Overview

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

01 

 

Company Branding  

Filter Condition 

Light 

Lid  

Clips 

Air Inlets 


background image

Overview

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

02 

Main Isolation 

Switch 

Mains Inlet 

Fuse Compartment 

Air Outlet/Motor 

Cooling 


background image

S

afety Instructions

 

 

 

 

Important safety notes

 

Concerning symbols used on the extraction unit and referred 
to within this manual. 

 

Danger 
 

Refers to an immediately impending danger. If the danger is 
not avoided, it could result in death or severe (crippling) 
injury. Please consult the manual when this symbol is 
displayed. 

Warning 

 
Refers to a possibly dangerous situation. If not avoided it 
could result in death or severe injury. Please consult the 
manual when this symbol is displayed. 

Caution 
 

Refers to a possibly harmful situation. If not avoided, 
damage could be caused to the product or something in its 
environment. 

  
Important (Refer to manual) 
 

Refers to handling tip and other particularly useful 
information. This does not signify a dangerous or harmful 
situation. Refer to manual when this symbol is displayed. 
 

Electrical Safety 

The V250 has been designed to meet the safety 
requirements of the Low Voltage Directive 2006/95/EC 
(previously numbered 73/23/EEC) 

Warning 

When working with the pump/motor housing open, Live 
230/115 volt mains components are accessible. Ensure that 
the rules and regulations for work on live components are 
always observed.

 

 

 

Important 

To reduce the risk of fire, electric shock or injury: 

1.

 

Always isolate the system from the mains power 
supply before removing the pump/motor access 
panel. 

2.

 

Use only as described in this manual. 

3.

 

Connect the system to a properly grounded outlet. 

 
Dangers to eyes, breathing and skin 

Once used, the filter within the V250 system may contain a 
mixture of particulates, some of which may be sub-micron 
size. When the used filters are moved it may agitate some of 
this particulate, which could get into the breathing zone and 
eyes of the operative. Additionally, depending on the 
materials being used, the particulate may be an irritant to 
the skin.  

This unit should not be used on processes with sparks of 
flammable materials or with explosive dusts and gases, 
without implementation of additional precautions. 

 

Caution: When changing used filters always wear a mask, 
safety shoes, goggles and gloves.

 

 

Carbon selection 

Please note that the media within the filter fitted in the 
V250 is capable of adsorbing a wide range of organic 
compounds. However, it is the responsibility of the user to 
ensure it is suitable for the particular application it is being 
used on. 

 

BOFA Technical Service 

If problems arises with your V250 unit please contact us:  

 

Visit our website at www.bofa.co.uk for on-line help. 

 

Or contact the helpline on 

+44 (0) 1202 699444

Mon-Fri, 9am-5pm. 
Email: Technical@bofa.co.uk 

Serial Number 

For future reference, fill in your system details in the space 
provided. The serial number is on the rating label located on 
the side/rear of the unit.  
Serial Number: 

 

 

 

 

01 


background image

S

afety Instructions

 

 

 

 

 

 

 
Warning and Information labels 

The following listing details labels used on your V250 
extraction unit. 
 

Goggles, Gloves & Mask Label 

 
 
 

 

Location: Front face of filter. 
Meaning: Goggles, Gloves and Masks should be worn while 
handling used filters. 
 

Do Not Cover Label

 

 
 
 
 

Location: Bottom panel. 
Meaning: Do not cover any louvers or holes adjacent to the 
label. 
 

Electrical Danger 
 

 
 

 

Location: Bottom Panel 
Meaning: Removal of panels with this label attached will 
allow access to potentially live components. 
 

Warning Label 

 

Location: Next to release clips. 
Meaning: Power should be isolated before the panel with 
this label attached is opened/ removed. 
 
 
 
 

Serial Number Label 

 

Location:  Next to mains inlet. 
Meaning: This label contains a variety of information about 
the extraction unit, including.   
 

 

Company name, Address & Contact number 

 

Extractor model 

 

Unit serial number 

 

Operating voltage range 

 

Maximum current load 

 

Operating frequency 

 

Year of Manufacture 

 

Relevant approval markings/ logos 
 

PLEASE NOTE: 

If the equipment is used in a manner not 

specified by the manufacturer, the protection provided by 
the equipment maybe compromised. 

Fire Risk Warning 

In the very rare event that a burning ember or spark is 

drawn into the fume extraction unit, it may be possible that 

the filters will ignite. 

Whilst any resultant fire would typically be retained within 

the fume extraction unit, the damage to the extractor 

could

 

be significant. It is therefore essential to minimise the 

possibility of this occurring by undertaking an appropriate 

Risk assessment to determine:- 

a). Whether additional fire protection equipment should be 

installed.  

b). Appropriate maintenance procedures to prevent the risk 

of build-up of debris which could potentially combust. 

