ENGLISH

Fast Curing Cordless LED Light

Instructions For Use

Ortholux

 Luminous Curing Light

 


background image

 

English 

2

ENGLISH

ENGLISH

Ortholux

 Luminous Curing Light

Fast-Curing Cordless LED Light

Table of Contents 

Page

Safety   

2

Glossary of Symbols   

3

Product Description   

4

Fields of Application   

4

Technical Data   

4

Charger   

4

Handpiece   

5

Charger and Handpiece   

5

Transport and Storage Conditions   

5

Installation of the Unit   

5

Factory Settings   

5

Initial Steps   

5

  Charger   

5

  Light Guide/Handpiece   

5

Removing and Inserting the Light Guide 

  from/into the Handpiece   

5

  Inserting the Battery   

6

  Battery Charging   

6

Operating Status Display of the Charger   

6

Power Level Display of the Handpiece 

6

Operation   

7

Activating the Handpiece 

7

Selection of Exposure Time 

7

Activating and Deactivating the Light Emission  7

Tack Cure Function  

7

Recommended Cure Techniques with 

3M Unitek Adhesives 

7-8 

Sleep Mode   

8

Measurement of Light Intensity   

8

Acoustical Signals - Handpiece   

8

Troubleshooting   

9-10

Maintenance and Care   

10

Inserting/Removing the Battery   

10

Handpiece/Battery Care   

10

Cleaning the Light Guide   

10

Cleaning the Charger, Handpiece, and 

  Eye Shield   

10

Storage of the Handpiece during Extended 

  Periods of Non-Use   

10

Disposal   

10 

Customer Information   

11

Warranty   

11

Limitation of Liability   

11

3M Unitek Service Centers 

12

Safety

PLEASE NOTE! 

Prior to installation and start-up of the unit, please read 

these instructions carefully!
As with all electrical devices, the proper function and safe 

operation of this unit depend on the user’s compliance with 

the standard safety procedures as well as the specific safety 

recommendations presented in these Instructions for Use.
1.   The unit must be used in strict accordance with the 

following Instructions for Use. The manufacturer 

assumes no liability for any damage resulting from 

the use of this unit for any other purpose.

2.  Prior to start-up of the unit, ensure that the operating 

voltage stated on the rating plate is compatible with 

the available main voltage. Operation of the unit at a 

different voltage may damage the unit.

3.  Position the unit so that the power plug is accessible at 

all times. The power plug is used to turn the unit on and 

off. To disconnect the charger from the main voltage, 

remove the power plug from the electrical outlet.

4.  Use supplied 3M Unitek charger. The use of any 

other charger can result in damage to the battery.

5.  The Ortholux

 Luminous Curing Light must be aimed 

away from the eyes in order to avoid serious health 

consequences due to irradiation of the eyes. 

Use the 

eye shield. Exposure must be restricted to the area of 

the oral cavity in which clinical treatment is intended.

6.  

CAUTION! In cases where the Ortholux Luminous 

Curing Light will be used outside of the oral cavity 

additional eye protection may be needed. The 

reflected light from curing surfaces when not 

confined to the oral cavity may be more intense than 

intra-orally. It is recommended that protective orange 

glasses be worn by everyone in close proximity of 

the extraoral curing in addition to the use of the eye-

shield on the light.

7.  

CAUTION! The Ortholux Luminous Curing Light 

generates high intensity light. High intensity light 

always involves the production of heat. The light 

emitted should be aimed directly above the material 

to be cured - exposure of the soft tissues (gingiva, 

oral mucosa, and skin) to high-intensity light should 

be avoided as such exposure may cause damage 

or irritation. If applicable, cover such areas. If 

exposure of soft tissues cannot be avoided, adjust 

the polymerization process to the light level, e.g., by 

shortening the polymerization times or increasing the 

distance between the light guide exit and the material 

to be cured. When utilizing the light in Extended 

Mode (15s), do not keep the light in one position. 

Longer exposure in the region of the pulp can lead 

to injury or irritation of the pulp, which is why the 

specified exposure times must not be exceeded. 

 


background image

English

8.  Ortholux Luminous Curing Light may be operated 

only with the supplied light guide or 3M Unitek 

replacement light guide. The use of other light 

guides may result in a reduction or increase in the 

light intensity. The product’s warranty does not 

cover any damage resulting from the use of third-

party light guides.

9.   Condensation resulting from the unit being 

transferred from a cold to a warm environment may 

be a potential risk. Hence, the unit should be turned 

on only after it is completely equilibrated to ambient 

temperature.

10. In order to avoid electric shock do not introduce 

any objects into the unit with the exception of 

replacement parts handled in accordance with the 

Instructions for Use.

11. Use only genuine 3M Unitek parts when replacing 

defective components as directed in these 

Instructions for Use. The product’s warranty does 

not cover any damage resulting from the use of 

third-party replacement parts. 

