PDF User Manual

  1. Home
  2. Manuals
  3. AAB ABM-200 User Manual

AAB ABM-200 User Manual

Made by: AAB
Type: User Guide
Category: Measuring Instruments
Pages: 12
Size: 0.57 MB

 

Download PDF User Manual


Related Product Video

 

Full Text Searchable PDF User Manual



background image

 

 

 

 

 

 

 

 

 

ABM-200 

Airflow & Environmental Meter 

User Guide 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

Introduction to the ABM-200 

 

 

Table of Contents

 

(by page number)

 

Overview 

First connection 

Settings 

LED Indicators 

General Use 

Sending and Editing Reports 

Maintenance (Change Battery/Cleaning/Storing) 

10  Troubleshooting 
11  Technical Data 
11  Range & Accuracy 
12  Glossary 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

Overview 

 

 

 

 
 
 
 

 

 

 

Airspeed 

Temp/Humidity  

                Sensor 

Power Button 

¼-20 Threaded Port 

Battery Door 

 


background image

 

First Connection 

Step One: On your smartphone, go to Apple App Store 

  or Google Play 

 and 

download the AAB Smart ABM-200 app marked with a blue icon 

.  

When installing the app, your smart device will ask for you to approve certain 
permissions such as access to your contacts. This is necessary in order to email 
reports directly from the app. It will ask for access to your location, this is for 
geotagging your location. It will ask for access to your photos, this gives you the ability 
to add your company logo to reports. None of the items granted for access is used for 
any purpose other than your use of the meter. AAB does not gain access to, receive 
or retrieve any of your personal data or information. 

Step Two: In preparing your new ABM-200 for use, remove the white plastic 
tab/battery insulator. To do so, simply pull the white and green tab in the direction of 
the arrow. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

Step Three: Connecting the ABM-200 to the app.  

Start the app by selecting ABM-200 blue icon 

 in the app list. This will bring you to 

the start screen. 

 

On the ABM-200, press the power button marked with the AAB logo. The app screen 
will change to show the device is connected and ready to take readings. 

 

 

The ABM-200 is now paired with the smartphone and ready for use. 

 

Turning the app off: 

For android, press the phone’s back button twice. For Apple,  

 

Turning the ABM-200 off: Hold the power button 3 seconds. The ABM-200 will also 
shut off automatically after 30 seconds, when not connected to the app.  

 

 

 

 

 

 


background image

 

 

Settings 

Company Name: Used to display your company name on reports 

Measurement units: Allows the selection of Metric or Standard units depending on 
the regional and user needs. 

Send History To: This field can be used to relay the history log to an email address of 
your choosing.  

Manual Free Space Adjust (AK Factor): This setting is used to offset the amount of 
free area left by a grille or similar obstruction to the duct. Consult the manufacturer of 
the grille for the amount of area obstructed or open. 

 

LED Indicators 

The AAB logo is illuminated by an LED for confirmation that the unit is operational. 
The LED will flash when the button is pressed until connection. Once connected, the 
LED will stop flashing. The phone will inform the user of the connection to the ABM-
200.  

When shutting off the ABM-200 with the power button, the LED will fast flash for 1 
second to indicate shutdown. 

 

General Use 

The ABM-200 was designed for taking air speed, temperature, and humidity in order 
to aid the contractor in effectively balancing commercial and residential HVAC 
systems.  

 

 

 

 

 

 

 

 

 

 

 


background image

 

Sending and Editing Reports 

Reports can be accessed and sent with or without the phone connected to an ABM-

200. To access saved reports, press the three bar menu button 

 in the upper left 

hand corner of the app. In the dropdown menu, choose History. All of the reports 
saved are displayed on this page.  

Viewing Report Data: In order to see the data in a saved report, tap the report and 
a Report screen will load. Press the arrow to return to the History screen. 

Export Report: To export a report, tap Export in the lower left corner of the History 
screen. For Android, a “Complete action using” pop up menu will appear if multiple 
programs are available. Choose the program you would like to use and it will add the 
report in HTML format. For Apple, ………… 

Deleting Single Report: Select Edit in the top right corner of the History screen. 

This will place a Delete icon 

 next to each report. Choose the Delete icon to 

delete the associated report. When finished, select Done in the upper right corner of 
the screen. 