 

This unit should not be used on processes where sparks 

could occur, with explosive dusts and gases, or with 

particulates which can be pyrophoric (can spontaneously 

ignite), without implementation of additional precautions 

It is essential that nozzles or other extraction/ fume capture 

devices and hoses/pipework are cleaned regularly to prevent 

the build-up of potentially ignitable debris 

02 

 

02 


background image

Before installation

 

 

 

 

 

 

 
Packaging Removal & Unit Placement

 

 
Before installation, check the extraction unit for damage. 
All packaging must be removed before the unit is connected 
to the power supply.  

Please read all instructions in this manual before using 
this extractor. 

1. Move the unit to the location where it is going to be 
installed and remove the outer packaging. 

This unit should 

be installed in a well-ventilated area.

 

Ensure that 500 mm space is available around any vented 
panels on the extractor to ensure adequate airflow. 
 

2. Check the filter is located in its correct position before 
replacing the lid and securing the clips 

 

 

 

 

 

 

 

 

 

Caution 

Do not block or cover the airflow and motor 

cooling ports on the unit, as this severely restricts airflow 
and may cause damage to the unit. 

 

Caution 

Under no circumstances should the exhaust 

outlet/s be covered as this will restrict the airflow and cause 
overheating. 

 

 

 

01 


background image

Installation 

 

 

 
Specification V250  
 

Dimensions: Height 380mm Depth 260mm Width 260mm 
Weight: 10kg   (22lbs) 
Voltage: 115-230V 
Frequency: 50/60Hz 
Full load current:      230V:1A              115V:  1.2A 
Power: 140w 
Capacity: 180m³h   (106CFM) 
 

Connection to Power Supply 

Please follow the above specification when selecting the 
power supply outlet for the extraction system, ensure the 
power supply is suitable before connecting the V250 
system. 
 
Check the Integrity of the electrical power cable, if the 
supply cord is damaged the extraction unit should not be 
connected to the mains. The supply cord should only be 
replaced by a BOFA engineer as an electrical safety test may 
be required after replacement. 
 

The extraction unit 

MUST

 be connected to a 

properly earthed outlet. 

 
 

If your extraction system was ordered with any 
optional extras please read section 

4.03

 

before 

the power connection is made as additional connections 
may be required before power is connected to the 
extractor. 

 
Connect the power cable to an isolated electrical supply. 
 
The mains socket should be installed near the extractor it 
should be easily accessible and able to be switched On/ Off. 
The cable run should be arranged so as not to create a trip 
hazard.  
 
 

 

 

 

 

 

 

 

 

01 


background image

Operation

 

 
 
 

Turning extraction unit On 

 
The V250 features a fused IEC inlet for the mains cable as 
well as a main isolation switch. The unit can be powered on 
and off by pressing the red rocker switch to the right hand 
side. 
 
 
 
 
 
 
 

 

 

 

 
 
 
 
 

 

 

 

 
 

 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Remote Speed Control Optional Extra (230V ONLY) 

 
A remote to control the motor speed can be purchased as 
part of the V250 range at the ordering stage. 
 
This allows the user to manually set the airflow from the 
V250 system. Over time as the filters get blocked it is 
advised that the motor speed be increased to ensure the 
unit maintains good airflow. 
 
To increase the airflow, press and hold the ‘up’ arrow. The 
number of glowing blue LEDs will increase as the motor 
speed is increased. 
 
To decrease the airflow, press and hold the ‘down’ arrow. As 
the motor speed slows down, the number of glowing blue 
LEDs will decrease. 
 
 
 
 

 

 
 
 
 
 

 

 
 

 
 
 

 

 

 

 

 

 

 

 
 

 

01 

Main Isolation Switch 

Mains Inlet 

Fuse Compartment 


background image

Maintenance 

 

 
 
 
 
 
 
Maintenance UK 
 

It is a legal requirement, under regulation 9 of the COSHH 
regulations that all local exhaust ventilation systems are 
thoroughly examined and tested at least once every 14 
months (typically carried out annually). The approved code 
of practice recommends that a visual check should be 
carried out at least once a week. 
 
COSHH requires the annual inspection and testing to be 
carried out by a competent person and specifies that 
documentation results are recorded in a log.  
 
Contact the seller for more information about inspection 
and certification.  
 

Maintenance General 
 

User maintenance is limited to cleaning the unit and filter 
replacement, only the manufacturers trained maintenance 
technicians are authorised to carry out component testing 
and replacement. Unauthorised work or the use of 
unauthorised replacement filters may result in a potentially 
dangerous situation and/or damage to the extractor unit 
and will invalidate the manufacturer’s warranty. 
 

Cleaning the unit 

 
The powder coat finish can be cleaned with a damp cloth 
and non-aggressive detergent, do not use an abrasive 
cleaning product as this will damage the finish. 
 
The cooling inlets and outlets should be cleaned once a year 
to prevent build-up of dust and overheating of the unit.  
 

Filter Information

 

 
A log of filter changes should be maintained by the user. 
The filters require attention when the display shows the 
configuration shown on the next page or when the extractor 

no longer removes fume efficiently. 
 
All filters are tested to BS3928. A certificate of conformity 
for each filter is available on request. 
 
