected. 

4.

 

Avoid placing the sample line above the Trident.  If the sample line becomes 

disconnected, then a gravity siphon can occur. 

Connect 1LINK to Apex 

Connect the 1LINK cable to an available 1LINK port. 

For customers outside the United States and Canada or those without an 
available 1LINK port, connect the Aaquabus cable and power supply.  The 

power supply should 

not

 be plugged into an Energy Bar outlet.  If the power 

supply is connected to an Energy Bar outlet, that outlet 

must

be programmed 

to be always on. 

Upon connection the Trident status LED panel will turn blue, indicating that it 

is initializing.  The initialization will take 2-4 minutes to complete.  Upon 
completion, the Trident status LED panel will turn orange. 

 

 


background image

Set Waste Size Container 

1.

 

Click on the configuration button (the button with 3 gears)

 

2.

 

Select the Module icon 

3.

 

Select the Trident 

4.

 

Select Container or Drain 

5.

 

If you selected Container, enter the capacity of your waste container in 
milliliters 

6.

 

Update to the cloud 

Prime Reagent and Sample Lines 

Once the Trident has a solid orange status LED panel, then: 

1.

 

Select the Prime dropdown 

2.

 

Select the “Select All” option

 

3.

 

Update to the cloud 

Trident Tile 

The Trident Tile offers many different features.   Clicking on the gear: 

1.

 

Can initiate an alkalinity or combined test (Alk, Ca, and Mg) immediately. 

2.

 

Can turn your trident to “off” by switching it to “Manual.”

 

PLEASE NOTE: THAT IF TURNING THE TRIDENT OFF FOR MORE THAN 72 

HOURS THE 

SHUTDOWN PROCESS

 MUST BE INITIATED!

 

3.

 

Offers further configuration options 

 such as choosing a testing schedule. 

Clicking on the tile will: 

1.

 

Show the Trident mini-graphs 

2.

 

Illustrate the remaining volume in the reagent bottles 

3.

 

Illustrate how full the waste container is 

4.

 

Illustrate the number of tests remaining 

Schedule Tests 

After observing the Trident tile, schedule the testing interval desired for the 
Trident. 

1.

 

Select the gear 

2.

 

Select “Schedule”

 

3.

 

Choose the desired testing frequency 

4.

 

Update to Apex Fusion 

Please note the included reagent with the Trident purchase will last for 

approximately a two month period and allow for testing of alkalinity four 
times per day and calcium and magnesium twice per day.  Choosing a higher 


background image

testing frequency will use reagent at a greater rate and therefore the 
reagents will need to be replenished more frequently. 

Add Trident Error Notification 

Make sure to be notified if the Trident experiences an error that can be 
resolved in the 

Trident Troubleshooting Guide

 (below). 

1.

 

Select the gear associated with the email alert outlet 

2.

 

Type the following line to the email alert outlet 

“If Error Tri”

 

3.

 

Select the auto filled Trident name 

4.

 

Complete the coding line: 

“If Error Trident_#_3 Then ON” (# will be unique for your installation; it is 
the Trident’s Aquab

us address) 

5.

 

Update to Apex Fusion 

This completes the Trident Installation Guide. 

 

 


background image

Trident Troubleshooting: Preface 

This guide will help to resolve Trident errors than can occur with the 
operation of the Trident. 

It is important that you complete this troubleshooting before contacting 
Neptune Systems.  While Neptune Support would be happy to take your call 

if you are having trouble

 the troubleshooting steps required for the Trident 

take time.   The steps contained here will help achieve the best possible 

support for you as our customer. 

While you should rarely need to use this resource, if your Trident has a “red” 

status light, then that means that your Trident is currently in an error 

condition.  You can determine the error by simply looking at the Trident tile 
on your Apex Fusion dashboard. 

 

Error Condition 

Current 

Explanation of Error 

Current

 

The "current" error indicates that the peristaltic pump inside the Trident is malfunctioning. 

Recommended Course of Action 

This indicates an issue that can only be diagnosed by Neptune Systems Support. 

Error Unresolved 

We are sorry this troubleshooting did not resolve your Trident Error.   