Delete All: To delete all of the stored reports, tap Delete All in the lower left corner 
of the history screen. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

 

Maintenance  

Changing the battery 

Step one: Pull lip of blue sleeve out of fan groove. This may require the use of a small 
flathead screwdriver.  

 

Step two: Remove sleeve by pulling it over the top of the ABM-200 and off of the 
bottom. 

 

 

Step three: On the back side of the ABM-200, remove the battery door by pulling it 
downward. 

 

Step four: Replace battery with a CR2450 coin cell. The positive (+) faces away from 
the circuit board. 

Step Five: Assemble using steps one through three in reverse. When placing the 
sleeve on the ABM-200, align AAB logo with white button (as shown in picture X).  

 

 

 


background image

 

Cleaning 

To clean the ABM-200, start by removing the blue sleeve (see Changing the Battery). 
Use a damp rag with mild detergent and wipe down exterior. Do not clean with the 
sleeve on the ABM-200, as water could enter the sensor port, causing damage and 
voiding the warranty. 

 

Storage 

The ABM-200 is to be stored out of direct sunlight in temperatures between 32° and 
95° F (0° and 35° C).  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

10 

 

Troubleshooting 

PROBLEM 

POSSIBLE CAUSE 

CORRECTION 

ABM-200 not found 

Low/dead battery 

Replace battery 

Bluetooth off  

Turn on phone’s Bluetooth 

Pairing not performed 

through app 

Un-pair in OS and pair 

through app 

Incorrect air speed reading 

Low battery 

Replace battery 

Incorrect AK % 

Correct AK adjust 

percentage 

Meter not 

perpendicular to airflow 

Adjust fan directly into 

airflow 

Obstruction in front of 

meter 

Remove obstruction 

Move meter to side 

obstruction 

Slow temperature/humidity 

response 

Low air flow 

Move into air stream 

Ports blocked 

Remove obstruction from 

front and/or rear ports 

Incorrect 

temperature/humidity reading 

Low Battery 

Replace battery 

Meter acclimating 

Give meter time to adjust to 

current conditions 

Ports blocked 

Remove obstruction from 

front and/or rear ports 

Direct sunlight 

Move from direct sunlight 

Fan popped out 

Impact to fan area 

Align notch and groove on 

circumference. Press back 

in.  

Screw port doesn’t work 

Wrong thread size/type 

Ensure accessory has a  

¼-20 threaded screw 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

11 

 

Technical Data 

Operational temperature range: -40°F to +185°F (-40°C to +85°C) 

Operational temperature range, full accuracy: -32°F to +149°F (0°C to +65°C) 

Power: Panasonic 3.6V, CR 2450 Coin Cell Lithium Battery 

Battery Life: >500 hours 

Auto power: Shuts off 30 seconds after closing or minimizing app or no connection 

Units of Measure: inWC, inHg, psi, mmWC, mmHg, Pa, kPa, mBar, fpm, cfm, l/s, 
m/s, F, C, RH%, lb/ft3 

 

Range & Accuracy 

Airflow Velocity: 44 

– 12,320 fpm ± 0.5% 

Temperature: -40 to 185°F ± 0.5°F 

Humidity: 0-95% ± 3% 

Pressure: 300-1100 mbar ± 0.12 mbar 

 

*Please note that we are often asked the maximum volume of airflow, this is totally 
dependent on the size of the duct. For example: 12,320 fpm through a 20” x 24” grille is 
41,028 CFM. So the range of this meter goes high enough to satisfy any practical HVAC 
application. 

 

 

 

 

 

 

 

 

 

 

 

 

 


background image

12 

 

Glossary 

Airflow Velocity: speed in which the air is being moved into or out of the duct.  

Airflow Volume: amount of air being moved into or out of a duct in a given amount of 
time. 

Humidity: a quantity representing the amount of water vapor present in the air. 

AK-factor (K-factor): amount of space blocked by a grill or other duct cover. Typically 
supplied by the manufacturer.  

Relative Humidity: the amount of water vapor present in air expressed as a 
percentage of the amount needed for saturation at the same temperature. 

Temperature: the degree of measured heat in an environment. 

Temperature Differential: difference in temperature between two spaces in a 
building, or between indoor and outdoor temperature. 

Wet Bulb: Temperature acquired by use of a thermometer with the bulb enclosed in 
wetted material.