PDF User Manual

  1. Home
  2. Manuals
  3. MELAG MELAdest 65 Operating Manual

MELAG MELAdest 65 Operating Manual

Made by: MELAG
Type: Operating Manual
Category: Distilling equipment
Pages: 12
Size: 1.00 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

 

 
Dear Dr. 
 
We should like to extend our thanks for the expression of trust in our company which you have displayed through the 

purchase of this device. 
MELAG, a medium-sized family run company has specialized in the area of practice hygiene since 1951. During this 

time, we advanced to become one of the leading manufacturers of sterilizers. World sales of more than 450,000 MELAG 

devices gives testimony to the quality of our devices, which are manufactured exclusively in Germany. 
This device was manufactured and checked according to strict quality criteria.  
 
MELAG – management and employees 

Operating Manual 

MELAdest

®

65 

Water distiller 

 

 

 
 
 

 


background image

 

General notes 

Please read this operating manual carefully before commissioning the device. The instructions include important safety 

information. The functionality and value-retention of this autoclave depends on the care accorded to it. 
Please store this operating manual carefully and in close proximity to the device. It represents a component of the 

product. 

User group 

These instructions are addressed to doctors, their assistants and service departments. 

About this manual

 

Symbols used 

Symbol 

Explanation 

 

Indicates a dangerous situation, which if not avoided could result in injuries ranging in 

seriousness from light to life-threatening. 

 

Draws your attention to a situation, which if not avoided, could result in damage to the 

instruments, the practice fittings or the device. 

 

Draws your attention to important information. 

 

 


background image

Contents 

Contents

Chapter 1 – Device description ....................................................................................................... 4

 

Intended Use ................................................................................................................................................................... 4

 

Functional principle .......................................................................................................................................................... 4

 

Scope of delivery ............................................................................................................................................................. 4

 

Views of the device .......................................................................................................................................................... 4

 

Chapter 2 – Commissioning ............................................................................................................ 5

 

Fitting the glass flask ....................................................................................................................................................... 5

 

Before commissioning ..................................................................................................................................................... 7

 

Chapter 3 – Operation and care ...................................................................................................... 8

 

Producing distilled water .................................................................................................................................................. 8

 

Cleaning the distilling chamber ........................................................................................................................................ 9

 

Chapter 4 – Malfunctions .............................................................................................................. 10

 

Appendix A – Spare parts ............................................................................................................. 11

 

Technical Data .............................................................................................................................. 11

 

 

 
 
 

 


background image

Chapter 1 – Device description 

Chapter 1 – Device description 

 

Intended Use 

The MELA

dest 

65 water distiller produces distilled water. 

Functional principle 

This practical water distillator produces water of the highest quality in accordance with what represents 

one of the oldest methods of water treatment. When the device is activated, the water temperature 

increases and escaping gases are boiled, then led through an opening. The water temperature increases 

to 100°C and begins to boil. During evaporation, the chemicals, salts and soiling remain in the container. 

The stream flows through the cooling coil system in the upper section where it cools and condenses into 

distilled water.  

Scope of delivery 

Standard scope of delivery 

  MELA

dest

 65 water distiller 

  Power cable 

  Glass flask 

  Plastic guard ring 

  Adapter cover 

  Spout cap 

  Handle 

  Handle cover inc. bracket for adaptor cover 

Views of the device 

 

 






Hook for spout cap 
Bracket for adapter-cover 
Metal collar 
Glass flask 
Spout cap 
Handle 




10 
11 
12 

Cover 
Adapter cover 
Plastic guard ring 
Upper section of the distilling chamber 
Connection cable 
Power cable 

 


background image

Chapter 2 – Commissioning 

Chapter 2 – Commissioning 

 

Fitting the glass flask 

1. 

Remove all the individual parts from the glass flask carefully.  

2. 

Remove the packaging from the individual parts. 

3. 

Place the glass flask in the plastic guard ring. 

4. 

Loosen a screw on the metal collar on the neck of the glass flask. 

 

5. 

Place the collar on the neck of the glass flask in such a way that the other screws on the collar lie 

exactly opposite the spout cap (pos. a). 