12. Should you have any reason to suspect the safety 

of the unit to be compromised, the unit must be 

taken out of operation and labeled accordingly 

to prevent third parties from inadvertently 

using a possibly defective unit. Safety may be 

compromised, e.g., if the unit malfunctions or is 

noticeably damaged.

13. Keep solvents, flammable liquids, and sources 

of intense heat away from the unit as they may 

damage the plastic housing of the charger, the 

seals, or the cover on the operating buttons.

14. Do not operate the unit in the proximity of 

flammable mixtures.

15. Do not allow cleaning agents to enter the unit as 

this may short-circuit the unit or cause potentially 

dangerous malfunction.

16. Only service centers authorized by 3M Unitek may 

open the unit housing and repair the device.

17. Ortholux Luminous Curing Light must not be used 

for patients, or by users, with heart pacemaker 

implants who have been advised to be cautious with 

regard to their exposure to small electrical devices.

18. Do not use Ortholux Luminous Curing Light 

for patients with a history of photobiological 

reactions (including individuals with urticaria 

solaris or erythropoetic protoporphyria) or who are 

currently on photosensitizing medication (including 

8-methoxypsoralen or dimethylchlorotetracycline).

19. Individuals with a history of cataract surgery may 

be particularly sensitive to the exposure to light 

and should be discouraged from exposure to the 

Ortholux Luminous Curing Light unless adequate 

safety measures, such as the use of protective 

goggles to filter blue light, are undertaken.

20. Individuals with a history of retinal disease should 

seek advice from their ophthalmologist prior 

to operating the unit. In operating the Ortholux 

Luminous Curing Light, this group of individuals 

must take extreme care and comply with any and 

all safety precautions (including the use of suitable 

light-filtering safety goggles).

21. This unit has been developed and tested in 

accordance with the relevant EMC regulations 

and standards. It is in conformity with legal 

requirements. Since various factors such as power 

supply, wiring, and the ambient conditions at the 

place of operation can affect the EMC properties 

of the unit, the possibility that, under unfavorable 

conditions, there will be EMC disruptions cannot be 

completely excluded. If you should notice problems 

in the operation of this or other units, move the unit 

to a different location.

22. Prior to each use of the unit ensure that the emitted 

light intensity is sufficient to safely guarantee 

polymerization. Use the light testing area integrated 

into the charger for this purpose.

Glossary of Symbols

Attention, Consult Accompanying 

Documents

Type B Equipment -  

Protection against electric shock

Class II Equipment - double insulated

Establishes that the product conforms to all 

applicable legal requirements

NRTL approval applies only to  

USA version PN 704-460

Battery power level

Icon to identify electric and electronic 

devices. The unit must be collected and 

disposed of separately.

Temperature limitation

Humidity limitation

Atmospheric pressure limitation

 


background image

 

English 

4

ENGLISH

Product Description

Ortholux Luminous Curing Light, is a high-performance 

light source for intraoral polymerization of orthodontic 

adhesives. It consists of a charger and a cordless 

handpiece powered by a rechargeable battery.  

The charger is designed for use on a table and cannot  

be wall-mounted.
The light source is a high-performance light-emitting 

diode (LED). The emitted light covers the light 

wavelengths between 430 and 480 nm.
Settable exposure times:
• 3, 6, 9, or 12 seconds.
• Extended mode (15 seconds)
• Tack cure mode (1 second)
The charger is equipped with an integrated light 

intensity testing area.
The unit is shipped with an 8 mm diameter light guide. 

It is not permissible to use light guides from other 

manufacturers.
The handpiece is equipped with a sleep mode to minimize 

the unit’s energy consumption. The handpiece switches 

to sleep mode once it is placed in the charger or if left 

unused for approximately 5 minutes outside the charger.
NOTE: Do not discard these Instructions for the 

duration of product use.

Fields of Application

•   Polymerization of light-curing orthodontic  

adhesives with photoinitiators in the  

wavelength range of 430-480 nm.

  The high intensity light output of the Ortholux 

Luminous Curing Light allows for faster curing 

times than other available curing lights. Refer to 

the table under Recommended Cure Techniques 

with 3M Unitek Adhesives.

-    Though the majority of light-curing orthodontic 

adhesives are responsive in this range of 

wavelengths, you may wish to contact the 

manufacturer of the material in question.