 

Error Condition 

No Test in 24H

 

Explanation of Error 

No Test in 24H

 

The "No Test in 24H" error indicates that the Trident has been off for more than 24 hours.  For the Trident to 

continue to run appropriately and not require a Neptune Systems service, it must be run at least once every 24 

hours. 


background image

It is also possible to experience this error if the Trident had a previous error condition that has since been 

cleared, but a new test has not yet been initiated. 

Recommended Course of Action 

No Test in 24H

 

Check to make sure the Trident is not in Manual operation mode.  If it is set to manual, then change it to 

Automatic.  Next, run a test of your choosing. 

 

Error Condition 

No Water Sample 

Explanation of Error 

No Water Sample

 

The “no water sample” error indicates that no water is being drawn into the Trident.

 

Recommended Course of Action 

No Water Sample

 

Check to make sure the sample line is in the water and is not kinked. 

If the sample line was out of the water or it was kinked, then remove the 1LINK cable to the Trident and 

reconnect it to clear the error. 

 

Error Condition 

Reagent A Empty

 

Explanation of Error 

Reagent A Empty

 

The “Reagent A Empty” error indicates that the reagent A bottle is empty or nearly empty.

 

Recommended Course of Action 

Reagent A Empty

 

If the reagent A is nearly empty, then run the reagent A replacement task in Apex Fusion or see the guide to 

manually replace reagents 

 


background image

Error Condition 

Reagent B Empty 

Explanation of Error 

Reagent B Empty

 

The “Reagent

 

B Empty” error indicates that the reagent

 B bottle is empty or nearly empty. 

Recommended Course of Action 

Reagent B Empty

 

If the reagent B is nearly empty and the other reagents are close to empty too, then run reagent replacement 

task in Apex Fusion or see the guide to manually replace reagents 

 

Error Condition 

Reagent C Empty

 

Explanation of Error 

Reagent C Empty

 

The “Reagent

 

C Empty” erro

r indicates that the reagent B bottle is empty or nearly empty. 

Recommended Course of Action 

Reagent C Empty

 

If the reagent C is nearly empty and the other reagents are close to empty too, then run reagent replacement 

task in Apex Fusion or see the guide to manually replace reagents 

 

Error Condition 

Stirrer 

Explanation of Error 

Stirrer

 

The “Stirrer” error indicates that the mixing stir bar inside the Trident is not working.

 

Recommended Course of Action 

This indicates an issue that can only be diagnosed by Neptune Systems Support. 


background image

Error Unresolved 

We are sorry this troubleshooting did not resolve your Trident Error.   

 

Error Condition 

Test A Fail

 

Explanation of Error 

Test A Fail

 

The “Test A Fail” error indicates there is an error in the titration associated with reagent A.

 

Recommended Course of Action 

Test A Fail

 

This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level 

according to the Trident tile and Apex.  

To resolve this, run the reagent A replacement task in Apex Fusion or see the guide to manually replace 

reagents. 

After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident. 

 

Error Condition 

Test B Fail 

Explanation of Error 

Test B Fail

 

The “Test B Fail” error indicates there is an error in the titration associated with reagent B.

 

Recommended Course of Action 

Test B Fail

 

This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level 

according to the Trident tile and Apex.  

To resolve this, run the reagent B replacement task in Apex Fusion or see the guide to manually replace 

reagents. 


background image

After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident. 

 

Error Condition 

Test C Fail

 

Explanation of Error 

Test C Fail

 

The “Test C Fail” error indicates ther

e is an error in the titration associated with reagent C. 

Recommended Course of Action 

Test C Fail

 

This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level 

according to the Trident tile and Apex.  

To resolve this, run the reagent C replacement task in Apex Fusion or see the guide to manually replace 

reagents. 

After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident. 

 

Error Condition 

Volts 

Explanation of Error 

Volts

 

A "Volts" error usually indicates an issue with the 24VDC power supplied to the Trident from the 1LINK or 

power supply. 

Recommended Course of Action 

Volts- 1LINK

 

If you are using a 1LINK cable with the Trident, this error can be resolved by simply removing the 1LINK 
cable from the Trident and then reconnecting it.  

If the VOLT error continues, disconnect the far end of the 1LINK cable from the EB832 or 1LINK module 
then reconnect it.   

