Table of Contents

1.

Introduction

………………………………………………. 1

2.

Safety note

……………………………………………..… 2

3.

Explanation of Controls and Indicators …………………... 3

3-1. Product illustration

………………………………………. 3

3-2. Functional push button

……………………………………. 4

3-3. LCD display

………………………………………………. 5

4.

Specification

………………………………………………. 5

4-1. General Specification

……………………………………... 6

4-2. Electrical Specification

…………………………………. 7- 9

5.

Measurement operation

………………………………….. 9

5-1

DC voltage & AC voltage measurement

………………... 9

5-2.

Resistance measurement

…………………………….….. 10

5-3.

Diode & Continuity check

………………………..…..… 11

5-4.

Frequency Measurement

[ Only for ADM01 ]

…….…… 11

5-5.

DC/AC µA or mA measurement

………………….……. 12

5-6.

DC/AC 10A measurement

…………………………..…. 12

5-7 Temperature measurement

[ Only for ADM02 ]

………..…. 12

6. Maintenance

………………………………………………. 13

6-1. Replacing the battery

…………………………...………. 13

6-2. Replacing the fuse

………………………………………. 13

6-3. Cleaning and Decontamination

…………………………. 13

 


background image

1.

Introduction

This manual is used to ADM01 and ADM02 Digital Muti Meter only.
This Meter is a handheld and battery operated Digital
MultiMeter(DMM) with multi function. This Meter is designed to meet
IEC61010-1 &

CAT II 600V

over voltage category and double insulation.

The meter with holster that is giving the main body, though
downsized, high resistance against the shock of a drop.
This operating instruction covers information on safety and caution.
Please read relevant information carefully and observe all the
warnings and note strictly.
The DMM as general measurement tool and widely used in the school,
laboratory, factory and other social field.

2.

Safety note

Warning

To avoid possible electric shock or personal injury and to avoid
possible damage to the meter or to the equipment under test, adhere
to the following rule:

Do not apply more than the rated voltage, of marked

on the meter, between the input terminal and
grounding terminal..

Do not apply voltage between COM and OHM terminal,

in the resistance measuring state.

Do not measure current with test lead inserted into

voltage or OHM terminal.

Do not expose the instrument to the direct sun light,

extreme temperature and humidity or dew full.

Inspect the test lead for damaged insulation or

exposed metal.

Before measuring current, check the Meter’s fuses and

turn off power to the circuit before connecting the
meter to the circuit.

Disconnect circuit power and discharge all high

1

 


background image

voltage capacitors before testing continuity, diode,
resistance, capacitance or current.

Note international Electrical Symbol.

Dangerous

Voltage

Ground

AC Alternating
current

Warning

see explain in manual

DC (Direct

Current)

Double insulation

AC or DC

Fuse

Measurement category(over voltage category):

This instrument is meet the safety condition of

CAT II

. The

equipment is used for measurement in building facilities.
Examples are measurements on distribution boards, circuit
breaker and industrial equipment located in fixed facilities, as
a fixed motor.

3. Explanation of controls and indicators

3-1. Meter illustration

 

2

 


background image

 


background image

Fig. 1

Exterior view

1. LCD display

2. “MAX” push button

3. “BACK LIGHT” push

button

4. ‘HOLD” push button

5.

“FUNC” Push button

6. Rotary Switch (Knob) 7. “V/Ω/HZ/uA/mA/Temp”Input terminal

[‘Hz’ is for ADM01 and ‘Temp’ is for ADM02]

8. “COM” input terminal

9. “10A” input terminal

3-2. Functional push button

Push button

Function

Func

“FUNC” key is the function select key that acts
with trigger. Use the key as switch of

DC/AC

current, Diode/Continuity and

/

.

HOLD

Press “HOLD” to enter and exit the hold mode
in any mode. That act with trigger.

MAX

This key is act with trigger. Press this key once,
the maximum value is holding (Will displays
‘MAX’ symbol in the LCD).
After pressing the key, A/D will keep working,
and the display value are always up dated and
kept the maximum value.
NOTE: The actual gained value is not the peak
value.

This key is used control Backlight. This key is act
with trigger. When press and hold the key over
2 sec, will enable Backlight.
Press the key again , the backlight will disable.

3

 


background image

3-3.

Display indicators

Fig. 2

LCD

Indicator

Meaning

DC voltage or current

AC voltage or current

Diode

MAX

Maximum value

HOLD

Data hold

Low battery indicator

MKΩ

Ω KΩ MΩ is unit of resistance

/

The unit of temperature
(

: Centigrade;

: Fahrenheit)

4

 


background image

µmVA

mV ,V is unit of voltage

µA, mA, A is unit of current

Indicate negative reading

4

Specification

4-1. General Specification

Auto ranging DMM , that full scale is 2000 counts

Display : 3 1/2 digit LCD display..

Over load protection:

Used the PTC protection circuit for

Resistance,

temperature(

for

ADM02)

and

frequency

measurement for (AMD01).