Technical Data
Charger

Operating voltage:  

100 - 127 V 50/60 Hz  

 

230 V 50/60 Hz (see rating  

 

plate for factory-set voltage)

Power input:  

Nominal consumption  

 

0.045 A (230 V)  

 

0.1 A (100-127 V)

Dimensions: 

Length:  170 mm   

 

Width:   95 mm 

 

Height:  50 mm  

 

Weight:  650 g

Light guide

Handpiece

Charger

Power cord

Eyeshield

 


background image

English

Classification: Protection class II, 

Handpiece

Power supply:  

lithium-ion battery, nominal    

 

voltage 3.7 V

Utilizable wavelength  

range:  

430-480 nm

Wavelength peak: 

455 nm +/- 10 nm

Light intensity (between  

400 and 515 nm):  

1600 mW/cm

2

 (independent    

 

of battery power level)

Light emission area: 

typically 42 mm

2

  

 

(optically active)

Intermittent  

operation:  

1 minute on, 15 minutes off 

 

(ambient temperature of 40°C) 

 

typically 7 minutes operating time  

 

at ambient temperature (23°C)

Total exposure time 

with new, fully 

charged battery:  

typically 60 minutes

Dimensions: 

Diameter: 28 mm 

 

Length:  270 mm

Weight:  

250 g (including light guide)

Charger and Handpiece

Time to charge 

empty battery:  

approximately 2.0 hours

Operating temperature:   10°C to 40°C (59°F to 104°F)
Relative humidity:  

30% to 75%

Atmospheric pressure:   700 hPa to 1060 hPa
Total height with 

handpiece inserted in 

the charger:  

180 mm

Transport and Storage Conditions:

Ambient temperature  

range:  

-20°C to +40°C / 

 

-4°F to +104°F

Relative humidity:  

30% to 75%

Atmospheric  

pressure:  

700 hPa to 1060 hPa

Subject to technical modification without prior notice.

Installation of the Unit
Factory Settings

The factory settings of the unit are as follows:
•  When first turned on the unit is in Extended Mode, 

i.e. only the Power level LED is lit on the hand piece.

Initial Steps

Charger

•  Please ensure first that the voltage stated on the 

rating plate corresponds to the existing main supply 

voltage. The rating plate is attached to the bottom 

of the charger.

•  Place the charger on a level surface.
•  Connect the power cable of the charger to the 

power supply. 

-    The green LED on the charger showing that the unit 

is ready for operation comes on. This shows that 

the unit is ready for operation; please refer to the 

section ”Operating Status Display of the Charger.”

Light Guide/Handpiece

•  Never place the handpiece in the charger without 

the battery inserted in the handpiece!

•  Place the eye shield on the front of the unit.
•  Autoclave the light guide prior to first use.
•  Insert the light guide firmly into the handpiece.

Removing and Inserting the Light Guide 

from/into the Handpiece

•  The light guide has a magnetic holder. Remove the 

light guide from the handpiece by pulling it towards 

the front.

•  Attach the light guide to the handpiece until it is 

firmly in place.

 


background image

 

English 

6

ENGLISH

Insert the battery: 

• 

Remove the protective cap from the battery 

and keep with the unit packaging.

•  Slowly insert the battery with the thread side 

towards the metal housing into the handpiece 

until it stops. Screw the battery clockwise by hand 

until the seal is pressed firmly against the metal 

housing. The unit will not function properly if the 

battery is not screwed in completely!

• Iftheunitdoesnotfunctionproperly,remove

the battery and reinsert it as described above.

Power Level Display of the Handpiece

Battery Charging

•  The unit contains a powerful lithium-ion 

rechargeable battery. The battery does not have any 

memory effect and can therefore be recharged at 

any time by placing it in the charger (see “Power 

Level Display of the Handpiece” below).

•  Alternatively, the battery in the unit can be replaced 

by a spare battery (available separately) which has 

been charged. 

•  Place the handpiece on the charger for approximately 

1.5 hours before the first use, in order to charge the 

new battery completely for the first time.

-   The status LED showing the operating status of 

the charger flashes green during the charging 

operation; see below “Operating Status Display  

of the Charger.”

Operating Status Display of the Charger

Status LED

Steady green light

Flashes green

Steady red light

Flashes alternately red and green

Handpiece/battery outside the charger

Charger ready for operation

n/a

Charge contact pins are wet

Malfunction in the charger

Handpiece/battery in the charger

Charging has been completed

Battery is being charged

Charge contact pins are wet

Problem during charging

Status LED showing 

operating status

Power Level LED

Operating status

Power Level LED

Steady green light

Steady red light

Flashes red

Handpiece outside the charger

Battery is charged, handpiece ready for operation

Warning for low battery capacity, remaining capacity 

adequate typically for 20 3-second cycles 

Battery fully discharged, exposure cycle will be 

completed or, if in extended mode, stopped

Handpiece in the charger

Not possible, handpiece is in sleep mode

Not possible, handpiece is in sleep mode

Problem in charging, battery is defective 

or cannot be charged

Operating status

Seal

 


background image

English

•  For Extended Mode, an acoustical signal is emitted 

every 3 seconds for a total of 15 seconds.

•  If desired, press the START button to stop the light 

emission before the set time is over.

Tack Cure Function

Hold down the START button to activate the Tack Cure 

function regardless of the set Exposure Time. The unit 

emits a single one-second light pulse which allows for 

a short exposure to tack cure brackets. The light will 

return to standard operation when the START button is 

no longer depressed.