It is recommended that a spare set of filters are kept on site 
to avoid prolonged unit unavailability. Part numbers for 
replacement filters can be found on the filters fitted in your 
system.  
 
To prevent overheating, units should not be run with a 
blocked filter condition, or with dust obstruction of Inlets / 
Outlets. 

 

Fire Risk Warning 

In the very rare event that a burning ember or spark is 

drawn into the fume extraction unit, it may be possible that 

the filters will ignite. 

Whilst any resultant fire would typically be retained within 

the fume extraction unit, the damage to the extractor 

could

 

be significant. 

It is therefore essential to minimise the possibility of this 

occurring by undertaking an appropriate Risk assessment to 

determine:- 

a). Whether additional fire protection equipment should be 

installed.  

b). Appropriate maintenance procedures to prevent the risk 

of build-up of debris which could potentially combust. 

 

This unit should not be used on processes where sparks 

could occur, with explosive dusts and gases, or with 

particulates which can be pyrophoric (can spontaneously 

ignite), without implementation of additional precautions 

It is essential that nozzles or other extraction/ fume capture 

devices and hoses/pipework are cleaned regularly to prevent 

the build-up of potentially ignitable debris 

 
 

 
 

 

01 


background image

Maintenance 

 

 
 
 
 
 
Filter Replacement 
 

The V250 will alert the user when its filters needs to be 
replaced. When the filter becomes full the red ‘Filter 
Indicator ‘ light next on the left side of the label will glow. 
 
 
 
 
 
 
 
 
 
 
 
 
To remove and replace the combined filter follow the 
procedure detailed below. 
 

1.

 

Isolate the electrical supply to the extractor 

2.

 

Undo the clips on the sides of the unit and remove 
the lid. 

3.

 

Lift the filter out of the unit. Once removed it is 
recommend that the used filters are bagged and 
sealed. 

4.

 

Lower the new filter into position. 
Replace the lid, and fasten the clips. 
 

To remove and replace the pre filter follow the procedure 
detailed below. 
 

1.

 

Isolate the electrical supply to the extractor 

2.

 

Undo the clips on the sides of the unit and remove 
the lid. 

3.

 

Lift the pre filter out of the combined filter. Once 
removed it is recommend that the used filter is 
bagged and sealed. 

4.

 

Lower the new filter into the top of the combined 
filter. 

5.

 

Replace the lid, and fasten the clips. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Note: The filter MUST be fitted when the extractor is in 
use. 

 

 

 

02 


background image

Replacement Parts  

 
 
 
 
 

Consumable Spares 
 

The V250 contains a pre filter and combined filter. These 
should be replaced when instructed to do so by the V250 
system (see section 

6

 for replacing the filters) 

 
To maintain performance it is important that the filters are 
replaced with identical BOFA filters. To re-order please refer 
to the Filter number printed on the filter installed in your 
extraction unit. 
 

 
 
Maintenance Protocol 

 
Users can record changes in filter change intervals on the 
table below. 
 

Unit Serial Number: 

 

Pre Filter 

Combined Filter 

Date 

Engineer 

Date 

Engineer 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Filter disposal 
 

The combined filter is manufactured from non-toxic 
materials. Filters are not re-usable, cleaning used filters is 
not recommended. The method of disposal of the used 
filters depends on the material deposited on them. 
 
For your guidance 
 

Deposit 

EWC 
Listing* 

Comment 

Non 
Hazardous 

15 02 03 

Can be disposed of as non-
hazardous waste.  

Hazardous  15 02 02M 

The type of hazard needs to 
be identified and the 
associated risks defined. The 
thresholds for these risks 
can then be compared with 
the amount of material in 
the filters to see if they fall 
into the hazardous category, 
if so, the filters will need to 
be disposed of in line with 
the local/national 
regulations. 

*European Waste Catalogue 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

01 


background image

System Specifications 

 

 
 
 
 
 

Unit:  V250  
 

Capacity: 180 m³h (106 CFM) 
Weight: 10kg (22lbs) 
Motor: Centrifugal Fan 
Output: 140w 
Electrical supply: 115/230V 
Hertz: 50/60Hz 
Full Load Current: 230v: 1A         115v:1.2A 
Noise Level: Below  55db  (at typical operating speed) 
 
Size: 

 

Metric 
(mm) 

Imperial 
(inches) 

Height 

420 

16.7 

Depth 

260 

10.2 

Width 

260 

10.2 

 

Filter Type 

Carbon 
Type 

Amount 

Carbon filter 
(combined 
filter) 

Activated 
Carbon  

 

 
Filters: 

Filter Type 

Construction 

Efficiency 

Pre Filter 

Pad 

F7 (96% @ 2 microns

)

 

Combined 

Filter 

Maxi Pleat Construction 

with 

Webbing Spacers

 

99.997% @ 0.3 

microns

 

 
Environmental operating range: 
 
Temperature: +5

o

C to + 40

o

Humidity: Max 80% RH up to 31