DATA HOLD function

MAX value hold function

Back Light

Low battery indication

Auto Power- OFF. : If the meter is idle for 15 minutes (idle time),

the meter automatically turns the power off. After auto
power-off, pushing any of the push button or changing the rotary
switch can turn on the meter again.

NOTE:

1

After auto power off in the AC mode, if changing the

rotary switch to the DC mode, the Re-power on if
disabled.

(2)

The meter enters sleep mode after quto power off. If

press “HOLD” push button to re-power on in the sleep
mode, the auto power off function is disabled.

Operating temperature & Humidity:

0 ~ 40

(32 ~104

) &

< 80% RH

Storage temperature & Humidity:

-10 ~ 50

(14 ~ 122

) &

<70%RH

6 5

 


background image

Power Supply:

9V Battery(6F22 or 1604A Type) x 1pc.

Safety Class:

IEC 61010-1,

CAT II

600V.

Dimension(L x W x H) & Weight:140 x 67 x 30mm, Approx. 112g

4.1.1 Accessary:

1. User’s Manual ----------------------------

1pc

2. Test lead

----------------------------

1set

3.

9V

Battery

---------------------------

1pc

4.

K-type temperature probe----

1pc(For ADM02)

4-2. Electrical Specification

( at 23±5

; <75% RH)

4.2.1 DC Voltage

Range

Resolution

Accuracy

200mV

0.1mV

±

0.5% rdg + 2dgt

2V

0.001V

20V

0.01V

200V

0.1V

600V

1V

±

0.8% rdg + 2dgt

* Over load protection: SG(Spark Gap) used to protect that the
voltage overred 1500V

4.2.2 AC Voltage

Range

Resolution

Accuracy

2V

(

40Hz-400Hz

)

0.001V

±

0.9% rdg

+ 3dgt

20V

(

40Hz-400Hz

)

0.01V

200V (

40Hz-400Hz

)

0.1V

600V (40Hz-200Hz)

1V

±

1.2%

rdg + 3dgt

* SG(Spark Gap) used to protect High voltage that is

over 1500V.

7

 


background image

4.2.3

Resistance

Range

Resolution

Accuracy

200Ω

0.1Ω

±

0.8% rdg + 2dgt

2kΩ

0.001kΩ

20kΩ

0.01kΩ

200kΩ

0.1kΩ

2MΩ

0.001MΩ

±

1.0% rdg + 2dgt

20MΩ

0.01MΩ

4.2.4

Diode check

Range

Resolution

Function

0.001V

Will display the forward drop voltage.

* Operating current

about 1mA

* Open circuit voltage

about 1.48V

4.2.5 Continuity

Rang

e

Function

If measured resistance less than

100Ω

, will buzzer is

sounded.

* Open voltage: about 0.5V

4.2.6 DC Current

Range

Resolutio

n

Accuracy

200µA

0.1μA

±

1.5% rdg + 3dgt

88

 


background image

2000µ
A

1µA

20mA

0.01mA

200mA

0.1mA

10A

0.01A

* Over Load protection

use the fuse(

F250mA/250V

) at µA /mA range,

and use the fuse(F10A/250V) at 10A range.

* Max input current:

250mA

at ‘mA’ input terminal and 10A at ‘10A’

input terminal..

.

4.2.7 AC Current [40Hz-400Hz]

Range

Resolutio

n

Accuracy

200µA

0.1µA

1.5% rdg + 4dgt

2000µ
A

1µA

20mA

0.01mA

200mA

0.1mA

10A

0.01A

* Over Load protection

use the fuse(

F250mA/250V

) at µA /mA range,

and use the fuse(F10A/250V) at 10A range.

* Max input current:

250mA

at ‘mA’ input terminal

and 10A at ‘10A’ input terminal...

* Frequency response

40

400Hz

4.2.8

Frequency

(For ADM01)

Range

Resolution

Accuracy

20kHz

10 KHz

±

1.5%rdg+5dgt

* Sensitivity

0.8V

4.2.9 Temperature

(For ADM02)

 


background image

You can selecting [

] or [

] by “FUNC” key.

Range

-20

~ 1000

Resolutio
n

1

-20

~0

(

5% rdg + 4dgt

Accuracy

0

~400

(

2% rdg + 3dgt

400

~1000

(

3% rdg + 3dgt

Fahrenheit Temperature [

]

Range

0

~1800

Resolutio
n

1

-0

~50

(

5% rdg + 4dgt

Accuracy

50

~750

(

2% rdg + 3dgt

750

~1800

(

3% rdg + 3dgt

5. Measurement operation

5-1

DC & AC voltage measurement

Warning:

To avoid harms to you or damage to the meter from electric

shock. Please do not attempt to measure voltage higher than

DC/AC 1000V although readings may be obtained.
The DC voltage range are 200.0mV, 2.000V, 20.00V , 200.0V and
600V and then. The AC voltage ranges are 2.000V, 20.00V,
200.0V and 600V.