Recommended Cure Techniques with  

3M Unitek Adhesives

•    Rotate the light guide into the desired position  

for polymerization.

•     For most 3M Unitek adhesives, reference the table 

below for recommended cure technique. 

•  For other 3M Unitek adhesives not specifically 

mentioned, please refer to the adhesive Instructions 

for Use.

•   If using other light cure adhesive systems, consult 

the adhesive manufacturer’s instructions.

•   The light guide tip should be placed approximately  

1-2 mm above the appliance and pointed 

perpendicular towards the tooth surface.

Operation

Activating the Handpiece

If the Power Level LED is off, briefly press the START 

button to activate the hand piece. The Power Level LED 

will come on and a short audible signal is emitted. The 

hand piece is now ready for operation; the handpiece 

displays the latest set exposure time setting.

Selection of Exposure Time

Press the TIME button to cycle through the following 

settings from left to right. Press the TIME button again 

to go back to 3s settings from the Extended Mode.

Caution! High intensity light always involves the 

production of heat. When utilizing the light in  

Extended Mode (15s), do not keep the light in  

one position. Longer exposure in the region of the pulp 

can lead to injury or irritation of the pulp, which is why 

the specified exposure times must not be exceeded.

Activating and Deactivating the  

Light Emission

Briefly press the START button to activate the light 

emission for the Exposure Time set as described above. 
•  Once the emission has begun, the active Exposure 

Time LEDs will count down every 3s with an acoustical 

signal indicating the elapsed time, except for the 

Extended Mode. When the emission is completed, the 

Exposure Time LEDs will return as set before. 

•  Example with 12s Exposure Time setting:

START button

Power Level LED

TIME button

LEDs showing 

exposure times

3s

6s

9s

12s

Extended  

Mode (15s)

Beginning 

of Emission

3 sec 

Elapsed

6 sec 

Elapsed

9 sec 

Elapsed

End of 

Emission

ON Signal

1 beep

2 beeps

3 beeps OFF Signal

Appliance

Adhesive

Ortholux

 

Luminous 

Curing Light 

Metal  

Brackets

Transbond

 XT Adhesive,  

Transbond

 PLUS Color 

Change Adhesive,  

APC

 II Adhesive  

Coated Brackets,  

APC

 PLUS Adhesive 

Coated Brackets

3 seconds mesial x 

3 seconds distal

Ceramic  

Brackets

3 seconds through  

the bracket

Buccal Tubes  

(direct bond)

6

 

seconds mesial x 

6 seconds occlusal

Lingual  

Retainers

Transbond

 LR Adhesive

3 seconds mesial x 

3 seconds distal

Molar  

Bands

Transbond

 Plus  

Band Adhesive

12 seconds  

(3 seconds  

per cusp)

Unitek

 Multi-Cure Glass 

Ionomer Band Cement

Indirect Trays  

(Metal Brackets)

Transbond

 Supreme  

LV Adhesive

6 seconds mesial 

x 6 seconds distal 

through the tray

Indirect Trays  

(Ceramic 

Brackets)

6 seconds through  

the bracket  

and tray

 


background image

 

English 

8

ENGLISH

•   The recommendations are applicable with 100% 

light output intensity, measured at the intensity test 

area on the charging base.

•  A technique tip for faster light curing of metal 

brackets is to position the light guide interproximal 

to two brackets. In order for the bracket to fully 

cure, both sides of the bracket must be exposed to 

the light.

Keep the light guide clean at all times to 

obtain full light intensity. Cured primer/

adhesive on the light guide will reduce the 

intensity resulting in incomplete cure.

Damaged light guides substantially reduce 

the light intensity and must be replaced 

immediately,sharpedgesmaycause

serious injury!

NOTE: Holding the light away from the tooth, moving 

the light guide and turning the light guide tip at an 

angle may cause incomplete cure of the material.

Sleep Mode

Once the handpiece is placed in the charger, all 

internal functions and LEDs are automatically turned 

off as the handpiece switches to sleep mode. This 

reduces the power consumption of the battery to a 

minimal level. Outside the charger, the handpiece 

is also switched to sleep mode if it is not used 

for approximately 5 minutes. See “Activating the 

Handpiece” to wake up the unit.

Measurement of Light Intensity

The light intensity of the Ortholux Luminous Curing 

Light handpiece can only be reliably determined 

using the light intensity testing area on the Ortholux 

Luminous Curing Light charger! The circular testing 

area is located below the Status LED on the charger. 

Measurements carried out using other units will 

produce incorrect results due to the different 

light sources and the varying arrangement of the 

components of the units.

NOTE: The light intensity is independent of the 

battery charge level.
Caution!
•  Clean the testing area with a damp cloth.
•  Without applying pressure, place the light guide 

on the testing area such that the light-emitting 

opening of the light guide is flush with the 

testing area.