 

 

6. 

Refasten the metal collar with both screws. 

 

PLEASE NOTE 

Do not overtighten the screws when tightening.  

 

7. 

Fill the glass flask 3/4 full with warm tap water. Add 1-2 drops of washing-up liquid and clean the 

inside of the glass flask. 

8. 

Afterwards, rinse and clean the glass flask thoroughly with clear water. 

9. 

Place the cover (pos. b.) on the glass flask in such a way that the spout remains open. 

 

 

 


background image

Chapter 2 – Commissioning 

10. 

Place the adapter cover (pos. c) and the spout cover cap (pos. d) on the cover. 

 

 

11. 

Insert the handle cover (pos. e) and the small hook (pos. f) on the end of the handle end so that 

they snap closed. 

     

 
A bracket for the adapter cover is integrated in the handle cover. To this end, the adapter opening 

is pushed onto the handle cover snap tab. 
The spout cap is held on the small hook. 

 

 

 

 


background image

Chapter 2 – Commissioning 

Before commissioning 

Cleaning the distilling chamber 

Although the distilling chamber of the MELA

dest

 65 has been cleaned with the greatest of care, 

manufacturing, packaging and dispatch procedures could result in the presence of harmless residues of 

various substances. As a result, the distilled water produced during initial commissioning could assume an 

undesired odour. To eliminate all odour-generating factors, proceed as follows: 

1. 

Setup the water distiller.  

2. 

Remove the connection cable from the distilling chamber. 

3. 

Remove the upper section of the water distilling chamber and fill with c. 0.5 L tap water. 

4. 

Return the upper section to the distilling chamber and connect the connection cable to the 

bushing on the side wall of the chamber. 

5. 

Replace the glass flask laterally rotated without the adapter cover under the distilling chamber 

discharge opening and plug the plug into the socket.  

                                   

 

6. 

Press the START/RESET button to activate the water distiller.  

The ventilator motor begins to run. 

The water distiller deactivates automatically after the water has boiled (no steam egress from the 

upper section discharge opening.  

7. 

Remove the mains plug and empty the glass flask.  

The distilling chamber is now flushed and ready for commissioning. 

 

Checking the cover seal in the upper section of the distilling chamber  

Before initial commissioning and every further use of the device, check the correct position of the cover 

seal in the upper section of the distilling chamber. 
The seal may not present any uneven surfaces and must lay entirely flat so that it maintains the seal. 
Otherwise, boiling water could leave the upper section and e.g. flow onto the work surface. 
If the seal does not lay flat, press with the hand. Do not use any sharp objects. 

 

 

incorrect position of the seal

 

sealing lip

 

sealing lip

 

seal evenly fitted at all points

 

 

Discharge opening 

of the distilling 

chamber 

 


background image

Chapter 3 – Operation and care 

Chapter 3 – Operation and care 

 

Producing distilled water 

 

  DANGER! 

The water is very hot. Steam can be emitted from the upper section. 

Danger of burns. 

 

Never remove the upper section if the ventilator motor is running. 

 

Allow the water distiller to cool for 30 minutes before removing the upper section. 

 

 
Proceed as follows to produce distilled water: 

1. 

Remove the mains plug and disconnect the connecting cable from the bushing of the distilling 

chamber. 

2. 

Remove the upper section of the water distiller. 

3. 

Fill the distilling chamber up to the upper marking with c. 4 L tap water. 

 

PLEASE NOTE 

The use of hot water reduces the length of the process.  

 

 

4. 

Replace the upper section on the distilling chamber and reconnect the connection cable to the 

distilling chamber bushing.  

 

PLEASE NOTE 

Check the seal in the upper section of the distilling chamber before every use. See: 

Chapter 2 – 

Commissioning

, p. 7. 

 

 

5. 

Place the glass flask (laterally rotated) under the distilling chamber discharge opening. 

                                   

 

6. 

Plug the mains plug into the mains socket.  

7. 

Press the START/RESET button. The ventilator motor in the upper section of the distilling 

chamber begins to run.  