To measure DC or AC voltage:

① Insert the red test lead into the “” input terminal and the

black test lead into the COM terminal.

② Set the rotary switch to

DC or AC range.

③ Connect the test lead across with the object under testing.

The measured value will be show on the LCD display.

Note:

9

 


background image

When DC or AC voltage measurement has been completed,
disconnect the connection between the testing lead and the circuit
under testing.

5-2. Resistance measurement

The resistance range are: 200.0Ω, 2.000KΩ,20.00KΩ, 200.0KΩ ,

2.000MΩ.20.00MΩ.

To measure resistance, connect the meter as follows:

① Insert the red test lead into the ”” terminal and the black

test lead into the COM terminal.

② Set the rotary switch to proper resistance range..
③ Connect the test lead across with the object under testing.

The measured value will be show on the LCD display.

Note:

The test lead can add 0.1Ω to 0.2Ω of error to resistance

measurement. To obtain precision reading in low-resistance
measurement, that is the range of 200.0Ω, short the input
terminal before measuring. In this time, the contact
resistance displayed on the LCD. You can subtract the contact
resistance value from the measured value.

For high-resistance measurement (>10MΩ), it is normal

taking several second to obtain stable reading.

The LCD display “OL” indicating open-circuit for the tested

resistor or the resistor value is higher than the maximum
range of the meter.

5-3. Diode/Continuity check

5.3.1.

Diode

① Set the rotary switch to “

” position.

First time, default

mode is diode check mode. You can enter the continuity check

mode by the “ FUNC” Key.

② insert the red test lead into the “VΩ” terminal and the black test

lead into the “COM” terminal.

10

 


background image

③ Use the diode test mode to check diodes, transistors and other

semiconductor device. In the diode test mode sends a current
through the semiconductor junction, and the measure the
voltage drop across the junction. A good silicon junction drop
between 0.5V and 0.8V.

④ For forward voltage drop reading on any semiconductor

component, place the red test lead on the component anode
and place the black test lead on the component cathode. The
measured value show on the display.

⑤ Reverse the test lead and measure the voltage across the diode

again.

If diode is good, the display shows “OL”.

If diode is shorted, the display shows 0 (zero) in both

direction.

If display shows “OL” in both direction, the diode is open.

5.3.2. Continuity Check:

① Press the “FUNC” key to enter to the continuity mode.
② The buzzer sound if the resistance of a circuit under test is less

than

100Ω

.

5-4. Frequency measurement

[ Only for ADM01]

① Set the rotary switch to “Hz” position.
② Insert the red test lead into the “ VΩHz” input terminal and the

black test lead into the “COM” terminal.

③ Connect the test leads across with the circuit under testing. The

measured value shown.on the LCD display.

NOTE:

Input signal level must be higher than 0.5V (it is sensitivity).

5-5. DC/AC µA or mA measurement

DC Current range is 200.0µA/2000µA and 20.00mA,/200.0mA and

then 10A range.

AC Current range is 200.0µA/2000µA and 20.00mA,/200.0mA and

then 10A range.

① Turn off power to the circuit. Set the rotary switch to the

11

 


background image

properDC/AC µA or

DC/AC mA position.

② Break the current path to be tested. Connect the red test

lead to the more positive side of the break and the black
test lead to the more negative side of the break.

③ Turn on power to the circuit. The measured value show on

the display.

5-6. DC/AC 10A measurement

① Insert the red test lead into the input terminal marked as

“10A”.

② The measuring procedure is same as that of

5-5 section..

Note:

For safety’s sake, the measuring time for high current should be

≤10 second for each measurement and the interval time between
two measurement should be greater than 5 minutes.

When current measurement has been completed, disconnect the

connection between the testing lead and the circuit under test.

5-7. Temperature measurement

[ Only for ADM02]

To measuring temperature shuld be use the

K-type

probe :P3400.

① Set the rotary switch to the “TEMP” range. In this time, The

environment temperature value displayed on the LCD.

② Insert the K-type probe to the two “COM” and “TEMP”

terminals. The two “+” and “-“ pins of temperature probe(P3400)
must be direct at the “COM” & “

”terminal respectively.

③ The measured temperature value will be displayed on the LCD.

6.

Maintenance

6-1. Replacing the battery

When meter display

the battery must be replace to

maintain normal operation.

① Disconnect and remove all test probes from any live source

and meter.

② Open the battery cover on the bottom case by screwdriver.

12

 


background image

③ Remove old battery and snap new one into battery holder

6-2. Fuse replacement

Replacing the defective fuse should the done according to the
following procedure.

① To avoid electrical shock, remove the test lead and any input

signal before opening the bottom case.

② Open the botton case and then remove the defective fuse

and insert a new fuse of the same size and rating.

③ Replace the bottom case and reinstall all the screw.

6-3.Cleaning and Decontamination

The meter can be cleaned with soft clean cloth to remove any oil,

grease or grim. Do not use liquid solvent or detergent.

13