•  Activate the lamp by pressing the START button.

-  The number of illuminated LEDs on the 

charger is indicative of the measured light 

intensity: 5 LEDs = 100%, 4 LEDs = 90%, 3 

LEDs = 80%, 2 LEDs = 70%, 1 LED = 60%.

• If the light intensity is below 80% (fewer 

than3LEDsareON),checkthelightguide

and handpiece for contamination or defects.

•  Remove and clean the light guide of any 

contamination. Replace a defective light guide 

with a new light guide. If the measures described 

above do not result in any improvement, call  

3M Unitek Customer Service.

Acoustical Signals - Handpiece

An acoustical signal is emitted
•  every time a button is pressed
•  every time the light is turned ON or OFF
•  1 time after 3 seconds exposure time, 2 times 

after 6 seconds, 3 times after 9 seconds 

Exception: in Extended Mode; an acoustical 

signal is emitted every 3 seconds

•  every time the sleep mode is terminated by 

pressing the START button.

A 2 second-error signal is emitted, if
•  the handpiece overheats
•  the battery lacks sufficient charge

The acoustical signals from the handpiece can  

be turned off (except for the 2-second error signal). 

Put the handpiece in sleep, e.g., by setting it in the 

charger. Take the device from the charger; press 

the TIME button, then press the START button. This 

changes the sleep mode of the handpiece back to 

active and the status “Acoustical signals activated” 

to “Acoustical signals deactivated”. The acoustical 

signals can be reactivated by following the  

same procedure. 

Testing area

LEDs showing 

light intensity

 


background image

English

Error

Cause

•  Solution

The LED showing the 

operating status in the charger 

flashes alternately red and 

green. The handpiece is in the 

charger.

Malfunction during charging.
•  Remove and re-insert the 

battery to reset the unit.

•  The battery is damaged.
•  Replace the battery.

The LED showing the 

operating status in the 

charger does not come  

on although the plug has 

been connected to an 

electrical outlet.

Power outlet carries no 

voltage.
•  Use a different power 

outlet. 

Charger is defective.
•  Have the charger 

repaired.

The light intensity is too low.

•  Clean the light guide 

and the protecting 

glass in the light guide 

mounting hole (please 

refer to “Cleaning the 

Light Guide”).

The LED showing the 

operating status in the charger 

flashes alternately red and 

green. The handpiece is not in 

the charger.

Charger is defective.
•  Unplug and plug in the 

charger to reset.

•  Have the charger 

repaired.

During light emission in 

extended mode, an error 

signal sounds for 2 seconds,  

the emission is stopped,  

and the handpiece changes 

to sleep mode.

The handpiece has become 

overheated in the course 

of use. The handpiece can 

be used again once it has 

cooled down.
•  Allow the handpiece to 

cool down.

-   Immediately after the 

temperature control 

has turned off the 

handpiece, it must cool 

down a minimum of 1.5 

minutes in the air before 

a new light emission can 

be started.

-   The handpiece can 

be used again, once 

pressing the START 

button successfully 

activates the light.

-   During the cooling-off 

period immediately after 

the acoustical signal 

has gone off, 2 minute 

cooling-off time in the air 

typically corresponds to 

1 minute operating time.

The light emission does not 

start when the START button 

is pressed; an error signal 

sounds for 2 seconds.

The handpiece has become 

overheated in the course 

of use. The handpiece can 

be used again once it has 

cooled down briefly.
•  Allow the handpiece to 

cool down.

-   The handpiece can 

be used again, once 

pressing the START 

button successfully 

activates the light.

-   During the cooling-off 

period immediately after 

the acoustical signal 

has gone off, 2 minute 

cooling-off time in the air 

typically corresponds to 

1 minute operating time. 

The power level display of 

the handpiece flashes red 

while the handpiece is in  

the charger.

Charging problem. The 

battery is defective or at the 

end of its useful life.
•  Remove and re-insert the 

battery to reset the unit.

•  Replace the battery.

The handpiece has not been 

used for a long time and 

now it cannot be turned on.

There is not enough charge 

in the battery to turn on the 

handpiece.
•  Place the handpiece  

in the charger and  

re-charge the battery.

The power level display of 

the handpiece flashes red.
The ongoing exposure is 

interrupted (light off signal 

is emitted) followed by 

a 2 second-error signal; 

the handpiece switches to 

“sleep” mode and resists 

further activation.

The battery lacks sufficient 

charge.
•  Place the handpiece  

in the charger and  

re-charge the battery.

The charge level display  

on the handpiece glows  

red steadily. 

The remaining battery 

charge is adequate only 

for typically 20 3-second 

exposure cycles.
•  Place the handpiece  

in the charger and  

re-charge the battery.

Error

Cause

•  Solution

Troubleshooting

 


background image

 

English  10

ENGLISH

sterilization, the spots of dried liquid should be 

wiped off the ends of the device.