Volts- Aquabus and Power Supply

 


background image

Ensure that the power supply cord is properly connected to  the Trident power input jack.  Also verify that 
the power supply has AC input power.  

Neptune Systems does not recommend plugging the Trident power supply into an EneryBar outlet.  But if 
you have done so, the EnergyBar outlet must be programmed to be always on, like this: 

Fallback ON 

Set ON 

 

Error Condition 

Waste Full

 

Explanation of Error 

Waste Full

 

The “Waste Full” error occurs because the waste container is completely full.

 

Recommended Course of Action 

Waste Full- Container

 

To resolve this: 

1.  Empty the waste container 

2.  From the Trident tile click the gear and select "Configure Trident" 

3.  Select "Reset" 

4.  Select "Waste" 

5.  Sent the update to Apex Fusion 

The error should clear in about a minute. 

Waste Full- Drain

 

To resolve this: 

1.  From the Trident tile click the gear and select "Configure Trident" 

2.  Select "Waste lne" 

3.  Select "Drain" 

4.  Sent the update to Apex Fusion 

The error should clear in about a minute. 


background image

Trident Controlled DŌS: Preface

 

This guide will walk through setup of Trident controlled dosing using the 

Neptune Systems DŌS. Controlled dosing allows you to set desired levels for 

alkalinity, calcium, and magnesium, and have automatic control of those 
parameters.  The best way to configure this will be with the task found in 

Apex Fu

sion called “Trident Controlled Dosing.”

  Optionally, you may use this 

guide in lieu of the Trident Controlled Dosing Task. 

Neptune Systems does not recommend running the Trident Controlled 
Dosing Task or going through this guide until the Trident has been in 
operation for at least 7 to 10 days

You, as the aquarist, must generally understand the amount of each additive 
needed to maintain stable alkalinity, calcium, and magnesium parameters in 

the aquarium for this guide to be successful. 

Select DŌS Pump

 

This guide assumes the Neptune Systems DŌS has already been connected 

to the Apex system and that the DŌS has been properly calibrated.

  If that 

still needs to be completed, then please see the 

DŌS Get Started Guide

 to 

physically install the DŌS

.

 

Provided that is complete, select the DŌS pump you wish to configure with 

the Trident by selecting the gear above the dashboard slider control for that 

DŌS pump

.

 

Configure Amount to Dose 

1.

 

Select Modify Interval

 

2.

 

Enter the Start time as “0:00”

 

3.

 

Enter the End time as “23:59”

 

4.

 

Make sure Action is set to “Add”

 

5.

 

Enter the volume of additive to dose to maintain the parameter.  In this 

example, that is calcium.  

This is called the “normal dosing amount.”

 

6.

 

Select “OK”

 

For Trident Controlled Dosing of calcium and magnesium supplements, 

Neptune Systems recommends as few intervals as possible to experience the 
best results; a single interval spanning the full 24-hour day is suggested for 

calcium and magnesium additives, and also for alkalinity dosing for 
aquariums with a relatively low to moderate rate of alkalinity consumption. 

Typically in aquariums with high rates of calcification, the majority (about 

75%) of the daily alkalinity consumption occurs during the primary lighting 


background image

period of the aquarium.  Consequently, when configuring Trident Controlled 
Dosing of alkalinity supplements, the use of three intervals is 

suggested.  Here is an example: 

1.

 

the first interval will cover the period from 0:00 until the start of the primary 

lighting period; the volume configured for this interval will be approximately 

15% of the total nominal daily amount of alkalinity supplement 

2.

 

the second interval will cover the primary lighting period, for example, 08:00 

to 19:00; the volume configured for this interval should be approximately 
70% of the total nominal daily amount of alkalinity supplement 

3.

 

the third interval will cover the period from the end of the primary lighting 

period until 23:59; the volume configured for this interval will be 
approximately 15% of the total nominal daily amount of alkalinity 

supplement 

Optionally, distribute the volume proportionally (rather than equally) 
between the first and third intervals based on the number of hours of those 

intervals. 

Enable Trident Control

.

 

1.

 

Select the “Configuration Gear”

 

2.

 

Select “Control Interval”

 

3.

 

Check the “Automatically adjust dosing using Trident measurements.”

 

Select Parameter

 

1.

 

Select “Test”

 

2.