Once the cycle has finished, the water distiller will switch off automatically. The distilled water in 

the glass flask is ready to use. Close the lateral opening with the adapter cover. 

8. 

Remove the mains plug from the socket after use. 

 

 

Discharge opening 

of the distilling 

chamber 

 


background image

Chapter 3 – Operation and care 

9. 

Dry the inside of the distilling chamber with a cloth to prevent corrosion. 

 

The conductivity meter MELAtest

®

60 from MELAG can be used to check the conductivity 

of the water thus produced (aqua dest). 

Cleaning the distilling chamber 

Before cleaning the water distilling chamber, please observe the following: 

 

Clean the inside of the distilling chamber with lukewarm tap water and a sponge to remove the 

deposits from the distilling chamber. With stubborn soiling / deposits on the inside walls, use only a 

mild stainless steel cleaning medium, with a pH value between 5 and 8 e.g. Sidol. 

 

Clean the inside of the distilling chamber following heavy limescale deposits or after 6 months at the 

latest with MELAd

est

 cleaning fluid. 

 

  WARNING 

If water comes into contact with the electric connections, this can result in the short-

circuiting of the water distiller. 

 

Never immerse the distillation chamber in water. 

 

 

 


background image

Chapter 4 – Malfunctions 

10 

Chapter 4 – Malfunctions 

 

Incident 

Possible cause 

What you can do 

No distilled water is produced 

The mains plug is not in the 

mains socket 

 

Mains socket is defective 

 

 

Upper section connector cable is 

not on the distilling chamber 

 

START/RESET button is not 

depressed 

Check the mains plug 

 

 

Check the electricity supply on 

the mains socket 

 

Connect the connection cable to 

the distilling chamber 

 

Press the START/RESET button 

Too little distilled water produced  Upper section connection cable is 

not in the distilling chamber 

 

Distilling chamber is not filled 

sufficiently 

 

The cover seal does not seal 

sufficiently, water escapes 

 

 

Glass flask positioned incorrectly 

under the water distiller 

 

 

 

Cooling system soiled 

Connect the connection cable to 

the distilling chamber 

 

Refill with tap water 

 

 

Replace the cover seal in the 

upper section of the distilling 

chamber  

 

Check whether water flows past 

the glass flask and place the 

glass flask under the water 

distiller correctly 

 

Clean the cooling system, blow 

out with compressed air if 

necessary 

Steam escapes under the 

discharge opening of the upper 

section 

Upper section connection cable is 

not in the distilling chamber 

 

Connect the connection cable to 

the distilling chamber 

 

The water distiller does not heat  The water distiller overheats 

Allow the water distiller to cool for 

5-10 minutes, then restart 

The water distiller does not 

deactivate automatically 

Overheating switch defective 

 

 

 


background image

Appendix A – Spare parts 

11 

Appendix A – Spare parts 

 

Article 

Order no. 

Glass flask 

01067 

MELA

dest

 65 cleaning fluid 

01068 

MELA

test

 60, conductivity meter (for checking water quality) 

01060 

MELA

dest

 65 housing lower section  

65410 

MELA

dest

 65 housing upper section 

65420 

MELA

dest

 65 cover seal (for housing upper section) 

65510 

 
 

Technical Data 

 

Model name 

MELAdest

®

65 

Exterior dimensions of the water distiller (Ø x H) 

Ø 23 cm x 38 cm 

Exterior dimensions glass flask (Ø x H) 

Ø 18 cm x 18.3 cm 

Empty weight of the water distiller 

3 kg 

Electrical connection 

230 V, 800 W 

Capacity 

c. 0.7 litres/hr. 

Conductivity of the distilled water 

2-3 µS 

 

 
 

 


background image

 

MELAG Medizintechnik oHG 

Geneststraße 6-10 

10829 Berlin 

Germany 
E-Mail: info@melag.de 
Web: www.melag.de 

Responsible for content: Technical Office 
Subject to technical alterations 
 
 
 
 
 
 
 
 
 

Your stockist: 
 
 
 
 
 

B

A

_E

N

_des

t65_

v2.

do

cx

 

R

ev

.: 

– 

13/

1344