•  Adhering polymerized material should be removed 

with alcohol, a plastic spatula may help in removing 

the material.

-  Do not use any sharp or pointed tools for 

cleaning in order to protect the surface of the 

unit from scratching.

•  The protective glass of the handpiece can be 

cleaned with a soft and fluff-free cloth.

CleaningtheCharger,Handpiece,and 

Eye Shield

•  To disinfect all components, spray the disinfectant 

on a towel and use it to disinfect the unit. 

Do 

not spray the disinfectant directly on the 

handpiece or the charger.

-  Disinfection agents must not enter the unit!

•  Dry residual disinfectants on the charger, the 

handpiece and the eye shield with a soft and fluff-

free cloth, as they damage the plastic components.

•  Clean the charger, the handpiece and the eye 

shield with a soft cloth and, if required, a mild 

cleaning agent.

-  Solvents or abrasive cleaners may not be used in 

any case, as they damage the plastic components!

-  Cleaning agents must not enter the unit!
-  Make sure that charge contact pins remain dry 

and are not contacted by metallic or greasy 

parts. Do not bend the charge contact pins 

during drying. Wet charge contact pins will 

cause an operating error (malfunction message: 

the LED showing the operating status of the 

charger flashes red).

Storage of the Handpiece during  

Extended Periods of Non-Use

•  If the handpiece is not to be used for an extended 

period of time (e.g., during vacation) fully 

charge the battery prior to departure or keep the 

handpiece inserted in the operational charger.  

A safety switch within the battery prevents a  

total discharge.

•  Discharged or nearly discharged batteries must be 

recharged as soon as possible.

Disposal

Your new unit contains a powerful lithium-ion 

rechargeable battery. Dispose of defective batteries  

and units in accordance with local legal regulations!

Maintenance and Care

The Ortholux Luminous Curing Light is maintenance-

free with the exception of cleaning the light guide and 

replacing batteries as needed. The light output should 

be checked on a regular basis, i.e. daily, using the 

integrated light intensity testing area on the charging 

base. See the information contained in this chapter 

to secure problem-free operation. There are no user 

serviceable parts on the unit. If the unit malfunctions 

and the trouble shooting steps above do not remedy the 

situation, call 3M Unitek Customer Service.

Inserting/Removing the Battery 

Never place the handpiece in the charger without the 

battery inserted in the handpiece!
Use 3M Unitek batteries only! The use of other 

manufacturer batteries or non-rechargeable batteries/

primary batteries is a potential hazard and may 

damage the unit.
•  Remove the battery from the handpiece by turning 

it counter-clockwise as shown. (Page 6)

•  Insert the (new) battery into the handpiece and turn 

it clockwise until the seal is firmly pressed against 

the metal housing.

•  Place the handpiece in the charger for 1.5 hours in 

order to fully charge the new battery for the first time.

-  The LED showing the operating status on the 

charger flashes green; see also “Operating 

Status Display of the Charger.”

Handpiece/Battery Care

•  Use only the 3M Unitek charger which is provided 

with the unit; failure to do so could lead to damage 

to the battery!

•  Do not immerse the battery in water or incinerate! 

Please also observe the chapter on ”Safety.”

Cleaning the Light Guide

The light guide can be autoclaved. 
Do not sterilize by chemical means or hot-air!
•  The light guide should be regularly wiped clean 

with a soft cloth. Especially before and after steam 

Error

Cause

•  Solution

The LED showing the 

operating status in the 

charger glows red steadily.

The charge contact pins in 

the handpiece or the charger 

are wet.
•  Dry the charge contact 

pins. Be careful not to 

bend the flexible  

charge contact pins in 

the charger.

 


background image

11  English

Customer Information

No person is authorized to provide any information 

that deviates from the information provided in this 

instruction sheet.

Warranty

3M Unitek warrants this product will be free from 

defects in material and manufacture for a period of 

two years from date of purchase. 3M UNITEK MAKES 

NO OTHER WARRANTIES INCLUDING ANY IMPLIED 

WARRANTY OF MERCHANTABILITY OR FITNESS 

FOR A PARTICULAR PURPOSE. User is responsible 

for determining the suitability of the product for 

user’s application. If this product is defective within 

the warranty period, your exclusive remedy and 3M 

Unitek’s sole obligation shall be repair or replacement 

of the 3M Unitek product.

Limitation of Liability

Except where prohibited by law, 3M Unitek will not 

be liable for any loss or damage arising from this 

product, whether direct, indirect, special, incidental 

or consequential, regardless of the theory asserted, 

including warranty, contract, negligence or strict 

liability.

Information valid as of March 2009

 


background image

© 2009 3M.  All rights reserved.  
REF 011-635  0903

3

3M Unitek

Orthodontic Products
2724 South Peck Road

Monrovia, CA 91016 USA

European Representative:
3M Unitek
Ohmstrasse 3

86899 Landsberg, Germany

+49/(0)8191/ 9474-0

Denta n.v. 