 

Choose the parameter that will influence the total dosing volume 

Select Control Value

.

 

Enter the “Value” you wish to maintain for the selected parameter

 

The dosing volume will change up or down in order to maintain this desired 

value. 

Select the Control Range 

Enter the control range for the desired parameter. 

This is the range in which the Trident will have influence over the dosing 

volume.  

Outside the range the Trident will return to the “normal dosing 

amount.”

 


background image

For example, in the situation illustrated here with a control value of 425 and 
a range of +/-50, the Trident will have influence over the calcium level 

between 375 and 475.  The greatest degree of influence will be 375 ppm and 
the least degree of influence at 475 ppm. 

Parameter 

Recommended Range 

Alkalinity (dKH) 

0.4 

Calcium 

50 

Magnesium 

50 

Select the Control Limit 

Enter the control limit for the dosing pump of the controlled parameter. 

This is the degree in which the Trident will have influence over the dosing 
volume.  When the maximum or minimum dosing amount is reached the 

Trident will no longer influence and the dosing will return to the “normal 

dosing amount.”

 

For example, in the situation illustrated here with a control limit of 40% and 

a “normal dosing amount” of 25 mL, the maximum volume of additive that 

could be added over 24 hours is 35 mL(25*1.4 = 35) and the minimum 

amount is 15 mL (25*0.6=15). 

Parameter 

Recommended Limit (%) 

Alkalinity (dKH) 

35 

Calcium 

40 

Magnesium 

40 

Update to Apex Fusion 

1.

 

Select OK

 

2.

 

Update to Apex Fusion 

3.

 

Review the dosing settings by selecting the “Interval Summary”

 


background image

What Values Should I Enter? 

This is an involved question and Neptune Systems recommends to use the 
starting values and fine-

tune those values as the Trident and DŌS become 

dialed in. 

The best analogy for how Trident Controlled Dosing works is using that of a 

car speeding up and slowing down.  Imagine there is a car and the driver 
wants to maintain a certain speed. 

When the car is going far below the speed limit, the driver will push hard on 
the gas to achieve that speed.  This is what the Trident will do when the 
parameter is near the minimum allowable range- it will add the highest 

allowable amount of additive according to the control limit. 

When the car is only going a couple miles below the speed limit, then the 

driver will only slightly press on the gas to reach the desired speed.  The 

Trident will only influence the DŌS to increase small percentage over the 

“normal dosing amount”

 

The inverse of both examples holds true when the car is going above the 

desired speed.  If the driver is going too fast, then a hard break press is 
warranted.  Only a couple miles over the limit, the driver will lightly press 

the brakes. 

 

How do I fine tune this? 

As illustrated above, there are several variables that influence Trident 

Controlled Dosing.  Take a look at each variable to fine-tune the settings: 

Normal Dose Amount 

This is the value to always look at first.  If the desired parameter is not being 
maintained appropriately, then the dosing volume is likely too much or too little to 
maintain that desired parameter.

 

Control Value

 

This is the second value to look at.  

It’s possible the “control value” that has 

been selected is an “ideal value.”

  

However, if the “control value” is too far 

above or below the “actual value” the Trident is reading, then temporarily 

adjust that value up or down closer to the level the Trident is 


background image

reporting.  Make small changes.  

For example, if the “ideal” alkalinity is 9.0 

dKH and the “actual” alkalinity is 6.0 dKH, then a good “control value” would 

be 6.7.  

Then as the “actual value” becomes closer to the “control value” 

m

ove the “control value” closer to the desired, “ideal value.”

 

Control Range

 

In most situations, it is best to keep the control range consistent and 
unchanged.  Too narrow of a range is like being behind a driver that speeds 

up and slows down frequently and behaves erratically.  It best to keep this 
around the recommended values listed above. If using too narrow of range, 
then the Trident will not be able to effectively influence controlled dosing. 

Control Limit

 

In most situations, this is the variable to change if fine-tuning the desired 
level.  If overshooting or consistently too high, then decrease the control 

limit.  If undershooting or consistently too low, then increase the control 
limit. 

Closing

 

If adjustments are being made, then only adjust one variable at time and 

wait at least 24 hours to see what kind of affect that has on the controlled 
parameter before changing another parameter.