Eddy Dirk 

Van den Eynde Backsteen

Heiveldekens 2 

B – 2550  

Kontich

Belguim 

Tel.: 0032-3-450 93 20 

Fax: 0032-3-457 85 73 
3M Santé 

Patrick Lamotte 

Avenue du 11 Novembre

F - 45312 

Pithiviers Cedex

France

Tel.: 0033-2-38324483 

Fax: 0033-2-38324482
J & S DavisDavis Healthcare 

Services Ltd. 

Pete Harris 

Summit House Summit Road 

GB - Potters Bar, Hertfordshire  

EN6 3EF   

Great Britain 

Tel.: 0044-1707-64 63 30 

Fax: 0044-1707-64 64 29
3M ESPE AG 

Technical Service 

ESPE Platz 

D-82229 

Seefeld

Germany 

Tel.: 0049-8152-7000 

Fax: 0049-8152-7002094
Newtron 

Francesco Carra 

Via Emilia 22 

I-20097 

San Donato, Milanese (MI) 

Italy 

Tel.: 0039-025276252 

Fax: 0039-0255604379
Dental Serviceteknik AB

Lennart Hansson 

Carl Gustafs väg 34 

SE – 21410 

Malmö

Sweden

Tel.: 0046-70 55 46 465 

Fax: 0046-40 26 60 44

Tedensa 

Lorenzo Raul 

Corrales Marques 

C/López Recuero, 3 - local B

E - 28038 

Madrid

Spain 

Tel.: 0034-91-328 16 72 

Fax: 0034-91-328 16 72
HSC Industry 

Jan Hofmann 

Gogolova 275/24 

400 04  

Trmice 

Tschechien 

Tel.: 00420-47-5620808 

Fax: 00420-47-5620299
DSD Dental Service Dienst

Jaap Wim Hr. John 

Wolters de Jong Bosma Ketting 

Markerkant 13-03 i 

NL – 1314 

AL Amere

Netherlands

Tel.: 0031-36-53 4 32 12 

Fax: 0031-36-534 59 42
3M (Schweiz) AG

Eggstrasse 93 

8803 Ruschlikon 

Zürich

Switzerland 

Tel.: 0041-1-724-9251 

Fax: 0041-1-724-9238
3M Unitek

9-15 Chilvers Road

Thornleigh, N.S.W. 2120

Sydney, Australia

Free Call: 1300 363 484 Australia

Free Call: 0800 441 622 New Zealand

Fax: 61 2 9875 6418 Australia

Fax: 0800 441 620 New Zealand
X-tron 

John Thomas 

Bojlund Sjögren 

Kjeldsgardsvej 39-41 

DK-2500 Valby

Denmark

Tel.: 0045-36176517 

Plandent oyi 

Kari Sami 

Hämäläinen Oksanen

Asentajankatu 6 

810 

Helsinki

Finnland 

Tel.: 00358-204-595-294 

Fax: 00358-204-595-206
Infocare ASA 

Steinar Eriksen 

Breijvollveien 25 

614 

Oslo

Norway 

Tel.: 0047-23171200 

Fax: 0047-23171201
T.A.S 

Chris Papamihail 

Tzavella 3 

14231 

Nea Ionia Athens 

Greece 

Tel.: 030-10-2715937 

Fax: 030-10-2715937
3M Heathcare Center 

Hiroshi Yajima

3-8-8 Minamihasimoto

Sagamihara City

Kanagawa   

Japan 

Tel.: 0081-42-770-3453 

Fax: 0081-42-770-3453
For service in the U.S.

3M Health Care Service Center,  

3M ESPE Dental Products,  

Suite 200 Bldg.  

502-1W-01, 3350 Granada Avenue N, 

Oakdale, MN  55128

1-800-292-6298
For service in Canada

3M Canada Company,  

Service Support Centre,  

109 Meg Dr., Unit 9, London, Ontario 

N6E 3Z7

3M Unitek Service Centers

 

English  12

 


background image

 

Guidance and manufacturers declaration – electromagnetic emission 

The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment 
specified below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that 
it is used in such an environment. 

E m issio n s te st 

C o m p lia n ce 

E le ctro m a g n e tic e n viro n m e n t - g u id a n ce  

RF emissions 

CISPR 11 

Group 1 

The Ortholux Luminous Curing Light unit uses RF energy only 
for its internal function. Therefore, its RF emissions are very 
low and are not likely to cause any interference in 
nearby electronic equipment. 

RF emissions 
 
CISPR 11 

Class B 

Harmonic emissions 
 
IEC 61000-3-2 

Class A, 

Voltage fluctuations / 
flicker emissions 

IEC 61000-3-3 

Complies  

 

The Ortholux Luminous Curing Light unit is suitable for use in 
all establishments, including domestic establishments and 
those directly connected to the public low-voltage power 
supply network that supplies buildings used for domestic 
purposes. 

 

 

 


background image

 

Guidance and manufacturer´s declaration – electromagnetic immunity 

The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified 
below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in 
such an environment. 

Im m u n ity te st 

IE C  6 0 6 0 1  

te st le ve l 

C o m p lia n ce  le ve l 

E le ctro m a g n e tic e n viro n m e n t - 

gu id a n ce 

Electrostatic 
discharge (ESD) 
 
IEC 61000-4-2 

± 6 kV contact 

± 8 kV air 

± 6 kV contact 

± 8 kV air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %. 

Electrostatic 
transient / burst 
 
IEC 61000-4-4 

± 2 kV for power 
supply lines 
 
± 1 kV for input/output 
lines 

± 2 kV for power 
supply lines 
 
± 1 kV for input/output 
lines 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge 
 
IEC 61000-4-5 

± 1 kV differential 
mode 
 
± 2 kV common mode 

± 1 kV differential 
mode 
 
± 2 kV common mode  

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
 
IEC 61000-4-11 

< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 0,5 cycle 
 
40 % U

 

(60 % dip in U

T

 ) 

for 5 cycles 
 
 70 % U

 

(30 % dip in U

T

 ) 

for 25 cycles 
 
< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 5 sec  

< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 0,5 cycle 
 
40 % U

 

(60 % dip in U

T

 ) 

for 5 cycles 
 
 70 % U

 

(30 % dip in U

T

 ) 

for 25 cycles 
 
< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 5 sec 

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the Ortholux 
Luminous Curing Light unit requires 
continued operation during power mains 
interruptions, it is recommended that 
the Ortholux Luminous Curing Light unit 
be powered from an uninterruptible 
power supply or a battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 

NOTE         U

T  

is the a. c. mains voltage prior to application of the test level. 

 

 

 

 


background image

 

Guidance and manufacturer´s declaration – electromagnetic immunity 

The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified 
below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in such 
an environment. 

Im m u n ity te st 

IE C  6 0 6 0 1  te st 

le ve l 

C o m p lia n ce  

le ve l 

E le ctro m a g n e tic e n viro n m e n t - 

gu id a n ce 

 

 

 

 

 

Conducted RF 

IEC 61000-4-6 

 

Conducted RF 

IEC 61000-4-3 

 

 

 

 

 

 

3 Vrms 
 
150 kHz to 80 MHz 

 

3 V/m 

80 MHz to 2,5 GHz 

 

 

 

 

 

 

10 V 

 

 

3 V/m 

 

 

 

Portable and mobile RF communications 
equipment should be used no closer to any 
part of the Ortholux Luminous Curing Light 
unit, including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of the 
transmitter. 

R e co m m e n d e d  se p a ra tio n  d ista n ce  

 

   

80 MHz to 800 MHz 

   

800 MHz to 2,5 GHz 

where p is the maximum output power rating 
of the transmitter in watts (W) according to 
the transmitter manufacturer and d is the 
recommended separation distance in metres 
(m).

b

 

Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site 
survey,

should be less than the compliance 

level in each frequency range.

Interference may occur in the vicinity of 
equipment marked with the following symbol: 

 

NOTE 1      At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2      These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people. 

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic 
site survey should be considered. If the measured field strength in the location in which the Ortholux Luminous 
Curing Light unit is used exceeds the applicable RF compliance level above, the Ortholux Luminous Curing 
Light unit should be observed to verify normal operation. If abnormal performance is observed, additional 
measures may be necessary, such as reorienting or relocating the Ortholux Luminous Curing Light unit. 

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.  

 


background image

Recommended separation distances between portable and mobile RF 

communications equipment and the Ortholux™ Luminous Curing Light unit

 

The [EQUIPMENT or SYSTEM] is intended for use in an electromagnetic environment in which radiated RF 
disturbances are controlled. The customer or the user of the Ortholux™ Luminous Curing Light unit can help 
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the Ortholux Luminous Curing Light unit as recommended 
below, according to the maximum output power of the communications equipment 

 

S e p a ra tio n  d ista n ce  a cco rd in g  to  fre q u e n cy o f tra n sm itte r 

m  

R a te d  m a xim u m  

o u tp u t o f 

tra n sm itte r 

 

1 5 0  kH z to  8 0  M H z  

 

8 0  M H z to  8 0 0  M H z  

 

8 0 0  M H z to  2 ,5  G H z  

 

0,01 

0,04

 

0,12

 

0,40

 

0,1 

0,11

 

0,38

 

1,26

 

0,35

 

1,20

 

4,00

 

10 

1,11

 

3,79

 

12,65

 

100 

3,50

 

12,00

 

40,00

 

For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) 
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power 
rating of the transmitter in watts (W) according to the transmitter manufacturer. 

NOTE 1         At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 

NOTE 2         These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.